Fixed, com2019, Wed Apr 3 10:52:59 2019, OttoRun, RUN, General, N/A, Run111 
*************************
  [SIZE=3][B]Run  : 111[/B][/SIZE]
*************************
Fixed, com2019, Wed Apr 3 11:03:21 2019, OttoRun, RUN, General, N/A, Run112 
*************************
  [SIZE=3][B]Run  : 112[/B][/SIZE]
*************************
Fixed, com2019, Wed Apr 3 11:03:53 2019, Adrien Matta, CATS, Hardware, N/A, Cats mask will be shifted 
The propulsif supporting the cats mask is leaking.

It will be tighten to fix the leak, this lead to a small shift toward the left of the beam when looking at vamos.
Fixed, com2019, Wed Apr 3 11:05:36 2019, Adrien Matta, DAQ, Software, N/A, MFM scaler reading 
MfM scaler are read correctly for Must2 U2M but output some -1 for the numexo2 integrated scaler.
Fixed, com2019, Wed Apr 3 11:14:15 2019, OttoRun, RUN, General, N/A, Run113 
*************************
  [SIZE=3][B]Run  : 113[/B][/SIZE]
*************************
Fixed, com2019, Wed Apr 3 11:18:32 2019, OttoRun, RUN, General, N/A, Run114 
*************************
  [SIZE=3][B]Run  : 114[/B][/SIZE]
*************************
Fixed, com2019, Wed Apr 3 11:26:34 2019, OttoRun, RUN, General, N/A, Run115 
*************************
  [SIZE=3][B]Run  : 115[/B][/SIZE]
*************************
Fixed, com2019, Wed Apr 3 14:18:25 2019, OttoRun, RUN, General, N/A, Run120 
*************************
  [SIZE=3][B]Run  : 120[/B][/SIZE]
*************************
Fixed, TEST, Wed Apr 3 15:49:02 2019, Valerian and Jenn, Chloé and François , OTHER, Hardware, N/A, Electronic scheme of VAMOS and MUGAST VAMOS_electronic_scheme.pdfVAMOS_electronic_scheme.pptxMUGAST_Electronic_scheme_v2.pptxMUGAST_Electronic_scheme_v3.pdfMUGAST_Electronic_scheme_v3.pptx
Edit : 4/04/19 Valerian and Jenn small correction of MUGAST scheme
Fixed, com2019, Wed Apr 3 15:49:39 2019, mugast, GRIT, Hardware, N/A, Time calibration files for Mugast and MUST2 
Time calibration for Mugast and MUST2 has been done using Calimero run 173 to 178. Path for files is 

