MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 1 of 132  Not logged in MUGAST
New entries since:Thu Jan 1 01:00:00 1970
Status ExpNbr Date Author Type Categorydown Target-Source Subject
FixedTEST Tue Feb 26 13:21:32 2019 iulian freddy GRITSoftwareN/Arun 42 Calimero run MUVI 1 on X sectors
run #42
Calimero MUVI1 X


run #41
Calimero MUVI secttor 1X
FixedTEST Tue Feb 26 13:50:36 2019 iulian freddy GRITSoftwareN/Arun 43 Calimero run MUVI 1 on Y sectors
run #43
Calimero MUVI1 Y
FixedTEST Tue Feb 26 14:52:49 2019 iulian freddy GRITSoftwareN/Arun 44 Calimero run MUVI 2 on X sectors (bad run)
run #44 bad run
Calimero MUVI2 X

During the run we noticed that time calibrator is on MUVI1 only .
Attachment 1: run44_error_calimero.png
run44_error_calimero.png
FixedTEST Tue Mar 5 10:16:03 2019 iulianMUST2SoftwareN/Arun 57 Gamme 300 ns Time Calibration run (Calimero) on MUVI 3 (MUST) X sectors more stats (10000)
RUNS 53,54,55,56 are bad runs

Run 57


Calimero on Muvi3 MM1(X)

Gamme 300 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
Attachment 1: run57_calimero_config.png
run57_calimero_config.png
Attachment 2: run57_calimero_mm1.png
run57_calimero_mm1.png
FixedTEST Tue Mar 5 12:13:41 2019 iulian MarleneMUST2SoftwareN/Arun 60 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X
RUNS 58,59,60 are bad runs

Run 61


Calimero on Muvi3 MM1(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
Attachment 1: run60_calimero_mm1X.png
run60_calimero_mm1X.png
Attachment 2: run60_calimero_mm1X_spectra.png
run60_calimero_mm1X_spectra.png
FixedTEST Tue Mar 5 15:53:04 2019 iulian MarleneMUST2SoftwareN/ACalimero problems
problems with Calimero solved by killing some processes on lpcgrit related with Narval.
to find the processes on lpcgrit:
ps aux | grep Cali*
ps aux | grep Narval

kill -9 all the ids

Other possible solution:

start_acq on lpcgrit
tk followed by kall
exit
FixedTEST Tue Mar 5 16:20:11 2019 iulian MarleneMUST2SoftwareN/Arun 77 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X . good run
Run 76


Calimero on Muvi3 MM1(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
Attachment 1: run77_calimero_mm1X_spectra.png
run77_calimero_mm1X_spectra.png
FixedTEST Tue Mar 5 16:42:56 2019 iulian MarleneMUST2SoftwareN/Arun 79 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM2 X . good run
Run 79

Run 78 bad
Calimero on Muvi3 MM2(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
Attachment 1: run79_calimero_mm2X_spectra.png
run79_calimero_mm2X_spectra.png
FixedTEST Tue Mar 5 17:11:29 2019 iulian MarleneMUST2SoftwareN/Arun 81 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM3 X . good run
Run 81

Run 80 bad
Calimero on Muvi3 MM3(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
Attachment 1: run81_calimero_mm3X_spectra.png
run81_calimero_mm3X_spectra.png
FixedTEST Tue Mar 5 17:48:38 2019 iulian MarleneMUST2SoftwareN/Arun 82 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM4 X . good run
Run 82

Calimero on Muvi3 MM4(X)

Gamme 600 ns
Attachment 1: run82_calimero_mm4X_spectra.png
run82_calimero_mm4X_spectra.png
FixedTEST Tue Mar 5 18:03:24 2019 iulian MarleneMUST2SoftwareN/ACalimero problem on Y side
Calimero not working properly on Y side of any mm detector
Attachment 1: run83_calimero_mm4Y_spectra.png
run83_calimero_mm4Y_spectra.png
Attachment 2: run834calimero_mm2Y_spectra.png
run834calimero_mm2Y_spectra.png
FixedTEST Thu Mar 21 13:28:06 2019 Valerian and AdrienDAQSoftwareN/AFirst fully functionnal time calibration run on X and Y of MG1
Our program is able to perform the time calibration of our detector for both X and Y, see picture attached.
Attachment 1: CALIMEROY1.png
CALIMEROY1.png
Attachment 2: FIRST_TIMECALIBRATION_WITHOUT_CALIMERO.png
FIRST_TIMECALIBRATION_WITHOUT_CALIMERO.png
FixedTEST Sat Mar 23 15:26:06 2019 Adrien MattaOTHERSoftwareN/ATo do list
FSC:

