Status |
ExpNbr |
Date |
Author |
Type |
Category |
Target-Source |
Subject |
Fixed | e748 |
Mon Jul 3 21:13:18 2017 |
e748 | OFFLINE | General | N/A | cats |
run222 cats |
Attachment 1: Capture.png
|
|
Fixed | e748 |
Mon Jul 3 21:41:57 2017 |
Valérie L | RUN | General | CD2 | Runs 225-227 |
---------------------------------------------------------------------------------
3 µA at 21h25 ; CAVIAR D6 28000/s
CATS 20000/s
DIV = 100 (CATS1, CAT2, PLAST)
TRIGGERS MM1, MM8
RUN 225 3 JULY start 21h29 STOP at 2&h51 due to problem of DAD blocking of DAQ.FAG at 0
(start ok checked with Ganil-Tape)
-r--r--r-- 1 e748 Exxx 7143424 3 juil. 21:51 run_0225.dat.03-07-17_21h29m23s
RUN 226 227 JUNK
---------------------------------------------------------------------------------
Run 224 JUNK buffer header (stop due to problem -->checked with Ganil Tape)
rw-r--r-- 1 e748 Exxx 1835008 3 juil. 21:26 run_0224.dat.03-07-17_21h24m55s
--------------------------------------------------------------------------------- |
Attachment 1: Run225l3julyScalerAnalyzer3µA.png
|
|
Fixed | |
Mon Jul 3 22:10:28 2017 |
Valérie L | CATS | Hardware | CD2 | VXI CATS crate problem solved |
We had problems with CATS VME/VXI crate.
It was idle and it was not possible to restart it by breakup and init.
We enter in the room to reset the ganlx7 of CATS.
After the crate was correct and we could do INIT and start.
The visu in the CATS terminal was ok. |
Attachment 1: VisuCATproblemTerm.png
|
|
Attachment 2: VXICATSreset.png
|
|
Fixed | e748 |
Mon Jul 3 22:24:05 2017 |
Valérie L | RUN | General | CD2 | Run series of problems 3rd July |
The daq was unstable with several problems of the merge from LISE.
We did tk kna (kill narval process) and na.
but the next run was starting still unstable with Error from the merge
It qeems that the counting rate is too high due to low division on the triggers.
For the moment CATS div is 100, we will put 300.
plast div is 100.
We will put it at 1000 for the next runs.
Enter in the room to change division.
r--r--r-- 1 e748 Exxx 7143424 3 juil. 21:51 run_0225.dat.03-07-17_21h29m23s
-r--r--r-- 1 e748 Exxx 720896 3 juil. 21:55 run_0226.dat.03-07-17_21h54m51s STOP DUE TO CATS VXI problems
-rw-r--r-- 1 e748 Exxx 786432 3 juil. 21:56 run_0227.dat.03-07-17_21h55m56s STOP DUE TO DAQ blocking /MERGE message LIE
-r--r--r-- 1 e748 Exxx 57016320 3 juil. 22:16 run_0228.dat.03-07-17_22h12m59s STOP DUE TO MERGE message LISE
-r--r--r-- 1 e748 Exxx 52232192 3 juil. 22:22 run_0229.dat.03-07-17_22h19m12s |
Fixed | |
Mon Jul 3 22:40:33 2017 |
e748 | RUN | General | | Plastic_Cats_T |
Tac run_223
Range: T_CATS1_CAV 1 µs
T_CATS1_2 200ns
T_MUVI_CATS1 500ns
T_PL_CATS1 1µs
T_PLch_CATS1 1µs
T_PL_CATS2 1µs
T_Pl_CHIO 2µs
|
Attachment 1: run_223_tac2.pdf
|
|
Attachment 2: run_223_tac1.pdf
|
|
Fixed | e748 |
Mon Jul 3 23:06:24 2017 |
Valérie L | RUN | General | CD2 | Problem during Run 230-231 with CATS |
r-- 1 e748 Exxx 34013184 3 juil. 22:52 run_0230.dat.03-07-17_22h46m03s
-rw-r--r-- 1 e748 Exxx 97910784 3 juil. 23:09 run_0231.dat.03-07-17_22h54m42s |
Attachment 1: ResetEventCounterAfterRun230Centrum.png
|
|
Attachment 2: PRoblemRun230.png
|
|
Attachment 3: ScalersRUn230.png
|
|
Fixed | e748 |
Mon Jul 3 23:16:33 2017 |
Valérie L | DAQ | General | CD2 | Change of TAC RANGES (from run 230) |
TAC range changed after run 229.
NB: Change applied for Runs 230 and 231
BUT these runs are not working due to problems of Merger
(CATS not sending events in the branch of the DAQ).
The range was changed for all the TAC previously with 1µs, now put to 500ns
CATS1_FAG CAVIAR (del 550ns) 500ns T_CATS1_CAV ADC VXI MUST2 add=19
CATS1_FAG CATS2_FAG (del 33ns) 200ns T_CATS1_2 ADC VXI MUST2 add=28
OU_MUVI CATS1_FAG 500ns T_MUVI_CATS1 ADC VXI MUST2 add=21
CATS2_FAG PLAST (val) 500ns T_PL_CATS2 ADC VXI MUST2 add=22
CATS1_FAG PLAST (no val) 500ns T_PL_CATS1 ADC VXI MUST2 add=23
-------------------------------------------------------------------------------------------
CATS1_FAG PLAST (no val) 500ns T_PLch_CATS1 ADC VXI CHIO add=19
this TAC was changed, with Plastic with no validation by FAG; Range was 2µs before, now 5 µs
start:PLAST ChIo CHIO (no val) 5µs T_PL_CHIO ADC VXI CHIO add=3
-------------------------------------------------------------------------------------------
TAC EXL HF 100ns EXL_HF ADC VXI MUST2 add=18 |
Fixed | |
Tue Jul 4 00:53:18 2017 |
Valérie L | RUN | General | CD2 | Runs with merge and time out problems |
-r--r--r-- 1 e748 Exxx 720896 3 juil. 21:55 run_0226.dat.03-07-17_21h54m51s
-rw-r--r-- 1 e748 Exxx 786432 3 juil. 21:56 run_0227.dat.03-07-17_21h55m56s
-r--r--r-- 1 e748 Exxx 57016320 3 juil. 22:16 run_0228.dat.03-07-17_22h12m59s
-r--r--r-- 1 e748 Exxx 52232192 3 juil. 22:22 run_0229.dat.03-07-17_22h19m12s
-rw-r--r-- 1 e748 Exxx 34013184 3 juil. 22:52 run_0230.dat.03-07-17_22h46m03s
-r--r--r-- 1 e748 Exxx 241238016 3 juil. 23:37 run_0231.dat.03-07-17_22h54m42s |
Fixed | |
Tue Jul 4 00:54:41 2017 |
Valérie L | RUN | General | CD2 | Runs after reset of VXI |
VXI CHIO was offline --> reset of ganlx9 (entering in D6) on the VXI crate
-r--r--r-- 1 e748 Exxx 34013184 4 juil. 00:14 run_0232.dat.04-07-17_00h10m44s
-rw-r--r-- 1 e748 Exxx 5177344 4 juil. 00:17 run_0233.dat.04-07-17_00h16m00s
-rw-r--r-- 1 e748 Exxx 2162688 4 juil. 00:26 run_0234.dat.04-07-17_00h19m48s
-rw-r--r-- 1 e748 Exxx 2097152 4 juil. 00:30 run_0235.dat.04-07-17_00h28m10s
-------------------------------------------------------------------------------------------------
-rw-r--r-- 1 e748 Exxx 80609280 4 juil. 00:41 run_0236.dat.04-07-17_00h30m48s
-------------------------------------------------------------------------------------------------
-rw-r--r-- 1 e748 Exxx 111411200 4 juil. 00:46 run_0237.dat.04-07-17_00h41m02s BEAM ON
All triggers MM1 to MM8, CATS1 div = 300, CATS2 div = 300, PLAST = 1000
-------------------------------------------------------------------------------------------------
same conditions for run 238
-rw-r--r-- 1 e748 Exxx 120061952 4 juil. 00:58 run_0238.dat.04-07-17_00h47m14s
end of run 238
Level : ERROR
Date : 04/07/2017 00:48:05
Logger : vme
Remote Info : localhost:60507
- - - - - - - - - -
Message :
VXCAMEXE on ganlx1 : - TGV Watch Dog - 25 errors |
Fixed | e748 |
Tue Jul 4 01:00:26 2017 |
Valérie L | RUN | General | CD2 | Run 239 test with trigger MUST2 |
Started run #239 4JULY start 00h59 stop 1h15
PROBLEM junk header checked with Ganil Tape
triggers only MUST2 IN1 to IN8
rw-r--r-- 1 e748 Exxx 55181312 4 juil. 01:15 run_0239.dat.04-07-17_00h59m21s
triggers only MUST2 IN1 to IN8 |
Fixed | e748 |
Tue Jul 4 01:17:35 2017 |
Valérie L | RUN | General | CD2 | Run 240 |
Run #240 4JULY start 1h17
(run ok checked with Ganil_Tape)
All triggers MUST2 IN1 to IN8,
IN9 CATS1, IN10 CATS2 div = 300; IN11 PLASTIC |
Fixed | e748 |
Tue Jul 4 01:21:04 2017 |
e748 | OFFLINE | General | CD2 | run 236 |
cats > target
cats > target, gated by must |
Attachment 1: Capture.png
|
|
Fixed | e748 |
Tue Jul 4 02:33:36 2017 |
Valérie L, shifts U+S Yorick | RUN | Hardware | | GMT channel 8 |
Apparently Channel 8 of the GMT is not operational. We put telescope T8 in entry 15.
