Status |
ExpNbr |
Date |
Author |
Type |
Category |
Target-Source |
Subject |
Fixed | TEST |
Fri Mar 22 18:15:39 2019 |
Otto | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 122
- MUVI: MUVI3
- Side : MUFEE Y (Ohmic)
- Telescope: 1, 2, 3, 4 |
Fixed | TEST |
Fri Mar 22 18:06:11 2019 |
Otto | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 121
- MUVI: MUVI3
- Side : MUFEE X (Junction)
- Telescope: 1, 2, 3, 4 |
Fixed | TEST |
Fri Mar 22 17:56:37 2019 |
Otto | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 120
- MUVI: MUVI2
- Side : MUFEE Y (Ohmic)
- Telescope: 1, 2, 3, 4 |
Fixed | TEST |
Fri Mar 22 17:47:04 2019 |
Otto | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 119
- MUVI: MUVI2
- Side : MUFEE X (Junction)
- Telescope: 1, 2, 3, 4 |
Fixed | TEST |
Fri Mar 22 17:37:37 2019 |
Otto | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 118
- MUVI: MUVI1
- Side : MUFEE Y (Ohmic)
- Telescope: 1, 2, 3, 4 |
Fixed | TEST |
Fri Mar 22 17:28:09 2019 |
Otto | RUN | General | N/A | TimeCalibratorRun |
*** Run Time Calibrator FSC ***
- Run: 117
- MUVI: MUVI1
- Side : MUFEE X (Junction)
- Telescope: 1, 2, 3, 4 |
Fixed | TEST |
Thu Mar 21 13:28:06 2019 |
Valerian and Adrien | DAQ | Software | N/A | First fully functionnal time calibration run on X and Y of MG1 |
Our program is able to perform the time calibration of our detector for both X and Y, see picture attached. |
Attachment 1: CALIMEROY1.png
|
|
Attachment 2: FIRST_TIMECALIBRATION_WITHOUT_CALIMERO.png
|
|
Fixed | TEST |
Thu Mar 21 10:37:47 2019 |
Adrien and Valerian | MUST2 | General | N/A | Test TOF must2 |
To make sure the mate work correctly we took an alpha run with a delayed fag on the MUVI stop:
- all mate register a time in the the middle of the range as expected (fag delayed by 350ns, range is 600ns)
Test with time calibrator, all channel are enable, pulser on channel 3 of each mate T1X:
- energy spectrum: pulser where we expect it.
- we expect to see a time calibrator spectrum on channel 3 of each mate. instead we see times spectrum on Channel 3 of mate 1 and 2, channel 12 of mate 1, channel 14 of mate2. nothing else
Conclusion
The problem was that the synctst had to be checked and ord unchecked |
Attachment 1: Time_behavior_withDAS.png
|
|
Fixed | |
Tue Mar 19 19:35:53 2019 |
marlene | GRIT | General | | RUN 98 3alpha source all detectors |
RUN 98
3 alpha source
all detectors (trap,square, MUST2,annular)
Temp=15deg
Acquisition mugast |
Fixed | TEST |
Mon Mar 18 18:31:18 2019 |
valerian marlène | GRIT | General | 3-alpha | RUN 20 (com2019) 3 alpha overnight |
RUN20
Run with double 3alpha source MUST2+MG1_7+MG11
temperature -4 deg on both cooling
Acquisition with VAMOS+MUGAST coupled (com2019)
Time range =600ns
pedestal aligned and suppressed
run start at 18h23 monday 18
run stop at 8h34 tuesday 19
about 12Go of data |
Attachment 1: RUN20_HV_MG_3lapha.png
|
|
Attachment 2: RUN20_3alpha_HV_MUST2.png
|
|
Fixed | |
Tue Mar 12 10:21:17 2019 |
Marlène | MUST2 | General | | Debug Calimero Y |
Debug of Calimero
test of channels (mate) masking : OK with DAS, OK with GECCO (see attachment) when done in the "disable channels" of GECCO
