Fixed, e744, Tue Apr 24 16:22:08 2018, e744 Valérie L., RUN, Software, , Narval Merger Parameters
|
|
Fixed, e744, Wed Apr 25 20:47:59 2018, e744 Valérie, MUST2, General, , MUST2 mate Temperature on the DAS interface 
|
tests
The plot:ConfigDASMuvi1Tel4mufeeX.png shows the situation on the Mufee card when the cooling system was blocked (ice in the tube! there was water inside
instead of alcohol) on Wednesday 25th. cf there was 47°C on MufeeX mate of Telescope 4.
|
Fixed, e744, Mon Apr 23 11:33:53 2018, e744 Valerian, OTHER, Hardware, N/A, Inspection signals
|
Inspection signals between D6 and LISE Acquisition room are connected as follow :
178: FAG
|
Fixed, e744, Thu May 24 22:38:44 2018, e744 Must2+lise teams , RUN, General, CH2, Run Beam Spiral 14N centered 7x
|
Shift teams: François, Iulan, Freddy, Fairouz, Valérian, Valérie, Mihail
LISE teams during evening: Omar Kamalou, Luc Perrot, Pierre Delahaye
Triggers Must2 Telescopes T1 to T5
|
Fixed, e744, Thu May 24 22:45:36 2018, e744 Must2+lise teams , RUN, General, CH2, Run 1001 Beam Spiral 14N centered 
|
Run 1001, cible n°6: 183 um
start @ 21h55; Stop @ 22h36
Shift teams: François, Iulan, Freddy, Fairouz, Valérian, Valérie, Mihail
|
Fixed, e744, Fri May 25 22:46:26 2018, e744 Must2+LISE teams, edit Valérie, RUN, General, CH2, RUN 2004 6x
|
Run 2004 Date : 25/05/2018 22:45:58 25/05/2018 23:17:39
End of Run, error Narval see snapshot ProblemNarvalRCCEndRun2004.png
|
Fixed, e744, Fri May 25 23:26:23 2018, e744 Must2+LISE teams, edit Valérie, RUN, General, CH2, START NARVAL, new version
|
the NARVAL version is now reinstalled by Fred Saillant.
see plot ResolProblemNarvalEndofRun25may23h.png
|
Fixed, e744, Fri May 25 22:03:12 2018, e744 Must2+LISE teams, RUN, General, CH2, RUNS 2000, 2001 6x
|
Beam 14O
Run 2000 start with all triggers, including LISE trigger, during 1 minute for tests.
start 22:00:58; stop
|
Fixed, e744, Thu May 24 19:20:40 2018, e744 Must2 + Lise teams (edit: Valérie), RUN, General, N/A, Run 460-Spiral 14N test  
|
Run 460
Triggers Must2 Telescopes T1 to T5
( MM1 to MM5)
|
Fixed, e744, Thu May 24 19:32:31 2018, e744 Must2 + Lise teams (edit: Valérie), RUN, General, N/A, Run 461-Spiral 14N test 
|
Run 461
Triggers Must2 Telescopes T1 to T5
(MM1 to MM5)
|
Fixed, e744, Thu May 24 20:31:30 2018, e744 Must2 + Lise teams (edit: Valérie), RUN, General, N/A, Run 463 Beam Spiral 14N test  
|
Run 463 focalisation on Must2 24May at 20h30pm storage
Triggers Must2 Telescopes T1 to T5
(MM1 to MM5)
|
Fixed, e744, Thu May 24 21:05:19 2018, e744 Must2 + Lise teams (edit: Valérie), MUST2, Hardware, N/A, change of STOP MUST2
|
The STOP of MUST2 with CATS is now directly the CATS "delayed signal".
No coincidence with FAG.
NB: the CATS "delayed signal"is the fast time signal from CATS in D4, which is sent to the DAQ area of D6, then delayed.
|
Fixed, e744, Thu May 24 21:28:45 2018, e744 Must2 + Lise teams (edit: Valérie), RUN, General, CH2, Run 464 Beam Spiral 14N test    
|
Run 464 focalisation on Must2 24May at 21h25m storage
Triggers Must2 Telescopes T1 to T5
(MM1 to MM5)
|
Fixed, e744, Thu May 24 19:34:40 2018, e744 Must2 + Lise teams, RUN, General, N/A, tuning Wien-Spiral 14N test
|
Run 24May at 19h50-20pm no storage
Triggers Must2 Telescopes T1 to T5
(MM1 to MM5)
|
Fixed, e744, Thu May 24 20:18:00 2018, e744 Must2 + Lise teams, RUN, General, N/A, tuning Wien-Spiral 14N test   
|
Run 24May at 19h50-20pm no storage
Triggers Must2 Telescopes T1 to T5
(MM1 to MM5)
|
Fixed, e744, Thu May 24 20:19:19 2018, e744 Must2 + Lise teams, RUN, General, N/A, Beam centering on MM5-Spiral 14N test  
|
Run 24May at 19h50-20pm no storage
Triggers Must2 Telescopes T1 to T5
(MM1 to MM5)
|
Fixed, e744, Thu May 24 20:21:57 2018, e744 Must2 + Lise teams, RUN, General, N/A, Run 462 Beam centering on MM5-Spiral 14N test   
|
Run 462 focalisation on target 24May at 20h15pm storage
Triggers Must2 Telescopes T1 to T5
(MM1 to MM5)
|
Pinned, e744, Thu May 24 23:20:40 2018, e744 Must2 + Lise teams, RUN, General, CH2, run 1002
|
[I][B]run 1002:[/B][/I] 14N beam, CH2 187 um
trigger: MM1,2,3,4,5 + CATS%, EXL%, PPT%
start: 22h41
|
Fixed, e744, Fri May 25 21:34:19 2018, e744 Iulan, Fairouz, Valerie, DAQ, Hardware, CH2, Problem of FAG validation +change of ADC parameter for TAC signal
|
The Time signal MM_PPT is now in the parameter MCP_HF of the ADC.
The problem was on the FAG signal validation (bad output of the discriminator)
|
Fixed, e744, Fri May 25 17:51:45 2018, e744 Iulan Valérie, RUN, General, CH2, Run 1008 starting runs with 14N  
|
[U][B]BAD RUN FOR 14N[/B][/U]
Run 1008 BAD RUN 0 Events
|
|