Fixed, e744, Fri May 4 23:58:07 2018, iulian, RUN, General, 3-alpha, Run #203 alfa source & 60co source run_203_ht.pngrun_203_ppt.pngrun_203_EXL.pngrun_203_must2.png
run #203

alpha source & 60co source
start 23h56
stop 11h07
Fixed, e744, Thu May 10 14:38:41 2018, iulian, EXL-PPT, Hardware, N/A, Mesytech settings 
Tried unsuccessfully to change the gain on EXL energies.
Looking at the inspection lines noticed that all the EXL signals were inverted (negative) and we were coding only the pole zero part of the signals.
Looking into the manual find out how to change polarity of the signals. Did that and now the signals look normal with normal gain change after polarity change.
EXL polarity is negative so Mesytech modules have to be set to use negative input signals.

    Pinned, e744, Fri May 11 14:00:48 2018, iulian, EXL-PPT, Hardware, N/A, Mesytech settings 20180511_132034_1526038557770.jpg

iulian wrote:
Tried unsuccessfully to change the gain on EXL energies.
Looking at the inspection lines noticed that all the EXL signals were inverted (negative) and we were coding only the pole zero part of the signals.
Looking into the manual find out how to change polarity of the signals. Did that and now the signals look normal with normal gain change after polarity change.
EXL polarity is negative so Mesytech modules have to be set to use negative input signals.


PPT HV = 650 V

This is happening when the pre-amplifiers for EXL are not alimented.. When they are alimented the EXL sigal polarity is positive.. The gains and thresholds for Mesytech modules are below
Fixed, e744, Sat May 12 08:54:42 2018, iulian, MUST2, General, 3-alpha, Run #114 T1-T5 trigger 3alpha source calib run  Must_spectra_start_run214.pngMust_spectra_stop_run214.png
Run #214
Trigger T1-T5 & gamma
3 alpha run
start @8h15m55s 12/05
stop @20h01 12/05

This run use new das config copied from e744_save04mai_fonctionne.das, updated and then re-saved to e744_save12mai_fonctionne.das
The previous config , saved now under the name e744_save12mai.das, gives no Y energies in the spectras and an error on Muvi2 (some register different)
Ongoing, e744, Sat May 12 20:09:09 2018, iulian, MUST2, General, 3-alpha, Run #115 3 alfa run Calibration T1-T5 trigger Must_spectra_start_run215.pngMust_spectra_stop_run215.pngerror_run215.pngerror_1minlater_run215.png
Run #215
Trigger T1-T5
3 alpha run
start @20h09 12/05
stop 11h40 13/05

Run crashed during the night (looks line NARVAL)...
Needed to stop and start to be able to init VME slave.... VME master was ok
Ongoing, e744, Sun May 13 12:09:55 2018, iulian, MUST2, General, 3-alpha, Run #116 3 alfa run Calibration T1-T5 trigger  Must_spectra_start_run216.png
Run #115
3 alfa run Calibration T1-T5 trigger
start 12h10


start_online gives a ilegal instruction fault from time to time:
Pending, e744, Sun May 13 13:32:44 2018, iulian, ONLINE, Software, N/A, start_online ilegal instruction fault 
From time to time the start_online code gives an ilegal instruction error and crash... Once the ctrl-c pressed it shows:


"Error in <TFile::cd>: No such file /home/e748/analysis/nptool/Outputs/Simulation.root:/"
Fixed, e744, Wed May 16 11:30:21 2018, iulian, EXL-PPT, General, 60Co, run 302 60Co source 
Run #302

60 Co source PPT & EXL trigger

1start 21h46m40s 15/05
stop 10h 16/05
Ongoing, e744, Wed May 16 21:45:47 2018, iulian, MUST2, Software, N/A, Run #323 Calimero T1-T4 strips X Calimerro_run323.png
Run #323
Calimero T1-T4 strips X
Ongoing, e744, Wed May 16 22:03:13 2018, iulian, EXL-PPT, Software, 60Co, run #324 60Co  EXL_run324.pngPPT_run324.png
run #324
60Co

start 22h02 16/05
stop 10h12 17/05
Pinned, e744, Thu May 17 12:04:45 2018, iulian, EXL-PPT, Hardware, 60Co, EXL delay of ADC Codding window  Exl_timing_mesy_1_2.pdf
EXL timing delays for Mesitech module 2 & 1

