Fixed, e793s, Wed Mar 17 14:59:17 2021, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
- Run: 649
- MUVI: MUVI3
- Side : MUFEE Y (Ohmic)
- Telescope: 1, 2, 3, 4
Fixed, e793s, Wed Mar 17 15:11:37 2021, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
- Run: 650
- MUVI: MUVI4
- Side : MUFEE X (Junction)
- Telescope: 1,
Ongoing, e793s, Wed Mar 17 15:17:39 2021, Emmanuel C, OFFLINE, General, , New gamma gamma analysis 48K.pdf
The level scheme is based on coherent check of
1- gamma gamma
2- gamma gated by Ex
3- Ex gated by gamma
Fixed, e793s, Wed Mar 17 15:20:08 2021, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
- Run: 651
- MUVI: MUVI4
- Side : MUFEE Y (Ohmic)
- Telescope: 1,
Fixed, e793s, Wed Mar 17 19:19:05 2021, Franco, OFFLINE, General, CD2, Mugast detectors disaligned in Ex detectors_disalignment_Ex.pngEx_aligned_vs_unaligned.png
The Ex distributions gated on the different MG detectors show large differences, up to 500 keV. In the attached figure I show as an example the comparison between two nearby detectors, MG3 and MG4.
I tried a rough alignment of the Ex by shifting it manually and setting the first main peak at Ex=0 (essentially taking MG5 as the reference, since his first peak is at 0).
These are the shifts applied:
MG1 -> -580 keV
MG2 -> -280 keV
MG3 -> -550 keV
MG4 -> -100 keV
MG5 -> 0 keV
MG7 -> -400 keV
The resulting Ex distribution is plotted in the second attachment, compared with the original below. The resolution of the first peak (assuming there's a single level there) is 420 keV FWHM.
I think it would be worth having the position of the MUGAST detectors re-measured.
    Fixed, e793s, Thu Mar 18 16:25:56 2021, Emmanuel C, OFFLINE, General, , Efficiency gamma AGATA --- real measurement run70 

Emmanuel C wrote:
TF1 *fit_1 = new TF1("fit_1","TMath::Exp([0]+[1]*TMath::Log(x)+[2]*pow(TMath::Log(x),2.0)+[3]*pow(TMath::Log(x),3.0)+[4]*pow(TMath::Log(x),4.0))",10,5000); //NRJ traite en keV et efficiency en %


Data

40.000000 3.404000 0.070000
45.000000 6.512000 0.150000
121.782997 14.250000 0.300000
244.692001 11.880000 0.270000
295.938995 11.870000 0.500000
344.276001 11.170000 0.240000
411.114990 9.926000 0.330000
443.976013 11.050000 0.290000
778.903015 8.007000 0.180000
867.388000 7.562000 0.240000
964.130981 7.492000 0.180000
1085.800049 7.110000 0.160000
1112.115967 6.956000 0.150000
1212.949951 6.113000 0.190000
1408.010986 6.157000 0.130000


p0 -9.87216e+01
p1 6.98517e+01
p2 -1.78362e+01
p3 2.00526e+00
p4 -8.41808e-02
Fixed, e793s, Fri Mar 19 09:54:58 2021, Adrien, Freddy and Franco, GRIT, General, 3-alpha, Alpha calibration and noise on MG3 and MG4 after the experiment 
While starting the alpha source runs after the experiment we found unexpected noise on MG3 and MG4 which cannot be excluded just by masking some Y mates (The X mates were already masked for both of them). We cannot figure out the origin of this noise.

Run 652 and 653 -> alpha runs without pedestal suppression
Run 655 -> alpha run during the night with pedestal suppression. Here MG3 and MG4 are not in the trigger.
Run 658 -> alpha run after an attempt to mask some noisy strips. The number of ORD is anyway high (some 10^4) but we see some alphas on MG3 and MG4.
Run 660 -> same as 658 but with higher thresholds on X (8250) and Y (7600 on MG3 and 7950 on MG4) to write less on disk but actually nothing changes. Therefore we are writing about 2.7 GB per hour.

The other runs (654, 656, 657 and 659) can be discarded.
Fixed, e793s, Mon Mar 22 16:35:46 2021, OttoRun, RUN, General, N/A, Run67 
*************************
Run : 67
*************************

Start at 2021-03-22 16:34:11.552975

run67. Time calibrator in TDC VAMOS

PERIOD: 80 ns

Trigger VAMOS


Raw Data: /data/mugastX/e793s/acquisition/run/run_0067.dat.22-03-21_16h34m11s
Raw RootFile: /home/mugast/Analysis_e793s/RootR/r0067_000r.root
Fixed, e793s, Mon Mar 22 18:15:50 2021, mugast, RUN, General, N/A, TACS calibration - run 662 
TACS:

FAG-CATS2 (range 500 ns)
VAMOS-AGATA (range 500 ns)
CATS2-CATS2 (not in modular label) (range 500 ns)
AGATA-CATS2 (range 200 ns)

Time calibrator settings:

640 ns range
20 ns steps
Fixed, e793s, Mon Mar 22 19:04:55 2021, mugast, RUN, General, N/A, TACS calibration - run 653 
TACS:

MUGAST-AGATA (range 2 us)
MUGAST-CATS2 (range 1 us)
MUGAST-CATS2b (range 2 us)
MUGAST-VAMOS (range 2 us)
VAMOS-CATS2 (range 2 us)

Time calibrator settings:

