Fixed, TEST, Wed Dec 4 19:19:38 2019, Marlène, GRIT, Hardware, , Modifications of set-up
|
We have installed the new MUFEE board in the reaction chamber for MG2. This board is a new version with ATHED.
We have moved the MG9 board to become the board to read the future MUST2 detector at 90 deg. This board is now perpendicular to the other ones. We have |
Pending, TEST, Tue Jan 21 17:55:12 2020, Marlène, MUST2, Hardware, , Ovv error on the -5V of CAEN boards   
|
The Ovv error shows on the -5V (first one) of the 3 board (5,3.3,10V for MUST2§squares)
It means that the voltage is over the asked value.
If we ask 5V, we get 5.5V in reality
|
Fixed, TEST, Wed Jan 22 15:53:25 2020, Marlène, GRIT, Hardware, , Changes for MUGAST2020
|
Changes were made for MUGAST 2020 :
1- the board for MG10 before (2nd square) is now for MG6 new trap
2- the board for MG9 is put vertical and used for the MUST2 at 90 deg
|
Fixed, TEST, Thu Jan 23 14:06:51 2020, Franco&Marlène, GRIT, Hardware, 3-alpha, Status of detectors on 23-01-20 6x
|
Short alpha source run (not saved on disk) to test the detectors.
Summary :
|
Fixed, TEST, Mon Feb 10 18:42:23 2020, Franco&Marlène, GRIT, Hardware, , New MUGAST detectors MG2 MG6
|
MG2 and MG6 have been mounted in the chamber.
MG2 reference number is 3227_10 thickness=468um
|
Fixed, TEST, Wed Feb 12 09:32:17 2020, mugast, RUN, General, 3-alpha, Run 500 - 3alpha - mugast standalone 
|
RUN 500
start: 11/02/2020 @ 19h00
|
Fixed, TEST, Wed Feb 12 16:24:52 2020, Marlène, Freddy & Franco, MUST2, Hardware, , Modifications on the electronic scheme
|
We cleaned the electronic scheme.
Now the HF CLEAN VAMOS is taken directly from VAMOS and not re-cleaned by us.
TACs are all ok. Inspection lines have been checked and labeled.
|
Fixed, TEST, Fri Feb 14 10:49:22 2020, Marlène & Franco, RUN, General, 3-alpha, MG11 issues
|
We analyzed the behaviour of the annular. We realized that the 4-peaks behaviour on X was already present last year but only on one sector, now it's on
two sectors. We noticed that there is a correlation between the malfunctioning sectors and the missing strips on Y. Last year there was only a missing
strip on Y, affecting only one sector on X; now 3 strips are missing on Y, affecting 2 sectors on X. We believe that the fact that the boundings are missing |
Fixed, TEST, Fri Feb 14 17:26:05 2020, mugast, OTHER, General, N/A, Run 517 - Pedestal with source 
|
RUN 517
Pedestal run with alpha source |
Fixed, TEST, Mon Feb 17 15:01:40 2020, OttoH.V., OTHER, Hardware, N/A, HVM_SUMMARY
|
[SIZE=5][COLOR=green][B]*** hvm watcher summary ***[/B] [/COLOR][/SIZE]
[SIZE=2][COLOR=green]+10v/mg [/COLOR] Slot: [COLOR=blue]5 [/COLOR] (A1516B) Channel:[COLOR=blue]5 [/COLOR] Vmon: [COLOR=blue] 10 V [/COLOR] Imon:
[COLOR=blue]0.406 A [/COLOR] Status: [COLOR=green]On[/COLOR]
|
Fixed, TEST, Mon Feb 17 15:01:41 2020, OttoH.V., OTHER, Hardware, N/A, HVM_SUMMARY
|
[SIZE=5][COLOR=green][B]*** hvm watcher summary ***[/B] [/COLOR][/SIZE]
[SIZE=2][COLOR=green] [/COLOR] Slot: [COLOR=blue]8 [/COLOR] (A1821H) Channel:[COLOR=blue]1 [/COLOR] Vmon: [COLOR=blue] 0 V [/COLOR] Imon:
[COLOR=blue] 0 uA[/COLOR] Status: [COLOR=pink]Off[/COLOR]
|
Fixed, TEST, Mon Feb 17 15:01:41 2020, OttoH.V., OTHER, Hardware, N/A, HVM_SUMMARY
|
[SIZE=5][COLOR=green][B]*** hvm watcher summary ***[/B] [/COLOR][/SIZE]
[SIZE=2][COLOR=green]+10V/t1234 [/COLOR] Slot: [COLOR=blue]3 [/COLOR] (A1516B) Channel:[COLOR=blue]5 [/COLOR] Vmon: [COLOR=blue] 9.99 V [/COLOR] Imon:
[COLOR=blue]0.159 A [/COLOR] Status: [COLOR=green]On[/COLOR]
|
Fixed, TEST, Mon Feb 17 18:40:25 2020, OttoRun, RUN, General, N/A, Run518
|
*************************
[SIZE=3][B]Run : 518[/B][/SIZE]
*************************
|
Fixed, TEST, Mon Feb 17 19:20:44 2020, OttoRun, RUN, General, N/A, Run519
|
*************************
[SIZE=3][B]Run : 519[/B][/SIZE]
*************************
|
Fixed, TEST, Mon Feb 17 19:26:41 2020, OttoRun, RUN, General, N/A, Run520
|
*************************
[SIZE=3][B]Run : 520[/B][/SIZE]
*************************
|
Fixed, TEST, Mon Feb 17 19:32:25 2020, OttoRun, RUN, General, N/A, Run521
|
*************************
[SIZE=3][B]Run : 521[/B][/SIZE]
*************************
|
Fixed, TEST, Tue Feb 18 09:20:51 2020, mugast, OTHER, General, N/A, Bilan Runs de Calibrations
|
---------
ENERGIE
---------
|
Fixed, TEST, Tue Feb 18 10:00:40 2020, mugast, RUN, General, 3-alpha, RUN 501 - missing info
|
RUN 501
alpha run with cooling during the night from 13-14/02/2020 |
Fixed, TEST, Tue Feb 18 17:50:34 2020, OttoRun, RUN, General, N/A, Run522
|
*************************
[SIZE=3][B]Run : 522[/B][/SIZE]
*************************
|
Fixed, TEST, Tue Feb 18 17:53:36 2020, Franco, GRIT, General, 3-alpha, Alpha calibration for exp e775s (run 0501)   
|
In attachment raw and calibrated spectra for run 0501 (alpha source) for MUST2 and Mugast.
Overall resolution of 36 keV on MUST2 and 40 keV on MUGAST detector. |
|