MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 106 of 132  Not logged in MUGAST
Status ExpNbr Date Authordown Type Category Target-Source Subject
Fixede768s Tue Jul 23 19:21:50 2019 NicolasOTHERSocialN/AWaiting end of canicule (and beam!)
Attachment 1: IMG_20190723_173257.jpg
IMG_20190723_173257.jpg
Fixede768s Wed Jul 24 16:24:18 2019 NicolasOTHERGeneral 19Ne / 19F analog connections
Attachment 1: ne19_f19_AnalogStates.pdf
ne19_f19_AnalogStates.pdf ne19_f19_AnalogStates.pdf
Fixede768s Sun Jul 28 16:23:48 2019 NicolasRUNGeneral Update on possible restart
Same situation as before lunch, the beam is still in C0 and not yet in CSS1. We may have beam in late evening....
Ongoinge768s Sun Jul 28 19:58:57 2019 NicolasRUNGeneral Update on possible restart
Beam is now at the exit of CSS2 though We may have beam around midnight....
Fixede768s Sun Jul 28 21:59:43 2019 NicolasRUNGeneral Update on possible restart
Beam is in G1 with 3e6 pps. PCP will optimize for about 1-1,5h before we take the beam.
Fixede768s Tue Jul 30 15:47:46 2019 NicolasBEAMGeneral Beam stopped since 11h30
vacuum problem in CSS1
Pinnede775s Thu Feb 27 12:44:37 2020 NicolasOFFLINESoftware How to produce control spectra
In /home/mugast/analysis/nptool/Projects/MUGAST/e775s directory there is the Control.C macro which draws: Ex, Ex vs Eg and Elab vs thetalab for each run (run number must be given in argument) or for chained runs (argument must be 0)
Pinnede775s Thu Feb 27 16:35:26 2020 NicolasOFFLINESoftware How to produce control spectra

Nicolas wrote:
In /home/mugast/analysis/nptool/Projects/MUGAST/e775s directory there is the Control.C macro which draws: Ex, Ex vs Eg and Elab vs thetalab for each run (run number must be given in argument) or for chained runs (argument must be 0)



