Fixed | TEST |
Mon Mar 4 14:34:31 2019 |
Valerian, Marlène, Iulian | MUST2 | General | N/A | MUST2 positions (based on e755) |
Based on experiment e755, the MUST2 config is the following :
T1, PCB10:
X128Y1.......X1
. .
. .
Y128.......
T2, PCB5:
Y128.......X128Y1
. .
. .
.......X1
T3, PCB4:
.......Y128
. .
. .
X1.......X128Y1
T4, PCB7:
X1.......
. .
. .
X128Y1.......Y128 |
Attachment 1: Screenshot_2019-03-04_at_14.39.02.png
|
|
Fixed | TEST |
Mon Mar 4 16:24:51 2019 |
valerian marlène | GRIT | General | | RUN52 3 alpha source 4MUST2 no pedestal |
RUN52
3 alpha source run with only 4MUST2 (in MUVI3) with pedestal suppression at 13 deg. |
Attachment 1: HV_MUST2_during_run52.png
|
|
Attachment 2: HV_MUST2_afternightrun52.png
|
|
Attachment 3: 3alpha_source_4MUST2_nightrun52.png
|
|
Fixed | |
Mon Mar 4 17:22:29 2019 |
Marlene | GRIT | General | | Temperature on MUFEE MUST2 13deg |
Temperature given by DAS on MUVI3 (MUST2)
screenshots before run
after run : same values |
Attachment 1: temperature_MM1_cooling13deg.png
|
|
Attachment 2: temperature_MM1_Y_cooling13deg.png
|
|
Attachment 3: temperature_MM2_X_cooling13deg.png
|
|
Attachment 4: temperature_MM2_Y_cooling13deg.png
|
|
Attachment 5: temperature_MM3_X_cooling13deg.png
|
|
Attachment 6: temperature_MM3_Y_cooling13deg.png
|
|
Attachment 7: temperature_MM4_X_cooling13deg.png
|
|
Attachment 8: temperature_MM4_Y_cooling13deg.png
|
|
Fixed | TEST |
Thu Mar 7 09:29:47 2019 |
Marlene | GRIT | General | | run 96 double alpha source run shifted 60mm |
RUN 96
double alpha source
MG1
MG3
MG4
MG5
MG7
-
-
MG11
MM1
MM2
MM3
MM4
Missing strips in Y the same as in run 40
Problem with the annular MG11 -> no current when bias is put... --> to be checked
Problem with the insertion of the target holder --> solved
I also checked again MG9 (still noisy)
and MG10 with 140V I=0,52uA and counting rate is the same order of magnitude as trapezoid. Spectra are OK (not saved). |
Attachment 1: run96_MG7_MG11notworking.png
|
|
Attachment 2: run96_3alphasource_shifted_MG1-5.png
|
|
Attachment 3: run96_HV2.png
|
|
Attachment 4: run96_HV1.png
|
|
Fixed | TEST |
Thu Mar 7 14:02:03 2019 |
Marlène Iulian | MUST2 | General | N/A | test of MUST2 electronics |
Test with pulser at 20 kHz
The OR of MUVI was scaled down by a factor of 100, to provoke some resets |
Fixed | |
Tue Mar 12 10:21:17 2019 |
Marlène | MUST2 | General | | Debug Calimero Y |
Debug of Calimero
test of channels (mate) masking : OK with DAS, OK with GECCO (see attachment) when done in the "disable channels" of GECCO
With calimero, when we are on Y, when we click right to mask a mate we see 'disable mate 2' instead of 'disable mate 10' !!! |
Attachment 1: test_GECO_masquage_mate2&10activees_seulement.png
|
|
Fixed | TEST |
Mon Mar 18 18:31:18 2019 |
valerian marlène | GRIT | General | 3-alpha | RUN 20 (com2019) 3 alpha overnight |
RUN20
Run with double 3alpha source MUST2+MG1_7+MG11
temperature -4 deg on both cooling
Acquisition with VAMOS+MUGAST coupled (com2019)
Time range =600ns
pedestal aligned and suppressed
run start at 18h23 monday 18
run stop at 8h34 tuesday 19
about 12Go of data |
Attachment 1: RUN20_HV_MG_3lapha.png
|
|
Attachment 2: RUN20_3alpha_HV_MUST2.png
|
|
Fixed | |
Tue Mar 19 19:35:53 2019 |
marlene | GRIT | General | | RUN 98 3alpha source all detectors |
RUN 98
3 alpha source
all detectors (trap,square, MUST2,annular)
Temp=15deg
Acquisition mugast |
Fixed | TEST |
Thu Mar 21 10:37:47 2019 |
Adrien and Valerian | MUST2 | General | N/A | Test TOF must2 |
To make sure the mate work correctly we took an alpha run with a delayed fag on the MUVI stop:
- all mate register a time in the the middle of the range as expected (fag delayed by 350ns, range is 600ns)
Test with time calibrator, all channel are enable, pulser on channel 3 of each mate T1X:
- energy spectrum: pulser where we expect it.
- we expect to see a time calibrator spectrum on channel 3 of each mate. instead we see times spectrum on Channel 3 of mate 1 and 2, channel 12 of mate 1, channel 14 of mate2. nothing else
Conclusion
The problem was that the synctst had to be checked and ord unchecked |
Attachment 1: Time_behavior_withDAS.png
|
|
Fixed | TEST |
Fri Mar 22 17:28:09 2019 |
Otto | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 117
- MUVI: MUVI1
- Side : MUFEE X (Junction)
- Telescope: 1, 2, 3, 4 |
Fixed | TEST |
Fri Mar 22 17:37:37 2019 |
Otto | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 118
- MUVI: MUVI1
- Side : MUFEE Y (Ohmic)
- Telescope: 1, 2, 3, 4 |
Fixed | TEST |
Fri Mar 22 17:47:04 2019 |
Otto | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 119
- MUVI: MUVI2
- Side : MUFEE X (Junction)
- Telescope: 1, 2, 3, 4 |
Fixed | TEST |
Fri Mar 22 17:56:37 2019 |
Otto | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 120
- MUVI: MUVI2
- Side : MUFEE Y (Ohmic)
- Telescope: 1, 2, 3, 4 |
Fixed | TEST |
Fri Mar 22 18:06:11 2019 |
Otto | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 121
- MUVI: MUVI3
- Side : MUFEE X (Junction)
- Telescope: 1, 2, 3, 4 |
Fixed | TEST |
Fri Mar 22 18:15:39 2019 |
Otto | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 122
- MUVI: MUVI3
- Side : MUFEE Y (Ohmic)
- Telescope: 1, 2, 3, 4 |
Fixed | TEST |
Thu Mar 28 14:22:25 2019 |
OttoRun | RUN | General | N/A | TEST |
Run : 146
Date : 28/03/2019
Start: 14:22:25
asdasdasdasd
Stop: 14:22:42
asdasdasdasd |
Fixed | TEST |
Thu Mar 28 18:27:59 2019 |
OttoRun | RUN | General | N/A | Run ??? |
Run : ????
*************************
Date : on all de
Start: alpha source
*WARNING ! FIFO for BLK GMT in 4x4 instead of 2x8 to get only dead time from MUGAST VXI
Stop: 09:34:54
* |
Fixed | |
Thu Mar 28 18:29:04 2019 |
marlene valerian | GRIT | General | 3-alpha | RUN 147 3alpha source overnight all detectors -4deg |
RUN 147 start@ 18:30
3 alpha source over night
cooling at -4 deg on all detectors
warning !!! FIFO for BLK signal has been put in 4x4 instead of 2x8, BLK is only coming from the 3 VXI of MUGAST |
Fixed | TEST |
Fri Mar 29 10:00:52 2019 |
Iulian | VAMOS | General | N/A | GMT channel4 corresponds to VAMOS now |
Voie4 GMT = trigger Vamos |
Fixed | TEST |
Fri Mar 29 13:44:49 2019 |
OttoRun | RUN | General | N/A | Run108 |
"*************************\n Run : 108\n *************************\nRun Conditions:[\B]
- Operator:
- Trigger:
- Target/Source:
- Beam:
- Intensity:
Comment:[\B]
" |