2.56 us range
80 ns steps
    Fixed, e793s, Tue Mar 23 10:22:48 2021, Emmanuel C, OFFLINE, General, , Efficiency gamma AGATA --- real measurement run70 effi.png

I have been working on the high energy response.
The point is that we do not have source measurement beyond 2 MeV to constrain the fit.
In the attached file, the experimental points are the measured from run 70 in 152Eu source. The fit is then extrapolated up to high energy.
I have simulated the efficiency at 3.5 MeV in GEANT4 including real geometry, missing channels, unefficiency of capsules etc.... The overall agreement was checked on the 1408 line. The 3.5 MeV simulated point is shown. It disagrees strongly with the extrapolation. The blue fit includes the simulated point. It doesn't change the efficiency evaluation < 1.5 MeV but correct the high energy response.
Below, updated parameters

p0 -6.34543e+01
p1 4.24746e+01
p2 -1.00304e+01
p3 1.03468e+00
p4 -3.97076e-02


TF1 *fit_1 = new TF1("fit_1","TMath::Exp([0]+[1]*TMath::Log(x)+[2]*pow(TMath::Log(x),2.0)+[3]*pow(TMath::Log(x),3.0)+[4]*pow(TMath::Log(x),4.0))",10,5000); //NRJ traite en keV
Pinned, e786s, Wed Apr 10 13:30:48 2019, Valerian, Adrien and Antoine (edit by Louis), OFFLINE, Software, N/A, How to perform analysis for e786s 

Here is the updated procedure to convert a run, with VAMOS data and add those data to the Kinematic plot.


Every 4Go there will be a file change, to see if there is a file change do :

  • ls -ltrh /data/mugastX/e786s/acquisition/run/

convert the last 4Go file and add it to the chain using the following step:

Step O: Take your favorite ortholinear keyboard (optional)

Step 1: Login the VAMOS environment and convert the run :

Open a terminal from one of the mugast computer:
  • Type: source $HOME/GoToVamosAna
  • Type: MyVAna -d2Ar -f /data/mugastX/e786s/acquisition/run/run_$$$$-DD-MM-YY.HHhMMmSSs /data/mugastX/e786s/acquisition/RootA/r$$$$_000a.root
OR
  • Type: MyVAna -d2Ar -f /data/mugastX/e786s/acquisition/run/run_$$$$-DD-MM-YY.HHhMMmSSs.N /data/mugastX/e786s/acquisition/RootA/r$$$$_00Na.root
(If you encounter the error on Brho "Could not find the BrhoRef for the run <RunNr> ", Update the ./Calibs/ListBrho.dat file with relevant Brho for RunNr)

Step 2: Analyse the file using the npanalysis program:
  • Type: exit (<-this exit the environment of VAMOS)
  • Type: lpcgrit (<-this load the analysis environment )
  • Type: npp MUGAST
  • Type: edit the exe.sh to treat releverant run. command should be: npanalysis -D mugast.detector -C Calibration.txt -E ./Reaction/46Ar3Hed.reaction -T ./offline/root_e786svamos/r$$$$_00Na.root AD -O run_$$$$_N.root

(This could be used with any file, either from Offline conversion, or the VAMOS Analysis conversion or the Merged Tree from AGATA)

The run will be created in the NPOutputs folder
Fixed, e786s, Fri May 31 09:53:14 2019, Marlène, GRIT, General, , Alpha source --> status of detectors 
MG3T has been fixed by changing the X kapton.

I have removed aluminium foil on square detectors, the noise is slightly higher.

MG4T is not working properly : noisy on X but no signal on Y !! I see no triggers on the Y side. I tried to increase bias but it did increase noise. I checked in the last calibration of e744 it was working properly. Also I had to do pedestal realignement for this detector because everything was shifted.

MG7T seem to be slightly different than before. Strips X are hidden from the trigger because it became suddenly noisy.
Fixed, e786s, Mon Jun 3 18:00:10 2019, Marlène, GRIT, General, , MUGAST detectors status --> MG4T  
The problem of MG4T ha sbeen solved by changing the feedthrough.
The pulser on Y was not working and we were seeing no events (no ORD) because there was a problem with the bias of board. The issue is due to the feedthrough.
So now MG4T (HT &BT) are plugged in a free feedthrough (below MG1T) and I see pulser signals on Y.

The bias seems to apply normally on all detectors (in the dark).
The pulser is seen on each boards.
I have put back aluminium foil on square but I could not test it.
Fixed, e786s, Wed Jun 12 17:59:01 2019, mugast, OTHER, Software, , Online Energy Calibration x_E_cal.pngy_E_cal2.png
Updated Online Energy calibration for MUGAST, based on run_0231 data
    Fixed, e786s, Wed Jun 12 18:11:04 2019, mugast, OTHER, Software, , Online Energy Calibration 
Fixed, e786s, Wed Jun 19 10:09:10 2019, mugast, MUST2, Hardware, N/A, Time calibrator cabled 
range 640 ns
period 20 ns
Fixed, e786s, Wed Jun 19 12:02:07 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
- Run: 237
- MUVI: MUVI1
- Side : MUFEE X (Junction)
- Telescope: 1, 2, 3, 4
Fixed, e786s, Wed Jun 19 12:24:43 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
- Run: 238
- MUVI: MUVI1
- Side : MUFEE Y (Ohmic)
- Telescope: 1, 2, 3, 4
Fixed, e786s, Wed Jun 19 12:48:13 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
- Run: 239 - MUVI: MUVI2
- Side : MUFEE X (Junction)
- Telescope: 1, 2, 3, 4