  • Autoconf programme to set all regular parameter (PAC gain, TAC range, ORD, no mask, no pulser)


Sentinel:
  • Making the program
  • Preparing start script
  • Preparing stop script
  • Preparing init script


Online/offline:
  • Look into ZMQ to replace GRU
  • Datum through zmq
  • Switch NPSpectraServer/Client to ZMQ and datum
  • CsI Calibration with db scan
  • 3-alpha calibration using NPCalibration

HV:
  • Remote read
  • Auto logging
  • Alert on trip
  • Bt/HV/sector on/off
FixedTEST Thu Mar 28 15:39:01 2019 ValerianOFFLINESoftwareN/AStatus of the converion with VAMOS program
The conversion with VAMOS program was validated using run_0020 from com2019


Program output as a ref :

T1 : 9978458
T2 : 9978458
checking multiplicities
Reading: 99.9152 %MULTIPLICITIES OK!
checking values
Reading: 99.9152 %%
DETECTOR NUMBERS ARE OK!!!!!
STRIP NUMBERS ARE OK!!!!!
ENERGY VALUES ARE OK!!!!!
TIME VALUES ARE OK!!!!!
Fixedcom2019 Wed Apr 3 11:05:36 2019 Adrien MattaDAQSoftwareN/AMFM scaler reading
MfM scaler are read correctly for Must2 U2M but output some -1 for the numexo2 integrated scaler.
Fixedcom2019 Wed Apr 3 16:24:37 2019 AdrienCATSSoftwareN/ACalib file from T.Roger
Here are the calibration file provided by T. Roger.
Attachment 1: CATS2_Y.vamos2019
0.00000e+00 0.00000e+00 0.00000e+00 
2.71193e+02 3.51955e+01 9.20106e-01 
2.54820e+02 3.85477e+01 1.02407e+00 
2.70765e+02 9.63413e+01 1.01270e+00 
2.38067e+02 -1.94775e+02 2.02755e+00 
2.70950e+02 2.18371e+01 9.74726e-01 
2.71165e+02 1.20869e+02 1.07928e+00 
9.13653e+00 3.16800e+02 1.00754e+00 
5.66055e+02 -2.33425e+02 9.05174e-01 
2.87582e+02 4.24426e+01 9.10918e-01 
2.71272e+02 3.51955e+01 9.20106e-01 
2.87581e+02 1.47364e+01 9.01484e-01 
2.87546e+02 6.66734e+00 9.93090e-01 
2.05666e+02 1.43008e+02 9.85243e-01 
3.03492e+02 0.00000e+00 1.00000e+00 
3.03819e+02 -6.17854e+00 9.23957e-01 
7.46067e+01 2.16790e+02 8.97818e-01 
4.18173e+02 -1.88169e+01 8.82782e-01 
2.70940e+02 6.14310e+01 8.89892e-01 
2.38432e+02 9.12262e+01 8.87560e-01 
2.38441e+02 7.85597e+01 8.77800e-01 
2.38442e+02 8.48885e+01 8.70873e-01 
2.22035e+02 1.14424e+02 8.95446e-01 
2.54389e+02 8.12120e+01 8.90312e-01 
2.21670e+02 1.59730e+02 9.33857e-01 
2.70937e+02 7.20531e+01 8.66403e-01 
5.82149e+01 2.32593e+02 8.43761e-01 
2.87581e+02 -1.28631e+02 2.74796e+00 
Attachment 2: CATS2_X.vamos2019
0.00000e+00 0.00000e+00 0.00000e+00 
0.00000e+00 0.00000e+00 0.00000e+00 
3.44896e+02 5.63680e+01 9.41419e-01 
3.44871e+02 2.74924e+01 8.93518e-01 