/home/mugast/analysis/nptool/Projects/calibration/Time/Coeff/
Fixed, com2019, Wed Apr 3 16:24:37 2019, Adrien, CATS, Software, N/A, Calib file from T.Roger CATS2_Y.vamos2019CATS2_X.vamos2019CATS1_Y.vamos2019CATS1_X.vamos2019
Here are the calibration file provided by T. Roger.
Fixed, TEST, Wed Apr 3 19:15:42 2019, OttoH.V., OTHER, Hardware, N/A, HVM_SUMMARY 
[SIZE=5][COLOR=green][B]*** hvm watcher summary ***[/B] [/COLOR][/SIZE] 
 [SIZE=2][COLOR=green]+10v/mg     [/COLOR] Slot: [COLOR=blue]5  [/COLOR] (A1516B) Channel:[COLOR=blue]5 [/COLOR]  Vmon: [COLOR=blue] 9.99 V  [/COLOR] Imon:
[COLOR=blue]0.407 A [/COLOR] Status: [COLOR=green]On[/COLOR]
Fixed, TEST, Wed Apr 3 19:15:43 2019, OttoH.V., OTHER, Hardware, N/A, HVM_SUMMARY 
[SIZE=5][COLOR=green][B]*** hvm watcher summary ***[/B] [/COLOR][/SIZE] 
 [SIZE=2][COLOR=green]            [/COLOR] Slot: [COLOR=blue]8  [/COLOR] (A1821H) Channel:[COLOR=blue]1 [/COLOR]  Vmon: [COLOR=blue]    0 V  [/COLOR] Imon:
[COLOR=blue]    0 uA[/COLOR] Status: [COLOR=pink]Off[/COLOR]
Fixed, com2019, Wed Apr 3 21:02:18 2019, iulian ruxandra fairouz , ONLINE, General, N/A, Test correlation between different VXI acq test_corel_Pulesr_all.pngtest_corel.pngtest_corel_2000.pngtest_corel_Pulesr_Mugast_CATS.png
Test performed using a 100 Hz pulser signal send into the master trigger (entry 9 on GMTMASTER) and the analogical part send into a amplifier then split
in 3 signal send to VXI MUGAST , VXI CATS and VXI Vamos.
No problem observed. See first image below (test_corel_Pulesr_all.png)
    Fixed, com2019, Wed Apr 3 21:43:22 2019, Adrien, ONLINE, General, N/A, Test correlation between different VXI acq 
[quote="iulian ruxandra fairouz "]Test performed using a 100 Hz pulser signal send into the master trigger (entry 9 on GMTMASTER) and the analogical part
send into a amplifier then split in 3 signal send to VXI MUGAST , VXI CATS and VXI Vamos.
No problem observed. See first image below (test_corel_Pulesr_all.png)
Fixed, TEST, Thu Apr 4 03:58:50 2019, OttoH.V., OTHER, Hardware, N/A, HVM_SUMMARY 
[SIZE=5][COLOR=green][B]*** hvm watcher summary ***[/B] [/COLOR][/SIZE] 
 [SIZE=2][COLOR=green]+10v/mg     [/COLOR] Slot: [COLOR=blue]5  [/COLOR] (A1516B) Channel:[COLOR=blue]5 [/COLOR]  Vmon: [COLOR=blue] 9.99 V  [/COLOR] Imon:
[COLOR=blue]0.407 A [/COLOR] Status: [COLOR=green]On[/COLOR]
Fixed, TEST, Thu Apr 4 03:58:51 2019, OttoH.V., OTHER, Hardware, N/A, HVM_SUMMARY 
[SIZE=5][COLOR=green][B]*** hvm watcher summary ***[/B] [/COLOR][/SIZE] 
 [SIZE=2][COLOR=green]            [/COLOR] Slot: [COLOR=blue]8  [/COLOR] (A1821H) Channel:[COLOR=blue]1 [/COLOR]  Vmon: [COLOR=blue]    0 V  [/COLOR] Imon:
[COLOR=blue]    0 uA[/COLOR] Status: [COLOR=pink]Off[/COLOR]
Fixed, TEST, Thu Apr 4 05:10:48 2019, OttoH.V., OTHER, Hardware, N/A, HVM_ERROR 
[SIZE=5][COLOR=red][B]*** hvm watcher error ***[/B] [/COLOR][/SIZE] 
 [SIZE=4][COLOR=green]FP_CATS0    [/COLOR] Slot: [COLOR=blue]2  [/COLOR] (A1821HMP) Channel:[COLOR=blue]6 [/COLOR]  Vmon: [COLOR=blue]  530 V  [/COLOR]
Imon: [COLOR=blue]0.003 uA[/COLOR] Status: [COLOR=red]INTERNAL TRIP[/COLOR]
Fixed, com2019, Thu Apr 4 08:31:58 2019, mugast, RUN, General, N/A, Run 129 
Run 129 overnight vamos calibration.
Multiple trip of FP_CATSXXX between and 5.09 and 6.35
Fixed, com2019, Thu Apr 4 09:02:26 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
  - Run: 131
  - MUVI: MUVI1
Fixed, com2019, Thu Apr 4 09:25:13 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
  - Run: 133
  - MUVI: MUVI2

ELOG Home Page