A new file has been created to look into the CD2 target runs : Control_CD2.C
Fixede793s Sun Mar 14 18:13:28 2021 Miguel, Wilton, Nishu, Dieter, AblaihanRUNGeneralCD2run62 screenshots at 18h14
Attachment 1: Screenshot_from_2021-03-14_18-01-54_IM_MG.png
Screenshot_from_2021-03-14_18-01-54_IM_MG.png
Attachment 2: Screenshot_from_2021-03-14_18-02-54_IM_MM.png
Screenshot_from_2021-03-14_18-02-54_IM_MM.png
Attachment 3: Screenshot_from_2021-03-14_18-03-57_ETOF_MM.png
Screenshot_from_2021-03-14_18-03-57_ETOF_MM.png
Attachment 4: Screenshot_from_2021-03-14_18-04-25_ETOF_MG.png
Screenshot_from_2021-03-14_18-04-25_ETOF_MG.png
Attachment 5: Screenshot_from_2021-03-14_18-05-33_EdE_MM.png
Screenshot_from_2021-03-14_18-05-33_EdE_MM.png
Attachment 6: Screenshot_from_2021-03-14_18-06-44_EdE_MG.png
Screenshot_from_2021-03-14_18-06-44_EdE_MG.png
Attachment 7: Screenshot_from_2021-03-14_18-09-32_MW_TDC.png
Screenshot_from_2021-03-14_18-09-32_MW_TDC.png
Attachment 8: Screenshot_from_2021-03-14_18-08-55_TAC.png
Screenshot_from_2021-03-14_18-08-55_TAC.png
Attachment 9: Screenshot_from_2021-03-14_18-11-08_Scaler1.png
Screenshot_from_2021-03-14_18-11-08_Scaler1.png
Attachment 10: Screenshot_from_2021-03-14_18-11-28_Scaler2.png
Screenshot_from_2021-03-14_18-11-28_Scaler2.png
Fixede793s Sun Mar 14 20:11:17 2021 Miguel, Wilton, Nishu, Dieter, AblaihanRUNGeneralCD2run62 screenshots at 20h00
Beam progressively down from 5 to 4 10^5. Went to PCP, they were losing intensity in the primary beam, from 500 to 400 nA. Working on it (21:30)...
Attachment 1: run62_Impact_Matrix_MM_Screenshot_from_2021-03-14_20-02-49.png
run62_Impact_Matrix_MM_Screenshot_from_2021-03-14_20-02-49.png
Attachment 2: run62_Scaler2_Screenshot_from_2021-03-14_20-10-24.png
run62_Scaler2_Screenshot_from_2021-03-14_20-10-24.png
Attachment 3: run62_Scaler1_Screenshot_from_2021-03-14_20-09-59.png
run62_Scaler1_Screenshot_from_2021-03-14_20-09-59.png
Attachment 4: run62_MW_TDC_Screenshot_from_2021-03-14_20-09-30.png
run62_MW_TDC_Screenshot_from_2021-03-14_20-09-30.png
Attachment 5: run62_Raw_TAC_Screenshot_from_2021-03-14_20-08-52.png
run62_Raw_TAC_Screenshot_from_2021-03-14_20-08-52.png
Attachment 6: run62_EdE_MG_Screenshot_from_2021-03-14_20-08-17.png
run62_EdE_MG_Screenshot_from_2021-03-14_20-08-17.png
Attachment 7: run62_EdE_MM_Screenshot_from_2021-03-14_20-06-09.png
run62_EdE_MM_Screenshot_from_2021-03-14_20-06-09.png
Attachment 8: run62_E_TOF_MG_Screenshot_from_2021-03-14_20-05-42.png
run62_E_TOF_MG_Screenshot_from_2021-03-14_20-05-42.png
Attachment 9: run62_E_TOF_MM_Screenshot_from_2021-03-14_20-04-58.png
run62_E_TOF_MM_Screenshot_from_2021-03-14_20-04-58.png
Attachment 10: run62_Impact_Matrix_MG_Screenshot_from_2021-03-14_20-04-03.png
run62_Impact_Matrix_MG_Screenshot_from_2021-03-14_20-04-03.png
Fixede793s Sun Mar 14 22:01:04 2021 Miguel, Wilton, Ablaihan,nishuRUNGeneralCD2summary of run 62@22:00hrs
Attachment 1: Screenshot_from_2021-03-14_22-03-45.png
Screenshot_from_2021-03-14_22-03-45.png
Attachment 2: Screenshot_from_2021-03-14_22-03-31.png
Screenshot_from_2021-03-14_22-03-31.png
Attachment 3: Screenshot_from_2021-03-14_22-03-19.png
Screenshot_from_2021-03-14_22-03-19.png
Attachment 4: Screenshot_from_2021-03-14_22-03-01.png
Screenshot_from_2021-03-14_22-03-01.png
Attachment 5: Screenshot_from_2021-03-14_22-02-49.png
Screenshot_from_2021-03-14_22-02-49.png
Attachment 6: Screenshot_from_2021-03-14_22-02-20.png
Screenshot_from_2021-03-14_22-02-20.png
Attachment 7: Screenshot_from_2021-03-14_22-02-06.png
Screenshot_from_2021-03-14_22-02-06.png
Attachment 8: Screenshot_from_2021-03-14_22-08-49.png
Screenshot_from_2021-03-14_22-08-49.png
Attachment 9: Screenshot_from_2021-03-14_22-02-34.png
Screenshot_from_2021-03-14_22-02-34.png
Fixede793s Sun Mar 14 17:16:01 2021 Miguel, Wilton, AblaihanBEAMHardwareCD2Beam went down for a couple of minutes at 17:13. Now it is ok, screenshot attached.
Beam went up to 2x10^6 pps in CATS.