3.12042e+02 2.41995e+02 9.38371e-01 
2.79148e+02 3.23114e+02 1.10331e+00 
1.48358e+02 3.41540e+02 9.18807e-01 
1.80540e+02 4.53841e+02 9.26752e-01 
4.10458e+02 -5.02746e+01 9.52866e-01 
3.44871e+02 -2.17129e+02 1.19034e+00 
3.77669e+02 3.74782e+01 9.33222e-01 
3.77669e+02 8.51246e+01 8.96266e-01 
3.77644e+02 1.11250e+02 9.87395e-01 
3.44871e+02 -1.76949e+01 9.49754e-01 
3.77669e+02 0.00000e+00 1.00000e+00 
3.77663e+02 7.06397e+01 9.18807e-01 
2.13851e+02 1.68237e+02 9.52866e-01 
3.11592e+02 2.51688e+02 9.15915e-01 
2.79820e+02 1.11250e+02 9.87395e-01 
2.46601e+02 1.27417e+02 1.03243e+00 
2.79372e+02 1.46788e+02 8.79335e-01 
2.79380e+02 2.32394e+02 8.93518e-01 
3.12042e+02 1.81364e+02 9.23798e-01 
2.79366e+02 1.00739e+02 9.18807e-01 
2.46616e+02 1.03641e+02 9.03826e-01 
2.46587e+02 1.39501e+02 9.36753e-01 
3.44871e+02 2.12105e+02 9.11509e-01 
2.79146e+02 1.51065e+02 2.75169e+00 
Attachment 3: CATS1_Y.vamos2019
0.00000e+00 0.00000e+00 0.00000e+00 
0.00000e+00 0.00000e+00 0.00000e+00 
1.13931e+03 -3.52970e+02 9.68936e-01 
8.44459e+02 -1.44555e+02 1.07619e+00 
4.01828e+02 2.94406e+02 8.99414e-01 
3.53072e+02 3.83002e+02 1.07128e+00 
4.18602e+02 2.07103e+02 9.97538e-01 
4.18604e+02 2.22703e+02 9.35567e-01 
3.20322e+02 4.63142e+02 1.12991e+00 
3.69469e+02 3.68467e+02 9.91268e-01 
3.69466e+02 3.40471e+02 1.00655e+00 
3.69533e+02 3.23507e+02 9.85421e-01 
3.20391e+02 4.16747e+02 1.00459e+00 
3.03942e+02 3.77691e+02 9.26262e-01 
7.29833e+02 -1.81899e-12 1.00000e+00 
3.36702e+02 4.72202e+02 1.07558e+00 
3.20323e+02 5.25219e+02 1.01355e+00 
2.71183e+02 3.20537e+02 8.57953e-01 
3.20324e+02 4.53551e+02 9.49811e-01 
2.38494e+02 5.25479e+02 9.09205e-01 
3.20349e+02 4.37876e+02 8.58610e-01 
4.18602e+02 3.18712e+02 9.32195e-01 
3.85855e+02 2.52608e+02 9.56052e-01 
4.34843e+02 2.54458e+02 9.89716e-01 
4.50450e+02 1.87712e+02 8.76905e-01 
3.20254e+02 8.14065e+02 1.60565e+00 
3.53085e+02 2.57256e+02 9.52991e-01 
0.00000e+00 0.00000e+00 0.00000e+00 
Attachment 4: CATS1_X.vamos2019
0.00000e+00 0.00000e+00 0.00000e+00 
0.00000e+00 0.00000e+00 0.00000e+00 
1.40161e+02 9.92071e+01 9.76396e-01 
1.89289e+02 -2.69023e+00 1.68488e+00 
3.20343e+02 2.87366e+02 9.03085e-01 
1.56522e+02 6.72206e+01 9.76396e-01 
1.07402e+02 4.08998e+01 8.72104e-01 
1.40150e+02 8.09504e+01 1.02684e+00 
1.89310e+02 -5.35298e+01 1.48523e+00 
3.20322e+02 -1.42885e+02 9.20481e-01 
2.22045e+02 -2.12294e+01 1.00833e+00 
1.23762e+02 4.60139e+01 8.42356e-01 
1.40150e+02 1.46407e+02 1.09613e+00 
2.70859e+02 -2.25241e+02 1.59755e+00 
