MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 1 of 132  Not logged in MUGAST
Status ExpNbr Date Author Type Category Target-Source Subject Text Attachments
FixedTEST Fri Dec 14 09:08:46 2018 Adrien MattaOTHERSocialN/ATest PostMUGAST ELOG is now ONLINE   
FixedTEST Wed Jan 16 16:33:28 2019 MarlèneGRITGeneral3-alphaStatus of detectors on 16/01/19MG1 has some issue with the HT cable (for
the BT bias). There is a shortcut only under
vacuum. We checked the connector but we have
 Detectors_status_160119.xlsxCapture_d’eÌcran_2019-01-16_aÌ€_17.05.57.png 
FixedTEST Wed Jan 16 17:36:30 2019 Valérian-MarlèneGRITGeneralN/ACalimero tests - How to configure Calimero in GECOTests of Calimero time calibration with MUVI1
channel 2 (MM2)
 Calimero_MUVI1_MM2_voie3_allmates_bidim.pngCalimero_MUVI1_MM2_voie1_allmates_bidim.pngCalimero_MUVI1_MM3_voie1_allmates_bidim_withchannelsenabled.pngCalimero_MUVI1_MM3_voie3_allmates_bidim_withchannelsenabled.pngCalimero_MUVI1_MM2_allchannels_allmates_bidim_stat100000_withchannelsenabled.png 
FixedTEST Fri Jan 18 15:32:50 2019 Adrien ValérianGRITGeneralN/ACalimero tests Muvi2 MM7 MG10CAfter disabling nptool Multithreading, online
is working fine again.
 Calimero_MUVI2_MM7_allchannels_allmates__stat100000_withchannelsenabled.pngCalimero_MUVI2_MM7_allchannels_allmates_withchannelsenabled.png 
Fixed  Tue Jan 22 11:15:15 2019 MarlèneGRITGeneralN/ASquare detectorsMG9C (thin square detector coupled with thick
square detector unplugged) noisy
On X channels --> no noise problem ( about
 MG9C_disabled_channels__no_crazy_ORD.pngMG9C_disabled_channels__no_crazy_ORD_moredetailed.pngMG9C_piedestal_of_disabled_channels_Y.png 
OngoingTEST Tue Jan 22 18:52:23 2019 Valérian-MarlèneGRITGeneral3-alpha3 alpha source run on MG10C & MG3T3 alpha source (not centered)
MG10C and MG3T ON
MG10C is connected to MG5T HT and BT and
  
FixedTEST Wed Jan 23 09:11:26 2019 Valérian-MarlèneGRITGeneral3-alphaRUN9 : 3 alpha source run on MG10C & MG3T[B]RUN 9 (over night 22/01 ->23/01)[/B]
3 alpha source (not centered)
MG10C and MG3T ON
 3alpha_sourceMG10C-MM2_MG3T-MM3.pngHV_squares_plugged_on_MG4T-MG5T_andMG3T.pngHV_squares_plugged_on_MG4T-MG5T_andMG3T_afternightrun.pngMG10C_7missingstrips.png 
Fixed  Wed Jan 23 12:24:10 2019 MarlèneGRITGeneral Change cable MG1T - Exchange of detectors named MG9C & MG10C23/01/19

Change of the HT/BT cable of MG1T (the one
  
Fixed  Fri Jan 25 17:02:25 2019 MarlèneGRITGeneral Detectors numbering   MUGAST_detectors.pdf 
FixedTEST Thu Feb 7 09:52:08 2019 Freddy, MarlèneRUNGeneral3-alphaRUN10 : 3 alpha source on MG1-3-4-5-7Test run with alpha source in front of trapezoid
detectors.
 alpha_source_5trap_rawhistos_2.pngalpha_source_5trap_rawhistos.png 
Fixed  Thu Feb 14 08:50:58 2019 MarleneGRITGeneral3-alphaRUN 27RUN 27 on 14th of feb.
3 alpha source (one in front of trap., one
in front of square detectors)
 8x 
Fixed  Thu Feb 14 15:30:14 2019 Marlene Daniele JennyGRITGeneral3-alphaRUN 28RUN 28 14th Feb
3 alpha source on MG11A, MG3T, MG1T
 3alpha_sourceMG03XY._OKpng3alpha_sourceMG11XY.png3alpha_sourceMG01XY_OK.png 
Fixed  Fri Feb 15 15:31:28 2019 MarlèneGRITGeneralN/ADetectors bias    
Fixedcom2019 Mon Feb 18 10:41:36 2019 MarleneGRITGeneralN/ABias & current of detectors   bias.png 
FixedTEST Thu Feb 21 09:08:49 2019 FreddyRUNGeneral3-alphaRUN 30 - alpha on MG 1,3,5,7 [U][B]Run 30[/B][/U]

Alpha source run for trapezoid detectors
 3alpha_source_MG_1357_20Feb.pngLowV_bis_run30.pngLowV_run30.pngHV_run30.png 
FixedTEST Thu Feb 21 14:37:26 2019 iulian, valerian &FreddyGRITHardware3-alphaResults Calibration 3-alpha RUN30 MG1, MG3,MG5 & MG7Results of the calibration run 30
In order:
 Telescope1 X,Y (MG5 X,Y)
 8x 
FixedTEST Thu Feb 21 16:43:45 2019 Valérian, Freddy, Iulian and MarlèneGRITHardwareN/ACurrent measurementsJoined are the the I(V).

Detectors seems to respond accordingly with
 Depletion_Mugast_MG.pdfDepletion_Mugast_M2.pdfDepletion_Mugast.xlsx 
FixedTEST Thu Feb 21 17:32:42 2019 iulian, freddyGRITHardware3-alpharun 33 3 alfa with MG1,MG3,MG4,MG5,MG7,MG9 short runFirst test of the repaired 2xcables. The repaired
cables were placed between MUVI1_2 <=> MG4
& MUVI1_4 <=>MG11 (longer cable)
 3alpha_sourceMG1MG4MG7MG9XY_run33.png3alpha_sourceMG3MG5XY_run33.png 
FixedTEST Thu Feb 21 18:14:48 2019 iulianGRITHardware3-alphaACQ problem run 34 3 alfa with MG1,MG3,MG4,MG5,MG7,MG9
Alpha source run for trapezoid detectors
+ anulaire detector with pedestals aligned
 acq_problem_21and22fevrier.png 
FixedTEST Fri Feb 22 11:56:53 2019 iulian GRITHardware3-alpharun 38 3-alpha with reordering of the MUGAST cables for the logic and analogic signals[U][B]Run 38
[/B][/U] 
 Alpha source run for trapezoid detectors
 run38_3alpha_mg1345.pngrun38_3alpha_mg711.png 

ELOG Home Page