  MUGAST, Page 4 of 132  Not logged in MUGAST
Status ExpNbr Date Author Type Categorydown Target-Source Subject
Ongoinge775s Sat Feb 29 14:27:48 2020 mugast ; ValerianOFFLINESoftwareCD2Issues in the EX reconstruction

mugast ; valerian irene wrote:
We are currently working on the EX reconstruction which is off for detector 1, 3 and 7... One of our guess might be that there is an issue with the position of the detectors which gives error in theta... As the energy calibration is ok. Joined are :

- the Excitation energy for good detectors compared to all detectors,
- the theoretical kinematic line for the 3.57 state for each detectors,
- the excitation energy as a function of the laboratory angle (for this last spectra a clear dependency is visible for telescopes 1,3 and 7)
- the 3D view, which doesn't show clear problem in position...


Testing changing the beam impact :

-Kine for beam impact = (0,0,0) for reference
-Kine for beam impact = (0,0,-5mm)
-Kine for beam impact = (0,0,+5mm)

-Ex vs ThetaLab for beam impact = (0,0,0) for reference
-Ex vs ThetaLab for beam impact = (0,0,-5mm)
-Ex vs ThetaLab for beam impact = (0,0,+5mm)
Red lines corresponds to Ex = 3.570 MeV and Ex = 4.070 MeV

To try to correct the problem i also offseted the beam in X and Y, but this doesn't solve the problem quite the opposite as it make the result worse for other detectors also... The fact that it is better at lower angle and worse at higher angle might mean that the detectors angles are tilted compared to the
surveyors measurments... I also tried to use the CAO positions for MG3 (see Detectors/mugast.detector) but it didn't solve the problem.
Ongoinge775s Sat Feb 29 15:53:12 2020 mugast ; ValerianOFFLINESoftwareCD2-AuIssues in the EX reconstruction

mugast ; Valerian wrote:

mugast ; valerian irene wrote:
We are currently working on the EX reconstruction which is off for detector 1, 3 and 7... One of our guess might be that there is an issue with the position of the detectors which gives error in theta... As the energy calibration is ok. Joined are :

- the Excitation energy for good detectors compared to all detectors,
- the theoretical kinematic line for the 3.57 state for each detectors,
- the excitation energy as a function of the laboratory angle (for this last spectra a clear dependency is visible for telescopes 1,3 and 7)
- the 3D view, which doesn't show clear problem in position...


Testing changing the beam impact :

-Kine for beam impact = (0,0,0) for reference
-Kine for beam impact = (0,0,-5mm)
-Kine for beam impact = (0,0,+5mm)

-Ex vs ThetaLab for beam impact = (0,0,0) for reference
-Ex vs ThetaLab for beam impact = (0,0,-5mm)
-Ex vs ThetaLab for beam impact = (0,0,+5mm)
Red lines corresponds to Ex = 3.570 MeV and Ex = 4.070 MeV

To try to correct the problem i also offseted the beam in X and Y, but this doesn't solve the problem quite the opposite as it make the result worse for other detectors also... The fact that it is better at lower angle and worse at higher angle might mean that the detectors angles are tilted compared to the
surveyors measurments... I also tried to use the CAO positions for MG3 (see Detectors/mugast.detector) but it didn't solve the problem.


By looking at the EnergyCalibrator.C, I saw that the extrapolation used to find the dead layer of aluminum is commented... This might be an answer to the problem...
Ongoinge775s Sat Feb 29 17:18:02 2020 mugast ; Valerian, Marlène , PierreOFFLINESoftwareCD2-AuIssues in the EX reconstruction

mugast ; Valerian wrote:

mugast ; Valerian wrote:

mugast ; valerian irene wrote:
We are currently working on the EX reconstruction which is off for detector 1, 3 and 7... One of our guess might be that there is an issue with the position of the detectors which gives error in theta... As the energy calibration is ok. Joined are :

- the Excitation energy for good detectors compared to all detectors,
- the theoretical kinematic line for the 3.57 state for each detectors,
- the excitation energy as a function of the laboratory angle (for this last spectra a clear dependency is visible for telescopes 1,3 and 7)
- the 3D view, which doesn't show clear problem in position...


Testing changing the beam impact :

-Kine for beam impact = (0,0,0) for reference
-Kine for beam impact = (0,0,-5mm)
-Kine for beam impact = (0,0,+5mm)

-Ex vs ThetaLab for beam impact = (0,0,0) for reference
-Ex vs ThetaLab for beam impact = (0,0,-5mm)
-Ex vs ThetaLab for beam impact = (0,0,+5mm)
Red lines corresponds to Ex = 3.570 MeV and Ex = 4.070 MeV

To try to correct the problem i also offseted the beam in X and Y, but this doesn't solve the problem quite the opposite as it make the result worse for other detectors also... The fact that it is better at lower angle and worse at higher angle might mean that the detectors angles are tilted compared to the
surveyors measurments... I also tried to use the CAO positions for MG3 (see Detectors/mugast.detector) but it didn't solve the problem.


By looking at the EnergyCalibrator.C, I saw that the extrapolation used to find the dead layer of aluminum is commented... This might be an answer to the problem...