With calimero, when we are on Y, when we click right to mask a mate we see 'disable mate 2' instead of 'disable mate 10' !!! |
Attachment 1: test_GECO_masquage_mate2&10activees_seulement.png
|
|
Fixed | TEST |
Thu Mar 7 14:02:03 2019 |
Marlène Iulian | MUST2 | General | N/A | test of MUST2 electronics |
Test with pulser at 20 kHz
The OR of MUVI was scaled down by a factor of 100, to provoke some resets |
Fixed | TEST |
Thu Mar 7 09:29:47 2019 |
Marlene | GRIT | General | | run 96 double alpha source run shifted 60mm |
RUN 96
double alpha source
MG1
MG3
MG4
MG5
MG7
-
-
MG11
MM1
MM2
MM3
MM4
Missing strips in Y the same as in run 40
Problem with the annular MG11 -> no current when bias is put... --> to be checked
Problem with the insertion of the target holder --> solved
I also checked again MG9 (still noisy)
and MG10 with 140V I=0,52uA and counting rate is the same order of magnitude as trapezoid. Spectra are OK (not saved). |
Attachment 1: run96_MG7_MG11notworking.png
|
|
Attachment 2: run96_3alphasource_shifted_MG1-5.png
|
|
Attachment 3: run96_HV2.png
|
|
Attachment 4: run96_HV1.png
|
|
Fixed | TEST |
Wed Mar 6 15:02:51 2019 |
Valerian | MUST2 | Hardware | 3-alpha | Cooling temperature is relevant for the resolution |
Using run 52 -> 13deg and run 87 -> -5deg.
We see that the resolution improves at -5 degree. See attached picture as an example for MM2 X. |
Attachment 1: MM2X_13deg.png
|
|
Attachment 2: MM2X_minus5deg.png
|
|
Fixed | TEST |
Tue Mar 5 18:18:46 2019 |
iulian Marlene | MUST2 | Hardware | 3-alpha | run 97 3a run Muvi3 night run |
Run 87
stop@ 5:42 am
3a run Muvi3 (MM)
-4 deg
pedestals aligned and suppressed |
Attachment 1: run87_3alpha_4mm_spectra.png
|
|
Fixed | TEST |
Tue Mar 5 18:03:24 2019 |
iulian Marlene | MUST2 | Software | N/A | Calimero problem on Y side |
Calimero not working properly on Y side of any mm detector |
Attachment 1: run83_calimero_mm4Y_spectra.png
|
|
Attachment 2: run834calimero_mm2Y_spectra.png
|
|
Fixed | TEST |
Tue Mar 5 17:48:38 2019 |
iulian Marlene | MUST2 | Software | N/A | run 82 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM4 X . good run |
Run 82
Calimero on Muvi3 MM4(X)
Gamme 600 ns |
Attachment 1: run82_calimero_mm4X_spectra.png
|
|
Fixed | TEST |
Tue Mar 5 17:11:29 2019 |
iulian Marlene | MUST2 | Software | N/A | run 81 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM3 X . good run |
Run 81
Run 80 bad
Calimero on Muvi3 MM3(X)
Gamme 600 ns
Procedure:
0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco) |
Attachment 1: run81_calimero_mm3X_spectra.png
|
|
Fixed | TEST |
Tue Mar 5 16:42:56 2019 |
iulian Marlene | MUST2 | Software | N/A | run 79 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM2 X . good run |
Run 79
Run 78 bad
Calimero on Muvi3 MM2(X)
Gamme 600 ns
Procedure:
0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco) |
Attachment 1: run79_calimero_mm2X_spectra.png
|
|
Fixed | TEST |
Tue Mar 5 16:20:11 2019 |
iulian Marlene | MUST2 | Software | N/A | run 77 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X . good run |
Run 76
Calimero on Muvi3 MM1(X)
Gamme 600 ns
Procedure:
0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco) |
Attachment 1: run77_calimero_mm1X_spectra.png
|
|