Fixede793s Fri Feb 12 18:12:00 2021 AdrienS1HardwareN/AKapton changeArnaud open today and change the Y Kapton
(YA5) of the S1. Everything under vacuum/cooled
again so I ran FSC, hopefully this will fix
  
Fixede793s Fri Feb 12 18:07:24 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 614
  - MUVI: MUVI2
  
Fixede793s Fri Feb 12 08:52:30 2021 mugastRUNGeneral3-alpharun 613: long alpha run all Muvi in trigger, mugast stand alone. all
detector Bias on.
  
Fixede793s Mon Feb 8 16:33:22 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 612
  - MUVI: MUVI2
  
Fixede793s Mon Feb 8 16:20:42 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 611
  - MUVI: MUVI2
  
Fixede793s Mon Feb 8 11:28:38 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 610
  - MUVI: MUVI3
  
Fixede793s Mon Feb 8 11:25:29 2021 Adrien and FreddyS1GeneralN/AFSC runwe ran  FSC on all card. S1 is dismounted.
Pulser goes on all channel and works well
so electronic is fine.
 fscE_08022023.pngfscT_08022022.png 
Fixede793s Mon Feb 8 11:16:18 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 609
  - MUVI: MUVI3
  
Fixede793s Mon Feb 8 11:04:04 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 608
  - MUVI: MUVI2
  
Fixede793s Mon Feb 8 10:51:48 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 607
  - MUVI: MUVI2
  
Fixede793s Mon Feb 8 10:39:35 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 606
  - MUVI: MUVI1
  
Fixede793s Mon Feb 8 10:27:23 2021 OttoTimeRUNGeneralN/ATimeCalibratorRun*** Run Time Calibrator FSC ***
  - Run: 605
  - MUVI: MUVI1
  
FixedTEST Fri Feb 5 12:44:42 2021 FrancoGRITHardwareN/APulser on MG11 MUFEE YI managed to obtain the spectrum with the
pulser on the annular (Y side, namely sectors).
The raw strip numbers where I should have
 pulser_all_channels_MG11Y_05022021.pngpulser_all_strips_MG11Y_05022021.pngWorking_rings_annular_05022021.png 
FixedTEST Thu Feb 4 18:16:47 2021 FrancoGRITHardware3-alphaProblems with the annular persistWe opened the chamber to verify the connections
of the kaptons but they were all well connected.
We checked with the tester the conductivity
 8x 
FixedTEST Wed Feb 3 16:42:41 2021 FrancoGRITGeneral3-alphaStatus of the set-up (February 3rd 2021)We re-mounted the annular, put the source
and cooled down to 0 deg. Voltages and spectra
are attached.
 Voltages_03022021.pngMM_raw_03022021.pngMG_X_raw_03022021.pngMG_Y_raw_03022021.png 
FixedTEST Tue Feb 2 16:41:17 2021 FrancoVACUUMHardwareN/AMUGAST undervacuum and powered upI put back the supply of MM1 to the channel
1 of the MUST2 box (as in the normal configuration)
and polarized the detectors to make a rapid
  
Fixede793s Fri Jan 29 15:11:54 2021 AdrienVACUUMGeneralN/AMUGAST undervacuum and powered upMUGAST is undervacuum thanks to Arnaud and
Laurent. 
  
   Tue May 19 12:23:57 2020 mugastONLINE       
Fixede793s Thu Mar 12 17:25:19 2020 mugastGRITGeneral3-alpharun 0601 - alpha source with pedestalsAlpha source calibration run without pedestal
suppression and cooling at 0 deg.
  
Fixede793s Thu Mar 12 15:48:58 2020 Franco&MarlèneGRITGeneral Summary and TO DO list 12/03/2020What remains to be done
- Do the time calibration of all boards (time
calibrator has to be connected)
  