WE HAVE A SOLUTION!!! It seems that the order of the detectors in mugast.detector matters!! Now we start with a dummy MG8, then MG1, MG2, ... And everything is fine!! We think that is because it was taking the wrong calibration files (to be checked).

Joined are (for run 128,137,138,139,151,152,153):
-Ex vs ThetaLab
-Ex for all detectors
-Egamma vs Ex
Fixede793s Fri Mar 5 16:16:28 2021 mugastDAQSoftware Pedestals for E793S
Pedestal alignment was performed for Energy qnd Time for all cards with cooling at 0°.
The values are attached here and saved in : cd ganacq_manip/mugast/das-save/pedestal-save/pedestal_5march2021_15h54/
Pendinge793s Sat Mar 13 05:11:08 2021 Faïrouz, Juan LoisAGATASoftwareCD2Problem with AGATA acquisition started at 4h40

There is a problem with AGATA Acquisition. All the bars in the global status & control are yellow!!!
This problem started at 4h40.
We called the acquisition astreinte but he doesn't know how to fix the problem.
We called Emmanuel but we got on the answering machine. We will call him again at 7 am.

The count rate of AGATA in the scaler looks ok: ~ 1200
Fixede793s Sat Mar 13 07:29:56 2021 Faïrouz, Juan LoisAGATASoftwareCD2AGATA acquisition problem fixed by Emmanuel
FixedTEST Fri Dec 14 09:08:46 2018 Adrien MattaOTHERSocialN/ATest Post
MUGAST ELOG is now ONLINE
Fixede744s Wed Apr 10 06:45:53 2019 FrançoisBEAMSocial No beam yet
They are about to align the beam from the cyclotron CIME to the area G1.
Fixede744s Wed Apr 10 07:15:22 2019 FrançoisBEAMSocial no beam

They aligned the beam up to G1 Pr31 but the stripper foil was not inserted, so they have to start the work again.
Fixede744s Wed Apr 10 09:05:33 2019 FrançoisBEAMSocial No beam in G1 yet.
The beam is at the entrance of G1.
Fixede744s Wed Apr 10 10:15:38 2019 FrançoisBEAMSocial Beam stopped
A problem with the cooling water.
2 UT at least to restart the accelerator
Fixede744s Wed Apr 10 11:46:04 2019 iulianBEAMSocialN/AE744 Status
Refrigeration problem, beam stopped.
Optimistic estimation: they need 1 UT to setup CIME => 14N could be back around 20h.
Optimistic scenario still on track
Then need 2 UT's to set CSS1 -> 14O next day around 18h
No shifts until 24h00 (midnight) today.
Shift to resume 11/04 @00h00


18h30 14N beam is back.
Fixede744s Thu Apr 11 07:30:15 2019 mugastBEAMSocialN/ABeam stop
The beam will be back around 8:30.
Fixede744s Sat Apr 13 08:02:42 2019 Marion, Kseniia, Dieter, Didier, GuiseppeONLINESocialCH2Screenshots run 122 v1
Fixede744s Sat Apr 13 11:42:16 2019 Marion, Kseniia, Dieter, Didier, Guiseppe, IuliaHONLINESocialCH2Screenshots - spectra zeroed after previous screenshot taking
spectra zeroed after previous screenshot taking
Fixede744s Sat Apr 13 12:04:56 2019 Adrien MattaOTHERSocialN/ABlow some steam this afternoon (staturday13th)
If some of you want to blow some steam this afternoon Crossfit Caen ( located in Mondeville, you. Can find them on google map) is organising an initiation Wod this afternoon for their 4th anniversary. Here is the program of the day:
Les quatre ans de la boxe commence, on vous attend avec impatience et venez nombreux. 👌🏻💪🏻 Si vous voulez découvrir le Crossfit ou tout simplement essayer (wod initiation),ça sera avec plaisir ! 😁

4 ans de CrossfitCaen 🔥🔥
Le déroulement de la journée !! 👌🏻

10h-13h Floater 1
10h30 Wod Team 20 personnes
11h15 Initiation
12h Wod Élite
12h30 Wod Indiv
13h30 Kids
14h-17h Floater 2
14h15 Wod Indiv
15h Initiation
15h45 Wod Élite
16h30 Wod Team
20h Tenue Correcte Exigée
Fixede744s Sat Apr 13 13:13:18 2019 Marion, Kseniia, Dieter, Didier, Guiseppe, IuliaHONLINESocialCH2Screenshots run 122 v5
Fixede744s Sat Apr 13 15:31:32 2019 Marion, Kseniia, Dieter, Didier, Guiseppe, IuliaH, CatalinONLINESocialCH2Screenshots run 122.1 v6
"1_" added to screenshot filenames to indicate automatic file opening
Fixede744s Sun Apr 14 03:30:28 2019 Iulian IuliaOFFLINESocialCH2first calculation ECM ThetaCM for gamma-P emission
Fixede744s Sun Apr 14 08:41:52 2019 Nikola Dieter Nicolas Beyhan KseniiaONLINESocialCH2Log & Scalers -- Run 122 at 08:40
Spectra zeroed after screenshot taking