Fixed | e748 |
Mon Jul 10 00:23:41 2017 |
e748 | RUN | General | CD2 | Return of the beam (Jedi not available) |
Beam is back.
JCT refoc the beam on target (attached pictures).
All god beside that. Still no phone on campus! |
Attachment 1: CATS.png
|
|
Fixed | e748 |
Mon Jul 10 00:35:29 2017 |
e748 | RUN | General | CD2 | Run 380 |
Started at 0:02
18O 4.3 uAe, fluctuating
SiLi 1/T5: 370 V, 25.7 uA
Scalers attached, see below. |
Attachment 1: scalers_run380.png
|
|
Fixed | e748 |
Mon Jul 10 01:34:12 2017 |
e748 | RUN | General | CD2 | Run 380 (file change) |
Data file change at 1:18 -> run_0380.dat.10-07-17_00h02m00s.1
18O 4.5 uAe
220 events/s
SiLi 1/T5: 370 V, 25.8 uA
Scalers attached, see below |
Attachment 1: scalers_run380_1.png
|
|
Fixed | e748 |
Mon Jul 10 02:04:33 2017 |
e748 | CATS | General | CD2 | CATS raw strips and trajectories |
|
Attachment 1: CATS_raw_run380_1.png
|
|
Attachment 2: CATS_run380_1.png
|
|
Fixed | e748 |
Mon Jul 10 02:47:27 2017 |
e748 | RUN | General | CD2 | Run 380 (file change) |
2:26 data file change -> run_0380.dat.10-07-17_00h02m00s.2
18O 4.5 uAe
240 events/s
SiLi 1/T5: 370 V, 25.8 uA |
Attachment 1: Scalers_run_380_2.png
|
|
Fixed | e748 |
Mon Jul 10 03:34:24 2017 |
e748 | RUN | General | CD2 | Run 380 (file change) |
3:33 data file change -> run_0380.dat.10-07-17_00h02m00s.3
18O 4.5 uAe
240 events/s
SiLi 1/T5: 370 V, 25.9 uA |
Attachment 1: scalers_run380_3.png
|
|
Fixed | e748 |
Mon Jul 10 03:45:19 2017 |
e748 | MUST2 | General | CD2 | SiLi 1/T5 triped |
Triped between 3:34 and 3:44.
We turn it back on at 3:46 (370 V) -> 22.5 uA |
Fixed | e748 |
Mon Jul 10 04:33:32 2017 |
e748 | CATS | General | CD2 | CATS: raw strips, trajectories and pos on target |
|
Attachment 1: CATS_raw_run_380_3_4h35.png
|
|
Attachment 2: CATS_run380_3_4h35.png
|
|
Attachment 3: target_pos_run380_3.png
|
|
Fixed | e748 |
Mon Jul 10 04:46:11 2017 |
e748 | RUN | General | CD2 | Run 380 (file change) |
4:42 data file change -> run_0380.dat.10-07-17_00h02m00s.4
18O 4.4 uAe
240 events/s
SiLi 1/T5: 370 V, 25.7 uA
MUST2 temperatures: T1-T2 -1.8°, T3-T4 -0.3°, T5 0.4°, T6 1.6° |
Attachment 1: scalers_run380_4.png
|
|
Fixed | e748 |
Mon Jul 10 05:47:40 2017 |
e748 | RUN | General | CD2 | Run 380 (file change) |
5:46 data file change -> run_0380.dat.10-07-17_00h02m00s.5
18O 4.5 uAe
260 events/s
SiLi 1/T5 : 25.9 uA |
Attachment 1: scalers_run380_5.png
|
|
Fixed | e748 |
Mon Jul 10 06:53:40 2017 |
e748 | MUST2 | General | CD2 | SiLi 1/T5 triped |
Triped at ~ 6:50.
We turn it back on at 370 V -> I = 18 uA but increases rapidly to 21 uA |
Fixed | e748 |
Mon Jul 10 06:55:58 2017 |
e748 | RUN | General | CD2 | Run 380 (file change) |
6:48 data file change -> run_0380.dat.10-07-17_00h02m00s.6
18O 4.5 uAe
260 events/s |
Attachment 1: scalers_run380_6.png
|
|
Fixed | e748 |
Mon Jul 10 08:19:45 2017 |
e748 | RUN | General | CD2 | End run 380 |
We stop run 380 to enter D6 and select mask target for CATS reference data. |
Fixed | e748 |
Mon Jul 10 08:37:23 2017 |
e748 | RUN | General | N/A | Run 381 mask target |
Trigger = Plastic, div=1
Primary intensity = 1 uAe
CATS rate = 35 kcps
800 events/s
Run 381 started at 8:42
Changing beam spot position during run to scan the mask:
-DAC2 : passe de -3A a -3.5A
-DAC2 : passe de -3.5A a -2.5A
-DAC2 : passe de -2.5A a -2.0A
-DAC2 : passe de -2.0A a -1.5A |
Fixed | e748 |
Mon Jul 10 10:06:27 2017 |
e748 | RUN | General | N/A | end of experiment |
10h06 we stop the experiment. |
Fixed | e748 |
Mon Jul 10 17:43:37 2017 |
Adrien | MUST2 | General | N/A | Calimero time calibrator |
Run 1000 : MUVI 1 X
Time calibrator 20ns period, 512ns range |
Fixed | e748 |
Thu Jul 20 11:07:54 2017 |
e748 | CATS | General | N/A | Run 1035 calib cats1 |
After experiment CATS 1 pulser calibration
Pulser Quality: very bad - to be analyzed with care
Calibration almost unchanged |
Fixed | e748 |
Thu Jul 20 11:09:34 2017 |
e748 | CATS | General | N/A | Run 1036 calib cats 2 Y |
After experiment pulser calibration CATS 2 (Y)
1st amplitude slightly saturates CATS 2 X.
Use for CATS 2 Y recalibration
Strip Y 14 indeed has a smaller gain than before the experiment |
Pinned | e744 |
Thu Mar 15 10:50:00 2018 |
Adrien Matta | MUST2 | Hardware | | Distribution box LPC |
Cabling file by Fabien Naury (LPC). |
Attachment 1: cablage_panier_distribution_123_der.xlsx
|
Attachment 2: cablage_panier_distribution_456_der.xlsx
|
Attachment 3: prise_fischer_BT1_der.xlsx
|
Attachment 4: prise_fischer_BT3_der.xlsx
|
Attachment 5: prise_fischer_BT5_der.xlsx
|
Attachment 6: prise_fischer_HT1_der.xlsx
|
Attachment 7: prise_fischer_HT3_der.xlsx
|
Attachment 8: prise_fischer_HT5_der.xlsx
|
Pinned | e744 |
Mon Mar 26 11:44:27 2018 |
e744 : Valerian | OTHER | General | | Current Elog state |
Currently we do not have access to the experiment number. To write an entry for e744 the nickname TEST will be used. |