  MUGAST, Page 8 of 132  Not logged in MUGAST
Status ExpNbr Date Author Type Category Target-Source Subjectdown
Fixedcom2019 Sat Apr 6 06:24:04 2019 mugastRUNSoftwareCD2new data 5more statistics°
Attachment 1: Run_256_kin.png
Run_256_kin.png
Fixede768s Wed Jul 17 01:08:03 2019 mugastVAMOSGeneralLiFnew beam settings
The last values Q32=21.0, Q31=22.8
With these new optimzed values, we have:
Std dev-x=6.147
Std dev y=13.29

Beam intensity=10⁴ pps, no masque mobile
Attachment 1: run_60_target_thick_01_Vamos.png
run_60_target_thick_01_Vamos.png
Fixede744s Sat Apr 13 17:31:54 2019 mugastVAMOSHardware new Threshold on DC1 Channel 127
Channel 127 of DC1 had a reduction of gain from yesterday, I corrected the gain by calibration and I reduced the threshold from 2114 to 2083 in order to reach the same low level.
(We can still see the lower gain in the online because the watcher is not reloaded, but it is ok in the root files)
Fixede744s Mon Apr 15 15:36:48 2019 iulian Francois Fairouz CatalinOFFLINEGeneralCH2looking for gamma proton events
elab_Brho_gamma_thetaCond_int_ThinTar.pdf -> thin target ELab protons vs Brho 14O in M/Q for protons mult 1 in MUGAST. Orange are the same, but conditioned by Theta Lab protons vs Brho Vamos.
elab_Brho_gamma_thetaCond_int_ThickTar.pdf -> thick target ELab protons vs Brho 14O in M/Q for protons mult 1 in MUGAST. Orange are the same, but conditioned by Theta Lab protons vs Brho Vamos.
elab_Etot_gamma_thetaCond_int_ThinTar.pdf -> thin target ELab protons vs Etot 14O in M/Q for protons mult 1 in MUGAST. Orange are the same, but conditioned by Theta Lab protons vs Brho Vamos.
elab_Etot_gamma_thetaCond_int_ThickTar.pdf -> thick target ELab protons vs Etot 14O in M/Q for protons mult 1 in MUGAST. Orange are the same, but conditioned by Theta Lab protons vs Brho 14O Vamos.


+ simulations
Attachment 1: elab_Brho_gamma_thetaCond_int_ThinTar.pdf
elab_Brho_gamma_thetaCond_int_ThinTar.pdf
Attachment 2: elab_Brho_gamma_thetaCond_int_ThickTar.pdf
elab_Brho_gamma_thetaCond_int_ThickTar.pdf
Attachment 3: elab_Etot_gamma_thetaCond_ThinTar.pdf
elab_Etot_gamma_thetaCond_ThinTar.pdf
Attachment 4: elab_Etot_gamma_thetaCond_ThickTar.pdf
elab_Etot_gamma_thetaCond_ThickTar.pdf
Attachment 5: 1504b.jpg
1504b.jpg
Fixede744s Fri Apr 12 13:53:56 2019 diegoOTHERGeneral intervention
TAC FPMW_HF in VAMOS was still out of range for MUGAST physics
We increased the range from 1us to 2us and applied a delay of 200 ns to the stop of the TAC.
FixedTEST Mon Feb 24 10:26:29 2020 MarlèneGRITGeneral infos
We have changed the range of the TAC FAG-HF to 500ns (it was 200 ns before)

there is no delay on the STOP of MUVI, it is directly the HF clean.

The inspection lines of MUVI1 are sent to vamos acquisition room.
Fixede768s Fri Jul 19 13:24:01 2019 mugastOFFLINEGeneralLiFg-ray spectrum (run 112) requiring 19Ne in VAMOS
Attachment 1: 21.png
21.png
Fixede744s Sat Apr 13 17:03:14 2019 Jérémie & FrançoisOFFLINEGeneral first gamma spectrum
After a few hours run.
Attachment 1: 35gamma.png
35gamma.png
Fixede744s Sun Apr 14 03:30:28 2019 Iulian IuliaOFFLINESocialCH2first calculation ECM ThetaCM for gamma-P emission
Attachment 1: first_gammaP_rec.pdf
first_gammaP_rec.pdf
Attachment 2: Alias_iul.C
TTree *tTmp;
void SetAlias()
{
  tTmp = (TTree*) gROOT->FindObject("PhysicsTree");
  //  tTmp->SetAlias("mp","1.00866491582");
  // tTmp->SetAlias("m14o","14.008596706");
  // tTmp->SetAlias("Elab14o","71.");//MeV
  // tTmp->SetAlias("Erel14o","1.00866491582*14.008596706/(1.00866491582+14.008596706)/(1.00866491582+14.008596706)*71."); 
  tTmp->SetAlias("EPCM","ELab+4.4485492-2*sqrt(ELab*4.4485492)*cos(ThetaLab/57.4)");
  tTmp->SetAlias("ThetaPCM","57.4*(3.14-acos((EPCM+4.4485492-ELab)/2./sqrt(EPCM*4.4485492)))");


