Fixed | TEST |
Thu May 30 16:10:18 2019 |
OttoRun | RUN | General | 3-alpha | Run230 |
*************************
Run : 230
*************************
Start at 2019-05-30 16:09:44.953775
MUGAST 3alpha run pedestal supression
temperature -3 deg (trap) & -4 deg (must2) consigne 5 deg
no pedestal suppression (but alignement)
MG4T =95V
same as before except that MG7 had a mate hidden on Y (now it is on)
Stop at 2019-05-31 06:39:11.358615
Stop comment: type your message here |
Attachment 1: run229_3lapha_mugast_calibre.png
|
|
Attachment 2: run229_3alpha_must2.png
|
|
Attachment 3: run229_3alpha.png
|
|
Fixed | e786s |
Fri May 31 09:53:14 2019 |
Marlène | GRIT | General | | Alpha source --> status of detectors |
MG3T has been fixed by changing the X kapton.
I have removed aluminium foil on square detectors, the noise is slightly higher.
MG4T is not working properly : noisy on X but no signal on Y !! I see no triggers on the Y side. I tried to increase bias but it did increase noise. I checked in the last calibration of e744 it was working properly. Also I had to do pedestal realignement for this detector because everything was shifted.
MG7T seem to be slightly different than before. Strips X are hidden from the trigger because it became suddenly noisy. |
Fixed | e786s |
Mon Jun 3 18:00:10 2019 |
Marlène | GRIT | General | | MUGAST detectors status --> MG4T |
The problem of MG4T ha sbeen solved by changing the feedthrough.
The pulser on Y was not working and we were seeing no events (no ORD) because there was a problem with the bias of board. The issue is due to the feedthrough.
So now MG4T (HT &BT) are plugged in a free feedthrough (below MG1T) and I see pulser signals on Y.
The bias seems to apply normally on all detectors (in the dark).
The pulser is seen on each boards.
I have put back aluminium foil on square but I could not test it. |
Fixed | TEST |
Tue Jun 11 18:59:36 2019 |
OttoRun | RUN | General | 3-alpha | Run231 |
*************************
Run : 231
*************************
Start at 2019-06-11 18:58:31.909158
MUGAST 3alpha run pedestal supression
temperature -1 deg (trap) & -4 deg (must2) consigne 5 deg
no pedestal suppression (but alignement)
alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!
Stop at 2019-06-11 21:40:22.107913
Stop comment: |
Fixed | TEST |
Tue Jun 11 21:44:38 2019 |
OttoRun | RUN | General | 3-alpha | Run232 |
*************************
Run : 232
*************************
Start at 2019-06-11 21:44:02.397316
MUGAST 3alpha run pedestal supression
temperature -1 deg (trap) & -4.5 deg (must2) consigne 5 deg
with pedestal suppression
alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!
Stop at 2019-06-12 10:05:20.279209
Stop comment: type your message here |
Fixed | TEST |
Wed Jun 12 18:40:57 2019 |
OttoRun | RUN | General | 3-alpha | Run233 |
*************************
Run : 233
*************************
Start at 2019-06-12 18:40:34.500264
MUGAST 3alpha run pedestal supression
temperature -1 deg (trap) & -4.5 deg (must2) consigne 5 deg
without pedestal suppression
alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!
Stop at 2019-06-13 09:06:36.277090
Stop comment: |
Fixed | TEST |
Thu Jun 13 09:46:40 2019 |
OttoRun | RUN | General | 3-alpha | Run234 |
*************************
Run : 234
*************************
Start at 2019-06-13 09:45:49.383074
Discard run
alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!
Stop at 2019-06-13 09:47:06.006149
Stop comment: Discard this run |
Fixed | TEST |
Thu Jun 13 09:47:39 2019 |
OttoRun | RUN | General | 3-alpha | Run235 |
*************************
Run : 235
*************************
Start at 2019-06-13 09:47:19.893278
Discard run
alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!
Stop at 2019-06-13 14:23:23.430392
Stop comment: type your message here |
Fixed | e786s |
Wed Jun 19 12:02:07 2019 |
OttoTime | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 237
- MUVI: MUVI1
- Side : MUFEE X (Junction)
- Telescope: 1, 2, 3, 4 |
Fixed | e786s |
Wed Jun 19 12:24:43 2019 |
OttoTime | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 238
- MUVI: MUVI1
- Side : MUFEE Y (Ohmic)
- Telescope: 1, 2, 3, 4 |
Fixed | e786s |
Wed Jun 19 12:48:13 2019 |
OttoTime | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 239 - MUVI: MUVI2
- Side : MUFEE X (Junction)
- Telescope: 1, 2, 3, 4 |
Fixed | e786s |
Wed Jun 19 13:11:57 2019 |
OttoTime | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 240
- MUVI: MUVI2
- Side : MUFEE Y (Ohmic)
- Telescope: 1, 2, 3, 4 |
Fixed | e786s |
Wed Jun 19 13:36:41 2019 |
OttoTime | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 241
- MUVI: MUVI3
- Side : MUFEE X (Junction)
- Telescope: 1, 2, 3, 4 |
Fixed | e786s |
Wed Jun 19 14:02:04 2019 |
OttoTime | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 242
- MUVI: MUVI3
- Side : MUFEE Y (Ohmic)
- Telescope: 1, 2, 3, 4 |
Fixed | TEST |
Fri Jun 21 17:14:55 2019 |
OttoRun | RUN | General | N/A | Run8 |
*************************
Run : 8
*************************
Start at 2019-06-21 17:14:12.777585
test
Stop at 2019-06-21 17:16:36.895632
Stop comment: |
Fixed | TEST |
Fri Jun 21 17:20:29 2019 |
OttoRun | RUN | General | N/A | Run9 |
*************************
Run : 9
*************************
Start at 2019-06-21 17:20:14.372415
Valerian n'a pas a courir ouf!!
Stop at 2019-06-21 17:21:30.206657
Stop comment: he is really not fit!
Stop at 2019-06-23 08:27:58.124593
Stop comment: |
Fixed | e786s |
Sun Jun 23 08:29:14 2019 |
OttoRun | RUN | General | Cryo-empty | Run10 |
*************************
Run : 10
*************************
Start at 2019-06-23 08:03:31.361779
Firt run, VAMOS tunning
Brho0=1.16 Tm
DC pressure 5.8 mbar
IC pressure 120 mbar
Stop at 2019-06-23 10:08:04.932948
Stop comment: Entering to Change IC 5-6 Preamps |
Fixed | e786s |
Sun Jun 23 09:06:29 2019 |
Daniele M | VAMOS | General | N/A | VAMOS SETTINGS |
attached the energy measured in CIME and the calculated VAMOS settings |
Attachment 1: vamos_setttings.jpg
|
|
Fixed | e786s |
Sun Jun 23 10:08:07 2019 |
OttoRun | RUN | General | Cryo-empty | Run11 |
*************************
Run : 11
*************************
Start at 2019-06-23 08:29:17.562468
VAMOS tunning
Brho0=1.16 Tm
DC pressure 5.8 mbar
IC pressure 120 mbar |
Fixed | TEST |
Sun Jun 23 10:31:08 2019 |
OttoRun | RUN | General | 3-alpha | Run233 |
OttoRun wrote: | *************************
Run : 233
*************************
Start at 2019-06-12 18:40:34.500264
MUGAST 3alpha run pedestal supression
temperature -1 deg (trap) & -4.5 deg (must2) consigne 5 deg
without pedestal suppression
alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!
Stop at 2019-06-13 09:06:36.277090
Stop comment: |
|
Attachment 1: energy_cal_run0233.png
|
|