Module 1
ECL sygnal going to the ADC are delayed by 1.3 us
Trigger output is delayed by 350 ns
EXL input signal has a rise time of 3 us

Module 2
ECL sygnal going to the ADC is delayed by 554 ns
Trigger output is delayed by 350 ns
EXL input signal rise time is 3 us
Ongoing, e744, Thu May 17 18:29:16 2018, iulian, MUST2, General, 3-alpha, Run #333 Last alfa run before experiment Must_spectra_start_run333.png
Run #333
3a source
T1-T5 trigger

start 18h30


ps: If no HT on the detectors but BT on, you will have alfa E&T on X, but not on Y side (except some strips)
Pinned, e744, Wed May 23 15:10:15 2018, iulian, OTHER, General, N/A, Procedure to start stop RUNS fiche_start_run.pdffiche_stop_run.pdf
Fixed, e744, Thu May 24 01:53:28 2018, iulian, RUN, General, N/A, run 438 Lise trigger  run_439_d6_cats_hf.pngrun_439_transmisiond4_d6.png
run 438 D4 in
run 439 d6 in.

D4 D6 transmision with CATS in:

83425./106149.=7.85923560278476430e-01
Ongoing, e744, Thu May 24 02:30:52 2018, iulian, RUN, General, N/A, RUN 440 14N no target trigger Must & cats %100 run440.png
RUN 440
14N beam \
no target
trigger t1-t5
cats %100

some counts in t5. Very low transmission . max 30 while in cats we had 500
Fixed, e744, Thu May 24 03:21:37 2018, iulian, OTHER, Hardware, N/A, Position measurements Must2/PPT/EXL STP-2018-E054_E744_I-Stefan_J-Mrazek_LISE-D6.pdf
Fixed, e744, Thu May 24 03:57:34 2018, iulian, RUN, General, N/A, run 442 Lise trigger d6 cats_e1_d6_t1_D6_run442.png
Fixed, e744, Thu May 24 04:20:34 2018, iulian, RUN, General, N/A, run 443 T1-t5 triggers mm5_run443.pngmm5_2_run443.pngmm5_3_run443.pngmm5_4_run443.png
run 443
T1-T5 triggers

still problems to recover the counting on must2. Cats: 5000 pps , T5: 50 pps max


Below changes in tuning with little change in counting...

The energy for 14N7+ at the entrance of must2 should be 28.44 meV if the beam dump thickness is correct (34 um). If a dead layer of 0.5 um of Al is included, the energy became 27.849. This looks very close to what we see on X strip, and about 1 MeV off on Y strips.
Pinned, e744, Thu May 24 05:44:56 2018, iulian, RUN, General, N/A, Important::analysis of the run 443/ things to do t_mm_PPT_run443.pdft_mmhf_run443.pdft_mm_exlor_run443.pdfcorrelation_run443.pdf
No stop on t5, but TAC MM_CATS ok.. So two possibilities:
1) Stop Cable not connected on Muvi2
2) Problem with the validation of CATS by FAG

HF validated by FAG was not working

Dead time scaler was not working

We need to erase all the scalers at the beginning of the run, including the part from Muvi (ORD, DESC, etc). Atm we erase only the U2M not the other part

The big problem is the transmission between LISE silicon detector D6 and T5 must with CATS on or off. 3000 counts/s on silicon d6 or CATS -> maximum 300 on T5. Is there a way to miss counting in Must2?


TAC_CATS_HF ok
TAC_MM_CATS ok
T_MM_EXLor present with prompt from cosmics?!
T_MM_HF ok
T_MM_HFv not present
Ongoing, e744, Thu May 24 05:59:44 2018, iulian, MUST2, General, N/A, run 444 t1-t4 trigger no target 14n beam mm5_8_run445.pngmm5_7_run444.pngmm5_6_run444.pngmm5_5_run444.pngmm5_4_run444.png
run 444
t1-t4 trigger
no target
14n beam