  //tTmp->SetAlias("EPCM","ELab+Erel14o-2*sqrt(ELab*Erel14o)*cos(ThetaLab)");
  //tTmp->SetAlias("ThetaPCM","57.4*(3.14-atan((EPCM+Erel14o-ELab)/2./sqrt(EPCM*Erel14o)))");


}
Fixede768s Sat Jul 27 11:02:13 2019 mugastONLINEGeneral figures run 152
Attachment 1: Tab_8bis_run152_6.png
Tab_8bis_run152_6.png
Attachment 2: IC_IC_run152_6.png
IC_IC_run152_6.png
Attachment 3: REC_run152_6.png
REC_run152_6.png
Attachment 4: TACS_run152_6.png
TACS_run152_6.png
Attachment 5: IC_run152_6.png
IC_run152_6.png
Attachment 6: MW_run152_6.png
MW_run152_6.png
Attachment 7: DC_E_run152_6.png
DC_E_run152_6.png
Attachment 8: DC_Y_run152_6.png
DC_Y_run152_6.png
Attachment 9: DC_X_run152_6.png
DC_X_run152_6.png
Attachment 10: Tab_10_run152_6.png
Tab_10_run152_6.png
Attachment 11: Tab_9_run152_6.png
Tab_9_run152_6.png
Attachment 12: Tab_8_run152_6.png
Tab_8_run152_6.png
Attachment 13: TAC_run152_6.png
TAC_run152_6.png
Attachment 14: MG_ToF_run152_6.png
MG_ToF_run152_6.png
Attachment 15: MM_DeE_run152_6.png
MM_DeE_run152_6.png
Attachment 16: MM_ToF_run152_6.png
MM_ToF_run152_6.png
Attachment 17: MM_INPACT_MATRIX_run152_6.png
MM_INPACT_MATRIX_run152_6.png
Attachment 18: MG_IMPACT_MATRIX_run152_6.png
MG_IMPACT_MATRIX_run152_6.png
Fixede744s Tue Apr 16 00:04:08 2019 beyhanOFFLINEGeneral ewire evolution vs beam intensity (normalized to CATS intensity)
Attachment 1: EWIRE0_change_122_vs_130_vs_131.pdf
EWIRE0_change_122_vs_130_vs_131.pdf
Attachment 2: EWIRE1_change_122_vs_130_vs_131.pdf
EWIRE1_change_122_vs_130_vs_131.pdf
Attachment 3: EWIRE2_change_122_vs_130_vs_131.pdf
EWIRE2_change_122_vs_130_vs_131.pdf
Attachment 4: EWIRE3_change_122_vs_130_vs_131.pdf
EWIRE3_change_122_vs_130_vs_131.pdf
Fixed  Mon Apr 8 07:54:53 2019 iulianMUST2Hardware end alpha contamination run
Problem M3 EY , moving piedestal... See below
Alfa spectra have changed, no longer continuous excepting partially MM2.
Attachment 1: bilan_run262_bck_mugast4.png
bilan_run262_bck_mugast4.png
Attachment 2: bilan_run262_bck_mugast3.png
bilan_run262_bck_mugast3.png
Attachment 3: bilan_run262_bck_mugast2.png
bilan_run262_bck_mugast2.png
Attachment 4: bilan_run262_bck_mugast1.png
bilan_run262_bck_mugast1.png
Fixede793s Fri Mar 12 03:57:46 2021 Taras, Nishu, Chloe, CharlieVAMOSGeneralCD2dipole issue fixed
the alimentation on dipole (G1.G.D4.P) was turned off,
the value was resetted(322.6) and works fine now.
cour: 323.5 A
tens. 39.21 V
Attachment 1: run56_dipolefixed.png
run56_dipolefixed.png
Fixede744s Tue Apr 16 01:01:42 2019 beyhan, louis, serge, OFFLINEGeneral different parameters evolution as function of intensity
Attachment 1: mM2_change_122_vs_130_vs_131.pdf
mM2_change_122_vs_130_vs_131.pdf
Attachment 2: mM_Q2_change_122_vs_130_vs_131.pdf
mM_Q2_change_122_vs_130_vs_131.pdf
Attachment 3: mM_Q2_change_122_vs_130_vs_131_zoom.pdf
mM_Q2_change_122_vs_130_vs_131_zoom.pdf
Attachment 4: mV2_change_122_vs_130_vs_131.pdf
mV2_change_122_vs_130_vs_131.pdf
Attachment 5: mV2_change_122_vs_130_vs_131_zoom.pdf
mV2_change_122_vs_130_vs_131_zoom.pdf
Attachment 6: T_FPMW_CATS2_change_122_vs_130_vs_131.pdf
T_FPMW_CATS2_change_122_vs_130_vs_131.pdf
Fixede768s Sat Jul 27 21:12:47 2019 teodora & beyhanOTHERHardware cout rate with no beam & water level check
Around 9 pm, we observed an average count rate of 30kHz on the "or agata" scaler rates panel of GECO. The other count rates were ok.
In the agata acquisition monitoring the rate was given about 8k. one of the detector was showing a rate in red.

