Ongoing, e744, Mon Mar 26 15:07:06 2018, e744 : Valerian et Louis, MUST2, Hardware, 3-alpha, Energy vs Strip for T6 et T7 StripXE_vs_StripNbr_T6.pdfStripXE_vs_StripNbr_T6_zoom.pdfStripYE_vs_StripNbr_T6.pdf
Weird signal on StripX of T6... might be the copper oxyde projections making a short-circuit

[U][B]EDIT 27/03/18 : The problem seen on Telescope 6 might be due to either some Asics Problem, or due to some dirt, the current reason in not known currently.
Pinned, e744, Mon Mar 26 11:44:27 2018, e744 : Valerian, OTHER, General, , Current Elog state 
Currently we do not have access to the experiment number. To write an entry for e744 the nickname TEST will be used.
Fixed, e744, Mon Mar 26 11:46:43 2018, e744 : Valerian, RUN, General, 3-alpha, Tests runs Ganacq log 
(15:51:33 - 19/03/2018) INFO - rcc_gui - Started run #1

(16:22:01 - 19/03/2018) INFO - rcc_gui - Stopped run #1
Fixed, e744, Mon Mar 26 14:20:22 2018, e744 : Valerian, MUST2, Hardware, 3-alpha, Bidimensional graphs for Telescope 7 and 6 T6_3alpha.pdfT7_3alpha.pdf
WARNING T5 == T7
[U][B]
EDIT 27/03/18 : WARNING X and Y are inverted, this entry should be ignored see entry : [URL=https://elog.lpc-caen.in2p3.fr/MUST2/333]Bidim for T1 T3 T4
Ongoing, TEST, Tue Mar 27 10:47:39 2018, e744 : Valerian, MUST2, Hardware, 3-alpha, Choice of DSSSDs 
 
Ongoing, e744, Tue Mar 27 11:11:04 2018, e744 : Valerian, MUST2, Hardware, 3-alpha, Choice of DSSSDs 14x
Different plots of Strip channel vs Strip number for X and Y in order to make our choice
Fixed, e744, Tue Mar 27 16:26:29 2018, e744 : Valerian, MUST2, General, 3-alpha, Bidim for T1 T3 T4 T5 T6 T7 T8 7x
 
Ongoing, e744, Tue Mar 27 17:22:24 2018, e744 : Valerian, MUST2, Hardware, 3-alpha, Choosing the DSSSDs 
We can see that the DSSSDs on T1(CsI) T3(CsI) T4(CsI) and T5(SiLi) are in good shape (cf :[URL=https://elog.lpc-caen.in2p3.fr/MUST2/333]Bidim for T1 T3
T4 T5 T6 T7 T8 [/URL]).
    Ongoing, e744, Wed Mar 28 13:21:32 2018, e744 : Valerian, MUST2, Hardware, 3-alpha, Choosing the DSSSDs 
[quote="e744 : Valerian"]We can see that the DSSSDs on T1(CsI) T3(CsI) T4(CsI) and T5(SiLi) are in good shape (cf :[URL=https://elog.lpc-caen.in2p3.fr/MUST2/333]Bidim
for T1 T3 T4 T5 T6 T7 T8 [/URL]).
    Ongoing, e744, Wed Mar 28 16:35:01 2018, e744 : Valerian, MUST2, Hardware, 3-alpha, Choosing the DSSSDs TimeMust2X_MUVI2_uncalib.pngTimeMust2Y_MUVI2_uncalib.pngSpectresMUVI2stripXT300ns.png
[quote="e744 : Valerian"][quote="e744 : Valerian"]We can see that the DSSSDs on T1(CsI) T3(CsI) T4(CsI) and T5(SiLi) are in good shape (cf :[URL=https://elog.lpc-caen.in2p3.fr/MUST2/333]Bidim
for T1 T3 T4 T5 T6 T7 T8 [/URL]).
Fixed, e744, Tue Apr 3 10:44:38 2018, e744 : Valérian, MUST2, Hardware, , Update on Must2 telescopes 20180403_103732.jpg
For the next day, the plan is to check the DSSD's connections of T6 & T7 followed by an alpha run, if the problem seen on T6 are not resolved, we will change
the DSSD of T6 and re-do another alpha run, the last option would be to change the Muffee card.
Ongoing, e744, Tue Apr 3 15:30:49 2018, e744 : Valérian, MUST2, Hardware, , Cabling of T6 & T7 Inversion_between_telescopes.pdf
After checking the cable of T6 & T7 it seems that there were in fact an inversion during experiment e748. T6 is the bottom one, and T7 the top one. 

During e748, it seems that there were some difference regarding the position, between the hardware and software position...
Ongoing, e744, Wed Apr 4 17:49:05 2018, e744 : Valérian, EXL-PPT, Hardware, , Update on EXL 6x
The EXL detector has been mounted, it is very close to the target ~4cm and doesn't do any shadow on the M2 telescopes. But it isn't at 90° anymore. The
cable have been plugged-in (cf photos). With a new mapping of the detector (cf drawing).
Fixed, e744, Tue May 22 22:33:09 2018, e744, MUST2, Software, N/A, Run #363 Calimero T1-T4 strips X again with threshold at 50% (same as 40%) T1T4_Xstrips_run363.png
[U][B]Run #363[/B][/U]
NB: Thresholds  are the ones on the pulser amplitude to inject the signal (Calimero Pulser interface on Geco). see snapshots on the Calimero summary.
Calimero with 50% thresholds on T4. Same spectra as with the thresholds @40%. I think T4 is triggering on cross-talk, so when the thresholds it's high
Fixed, e744, Tue May 22 22:50:47 2018, e744, MUST2, Software, N/A, Run #365 Calimero strips Y T1-T4 100% Threshold T1T4_Ystrips_run364.png
[U][B]Run #365[/B][/U]
Calimero strips Y T1-T4
100% threshold on T4
Fixed, e744, Wed May 23 00:30:34 2018, e744, EXL-PPT, General, N/A, RUN #402, EXL PPT trigger 
RUN #402
start exl ppt run for checking coincidence
exl ch 17 put in to adc2 ch 32
Fixed, e744, Wed May 23 18:20:59 2018, e744, MUST2, General, N/A, Test Calimero Run 425 
Started run #425 Calimero  NEW XT  T1toT4
ampl Pulser 12MEV; thr for T4 1 MeV
problem on data taking
Fixed, e755, Sat Jun 30 20:08:27 2018, didier, takeshi, daisuke, RUN, General, Cryo-LH2, RUN 199 37C + liq. H 28x
 
Fixed, e738, Fri Jul 13 22:00:59 2018, didier, nori, takeshi, shumpei, hide, daisuke, RUN, General, Cryo-LH2, RUN 376: 12C first run 22x
12C + liquid H. 
start 21h44
stop 22h43
Fixed, e738, Tue Jul 10 21:26:48 2018, didier, daisuke, RUN, General, Cryo-LH2, RUN 334 17x
9C + liquid H. Probably nobody cares this post right now, but the beam is back. 

21h00 the beam back to LISE. 

ELOG Home Page