Fixede744 Fri May 4 11:20:50 2018 ValerianMUST2General3-alphaT5 : 0° telescope
After increasing the voltage on T5 DSSD from 80V to 100V the resolution increased from ~40keV sigma to 20keV sigma.

StripX vs StripY bidim and the 3 alpha peaks are joined to this entry.
Attachment 1: StripX_vs_StripY_T5_100V.pdf
StripX_vs_StripY_T5_100V.pdf
Attachment 2: Si_E_T5_100V.pdf
Si_E_T5_100V.pdf
Fixede744 Fri May 4 23:58:07 2018 iulianRUNGeneral3-alphaRun #203 alfa source & 60co source
run #203

alpha source & 60co source
start 23h56
stop 11h07
Attachment 1: run_203_ht.png
run_203_ht.png
Attachment 2: run_203_ppt.png
run_203_ppt.png
Attachment 3: run_203_EXL.png
run_203_EXL.png
Attachment 4: run_203_must2.png
run_203_must2.png
Fixede744 Sun May 6 17:52:21 2018 ValerianRUNGeneral3-alphaRun #205 T1 T2 T3 & T4
run #205
T1 T2 T3 & T4 T5 still in trigger but no HT/BT
start @15h43
stop @17h52
Fixede744 Mon May 7 10:42:58 2018 ValerianOTHERGeneral60CoUpdate on the cooling system and run gamma (signal problem on PPT)
Today, both cooling system are working fine and their levels are OK.
The vacuum are 7e-6 mbar in M2C and 2e-5 in the 0 degree chamber.

I start a new gamma run :

run #207
trigger : EXL + PPT
start @10h44 07/05/18
stop @11h14 07/05/18
No signal on PPT. EXL OK.

I join to this entry a screenshot of EXL HT and EXL + PPT spectra

I start a new gamma run :

run #208
Bad run

run #209
trigger : EXL
start @11h21 07/05/18
stop @17h44 9/05/18
Attachment 1: HT_EXL.png
HT_EXL.png
Attachment 2: EXL_spectra_run207.png
EXL_spectra_run207.png
Attachment 3: PPT_spectra_run207.png
PPT_spectra_run207.png
Attachment 4: EXL_spectra_run209.png
EXL_spectra_run209.png
Fixede744 Thu May 10 14:38:41 2018 iulianEXL-PPTHardwareN/AMesytech settings
Tried unsuccessfully to change the gain on EXL energies.
Looking at the inspection lines noticed that all the EXL signals were inverted (negative) and we were coding only the pole zero part of the signals.
Looking into the manual find out how to change polarity of the signals. Did that and now the signals look normal with normal gain change after polarity change.
EXL polarity is negative so Mesytech modules have to be set to use negative input signals.

Fixede744 Thu May 10 15:06:44 2018 iulian & valerianEXL-PPTGeneral60CoStarted run #211 & #212 #213 60Co source efficiency run
Run #211 BAD Run no BT
60Co source efficiency run EXL & PPT trigger. Gains and thresholds set
start @15h05 10/05
stop @10h38 11/05

Run #212 BAD Run threshold changed
60Co source efficiency run EXL & PPT trigger. Gains and thresholds set
start @12h19 11/05
stop @13h14 11/05

Run #213
60Co source efficiency run EXL & PPT trigger. Gains and thresholds set
start @13h20 11/05
stop @8h20 12/05

QDCTDC_VOIE_0 -> PPT_E

This run used e744_12mai_founctione.das config (looks corrupted for must2 as no y energies)
Attachment 1: PPT_spectra_run213.png
PPT_spectra_run213.png
Attachment 2: EXL_spectra_run213.png
EXL_spectra_run213.png
Fixede744 Sat May 12 08:54:42 2018 iulianMUST2General3-alphaRun #114 T1-T5 trigger 3alpha source calib run
Run #214
Trigger T1-T5 & gamma
3 alpha run
start @8h15m55s 12/05
stop @20h01 12/05

This run use new das config copied from e744_save04mai_fonctionne.das, updated and then re-saved to e744_save12mai_fonctionne.das
The previous config , saved now under the name e744_save12mai.das, gives no Y energies in the spectras and an error on Muvi2 (some register different)
Attachment 1: Must_spectra_start_run214.png
Must_spectra_start_run214.png
Attachment 2: Must_spectra_stop_run214.png
Must_spectra_stop_run214.png
Fixede748 Mon May 14 10:49:47 2018 FrançoisDAQSoftware TDC problem
Problem :
TDC labels are 0. After the compilation of action list the labels seem to appear (if I remember correctly), but after a save/reload of the config they are gone, and in all cases no event coded (on the oscilloscope a validation point is visible, but do not know what this means in the case of a TDC – even after looking in GANIL wiki related doc).