Then it went down to 0 for a couple of minutes at 17:13.

Now it is ok, screenshot attached.
Attachment 1: Screenshot_from_2021-03-14_17-21-50_Beam_is_ok.png
Screenshot_from_2021-03-14_17-21-50_Beam_is_ok.png
Fixede768s Tue Jul 23 14:33:03 2019 Marlène, Yorick, Faïrouz & NicolasBEAMGeneral Cheking beam transmission from G1 profiler to VAMOS
Goal:
Estimate beam transmission from profiler to VAMOS (FP)
We question whether the factor of two found yesterday (VAMOS/Profiler ~ 0.5) could be due:
1. to an incorrect measurement of the beam intensity with the profiler
2. a charge state equilibration in the target (LISE++ calculation indicates 8+/7+ = 85%/15%)

Same conditions as yesterday:
1. Beam intensity is reduced with the pepper pot 1/9 and by closing slits after the 15O target production and the accelerators so that the beam quality (emittance, etc...) is not affected.
2. VAMOS plate is removed to allow the direct beam detection

Results:
We measure in VAMOS with the LiF target and with the target ladder totally removed. In the latter case, direct beam 15O(8+) is sent in VAMOS and no charge state effect enter into the game.
1. Beam intensity at G1 profiler ~ 6.5e4 pps
VAMOS (FP) + no target ~ 3.8e4 pps
2. Beam intensity at G1 profiler ~ 3.8e4 pps
VAMOS (FP) + thick LiF (nb 1) ~ 2.0e4 pps

Discussion:
We find the same results as yesterday, e.g. a factor of two attenuation between the profiler and VAMOS. If part of this loss would happen in the pipe/target/MUST2 mask, this should be observed in AGATA which is not the case. Could it be that the PPAC in VAMOS is not as efficient as announced at the beginning of the campaign (98%)? This must be checked when we get beam back later on.

Note:
While doing the above measurement, we observed a factor of 3 loss between profiler AR.Gaz13 and G1.Gaz31. Numbers follows:
L3.GAZ42 AR.GAZ13 G1.GAZ31
2.1e4 1.5e4 4.1e3
1.4e5 1.0e5 2.5e4
??? 1.8e5 6.5e4
Attachment 1: 714D4921-2A75-44B7-AD8D-87048BDA17A8.jpeg
714D4921-2A75-44B7-AD8D-87048BDA17A8.jpeg
Attachment 2: 6A77930D-8992-4118-B59C-DBAA37289533.jpeg
6A77930D-8992-4118-B59C-DBAA37289533.jpeg
Attachment 3: 3CD66D1D-D6E6-44B4-A15A-C947F048D2A0.jpeg
3CD66D1D-D6E6-44B4-A15A-C947F048D2A0.jpeg
Fixede768s Mon Jul 15 19:59:30 2019 Marlène, Nicolas, SergeRUNGeneralN/ARun10 - fsc
*************************
Run : 10
*************************

Start at 2019-07-15 19:57:41.031967

fsc / calibT run

GMT: MUGAST (square detector not biased) + MUST2

Time calibrator:
range -> 640 ns
freq. -> 20 ns
Fixede768s Mon Jul 15 20:18:07 2019 Marlène, Nicolas, SergeRUNGeneralN/ARun13 - fsc
*************************
Run : 13
*************************

Start at 2019-07-15 20:18:02.650063

fsc / calibT run

internal pulser frequency now set to 300 Hz (was about 1 kHz before)

GMT: MUGAST (square detector not biased) + MUST2

Time calibrator:
range -> 640 ns
freq. -> 20 ns

Stop at 2019-07-15 20:23:54.304478
Stop comment: type your message here
FixedTEST Wed Feb 12 16:24:52 2020 Marlène, Freddy & FrancoMUST2Hardware Modifications on the electronic scheme
We cleaned the electronic scheme.
Now the HF CLEAN VAMOS is taken directly from VAMOS and not re-cleaned by us.
TACs are all ok. Inspection lines have been checked and labeled.
GMT and Scalers are like below.

