MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 1 of 132  Not logged in MUGAST
Status ExpNbr Date Author Type Category Target-Sourcedown Subject
Fixede744s Fri Apr 5 22:13:33 2019 OttoRunRUNGeneralTi-tRun254
*************************
Run : 254
*************************

Start at 2019-04-05 22:13:28.719268

Run Conditions:
  • Operator:
  • Trigger:
  • Target/Source:
  • Beam:
  • Intensity:

Comment:



Stop at 2019-04-05 22:38:50.168158
Stop comment: stop to put Vamos in the trigger
Fixede744s Fri Apr 5 22:39:39 2019 OttoRunRUNGeneralTi-tRun255
*************************
Run : 255
*************************

Start at 2019-04-05 22:39:35.688874

Run Conditions:
  • Operator:
  • Trigger:
  • Target/Source:
  • Beam:
  • Intensity:

Comment:



Stop at 2019-04-05 22:41:54.607652
Stop comment:
FixedTEST Fri Dec 14 09:08:46 2018 Adrien MattaOTHERSocialN/ATest Post
MUGAST ELOG is now ONLINE
FixedTEST Wed Jan 16 17:36:30 2019 Valérian-MarlèneGRITGeneralN/ACalimero tests - How to configure Calimero in GECO
Tests of Calimero time calibration with MUVI1 channel 2 (MM2)

1- channel 3 of all mates selected (see screeshots)
2- channel 1 of all mates
3- channel 1 of all mates - with channels enabled !!!!!
4- channel 1 of all mates - with channels enabled !!!!!
5- all channels telescope MM2

Number of counts = total number of counts for one channel on all mates

Conclusion :
Options to be chosen
1- with channels enabled
2- 10000 for the statistics


Remarks
To run calimero on MUVI1, caliWatcher should be set to 10202 and CaliSlave to 10223
The opposite for MUVI2, caliWatcher 10223, CaliSlave 10202
FixedTEST Fri Jan 18 15:32:50 2019 Adrien ValérianGRITGeneralN/ACalimero tests Muvi2 MM7 MG10C
After disabling nptool Multithreading, online is working fine again.

Concerning Calimero, it is currently working both on MUVI1 and MUVI2 for the X at least.

Joined here is the result of Calimero for MG10C on MUVI2 channel 3 (MM7).
Fixed  Tue Jan 22 11:15:15 2019 MarlèneGRITGeneralN/ASquare detectors
MG9C (thin square detector coupled with thick square detector unplugged) noisy
On X channels --> no noise problem ( about 100 ORD in the dark)
On Y channels --> MATE 10 & 11 noisy ! (see picture below with noisy Y piedestals)

With internal pulser (trigered by CK_TST) the counting rate is OK only if the thresholds are set at 1 MeV. If they are set at 500 keV, the counting rate is 10 times higher than the CK_TST

For MG10C, ORD is OK !
Fixed  Fri Feb 15 15:31:28 2019 MarlèneGRITGeneralN/ADetectors bias
Fixedcom2019 Mon Feb 18 10:41:36 2019 MarleneGRITGeneralN/ABias & current of detectors
FixedTEST Thu Feb 21 16:43:45 2019 Valérian, Freddy, Iulian and MarlèneGRITHardwareN/ACurrent measurements
Joined are the the I(V).

Detectors seems to respond accordingly with the previous measurements from Marlène.

Annular detector curve is quite strange…

Joined files :

Depletion_Mugast_MG.pdf -> summary for trapezoidal + annular detectors
Depletion_Mugast_M2.pdf -> summary for Must2 detectors

Depletion_Mugast.xlsx -> Excel file
PinnedTEST Tue Feb 26 09:30:21 2019 FreddyOTHERHardwareN/ACabling - MUVI / HT-BT / SA-SL
Cabling scheme - to update until final
FixedTEST Tue Feb 26 13:21:32 2019 iulian freddy GRITSoftwareN/Arun 42 Calimero run MUVI 1 on X sectors
run #42
Calimero MUVI1 X


run #41
Calimero MUVI secttor 1X
FixedTEST Tue Feb 26 13:50:36 2019 iulian freddy GRITSoftwareN/Arun 43 Calimero run MUVI 1 on Y sectors
run #43
Calimero MUVI1 Y
FixedTEST Tue Feb 26 14:52:49 2019 iulian freddy GRITSoftwareN/Arun 44 Calimero run MUVI 2 on X sectors (bad run)
run #44 bad run
Calimero MUVI2 X

During the run we noticed that time calibrator is on MUVI1 only .
FixedTEST Mon Mar 4 14:34:31 2019 Valerian, Marlène, IulianMUST2GeneralN/AMUST2 positions (based on e755)
Based on experiment e755, the MUST2 config is the following :


T1, PCB10:

X128Y1.......X1
. .
. .
Y128.......


T2, PCB5:

Y128.......X128Y1
. .
. .
.......X1



T3, PCB4:

.......Y128
. .
. .
X1.......X128Y1



T4, PCB7:

X1.......
. .
. .
X128Y1.......Y128
FixedTEST Tue Mar 5 10:16:03 2019 iulianMUST2SoftwareN/Arun 57 Gamme 300 ns Time Calibration run (Calimero) on MUVI 3 (MUST) X sectors more stats (10000)
RUNS 53,54,55,56 are bad runs

Run 57


Calimero on Muvi3 MM1(X)

Gamme 300 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
FixedTEST Tue Mar 5 12:13:41 2019 iulian MarleneMUST2SoftwareN/Arun 60 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X
RUNS 58,59,60 are bad runs

Run 61


Calimero on Muvi3 MM1(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
FixedTEST Tue Mar 5 15:53:04 2019 iulian MarleneMUST2SoftwareN/ACalimero problems
problems with Calimero solved by killing some processes on lpcgrit related with Narval.
to find the processes on lpcgrit:
ps aux | grep Cali*
ps aux | grep Narval

kill -9 all the ids

Other possible solution:

start_acq on lpcgrit
tk followed by kall
exit
FixedTEST Tue Mar 5 16:20:11 2019 iulian MarleneMUST2SoftwareN/Arun 77 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X . good run
Run 76


Calimero on Muvi3 MM1(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
FixedTEST Tue Mar 5 16:42:56 2019 iulian MarleneMUST2SoftwareN/Arun 79 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM2 X . good run
Run 79

Run 78 bad
Calimero on Muvi3 MM2(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
FixedTEST Tue Mar 5 17:11:29 2019 iulian MarleneMUST2SoftwareN/Arun 81 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM3 X . good run
Run 81

Run 80 bad
Calimero on Muvi3 MM3(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)

ELOG Home Page