Fixed, e768s, Wed Jul 17 01:51:10 2019, mugast, BEAM, General, , Faisceau perdu
|
Un Q-pole dans L3 est tombé. L'opérateur a du mal à le redémarrer. Bertrand et un technicien ont essayé de réparer mais sans succès.
L'astreinte a été appelée à 2h50. |
Fixed, e768s, Wed Jul 17 09:07:19 2019, mugast, BEAM, General, , beam settings
|
values as it is now |
Fixed, e768s, Wed Jul 17 09:50:44 2019, mugast, MUST2, General, N/A, MUST2 CAEN alim burned down
|
Yesterday at around 23h the CAEN alim stopped working.
- The 48V power supply is blown up
- The Main controller is still starting
- We can see all card from the powered crate so hopefully no damage on the card themselves
Nicolas left a 6am this morning to pickup a spare at Orsay.
In the mean time we run 15N with VAMOS-AGATA daq. |
Fixed, e768s, Thu Jul 18 14:38:04 2019, mugast, GRIT, General, N/A, time calibrator for MG10 X 
|
Here is the plot |
Fixed, e768s, Thu Jul 18 23:35:26 2019, mugast, RUN, General, N/A, run 102
|
VAMOS ref run:
Brho = 0.5854 Tm
P(CHIO) = 45 mbar
P(drift) = 6 mbar
hole = 7mm
I(15O) = 1.5e3 pps |
Fixed, e768s, Thu Jul 18 23:42:44 2019, mugast, RUN, General, , run 103 -> Brho scan
|
|
Fixed, e768s, Fri Jul 19 00:59:20 2019, mugast, RUN, General, , RUN 107 - Reference Target Run
|
Reference Target Run after beam tuning
Brho 0.5854 Tm |
Fixed, e768s, Fri Jul 19 01:02:54 2019, mugast, OTHER, General, , 15O beam parameters
|
Brho manip = Brho after target - 2% = 0.556 Tm
RMN = 2821.476 Gauss
Q32 = 15.2 A
Q31 = 16.8 A
Q41 = 148.070 A
Q42 = 217.452 A
Dipole D4p = 187.632 A |
Fixed, e768s, Fri Jul 19 01:05:26 2019, mugast, RUN, General, , RUN 108
|
Brho Change to 0.556 Tm
Moving Plate : Final Position +7mm
with 6 kHz
Slowly increasing |
Fixed, e768s, Fri Jul 19 09:32:45 2019, mugast, BEAM, General, LiF, Beam is fluctuating
|
There was a temporary beam loss during the run 113 |
Fixed, e768s, Fri Jul 19 09:38:05 2019, mugast, GRIT, General, LiF, Impact matrix for MUGAST
|
|
Fixed, e768s, Fri Jul 19 13:09:14 2019, mugast, BEAM, General, LiF, Counting rates as a function of beam intensity
|
checking counting rates as a function of increasing beam intensity |
Fixed, e768s, Fri Jul 19 13:24:01 2019, mugast, OFFLINE, General, LiF, g-ray spectrum (run 112) requiring 19Ne in VAMOS
|
|
Fixed, e768s, Fri Jul 19 15:11:10 2019, mugast, OFFLINE, General, , Procedure for run conversion and analysis
|
Here is the procedure to convert a run with VAMOS and AGATA data and add those data to the Kinematic plot.
Every 4GB there will be a file change, to see if there is a file change type:
ls -ltrh /data/mugastX/e768s/acquisition/run/
Convert the last 4GB file and add it to the chain through the following steps:
Open a terminal from one of the Mugast computer:
Type: source $HOME/GoToVamosAna
Type: MyVAna -d2Ar -f /data/mugastX/e768s/acquisition/run/run_$$$$-DD-MM-YY.HHhMMmSSs /data/mugastX/e768s/acquisition/run/RootA/r$$$$_000a.root
OR (in case the run is split in more files)
Type: MyVAna -d2Ar -f /data/mugastX/e768s/acquisition/run/run_$$$$-DD-MM-YY.HHhMMmSSs.N /data/mugastX/e768s/acquisition/run/RootA/r$$$$_00Na.root
(If you get the error "Could not find the BrhoRef for the run <RunNr> ", update the ./Calibs/ListBrho.dat file with relevant Brho for RunNr)
Now you would be already able to run npanalysis to analyze the data VAMOS+Mugast with the following procedure:
Type: exit (to exit the VAMOS environment)
Type: lpcgrit (to load the analysis environment)
Type: npp MUGAST
Edit the exe_e768s.sh to specify which runs to analyze. The command should be the following:
npanalysis -D mugast.detector -C Calibration.txt -E ./Reaction/15N7Lit.reaction -T /data/mugastX/e768s/acquisition/run/RootA/r$$$$_00Na.root AD -O run_$$$$_N.root
Type: ./exe_e768s.sh
To convert a run including the AGATA data, the procedure is the following:
Go to the AGATA analysis computer (last on the right close to the window in the acquisition room), open a terminal and type the following commands:
> a4
> t e768s
> cd Replay
>./do_global_replay.sh -run ## (## is the number of the run you want to convert).
When it is finished, copy the root file to the GANIL world:
> mkdir /dataganil/e768s/acquisition/root_agata/run_00##/
> cp ./run_00##/Out/Analysis/Tree_00*.root /dataganil/e768s/acquisition/root_agata/run_00##/.
Now you can analyze the file using npanalysis.
The command in the exe_e768s.sh file should now look like:
npanalysis -D mugast.detector -C Calibration.txt -E ./Reaction/15N7Lit.reaction -T /data/mugastX/e768s/acquisition/root_agata/run_00##/Tree_000N.root TreeMaster -O run_00##_N_agata.root
The run will be created in the ./NPOutputs folder.
Bon courage! |
Fixed, e768s, Fri Jul 19 15:55:38 2019, mugast, RUN, General, , Beam is fluctuating 19x
|
Run 115 stopped at 16h40. |
Fixed, e768s, Fri Jul 19 17:15:46 2019, mugast, RUN, General, LiF, Run 116
|
|
Fixed, e768s, Fri Jul 19 18:57:12 2019, mugast, ONLINE, General, LiF, screenshots run 116 17x
|
|
Fixed, e768s, Fri Jul 19 22:16:43 2019, mugast, ONLINE, General, N/A, How to restart the vamos spectrum visualization (on the mugast daq computer)
|
On the mugast DAQ computer:
* Open a terminal
* Type "GoToVamosAna"
* Type "vigru"
* In vigru: File, Open config -> vigru_ganp154.xml
* Then: Setup, Sources, Host/FileName: "ganp154", port:9090
If a spectrum is missing -> Add it with "display one histo" |
Fixed, e768s, Sat Jul 20 17:03:23 2019, mugast, RUN, General, LiF, Run 117 stopped
|
Run 117 stopped because VAMOS VXI went offline |
Fixed, e768s, Sat Jul 20 17:06:12 2019, mugast, RUN, General, LiF, Run 118 Scaler
|
Start run 118 |
|