MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 99 of 132  Not logged in MUGAST
Status ExpNbr Date Author Type Category Target-Source Subject
Fixede768s Sat Aug 3 00:43:37 2019 OttoTimeRUNGeneralN/ATimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 161
- MUVI: MUVI2
- Side : MUFEE X (Junction)
- Telescope: 4
Fixede768s Sat Aug 3 01:48:22 2019 OttoTimeRUNGeneralN/ATimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 161
- MUVI: MUVI2
- Side : MUFEE Y (Ohmic)
- Telescope: 4
Fixede768s Sat Aug 3 02:53:09 2019 OttoTimeRUNGeneralN/ATimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 161
- MUVI: MUVI3
- Side : MUFEE X (Junction)
- Telescope: 1,
Fixede768s Sat Aug 3 03:56:54 2019 OttoTimeRUNGeneralN/ATimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 161
- MUVI: MUVI3
- Side : MUFEE Y (Ohmic)
- Telescope: 1,
Fixede768s Sat Aug 3 05:00:32 2019 OttoTimeRUNGeneralN/ATimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 161
- MUVI: MUVI3
- Side : MUFEE X (Junction)
- Telescope: 2,
Fixede768s Sat Aug 3 06:03:51 2019 OttoTimeRUNGeneralN/ATimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 161
- MUVI: MUVI3
- Side : MUFEE Y (Ohmic)
- Telescope: 2,
Fixede768s Sat Aug 3 08:13:33 2019 OttoTimeRUNGeneralN/ATimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 161
- MUVI: MUVI3
- Side : MUFEE Y (Ohmic)
- Telescope: 3,
Fixede768s Sat Aug 3 09:17:12 2019 OttoTimeRUNGeneralN/ATimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 161
- MUVI: MUVI3
- Side : MUFEE X (Junction)
- Telescope: 4
Fixede768s Sat Aug 3 10:21:04 2019 OttoTimeRUNGeneralN/ATimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 161
- MUVI: MUVI3
- Side : MUFEE Y (Ohmic)
- Telescope: 4
FixedTEST Thu Oct 17 12:01:09 2019 Franco & MarlèneGRITHardware3-alphaTest of configuration with 4 MUVI and 3 GAMERS
The electronics set-up for MUGAST2020 has been tested with 4 MUVI on 4 VXI crates, 1GMT for MASTER and 3 GAMERS for slaves.

Configuration file is named mugast.das
We have modified the mugast.detector and the parameters for MUVI4 with names:
- MUVI4_ch1 : MG2
- MUVI4_ch2 : MG6
- MUVI4_ch3 : MG8
- MUVI4_ch4 : MG12 (not used and not tested)
For the 3 first channels of MUVI4 we connected MG1, MG3 and MG4.

Cables for MUVI4_ch2 is not working (to be repaired).

With the alpha source (double side) we were able to run during all night and it was stable (attached are pictures of alpha source).
We notice that the rate of MG7 is lower (1/3) than the other ones (as usual). We tried to increase bias but leakage current is also increasing proportionally.
We also notice that 3alpha source shows lower statistics in the last (higher energy) peak but all detectors backward see the same.
The rates of MG2 and MG8 are consistent (there was no pedestal suppression).

We performed a pulser test of MUVI4 putting 5000Hz rate. The dead time is about 97%. It is consistent with MUVI1 in the same condition (5000 Hz and no pedestal suppression). The dead time is 90% when we suppress the pedestals.
Attachment 1: Mugast_alpha_17102019
Attachment 2: Mugast_alpha_17102020_2
FixedTEST Tue Dec 3 18:14:45 2019 Marlène & FrancoGRITHardware Pulser test of new board
The new board that will be installed for MG2 has been tested and validated with a pulser outside of the chamber.
FixedTEST Wed Dec 4 19:19:38 2019 MarlèneGRITHardware Modifications of set-up
We have installed the new MUFEE board in the reaction chamber for MG2. This board is a new version with ATHED.

We have moved the MG9 board to become the board to read the future MUST2 detector at 90 deg. This board is now perpendicular to the other ones. We have tested the kaptons connection to the detector and the placement of the detector.
The design will be refined on monday with Patrice.

The board that was reading MG10 is now used for MG6.

Every board has been tested again with pulser under vacuum. Everything seems OK.
PendingTEST Tue Jan 21 17:55:12 2020 MarlèneMUST2Hardware Ovv error on the -5V of CAEN boards
The Ovv error shows on the -5V (first one) of the 3 board (5,3.3,10V for MUST2§squares)
It means that the voltage is over the asked value.
If we ask 5V, we get 5.5V in reality

We changed for 4.5V and we get 5V. The current is different from the one on the MUGAST BT box but we don't know how many detectors are connected.
We exchanged cables and found the same current in the other board, so we guess that it is still working properly.
A check of stability should be done.
Attachment 1: BT_referenve_5V_cables_inv_a_comp_referenve_5V_partMUST2
Attachment 2: BT_referenve_5V_partMUST2_cables_inv_a_comp_referenve_5V
Attachment 3: BT_referenve_5V_partMUST2
Attachment 4: BT_referenve_5V
FixedTEST Wed Jan 22 15:53:25 2020 MarlèneGRITHardware Changes for MUGAST2020
Changes were made for MUGAST 2020 :
1- the board for MG10 before (2nd square) is now for MG6 new trap
2- the board for MG9 is put vertical and used for the MUST2 at 90 deg
3- a new board has been installed for MG2