2.05264e+02 -9.09495e-13 1.00000e+00 
2.71182e+02 -3.42651e+01 8.87728e-01 
3.36702e+02 1.07801e+02 8.72316e-01 
3.36720e+02 1.24392e+02 9.02291e-01 
3.03966e+02 1.23800e+02 9.90423e-01 
2.87562e+02 1.39729e+02 9.89467e-01 
3.36702e+02 1.30358e+02 8.86667e-01 
3.03668e+02 1.68313e+02 8.83800e-01 
3.20254e+02 1.38346e+02 7.81607e-01 
3.20392e+02 1.41100e+02 7.62264e-01 
3.69438e+02 -1.65316e+02 1.32820e+00 
3.53090e+02 8.71354e+01 7.81607e-01 
3.36451e+02 9.07214e+01 7.77743e-01 
2.87388e+02 1.41433e+02 6.18790e-01 
Fixedcom2019 Sat Apr 6 02:48:39 2019 Iulian RuxandraRUNSoftware Online Spectra run 256 @2h30 am 06/04 (for reference)
The online spectra (for reference)
No timing on Mg3 & Mg4 side X. Those correspond with the masked channels on the Muvi1
Attachment 1: Run_256_Impact_Matrix.png
Run_256_Impact_Matrix.png
Attachment 2: Run_256_Must_X_side.png
Run_256_Must_X_side.png
Attachment 3: Run_256_TACs.png
Run_256_TACs.png
Attachment 4: Run_256_Must_Y_side.png
Run_256_Must_Y_side.png
Attachment 5: Run_256_Must_X_side.png
Run_256_Must_X_side.png
Attachment 6: Run_256_Impact_Matrix.png
Run_256_Impact_Matrix.png
Attachment 7: Run_256_Vamos_1.png
Run_256_Vamos_1.png
Attachment 8: Run_256_Vamos_2.png
Run_256_Vamos_2.png
Attachment 9: Run_256_Vamos_3.png
Run_256_Vamos_3.png
Attachment 10: Run_256_Vamos_4.png
Run_256_Vamos_4.png
Attachment 11: Run_256_Vamos_5.png
Run_256_Vamos_5.png
Fixedcom2019 Sat Apr 6 02:52:51 2019 Iulian RuxandraRUNSoftware Online Scalers run 256 @2h30 am 06/04 for reference
Attachment 1: Run_256_Scaler1.png
Run_256_Scaler1.png
Attachment 2: Run_256_Scaler2.png
Run_256_Scaler2.png
Attachment 3: Run_256_Scaler3.png
Run_256_Scaler3.png
Attachment 4: Run_256_Scaler4.png
Run_256_Scaler4.png
Fixedcom2019 Sat Apr 6 03:22:30 2019 Iulian RuxandraRUNSoftware New Kinematics plots (4h of stats)
More statistics on the script written by Adrian and Valerian (first image)

Second image is a comparison between the excited states produced between 90-130 deg and 130-180 deg. Effects of angular distribution?


the kinematic line seen in the Elab vs Theta at backward angles can be observed in figure 3. Are those protons?
Attachment 1: Kine256_0-256_3.pdf
Kine256_0-256_3.pdf
Attachment 2: Kine256_0-256_3_comparison.pdf
Kine256_0-256_3_comparison.pdf
Attachment 3: Elab_vs_ThetaLab_backward.pdf
Elab_vs_ThetaLab_backward.pdf
Fixedcom2019 Sat Apr 6 06:24:04 2019 mugastRUNSoftwareCD2new data 5more statistics°
Attachment 1: Run_256_kin.png
Run_256_kin.png

ELOG Home Page