GMT:
1 - MUVI 1 (MG1, MG3, MG4, MG5)
2 - MUVI 2 (MG7, MG2, MG6, MG11)
3 - MUVI 3 (MM1, MM2, MM3, MM4)
4 - MUVI 4 (MG9 90-deg MUST2)
5 - COINCIDENCE VAMOS-AGATA
6 - VAMOS DIV
7 - MUST2 DIV
8 - broken
9 - SQUARE DIV
10 - AGATA DIV
11 - VAMOS
12 - empty
13 - empty
14 - empty
15 - broken
16 - empty

SCALERS:
1 - FAG
2 - OR MUGAST
3 - AGATA
4 - VAMOS
5 - empty
6 - DEAD TIME MUGAST
7 - 100 Hz PULSER
8 - DEAD TIME
9 - MUVI 1 (MG1, MG3, MG4, MG5)
10 - MUVI 2 (MG7, MG2, MG6, MG11)
11 - MUVI 3 (MM1, MM2, MM3, MM4)
12 - MUVI 4 (MG9 90-deg MUST2)
13 - empty
14 - empty
15 - CATS2
16 - FINGER
Fixede768s Mon Jul 22 11:42:22 2019 Marlène, François D, FranckBEAMGeneralLiFBeam profile and intensity
Intensity = 1.8e7 pps

Slight shifts in x and y compared to this morning 8:30
x : -1.1 mm
y : +1.7 mm
Attachment 1: IMG_20190722_113900.jpg
IMG_20190722_113900.jpg
Fixede768s Mon Jul 22 16:54:33 2019 Marlène, Franck, François & NicolasBEAMGeneral Cheking beam transmission from G1 profiler to VAMOS
Goal:
Estimate beam transmission from profiler to VAMOS (FP)

Method:
1. Beam intensity is reduced with the pepper pot 1/9 and by closing slits after the 15O target production and the accelerators so that the beam quality (emittance, etc...) is not affected.
2. VAMOS plate is removed to allow the direct beam detection
3. LiF (1.25 mg/cm2, number 1) is in position

Results:
1. beam intensity at G1 profiler ~ 4e4 pps
2. VAMOS FP ~ 2e4 pps

Discussion:
1. There is a factor of two loss between the profiler and VAMOS. Taken this factor into account in the g-ray yield estimate (see post ..) the numbers of 1297- and 1340-keV g-rays is in line with what is expected from cross-section calculation.
2. Loosing a factor of two means a substantial amount of 15O is implanted in the pipe/target frame/MUST2 masks. A dedicated AGATA activation measurement could be performed to cross-check this factor of two.
Fixede768s Mon Jul 22 17:07:39 2019 Marlène, Franck, François & NicolasBEAMGeneral Profiler conditions for measuring beam intensity
We checked that the measured beam intensity (present case 1.5e7 pps) is independent on the voltage on the profiler. Values from 93 V up to 177 V gave similar results.
Fixede768s Fri Jul 26 02:27:48 2019 Marlène, Faïrouz, Franck, Bertrand, Sylvain, Diego, NicolasBEAMGeneral G1 profiler efficiency
G1 profiler efficiency was checked against CATS1 for two beam intensities. Agreement between the two measurements is within 20%

Profiler -> CATS1
2.3e4 pps -> 1.8e4 pps
1.4e5 pps -> 1.16e5 pps

Conclusion:
G1 profiler is measuring correctly the beam intensity for these intensities
Goto page Previous  1, 2, 3 ... 105, 106, 107 ... 130, 131, 132   Next  