FixedTEST Thu May 30 16:10:18 2019 OttoRunRUNGeneral3-alphaRun230
*************************
Run : 230
*************************

Start at 2019-05-30 16:09:44.953775

MUGAST 3alpha run pedestal supression
temperature -3 deg (trap) & -4 deg (must2) consigne 5 deg

no pedestal suppression (but alignement)
MG4T =95V

same as before except that MG7 had a mate hidden on Y (now it is on)

Stop at 2019-05-31 06:39:11.358615
Stop comment: type your message here
Fixede786s Fri May 31 09:53:14 2019 MarlèneGRITGeneral Alpha source --> status of detectors
MG3T has been fixed by changing the X kapton.

I have removed aluminium foil on square detectors, the noise is slightly higher.

MG4T is not working properly : noisy on X but no signal on Y !! I see no triggers on the Y side. I tried to increase bias but it did increase noise. I checked in the last calibration of e744 it was working properly. Also I had to do pedestal realignement for this detector because everything was shifted.

MG7T seem to be slightly different than before. Strips X are hidden from the trigger because it became suddenly noisy.
Fixede786s Mon Jun 3 18:00:10 2019 MarlèneGRITGeneral MUGAST detectors status --> MG4T
The problem of MG4T ha sbeen solved by changing the feedthrough.
The pulser on Y was not working and we were seeing no events (no ORD) because there was a problem with the bias of board. The issue is due to the feedthrough.
So now MG4T (HT &BT) are plugged in a free feedthrough (below MG1T) and I see pulser signals on Y.

The bias seems to apply normally on all detectors (in the dark).
The pulser is seen on each boards.
I have put back aluminium foil on square but I could not test it.
FixedTEST Tue Jun 11 18:59:36 2019 OttoRunRUNGeneral3-alphaRun231
*************************
Run : 231
*************************

Start at 2019-06-11 18:58:31.909158

MUGAST 3alpha run pedestal supression
temperature -1 deg (trap) & -4 deg (must2) consigne 5 deg

no pedestal suppression (but alignement)

alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!

Stop at 2019-06-11 21:40:22.107913
Stop comment:
FixedTEST Tue Jun 11 21:44:38 2019 OttoRunRUNGeneral3-alphaRun232
*************************
Run : 232
*************************

Start at 2019-06-11 21:44:02.397316

MUGAST 3alpha run pedestal supression
temperature -1 deg (trap) & -4.5 deg (must2) consigne 5 deg

with pedestal suppression

alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!

Stop at 2019-06-12 10:05:20.279209
Stop comment: type your message here
FixedTEST Wed Jun 12 18:40:57 2019 OttoRunRUNGeneral3-alphaRun233
*************************
Run : 233
*************************

Start at 2019-06-12 18:40:34.500264

MUGAST 3alpha run pedestal supression
temperature -1 deg (trap) & -4.5 deg (must2) consigne 5 deg

without pedestal suppression

alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!

Stop at 2019-06-13 09:06:36.277090
Stop comment:
FixedTEST Thu Jun 13 09:46:40 2019 OttoRunRUNGeneral3-alphaRun234
*************************
Run : 234
*************************

Start at 2019-06-13 09:45:49.383074

Discard run
alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!

Stop at 2019-06-13 09:47:06.006149
Stop comment: Discard this run
FixedTEST Thu Jun 13 09:47:39 2019 OttoRunRUNGeneral3-alphaRun235
*************************
Run : 235
*************************

Start at 2019-06-13 09:47:19.893278

Discard run
alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!

Stop at 2019-06-13 14:23:23.430392
Stop comment: type your message here
Fixede786s Wed Jun 19 12:02:07 2019 OttoTimeRUNGeneralN/ATimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 237
- MUVI: MUVI1
- Side : MUFEE X (Junction)
- Telescope: 1, 2, 3, 4
Fixede786s Wed Jun 19 12:24:43 2019 OttoTimeRUNGeneralN/ATimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 238
- MUVI: MUVI1
- Side : MUFEE Y (Ohmic)
- Telescope: 1, 2, 3, 4
Fixede786s Wed Jun 19 12:48:13 2019 OttoTimeRUNGeneralN/ATimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 239 - MUVI: MUVI2
- Side : MUFEE X (Junction)
- Telescope: 1, 2, 3, 4
Fixede786s Wed Jun 19 13:11:57 2019 OttoTimeRUNGeneralN/ATimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 240
- MUVI: MUVI2
- Side : MUFEE Y (Ohmic)
- Telescope: 1, 2, 3, 4
Fixede786s Wed Jun 19 13:36:41 2019 OttoTimeRUNGeneralN/ATimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 241
- MUVI: MUVI3
- Side : MUFEE X (Junction)
- Telescope: 1, 2, 3, 4
Fixede786s Wed Jun 19 14:02:04 2019 OttoTimeRUNGeneralN/ATimeCalibratorRun
*** Run Time Calibrator FSC ***
- Run: 242
- MUVI: MUVI3
- Side : MUFEE Y (Ohmic)
- Telescope: 1, 2, 3, 4
FixedTEST Fri Jun 21 17:14:55 2019 OttoRunRUNGeneralN/ARun8
*************************
Run : 8
*************************

Start at 2019-06-21 17:14:12.777585

test

Stop at 2019-06-21 17:16:36.895632
Stop comment:
FixedTEST Fri Jun 21 17:20:29 2019 OttoRunRUNGeneralN/ARun9
*************************
Run : 9
*************************

Start at 2019-06-21 17:20:14.372415

Valerian n'a pas a courir ouf!!

Stop at 2019-06-21 17:21:30.206657
Stop comment: he is really not fit!

Stop at 2019-06-23 08:27:58.124593
Stop comment:
Fixede786s Sun Jun 23 08:29:14 2019 OttoRunRUNGeneralCryo-emptyRun10
*************************
Run : 10
*************************

Start at 2019-06-23 08:03:31.361779

Firt run, VAMOS tunning

Brho0=1.16 Tm
DC pressure 5.8 mbar
IC pressure 120 mbar

Stop at 2019-06-23 10:08:04.932948
Stop comment: Entering to Change IC 5-6 Preamps
Fixede786s Sun Jun 23 09:06:29 2019 Daniele MVAMOSGeneralN/AVAMOS SETTINGS
attached the energy measured in CIME and the calculated VAMOS settings
Fixede786s Sun Jun 23 10:08:07 2019 OttoRunRUNGeneralCryo-emptyRun11
*************************
Run : 11
*************************

Start at 2019-06-23 08:29:17.562468

VAMOS tunning

Brho0=1.16 Tm
DC pressure 5.8 mbar
IC pressure 120 mbar
FixedTEST Sun Jun 23 10:31:08 2019 OttoRunRUNGeneral3-alphaRun233

OttoRun wrote:
*************************
Run : 233
*************************

Start at 2019-06-12 18:40:34.500264

MUGAST 3alpha run pedestal supression
temperature -1 deg (trap) & -4.5 deg (must2) consigne 5 deg

without pedestal suppression

alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!

Stop at 2019-06-13 09:06:36.277090
Stop comment: