Status ExpNbr Date Author Type Categorydown Target-Source Subject
Fixede744s Sun Apr 14 15:04:39 2019 Nikola Dieter Nicolas Beyhan KseniiaONLINESocialCH2Log & Scalers -- Run 122 at 15:04 end of run
Fixede744s Mon Apr 15 08:34:36 2019 Nikola Marion Kseniia Damien Chloe Marlene FairouzONLINESocialCH2Log & Scalers -- Run 125 at 08:35
Fixede744s Mon Apr 15 08:45:05 2019 Nikola Marion Kseniia Damien Chloe Marlene FairouzONLINESocialCH2Log & Scalers -- Run 125 at 08:40
Spectra reset 8:50
Fixede744s Mon Apr 15 10:42:39 2019 Nikola Marion Kseniia Damien Chloe Marlene FairouzONLINESocialCH2Log & Scalers -- Run 126 at 10:40
spectra reset at 10:45
Fixede744s Mon Apr 15 12:22:40 2019 Faïrouz, Chloé, Sylvain, Damien & IuliaONLINESocialCH2Scaler -- Run 126 at 12:20
Fixede744s Mon Apr 15 18:07:43 2019 Marion Iulian Marlene Beyan Francois MarcoONLINESocialCH2Screenshots&Scalers Run 130 v1
Fixed  Mon Apr 15 21:34:11 2019 mugastONLINESocial Screenshots&Scalers Run 131
Fixede744s Tue Apr 16 08:30:58 2019 Kseniia, DieterAGATASocialCH2AGATA count rate over the last 12 hours
Fixede744s Tue Apr 16 10:23:06 2019 Marion, DieterONLINESocialCH2Log and scalers RUN 132 @10:20 (part5)
Fixede744s Tue Apr 16 11:48:07 2019 Marion, Dieter, Beyan, Kseniia, + everyone ONLINESocialCarbonRun 134. Carbon target.
Run #134
Target #4 Carbon @ 8.4 mg/cm**2
Brho=0.5807 Tm
Finger at 500 mm

Pepperpot removed. Closed slits.
PCP working to provide us with ~10**5 pps on CATS.
Fixede744s Tue Apr 16 15:53:42 2019 Marion, DieterONLINESocialCarbonLog and scalers RUN 134 @ 15h45
Fixed  Tue Jul 23 17:09:43 2019 UK refugeesOTHERSocial Kipper issues.
Fixede768s Tue Jul 23 19:21:50 2019 NicolasOTHERSocialN/AWaiting end of canicule (and beam!)
Fixed  Tue Jul 23 23:22:56 2019 UK refugeesOTHERSocial Kipper issues.
I would mail you some but the postage is nuts those days!
Fixede793s Fri Mar 12 09:57:30 2021 AdrienOTHERSocialN/ALost phone in agata counting room
Someone lost his/her phone in the agata counting room. Pictures attached.
FixedTEST Thu Feb 21 14:37:26 2019 iulian, valerian &FreddyGRITHardware3-alphaResults Calibration 3-alpha RUN30 MG1, MG3,MG5 & MG7
Results of the calibration run 30
In order:
Telescope1 X,Y (MG5 X,Y)
Telescope4 X,Y (MG3 X,Y)
Telescope7 X,Y (MG7 X,Y)
Telescope9 X,Y (MG1 X,Y)
FixedTEST Thu Feb 21 16:43:45 2019 Valérian, Freddy, Iulian and MarlèneGRITHardwareN/ACurrent measurements
Joined are the the I(V).

Detectors seems to respond accordingly with the previous measurements from Marlène.

Annular detector curve is quite strange…

Joined files :

Depletion_Mugast_MG.pdf -> summary for trapezoidal + annular detectors
Depletion_Mugast_M2.pdf -> summary for Must2 detectors

Depletion_Mugast.xlsx -> Excel file
FixedTEST Thu Feb 21 17:32:42 2019 iulian, freddyGRITHardware3-alpharun 33 3 alfa with MG1,MG3,MG4,MG5,MG7,MG9 short run
First test of the repaired 2xcables. The repaired cables were placed between MUVI1_2 <=> MG4 & MUVI1_4 <=>MG11 (longer cable)
Correspondance for the rest is the same as for run 30. The mapping is the same for all detectors (including anulaire detector).

Alpha source run for trapezoid detectors + anulaire detector with pedestals aligned and zero suppression. Cooling at 15°C.
Thresholds at 500 keV.

On MUVI1:

1--> MG5 (150V)
2--> MG4 (95V)
3--> MG3 (160V)
4--> MG11 (60V)

On MUVI2:
5--> MG7 (120V)
6--> MG1 (140V)
7--> --
8--> --

TRIGGERS are:
GMT ch1 --> OR MUVI1
GMT ch2 --> OR MUVI2
GMT ch3 --> OR MUVI3 (nothing on it)
FixedTEST Thu Feb 21 18:14:48 2019 iulianGRITHardware3-alphaACQ problem run 34 3 alfa with MG1,MG3,MG4,MG5,MG7,MG9

Alpha source run for trapezoid detectors + anulaire detector with pedestals aligned (with HV Frown ) and zero suppression. Cooling at 15°C.
Thresholds at 500 keV.

On MUVI1:

1--> MG5 (150V)
2--> MG4 (95V)
3--> MG3 (160V)
4--> MG11 (60V)

On MUVI2:
5--> MG7 (120V)
6--> MG1 (140V)
7--> --
8--> --

TRIGGERS are:
GMT ch1 --> OR MUVI1
GMT ch2 --> OR MUVI2
GMT ch3 --> OR MUVI3 (nothing on it)

PROBLEM ACQ , MUVI2 has no validations (DECS). MUVI1 OK!. Looks like a problem on the catchslave actor of Narval as no events seems to pass trough

The problem was caused by the cable used to triger the GMTSlave from the FanIn FanOut.
FixedTEST Fri Feb 22 11:56:53 2019 iulian GRITHardware3-alpharun 38 3-alpha with reordering of the MUGAST cables for the logic and analogic signals
Run 38

Alpha source run for trapezoid detectors during the night with pedestals aligned (with HV on) and zero suppression. Cooling at 15°C.
Thresholds at 500 keV.

Important:
The MUGAST signal cables (logic & analogic) have been reordered:


On MUVI1
ChannelSpectra LabelDetectorVoltage
1 MG1 MG1 140V
2 MG3 MG3 160V
3 MG4 MG4 95V
4 MG5 MG5 150V



On MUVI2:
ChannelSpectra LabelDetectorVoltage
1 MG7 MG7 120V
2 MG9 MG1160V
3 MG10 -- --
4 MG11 -- --


TRIGGERS are:
GMT ch1 --> OR MUVI1
GMT ch2 --> OR MUVI2
GMT ch3 --> OR MUVI3 (nothing on it)

ELOG Home Page