Solution :
Solved by Fred. He made "init" at the VME MUST2MASTER in the RCC.
Fixede744 Mon May 14 13:52:25 2018 FrançoisBEAMGeneral Profileur



Normalement, les profileurs devraient voir le faisceau.
En général, la fenêtre d’entrée est de 12µm de titane et il y a une plaque d’alu de 3mm en sortie.
Donc il y a une plaque d’arrêt sur le dernier profileur.
Mais même si c’est une feuille d’inox en sortie, le faisceau ne devrait pas sortir.
(Antony Dubois)
Attachment 1: Sans_titre.jpg
Sans_titre.jpg
Fixede744 Tue May 15 19:42:40 2018 ValérianEXL-PPTGeneral60CoCalibration of EXL from e748 data
The EXL calibration program is ready. Join is the result with previous data from e748.
Attachment 1: EXL_Sum_e748_Calib.png
EXL_Sum_e748_Calib.png
Fixede744 Wed May 16 11:30:21 2018 iulianEXL-PPTGeneral60Corun 302 60Co source
Run #302

60 Co source PPT & EXL trigger

1start 21h46m40s 15/05
stop 10h 16/05
Fixede744 Wed May 16 16:27:29 2018 e744, lama, beyhanOTHERHardware TACS from LISE to the exp DAQ
The following TACs signals from LISE have been cabled and send to the experimental acquisition (->D6)

TED4_HF (cable physically labeled 29 of the ADC)
TE2D6_HF (cable physically labeled 30 of the ADC)

--> They should be added to the DAS when possible

NOTE that TE1D6_HF (cable physically labeled 31 of the ADC) was already cabled
Fixede744 Wed May 16 17:03:18 2018 RuxandraEXL-PPTSoftware calibration
Calibration coefficients for the PPT detector using the 60Co and 40K lines: E=0.5514*ch - 83.322. Internal radioactivity at 1444 keV, not separated from 40K
Efficiency: During the run (21h46min40s) the number of gammas emitted from the 60Co source is 66796800; for the Co pics the integrals are: 2.71e05 for the 1173.2 and 1.87e05 for the 1333, respectively (0.4% and 0.3%) without geometrical factor
Fixede744 Wed May 16 17:14:20 2018 e744, lama, beyhanOTHERHardware setting of signals from LISE to exp. DAQ
We put the following signals from LISE on DAQ :

(counting channels from 1)

E_D4 channel 17
E1_D6 channel 18
E2_D6 channel 19
T_D4 channel 20
T1_D6 channel 21
T2_D6 channel 22

tomorrow : set individual triggers of the ADC (OR of T_D4, T1_D6 and T2_D6)
(need to build a cable...)
Fixede744 Wed May 16 18:01:47 2018 Iulian ValerianMUST2SoftwareN/AFirst successful Calimero run on T1 voie X run 316
First successful Calimero run on T1 voie X run 316
Attachment 1: Calimerro_run316.png
Calimerro_run316.png
Fixede744 Wed May 16 19:06:39 2018 Iulian ValerianMUST2SoftwareN/Arun #318 Calimero T1-T4 strips X
Run #318

Calimero T1-T4 X
Attachment 1: Calimerro_run318.png
Calimerro_run318.png
Attachment 2: Calimerro_T4_X_strip2_run319.png
Calimerro_T4_X_strip2_run319.png
Attachment 3: Calimerro_T2_X_strip12_run319.png
Calimerro_T2_X_strip12_run319.png
Fixede744 Wed May 16 20:26:48 2018 Iulian ValerianMUST2SoftwareN/Arun #319 calimero T1-T4 Y strips
run #319
calimero T1-T4 Y strips
Attachment 1: Calimerro_run319.png
Calimerro_run319.png
Fixede744 Wed May 16 20:36:40 2018 Iulian ValerianMUST2SoftwareN/Arun #320 Calimero T5 X
run #320

Calimero T5 X
Attachment 1: Calimerro_run320.png
Calimerro_run320.png
Fixede744 Wed May 16 20:42:43 2018 Iulian ValerianMUST2SoftwareN/ARun #321 Calimero T5 Y strips
Run #321

Calimero T5 Y strips
Attachment 1: Calimerro_run321.png
Calimerro_run321.png
Fixede744 Fri May 18 10:47:37 2018 RuxandraEXL-PPTHardware60CoCo60 caliibration
Thu May 17 . run 339
60Co calibration
start @19h26 17/05/18
stop @7h39 18/05/18
EXL + PPT trigger