We decided to stop : (1) to check the rates with no beam, (2) to check the water level in collection boxes and (3) the beam intensity.

Conclusions :

- The agata rate with no beam is about 4.5kHz. The rest seems ok.

- The level of water in the collection boxes did not change since 3 pm. So it should be ok for the night (see pictures)

- The beam profile and intensity are similar to what we got before

- The rate after we got the beam looks normal now.

So everything is going well Smile

PS : there is also a piscture about the water collection tank to empty the boxes if needed

Attachment 1: scalers-run154-nobeam.png
scalers-run154-nobeam.png
Attachment 2: IMG_3511.jpg
IMG_3511.jpg
Attachment 3: IMG_3512.jpg
IMG_3512.jpg
Attachment 4: IMG_3513.jpg
IMG_3513.jpg
Attachment 5: IMG_3514.jpg
IMG_3514.jpg
Attachment 6: scalers-run154-with_beam.png
scalers-run154-with_beam.png
Fixede793s Tue Mar 16 04:06:46 2021 Lynda, Cyril, NishuAGATAGeneralCD2counting rates AGATA
crystal 12B counting rate is too low ~ 5 / s compared to an average of ~ 80/s - 100/ s for the other crystals
and 05B counting rate is too high from time to time reaching ~2000/s
but every thing is "green" ?!

Issue solved By Emmanuel Clement at 10h45
Fixedcom2019 Sat Apr 6 13:10:11 2019 mugastVAMOSHardwareCD2check in VAMOS HV
target drift chamber HV checked. They are OK
Fixede744s Thu Apr 11 17:37:29 2019 Valérian, Chloé and FrançoisOTHERHardware changes in the electronic
Changes made in the electronic during the break between 14N and 14O

1) T_VAMOS-CATS1
it was not connected. We added 500 ns delay in CATS1 (CATS2 in reality) (48+150ns before)


2) T_CATS1_HF
range changed from 2µs to 200 ns
Fixede744s Thu Apr 11 18:08:12 2019 MarlèneOTHERHardware changes in the electronic

Valérian, Chloé and François wrote:
Changes made in the electronic during the break between 14N and 14O

1) T_VAMOS-CATS1
it was not connected. We added 500 ns delay in CATS1


2) T_CATS1_HF
range changed from 2µs to 200 ns


T_VAMOS-CATS1 is now connected in T_CATS1_CATS2 as posted this night. We have tried to add 500ns more delay (800ns in total) to CATS and it did not work, so we suspected the TAC or ADC. This is why we inverted T_VAMOS-CATS1 with T_CATS1_CATS2. For this TAC, CATS is the START (and is not delayed by 500ns).
Fixede744s Mon Apr 15 09:59:32 2019 MarlèneRUNGeneralCH2change thresholds in DSSD MUST2 and suppression CsI
From RUN126
change of MUST2 DSSD thresholds : thresholds are 583 keV now
change of CsI suppression : before it was suppressed up to 8210 now it is 8200