Ongoinge744 Tue May 29 01:33:58 2018 Phil, Valerian, FranckEXL-PPTGeneral Gamma-ray spectra from PPT
Attached are plots of gamma-ray energy spectra from PPT.

2-D plot is the PPT-CATS time (from PPT-MUST2 and MUST2-CATS time, obviously).

Two 1-D plots are PPT energy spectra gated on the two loci which can be seen in the 2-D plot.

Note that the 1.6-ish MeV line which may ordinarily expect to be due to the internal background from LaBr3 appears strongly in the prompt spectra and isn't really randomly time-correlated. We are suspicious, however, that the resolution is too good and that the peaks may be an artefact. We are checking this and will update this post accordingly.
Attachment 1: PPTSpectrum2D.png
PPTSpectrum2D.png
Attachment 2: PPTSpectrumPromptPeak1.png
PPTSpectrumPromptPeak1.png
Attachment 3: PPTSpectrumPromptPeak2.png
PPTSpectrumPromptPeak2.png
Ongoinge744 Tue May 29 05:07:33 2018 ValerianOFFLINEGeneralCH2Fusion evaporation background and physics data
Comparison between the fusion evaporation background and the data for the telescope 5 ~0degree angles and rough subtraction of this background.
Attachment 1: Ex_15F_and_Carbon.pdf
Ex_15F_and_Carbon.pdf
Attachment 2: rough_backgrund_suppresion.pdf
rough_backgrund_suppresion.pdf
Ongoinge744 Tue May 29 05:24:34 2018 ValerianEXL-PPTGeneralCH2Comparison of background and physics data for gamma ray detectors
Gamma ray spectrum for both targets seem extremely close. Subtraction is then made but no gamma show. More statistic is needed
Attachment 1: gamma_comparison_CH2_and_Carbon.pdf
gamma_comparison_CH2_and_Carbon.pdf
Attachment 2: background_substr_gamma_ray.pdf
background_substr_gamma_ray.pdf
Ongoinge744 Tue May 29 06:09:58 2018 Franck, PhilRUNGeneralCarbonRun 2106 - Target 1 Carbon
Run 2106

Beam 14O (CATS: 1.4e5 pps)

Start 6:08 29 May 2018
Stop 6:17 29 May 2018

Stopped because FAG rate low and ORDs low, "BAD MERGE" error messages

Target 1 Carbon
Attachment 1: Scaler_start_Run2106.png
Scaler_start_Run2106.png
Attachment 2: Scaler_stop_Run2106.png
Scaler_stop_Run2106.png
Ongoinge744 Tue May 29 07:39:53 2018 Franck, PhilRUNGeneralCarbonRun 2107 - Restart after Reboot of VXI - Same problem
Low FAG rate: 30 /s
Low ORD rates: 0-10 /s, 15 /s on MM5
Ongoinge744 Tue May 29 08:09:02 2018 e748RUNGeneralCarbonRUN 2108 - Target1 Carbon
Run 2019

Acq recovered

Beam: 14O
start: 8h04m40s
start: 8h15m07s

Target 1 Carbon

Stopped the run to enter and switch back to the CH2 target
Attachment 1: Scaler_start_Run2108.png
Scaler_start_Run2108.png
Attachment 2: ttac_Run2108.png
ttac_Run2108.png
Attachment 3: trigger_Run2108.png
trigger_Run2108.png
Attachment 4: mm8_Run2108.png
mm8_Run2108.png
Attachment 5: mm1_4_XE_Run2108.png
mm1_4_XE_Run2108.png
Attachment 6: mm1_4_YE_Run2109.png
mm1_4_YE_Run2109.png
Attachment 7: mm1_4_YE_Run2108.png
mm1_4_YE_Run2108.png
Attachment 8: mm1_4_T_Run2108.png
mm1_4_T_Run2108.png
Attachment 9: mm1_4_Run2108.png
mm1_4_Run2108.png
Attachment 10: impact_Run2108.png
impact_Run2108.png
Attachment 11: impact_Run2108.png
impact_Run2108.png
Ongoinge744 Tue May 29 08:59:18 2018 Freddy, Iulian, PierreRUNGeneralCH2Run 2109 - Target 4 CH2 - Tuning run
Run 2109

Beam: 14O (1.5*10^5 pps)

start: 8h37m01
stop: 9h26

Target nbr 4: CH2 (104 um)


Remark:
- Telesecope 4 is tripping all the time during this run. Impossible to put more than 50V. Even at 40 V, the leakage current is fluctuating around 8 uA. It seems like it is dying.
- LISE spectrometer adjustment during run - The attached plot show the final result of the beam tuning.
Attachment 1: impact_Run2109_end.png
impact_Run2109_end.png
Ongoinge744 Tue May 29 15:53:17 2018 Iulian MaurycyOFFLINEGeneralCH2 Angular distribution per telescope 1-4
Angular distribution per telescope 1-4
Attachment 1: ang_dist.pdf
ang_dist.pdf
Ongoinge744 Tue May 29 16:23:56 2018 Valerian Iulian MaurycyOFFLINEGeneralCH2Hint of a 2p peaks
Attachment 1: Two_proton_14Ovs12C.pdf
Two_proton_14Ovs12C.pdf
Ongoinge744 Tue May 29 18:49:21 2018 ValerianOFFLINEGeneralCH2Comparison of gamma prompt spectrum for EXL and PPT
Quick note, there is less data seen in EXL prompt. This should be investigated
Attachment 1: prompt_energyEXL_12C_14O.pdf
prompt_energyEXL_12C_14O.pdf
Attachment 2: prompt_energyPPT_12C_14O.pdf
prompt_energyPPT_12C_14O.pdf
Ongoinge744 Wed May 30 01:30:34 2018 ValerianOFFLINEGeneralCH2Checking the T4
After checking every telescope, it seems that the calibration of T4 is very off, I think that for now, it might be a good thing to consider this telescope out of the analysis ?
Ongoinge744 Wed May 30 03:09:09 2018 Valérie, LouisRUNGeneralCH2BEAM LINE PROBLEMS, alarm "water leakage"
The beam was stopped at 4h56, due to the alarm of "water leakage" on equipments, indicated in the D6 area.
All QPole in D6 were automatically put to 0.
Wien filter and DAP were put OFF to stop the beam and enter in the D6 area.
Between 5h and 5h30 PCP investigated the problem along the D3 and D6 areas, to check the water leakage problem indicated by the control command messages.
Since the problem could not be solved directly on the command boxes in PCP, it was not possible to put again the values on the QPoles.
then the area astreinte, Sébastien Le Moal, was called at 5h40.
At 6h15, he investigated the problem and check the cables in the D6 area. He saw some water condensation around the cable surveyor of the equipment (located behind the cooling system of MUST2).


Night/morning shift team: Louis Valérie.

[NB: in order to let them end up their night quietly, we will call later on the LISE experts for the beam line checking once the Qpoles are ready to put ON again.]

The T4 was increasing, from 16µA at 5h, then around 17 µA at 5h40 (see snapshot HT_MUST2_Wed30May_At5h52.png) we decided to put it OFF to let it end the night quietly.


Attachment 1: HT_MUST2_Wed30May_At5h52.png
HT_MUST2_Wed30May_At5h52.png
Ongoinge744 Wed May 30 09:24:28 2018 Valérie, Adrien, FreddyRUNGeneralCH2Run 2118
Wednesday 30th May.
Run 2118 started at 9h23 ;
Stop at: 11h25
Beam back at low rate and increase to 1.2*10^5 pps
target n°4. CH2 104 µm
All triggers, (no LISE)
Divisions CATS1 3. 10**4. PPT 3000 ; EXL 10**4
Morning team: Adrien, Freddy, François, Valérie ; LISE team: Jean-Charles.

cf current of T4 around 9.2 µA at 9h25. CHECK IT REGULARLY
Attachment 1: Run2118_scalersINIT.png
Run2118_scalersINIT.png
Attachment 2: Run2118_HT_EXLat9h29.png
Run2118_HT_EXLat9h29.png
Attachment 3: Run2118_scalersIat9h33.png
Run2118_scalersIat9h33.png
Attachment 4: Run2118_MMimpacts.png
Run2118_MMimpacts.png
Attachment 5: Run2118_CsIExlandETimePPT.png
Run2118_CsIExlandETimePPT.png
Ongoinge744 Wed May 30 11:17:00 2018 Valerian maurycy and OlivierOFFLINEGeneralCH2Condition on time T5
One should add the condition Si_TX < 150 for T5, problem from Ta?
Attachment 1: SiEvsSiT_T5_issue.png
SiEvsSiT_T5_issue.png
Ongoinge744 Wed May 30 11:22:16 2018 Valerian maurycy and OlivierOFFLINEGeneralCH2Checking the cuts
cut_DeE :

- T1 OK
- T2 OK
- T3 OK


cut ToF :

- T1 OK
- T2 OK
- T3 OK
Ongoinge744 Wed May 30 13:48:02 2018 PhilDAQGeneralN/APulser values
Here's a list of pulser values taken from the last few runs as a reference for what rate the "100-Hz" pulser is actually seeing.

All rates in Hz.

29 May:

6h52 62
9h36 96
9.39 96
10.35 50

30 May
12.15 76
12.16 84
3.01 48
3.21 96
3.27 48
4.57 48
7.15 48
8.39 48
9.26 48
9.33 66
11.35 48
1.29 48

Rate was 48 Hz when the run was stopped due to loss of beam (I looked but didn't take a picture).

I don't know if this problem is caused by the scaler counting behaving oddly but that seems like a more obvious solution than the pulser frequency randomly changing. If the pulser is actually vital for the experiment, someone should look in the offline data to see what's happening.
Ongoinge744 Thu May 31 16:42:30 2018 iulianEXL-PPTGeneral60CoRun 3000 60Co source PPT & EXL triggers
Run 3000
60Co source
PPT & EXL triggers

EXL 1 HT found tripped at the end of the run...

start 16h40 31/05
stop 17h35 31/05
Attachment 1: Scalers_start_run3001.png
Scalers_start_run3001.png
Attachment 2: Scalers_stop_run3000.png
Scalers_stop_run3000.png
Attachment 3: ppt_exl_run3000.png
ppt_exl_run3000.png
Attachment 4: ppt_exl2_run3000.png
ppt_exl2_run3000.png
Ongoinge744 Thu May 31 17:45:21 2018 iulianMUST2General3-alpharun #3001 alpha source T1-T5
run #3001
alpha source T1-T5
Attachment 1: scaler_startt_run3001.png
scaler_startt_run3001.png
Attachment 2: scaler_stop_run3002.png
scaler_stop_run3002.png
Ongoinge755 Sun Jun 17 11:12:57 2018 takeshiTARGETHardwareN/Acurrent cooling status of the cryogenic target
This is the temperature log of the cryogenic hydrogen target during the cooling test done last night.
The temperature of the heat shield (green line) and the cell (blue line) increased while it went the minimum values at once.

We could not try liquefying test this morning because the temperature was not enough low.
I would attach the boiling points of several gases in a vacuum. We suspect the rebound of the temperature is due to adsorption of vapour such as N2 at 20 K and H20 at 150 K.
Ongoinge755 Sat Jun 23 19:10:23 2018 NoriOFFLINEGeneralCryo-LH2 