Now there will be 13 detectors in the chamber: 7 trap + 1 square + 4 MUST2 forward.
Fixed  Wed Jan 22 18:39:54 2020 Marlène & FrancoGRITHardware Tests of the new box
The new box has been tested together with the new CAEN SY4527 power supply. We connected a laptop with Linux to the power supply by means of an ethernet cable and communicated with it by opening a browser (http://192.168.0.1, then username: admin and pwd: admin). To control the channels we open a terminal, go to the folder where we saved the file ChannelsController.jar and run the command: java -jar ChannelsController.jar 192.168.0.1 admin admin
The output biases were checked with a test cable connected to the flange output and are reported in the attached file. For comparison we did the same thing on 3 Mugast detectors (MG1, MG3 and MG4) We saw that the polarity of all the 3.3V in the new box was inverted with respect to the Mugast box so we opened the box and, like sappers, we followed the colored cables until we found the ones giving the 3.3V and inverted them. It worked! So now the biases should be correct. A reminder for possible future modifications: the coupled colours for the wires are blue-grey, white-brown, yellow-green.
Some low voltages seem to be unused because they are present in some BT outputs but not in others, but anyway the detectors which are supplied by these cables have always been working properly, so the small differences between different BT outputs might be irrelevant.
It is then important to notice that BT2 and BT3 have a -2.5V instead of a +2.5V so it's recommended not to use them unless the relative cables are first switched inside the box.
Attachment 1: box_pin_map.pdf
box_pin_map.pdf box_pin_map.pdf box_pin_map.pdf box_pin_map.pdf box_pin_map.pdf box_pin_map.pdf box_pin_map.pdf box_pin_map.pdf
FixedTEST Thu Jan 23 14:06:51 2020 Franco&MarlèneGRITHardware3-alphaStatus of detectors on 23-01-20
Short alpha source run (not saved on disk) to test the detectors.

Summary :
- MG1,3,4,5 and 7 are all OK (few more strips missing on Y)
- MM1,2,4 OK
- MM3 has a lot of strip missing both in X and Y and current has increased wince last campaign (4uA)--> katpon should be checked
Attachment 1: mg_impact_alpha_23-01-2020.png
mg_impact_alpha_23-01-2020.png
Attachment 2: mg_Y_raw_alpha_23-01-2021.png
mg_Y_raw_alpha_23-01-2021.png
Attachment 3: mg_X_alpha_23-01-2021.png
mg_X_alpha_23-01-2021.png
Attachment 4: mm_impact_alpha_23-01-2020.png
mm_impact_alpha_23-01-2020.png
Attachment 5: mm_Y_alpha_23-01-2020.png
mm_Y_alpha_23-01-2020.png
Attachment 6: mm_X_alpha_23-01-2020.png
mm_X_alpha_23-01-2020.png
FixedTEST Mon Feb 10 18:42:23 2020 Franco&MarlèneGRITHardware New MUGAST detectors MG2 MG6
MG2 and MG6 have been mounted in the chamber.

MG2 reference number is 3227_10 thickness=468um
MG6 3227_7 thickness=473um

Nominal depletion =35V
FixedTEST Wed Feb 12 09:32:17 2020 mugastRUNGeneral3-alphaRun 500 - 3alpha - mugast standalone
RUN 500

start: 11/02/2020 @ 19h00
stop: 12/02/2020 @ 10h00

alpha source test - cooling 14°C

Remarks:
- The two new detectors (MG2 and 6) seem functional;
- MG11 annular, ok for sectors (Y) but many rings have double peaks;
- MM1 current was oscillating yesterday evening with values around 3 uA, this morning it was stable at ~0.55 uA.
Attachment 1: MG_all_X_alpha_source_run500
Attachment 2: MG_all_Y_alpha_source_run500
FixedTEST Wed Feb 12 16:24:52 2020 Marlène, Freddy & FrancoMUST2Hardware Modifications on the electronic scheme
We cleaned the electronic scheme.
Now the HF CLEAN VAMOS is taken directly from VAMOS and not re-cleaned by us.
TACs are all ok. Inspection lines have been checked and labeled.
GMT and Scalers are like below.

GMT:
1 - MUVI 1 (MG1, MG3, MG4, MG5)
2 - MUVI 2 (MG7, MG2, MG6, MG11)
3 - MUVI 3 (MM1, MM2, MM3, MM4)
4 - MUVI 4 (MG9 90-deg MUST2)
5 - COINCIDENCE VAMOS-AGATA
6 - VAMOS DIV
7 - MUST2 DIV
8 - broken
9 - SQUARE DIV
10 - AGATA DIV
11 - VAMOS
12 - empty
13 - empty
14 - empty
15 - broken
16 - empty

SCALERS:
1 - FAG
2 - OR MUGAST
3 - AGATA
4 - VAMOS
5 - empty
6 - DEAD TIME MUGAST
7 - 100 Hz PULSER
8 - DEAD TIME
9 - MUVI 1 (MG1, MG3, MG4, MG5)
10 - MUVI 2 (MG7, MG2, MG6, MG11)
11 - MUVI 3 (MM1, MM2, MM3, MM4)
12 - MUVI 4 (MG9 90-deg MUST2)
13 - empty
14 - empty
15 - CATS2
16 - FINGER
Fixed  Fri Feb 14 10:26:46 2020 Franco&MarlèneGRITHardware HT BT screenshots before e775s
Detectors are cooled down to 0 deg.
Attachment 1: BT_CAEN2_MGnew.png
BT_CAEN2_MGnew.png
Attachment 2: BT_CAEN1_part1_MM.png
BT_CAEN1_part1_MM.png
Attachment 3: BT_CAEN1_part2_MG.png
BT_CAEN1_part2_MG.png
Attachment 4: HV_part1_MM_MG.png
HV_part1_MM_MG.png
Attachment 5: HV_part2_MG.png
HV_part2_MG.png