MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 2 of 127  Not logged in MUGAST
Fixed, TEST, Fri Feb 22 14:36:35 2019, iulian, MUST2, Hardware, 3-alpha, test new cable for Must2 detectors. Cables OK! 
Cables put on Muvi3.
Testing procedure. Use the numbered cables to conect the annular detector (MG11) and check if ok: GECO writting on MUVI3 , ORD Present, DECS present, STOP
present.
Fixed, TEST, Fri Feb 22 15:54:30 2019, iulian , GRIT, Hardware, 3-alpha, run 40 3-alpha night run  run40_3alpha_mg1345.pngrun40_3alpha_mg79.png
[CENTER][/CENTER][U][B]Run 40
[/B][/U] 
 Alpha source run for trapezoid detectors during the night with pedestals aligned (with HV off) and zero suppression. Cooling at 15°C.
Pinned, TEST, Tue Feb 26 09:30:21 2019, Freddy, OTHER, Hardware, N/A, Cabling - MUVI / HT-BT / SA-SL Cabling_2019_03_01.pdf
Cabling scheme - to update until final
Fixed, TEST, Tue Feb 26 13:21:32 2019, iulian freddy , GRIT, Software, N/A, run 42 Calimero run MUVI 1 on X sectors 
[U][B]run #42
[/B][/U]Calimero MUVI1 X
Fixed, TEST, Tue Feb 26 13:50:36 2019, iulian freddy , GRIT, Software, N/A, run 43 Calimero run MUVI 1 on Y sectors 
[U][B]run #43
[/B][/U]Calimero MUVI1 Y
Fixed, TEST, Tue Feb 26 14:52:49 2019, iulian freddy , GRIT, Software, N/A, run 44 Calimero run MUVI 2 on X sectors (bad run) run44_error_calimero.png
[U][B]run #44[/B][/U] bad run
Calimero MUVI2 X
Fixed, TEST, Fri Mar 1 12:07:44 2019, Freddy, GRIT, General, 3-alpha, Efficiency reconstruction comparison.pdfcomparison.Calpha_eff.root
blue: reconstructed efficiency using alpha run 40 (arbitratily normalized)
red: simulated efficiency
Fixed, TEST, Mon Mar 4 12:28:26 2019, Valerian and Iulian, MUST2, General, 3-alpha, No DECS and acq error run 47-48-49. Das Saved after problem fixed run49_error_acq.png
Fixed. Solution:
1) CENTRUM not configured on GENTRUM Master
2) GMT matrix on GMT3 (2nd slave) was all 0
Fixed, TEST, Mon Mar 4 13:54:24 2019, iulian & Valerian, GRIT, Hardware, 3-alpha, run 50 3-alpha short run run50_3a_Must.pngrun50_3a_Must2.png
[U][B]RUN 50[/B][/U]
First 3a test with  all Must2 detectors. HV on all set to 50 V for this test only to avoid any risks
Fixed, TEST, Mon Mar 4 14:34:31 2019, Valerian, Marlène, Iulian, MUST2, General, N/A, MUST2 positions (based on e755) Screenshot_2019-03-04_at_14.39.02.png
Based on experiment e755, the MUST2 config is the following :

Fixed, TEST, Mon Mar 4 15:30:31 2019, iulian Marlene & Valerian, MUST2, Hardware, 3-alpha, run 51 3-alpha run with Pedestals Aligned @ 13 deg Measured on cooling system  run51_HVs.pngrun51_3a_Must3_pedestalAligned.png
[U][B]RUN 51[/B][/U]
3a run with Pedestals Aligned[I][U] but not[/U][/I] suppressed 
Fixed, TEST, Mon Mar 4 16:24:51 2019, valerian marlène, GRIT, General, , RUN52 3 alpha source 4MUST2 no pedestal HV_MUST2_during_run52.pngHV_MUST2_afternightrun52.png3alpha_source_4MUST2_nightrun52.png
RUN52 
3 alpha source run with only 4MUST2 (in MUVI3) with pedestal suppression at 13 deg.
Fixed, , Mon Mar 4 17:22:29 2019, Marlene, GRIT, General, , Temperature on MUFEE MUST2 13deg 8x
Temperature given by DAS on MUVI3 (MUST2) 
screenshots before run
after run : same values 
Fixed, TEST, Tue Mar 5 10:16:03 2019, iulian, MUST2, Software, N/A, run 57 Gamme 300 ns Time Calibration run (Calimero) on MUVI 3 (MUST) X sectors more stats (10000) run57_calimero_config.pngrun57_calimero_mm1.png
RUNS 53,54,55,56 are bad runs

[U][B]Run 57[/B][/U]
Fixed, TEST, Tue Mar 5 12:13:41 2019, iulian Marlene, MUST2, Software, N/A, run 60 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X  run60_calimero_mm1X.pngrun60_calimero_mm1X_spectra.png
RUNS 58,59,60 are bad runs

[U][B]Run 61[/B][/U]
Fixed, TEST, Tue Mar 5 15:53:04 2019, iulian Marlene, MUST2, Software, N/A, Calimero problems 
problems with Calimero solved by killing some processes on lpcgrit related with Narval.
to find the processes on lpcgrit:
ps aux | grep Cali*
Fixed, TEST, Tue Mar 5 16:20:11 2019, iulian Marlene, MUST2, Software, N/A, run 77 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X . good run run77_calimero_mm1X_spectra.png
[U][B]Run 76[/B][/U]

Fixed, TEST, Tue Mar 5 16:42:56 2019, iulian Marlene, MUST2, Software, N/A, run 79 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM2 X . good run run79_calimero_mm2X_spectra.png
[U][B]Run 79[/B][/U]

Run 78 bad
Fixed, TEST, Tue Mar 5 17:11:29 2019, iulian Marlene, MUST2, Software, N/A, run 81 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM3 X . good run run81_calimero_mm3X_spectra.png
[U][B]Run 81[/B][/U]

Run 80 bad
Fixed, TEST, Tue Mar 5 17:48:38 2019, iulian Marlene, MUST2, Software, N/A, run 82 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM4 X . good run run82_calimero_mm4X_spectra.png
[U][B]Run 82[/B][/U]

Calimero on Muvi3 MM4(X)

ELOG Home Page