This operation gives good results, see pictures below of Gatconf with 2**(N-1) = 2**14 = 16384 for Run 248.
Triggers GMT are now:
IN1 to IN7 for T1 to T7.
Telescope T8 is on IN15.
Triggers IN9 CATS1 div =300; IN10 CATS2 div div =300
IN11 Plastic div (div=1000)
run_0248.dat.04-07-17_02h55m52s
Run 248 (header ok with GanilTape) start 2h55 STOP.. 3h56 584 MB -> apprx. 100 MB/10 min 1 GB ~ 100 min.
Switched the run to complete one data file for offline checking. no error no issue during the run. |
Attachment 1: GMTtriggerRun248.png
|
|
Attachment 2: GMTtriggerGatconfRun248.png
|
|
Fixed | e748 |
Tue Jul 4 04:01:50 2017 |
e748 | RUN | Hardware | | RUN249 |
RUN249
Same setting as RUN248.
run_0249.dat.04-07-17_03h56m24s
Start 3h56 Gtape OK
Stop 4h58
608 MB |
Fixed | e748 |
Tue Jul 4 05:01:21 2017 |
Yorick | RUN | General | CD2 | Runs 248-250 |
These are long runs with CD2
MUST8 is in GMT 15.
Missing Times: T7 X and Y; T5 Y.
Beam is very stable at 2.8 micro. |
Fixed | e748 |
Tue Jul 4 05:08:30 2017 |
e748 | MUST2 | Hardware | | MUST2 signql check in offline analysis |
RUN248
MUST2 DSSD XE signal for all telescopes T1-T8 confirmed.
MUST2 DSSD YE signal for all telescopes T1-T8 confirmed.
MUST2 DSSD XT signal for all telescopes T1-T6,T8 confirmed. No T7.
MUST2 DSSD XT signal for all telescopes T1-T4,T6,T8 confirmed. No T5,7.
MUST2 Csi E signal for all telescopes T1-T4,T6,T7 confirmed.
MUST2 Csi t signal for no telescopes T1-T8. This is normal.
MUST2 SiLi E signal for all telescopes T5,T8 confirmed.
MUST2 SiLi T signal for all telescopes T5,T8 confirmed. |
Fixed | e748 |
Tue Jul 4 05:11:20 2017 |
e748 | MUST2 | Hardware | | MUST2 signql check in offline analysis |
RUN248
MUST2 DSSD XE signal for all telescopes T1-T8 confirmed.
MUST2 DSSD YE signal for all telescopes T1-T8 confirmed.
MUST2 DSSD XT signal for all telescopes T1-T6,T8 confirmed. No T7.
MUST2 DSSD XT signal for all telescopes T1-T4,T6,T8 confirmed. No T5,7.
MUST2 Csi E signal for all telescopes T1-T4,T6,T7 confirmed.
MUST2 Csi t signal for no telescopes T1-T8. This is normal.
MUST2 SiLi E signal for all telescopes T5,T8 confirmed.
MUST2 SiLi T signal for all telescopes T5,T8 confirmed. |
Attachment 1: 2017-07-04_run248_gatconf.png
|
|
Attachment 2: 2017-07-04_run248_must2_energy_x.png
|
|
Attachment 3: run248_MUST2_xt.png
|
|
Attachment 4: run248_MUST2_yt.png
|
|
Attachment 5: run248_MUST2_CsIE.png
|
|
Attachment 6: run248_MUST2_CsIT.png
|
|
Attachment 7: run248_MUST2_SiLiE.png
|
|
Attachment 8: run248_MUST2_SiLiT.png
|
|
Attachment 9: run248_MUST2_MM1_SiEvsCsIE.png
|
|
Attachment 10: run248_MUST2_DSSDYE.png
|
|
Fixed | e748 |
Tue Jul 4 05:20:31 2017 |
e748 | RUN | Hardware | | RUN250 |
Same setting as RUN248.
run_0250.dat.04-07-17_04h58m50s
start 4h58 Gtape ok
Stop. The HT for SiLi T5 1 tripped off with the over current alert (current limit set 25 micro A for 400 V). Turned on the HT and restarted the run. The current was slowly increasing over time. Watch out the current read. |
Fixed | e748 |
Tue Jul 4 05:23:20 2017 |
e748 | CATS | Software | | CATS signal |
RUN248
The chqrge distribution of CATS1&2, X&Y
Gate:CATSdiv trigger |
Attachment 1: run248_CATS1XQ.png
|
|
Attachment 2: run248_CATS1YQ.png
|
|
Attachment 3: run248_CAT2XQ.png
|
|
Attachment 4: run248_CATS2YQ.png
|
|
Fixed | e748 |
Tue Jul 4 07:00:55 2017 |
e748 | RUN | Hardware | | RUN252 |
Same setting as RUN248.
RUN251 junk header
run_0252.dat.04-07-17_05h33m06s
Start 5h33 end 6h30.
No issue during the run. |