FixedTEST Fri Feb 22 14:36:35 2019 iulianMUST2Hardware3-alphatest new cable for Must2 detectors. Cables OK!Cables put on Muvi3.
Testing procedure. Use the numbered cables
to conect the annular detector (MG11) and
  
FixedTEST Fri Feb 22 15:54:30 2019 iulian GRITHardware3-alpharun 40 3-alpha night run [CENTER][/CENTER][U][B]Run 40
[/B][/U] 
 Alpha source run for trapezoid detectors
 run40_3alpha_mg1345.pngrun40_3alpha_mg79.png 
PinnedTEST Tue Feb 26 09:30:21 2019 FreddyOTHERHardwareN/ACabling - MUVI / HT-BT / SA-SLCabling scheme - to update until final  Cabling_2019_03_01.pdf 
FixedTEST Tue Feb 26 13:21:32 2019 iulian freddy GRITSoftwareN/Arun 42 Calimero run MUVI 1 on X sectors[U][B]run #42
[/B][/U]Calimero MUVI1 X
  
FixedTEST Tue Feb 26 13:50:36 2019 iulian freddy GRITSoftwareN/Arun 43 Calimero run MUVI 1 on Y sectors[U][B]run #43
[/B][/U]Calimero MUVI1 Y
  
FixedTEST Tue Feb 26 14:52:49 2019 iulian freddy GRITSoftwareN/Arun 44 Calimero run MUVI 2 on X sectors (bad run)[U][B]run #44[/B][/U] bad run
Calimero MUVI2 X
 run44_error_calimero.png 
FixedTEST Fri Mar 1 12:07:44 2019 FreddyGRITGeneral3-alphaEfficiency reconstructionblue: reconstructed efficiency using alpha
run 40 (arbitratily normalized)
red: simulated efficiency
 comparison.pdfcomparison.Calpha_eff.root 
FixedTEST Mon Mar 4 12:28:26 2019 Valerian and IulianMUST2General3-alphaNo DECS and acq error run 47-48-49. Das Saved after problem fixedFixed. Solution:
1) CENTRUM not configured on GENTRUM Master
2) GMT matrix on GMT3 (2nd slave) was all
 run49_error_acq.png 
FixedTEST Mon Mar 4 13:54:24 2019 iulian & ValerianGRITHardware3-alpharun 50 3-alpha short run[U][B]RUN 50[/B][/U]
First 3a test with  all Must2 detectors.
HV on all set to 50 V for this test only
 run50_3a_Must.pngrun50_3a_Must2.png 
FixedTEST Mon Mar 4 14:34:31 2019 Valerian, Marlène, IulianMUST2GeneralN/AMUST2 positions (based on e755)Based on experiment e755, the MUST2 config
is the following :
 Screenshot_2019-03-04_at_14.39.02.png 
FixedTEST Mon Mar 4 15:30:31 2019 iulian Marlene & ValerianMUST2Hardware3-alpharun 51 3-alpha run with Pedestals Aligned @ 13 deg Measured on cooling system [U][B]RUN 51[/B][/U]
3a run with Pedestals Aligned[I][U] but not[/U][/I]
suppressed 
 run51_HVs.pngrun51_3a_Must3_pedestalAligned.png 
FixedTEST Mon Mar 4 16:24:51 2019 valerian marlèneGRITGeneral RUN52 3 alpha source 4MUST2 no pedestalRUN52 
3 alpha source run with only 4MUST2 (in MUVI3)
with pedestal suppression at 13 deg.
 HV_MUST2_during_run52.pngHV_MUST2_afternightrun52.png3alpha_source_4MUST2_nightrun52.png 
Fixed  Mon Mar 4 17:22:29 2019 MarleneGRITGeneral Temperature on MUFEE MUST2 13degTemperature given by DAS on MUVI3 (MUST2)

screenshots before run
 8x 
FixedTEST Tue Mar 5 10:16:03 2019 iulianMUST2SoftwareN/Arun 57 Gamme 300 ns Time Calibration run (Calimero) on MUVI 3 (MUST) X sectors more stats (10000)RUNS 53,54,55,56 are bad runs

[U][B]Run 57[/B][/U]
 run57_calimero_config.pngrun57_calimero_mm1.png 
FixedTEST Tue Mar 5 12:13:41 2019 iulian MarleneMUST2SoftwareN/Arun 60 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X RUNS 58,59,60 are bad runs

[U][B]Run 61[/B][/U]
 run60_calimero_mm1X.pngrun60_calimero_mm1X_spectra.png 
FixedTEST Tue Mar 5 15:53:04 2019 iulian MarleneMUST2SoftwareN/ACalimero problemsproblems with Calimero solved by killing some
processes on lpcgrit related with Narval.
to find the processes on lpcgrit:
  
FixedTEST Tue Mar 5 16:20:11 2019 iulian MarleneMUST2SoftwareN/Arun 77 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X . good run[U][B]Run 76[/B][/U]

 run77_calimero_mm1X_spectra.png 
FixedTEST Tue Mar 5 16:42:56 2019 iulian MarleneMUST2SoftwareN/Arun 79 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM2 X . good run[U][B]Run 79[/B][/U]

Run 78 bad
 run79_calimero_mm2X_spectra.png 
FixedTEST Tue Mar 5 17:11:29 2019 iulian MarleneMUST2SoftwareN/Arun 81 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM3 X . good run[U][B]Run 81[/B][/U]

Run 80 bad
 run81_calimero_mm3X_spectra.png 
FixedTEST Tue Mar 5 17:48:38 2019 iulian MarleneMUST2SoftwareN/Arun 82 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM4 X . good run[U][B]Run 82[/B][/U]

Calimero on Muvi3 MM4(X)
 run82_calimero_mm4X_spectra.png 

ELOG Home Page