Fixed, TEST, Mon May 27 16:17:48 2019, OttoRun, RUN, General, 3-alpha, Run229
|
*************************
[SIZE=3][B]Run : 229[/B][/SIZE]
*************************
|
Fixed, TEST, Tue May 28 10:43:58 2019, OttoRun, RUN, General, 3-alpha, Run229
|
*************************
[SIZE=3][B]Run : 229[/B][/SIZE]
*************************
|
Fixed, TEST, Tue May 28 11:08:10 2019, mugast, MUST2, Hardware, 3-alpha, MG3 is tripping
|
MG3T is tripping, tryied to set max current to 20uA but it still trip around 30V. |
Fixed, TEST, Thu May 30 15:42:13 2019, OttoRun, RUN, General, 3-alpha, Run229
|
*************************
[SIZE=3][B]Run : 229[/B][/SIZE]
*************************
|
Fixed, TEST, Thu May 30 15:55:15 2019, OttoRun, RUN, General, 3-alpha, Run229
|
*************************
[SIZE=3][B]Run : 229[/B][/SIZE]
*************************
|
Fixed, TEST, Thu May 30 15:58:20 2019, OttoRun, RUN, General, 3-alpha, Run230
|
*************************
[SIZE=3][B]Run : 230[/B][/SIZE]
*************************
|
Fixed, TEST, Thu May 30 16:04:09 2019, OttoRun, RUN, General, 3-alpha, Run230
|
*************************
[SIZE=3][B]Run : 230[/B][/SIZE]
*************************
|
Fixed, TEST, Thu May 30 16:08:39 2019, OttoRun, RUN, General, 3-alpha, Run230
|
*************************
[SIZE=3][B]Run : 230[/B][/SIZE]
*************************
|
Fixed, TEST, Thu May 30 16:10:18 2019, OttoRun, RUN, General, 3-alpha, Run230  
|
*************************
[SIZE=3][B]Run : 230[/B][/SIZE]
*************************
|
Fixed, e786s, Fri May 31 09:53:14 2019, Marlène, GRIT, General, , Alpha source --> status of detectors
|
MG3T has been fixed by changing the X kapton.
I have removed aluminium foil on square detectors, the noise is slightly higher.
|
Fixed, e786s, Mon Jun 3 18:00:10 2019, Marlène, GRIT, General, , MUGAST detectors status --> MG4T
|
The problem of MG4T ha sbeen solved by changing the feedthrough.
The pulser on Y was not working and we were seeing no events (no ORD) because there was a problem with the bias of board. The issue is due to the feedthrough.
So now MG4T (HT &BT) are plugged in a free feedthrough (below MG1T) and I see pulser signals on Y.
|
Fixed, TEST, Tue Jun 11 18:59:36 2019, OttoRun, RUN, General, 3-alpha, Run231
|
*************************
[SIZE=3][B]Run : 231[/B][/SIZE]
*************************
|
Fixed, TEST, Tue Jun 11 21:44:38 2019, OttoRun, RUN, General, 3-alpha, Run232
|
*************************
[SIZE=3][B]Run : 232[/B][/SIZE]
*************************
|
Fixed, e786s, Wed Jun 12 17:59:01 2019, mugast, OTHER, Software, , Online Energy Calibration 
|
Updated Online Energy calibration for MUGAST, based on run_0231 data |
Fixed, TEST, Thu Jun 13 09:46:40 2019, OttoRun, RUN, General, 3-alpha, Run234
|
*************************
[SIZE=3][B]Run : 234[/B][/SIZE]
*************************
|
Fixed, TEST, Thu Jun 13 09:47:39 2019, OttoRun, RUN, General, 3-alpha, Run235
|
*************************
[SIZE=3][B]Run : 235[/B][/SIZE]
*************************
|
Fixed, e786s, Wed Jun 19 10:09:10 2019, mugast, MUST2, Hardware, N/A, Time calibrator cabled
|
range 640 ns
period 20 ns |
Fixed, e786s, Wed Jun 19 12:02:07 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun
|
*** Run Time Calibrator FSC ***
- Run: 237
- MUVI: MUVI1
|
Fixed, e786s, Wed Jun 19 12:24:43 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun
|
*** Run Time Calibrator FSC ***
- Run: 238
- MUVI: MUVI1
|
Fixed, e786s, Wed Jun 19 12:48:13 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun
|
*** Run Time Calibrator FSC ***
- Run: 239 - MUVI: MUVI2
- Side : MUFEE X (Junction)
|
|