MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST  Not logged in MUGAST
Entry  Wed Apr 3 21:02:18 2019, iulian ruxandra fairouz , com2019, ONLINE, General, Fixed, N/A, Test correlation between different VXI acq test_corel_Pulesr_all.pngtest_corel.pngtest_corel_2000.pngtest_corel_Pulesr_Mugast_CATS.png
    Reply  Wed Apr 3 21:43:22 2019, Adrien, com2019, ONLINE, General, Fixed, N/A, Test correlation between different VXI acq 
Message ID: 197     Entry time: Wed Apr 3 21:02:18 2019     Reply to this: 198
Author: iulian ruxandra fairouz  
ExpNbr: com2019 
Type: ONLINE 
Category: General 
Status: Fixed 
Target-Source: N/A 
Subject: Test correlation between different VXI acq 
Test performed using a 100 Hz pulser signal send into the master trigger (entry 9 on GMTMASTER) and the analogical part send into a amplifier then split in 3 signal send to VXI MUGAST , VXI CATS and VXI Vamos.
No problem observed. See first image below (test_corel_Pulesr_all.png)


Didn't work properly with a tac signal divided in 3 or from 3 tacs with same start /stop. See test_corel.png and test_corel_2000.png
This might be due to the time calibrator trigger as it is very random and pile-up effects can contribute. Maybe with a very low trigger rate might work
Attachment 1: test_corel_Pulesr_all.png  47 kB  Uploaded Wed Apr 3 22:05:16 2019  | Hide | Hide all
test_corel_Pulesr_all.png
Attachment 2: test_corel.png  43 kB  Uploaded Wed Apr 3 22:06:43 2019  | Hide | Hide all
test_corel.png
Attachment 3: test_corel_2000.png  35 kB  Uploaded Wed Apr 3 22:06:57 2019  | Hide | Hide all
test_corel_2000.png
Attachment 4: test_corel_Pulesr_Mugast_CATS.png  12 kB  Uploaded Thu Apr 4 11:19:05 2019  | Hide | Hide all
test_corel_Pulesr_Mugast_CATS.png