Status |
ExpNbr |
Date |
Author |
Type |
Category |
Target-Source |
Subject |
Text |
 |
Fixed | e768s |
Sat Jul 27 11:49:40 2019 |
Freddy | S1 | Hardware | N/A | Annular degradation | I posted a detailed entry during the night
about the evolution of the annular detector
but it was deleted somehow...
|
    |
Fixed | e793s |
Sat Mar 13 13:16:04 2021 |
Franco | OFFLINE | General | CD2 | Angular distributions | - Updated the macro for the angular distributions.
- For the angular distribution of the peak
at 4 MeV excitation energy I think we need |
 |
Fixed | e786s |
Fri May 31 09:53:14 2019 |
Marlène | GRIT | General | | Alpha source --> status of detectors | MG3T has been fixed by changing the X kapton.
I have removed aluminium foil on square detectors, |
|
Fixed | TEST |
Tue Feb 18 17:53:36 2020 |
Franco | GRIT | General | 3-alpha | Alpha calibration for exp e775s (run 0501) | In attachment raw and calibrated spectra for
run 0501 (alpha source) for MUST2 and Mugast.
|
   |
Fixed | e793s |
Wed Mar 11 19:02:54 2020 |
mugast | GRIT | General | 3-alpha | Alpha calibration for e793s - run 600 | ========================================
========================================
========================================
|
|
Fixed | e793s |
Fri Mar 19 09:54:58 2021 |
Adrien, Freddy and Franco | GRIT | General | 3-alpha | Alpha calibration and noise on MG3 and MG4 after the experiment | While starting the alpha source runs after
the experiment we found unexpected noise
on MG3 and MG4 which cannot be excluded just |
|
Fixed | e793s |
Wed Mar 11 22:01:18 2020 |
mugast | GRIT | General | 3-alpha | Alpha calibration - run 600 | After 2h30' of run we checked the temperatures
of the boards and we noticed that MG7 had
63 degrees on both X and Y, which is very |
|
Fixed | e793s |
Thu Mar 11 21:29:43 2021 |
Franco | RUN | General | CD2 | Alignment in time of the VAMOS sections | The aligned time is saved in a variable called
MWT in the output tree |
|
Fixed | e744s |
Sun Apr 14 12:32:32 2019 |
Beyhan, François | OTHER | General | | Alarm in G2 but no beam stop | Same issue as Saturday evening.
Alarm in G2 at 12:30 due to Exogam filling
with liquid nitrogen.
|
|
Fixed | e744s |
Sat Apr 13 20:37:16 2019 |
Beyhan | OTHER | General | | Alarm in G2 | Alarm signal from G2.
No beam since 20:30
PCP will have a look in G2, following the |
|
Fixed | e744s |
Sat Apr 13 22:45:27 2019 |
Beyhan, Sylvain, Jaromir | AGATA | Software | N/A | Agata yellow phase 1 minute | Agata yellow 1 minute
All the trigers emitted by (every) single |
|
Pending | e793s |
Wed Mar 17 08:03:15 2021 |
Laetitia, Charlie | AGATA | Hardware | CD2 | Agata issue | There is a problem on Agata (the dewars are
not being filled), and the Ge's start to
warm up. The beam has been stopped at ~08h05. |
|
Fixed | e793s |
Wed Mar 17 09:53:38 2021 |
Cyril | AGATA | Hardware | CD2 | Agata issue | Agata problem has been solved, a valve was
closed by mistake.
Beam returned at 9h40. |
|
Fixed | |
Tue Apr 16 12:18:46 2019 |
AGATA | AGATA | Hardware | | ATC4 - position 13 HV tripped | - HV tripped around 12h10.
- need access to check the origin |
|
Fixed | |
Tue Apr 16 13:15:48 2019 |
AGATA | AGATA | Hardware | | ATC4 - position 13 HV tripped |
set 1500 V at 13h15 |
|
Fixed | e768s |
Fri Jul 19 17:33:21 2019 |
Nicolas | OTHER | General | | AGATA: what if a crystal status get red | 1. Go to the AGATA Digitizer: Control tab
of the browser
2. Select the detector and crystal, then |
|
Fixed | e768s |
Fri Jul 19 17:52:24 2019 |
Nicolas | OTHER | General | | AGATA: replaying runs | 1. Go to the AGATA analysis computer, open
a terminal and execute the following commands:
> a4
|
|
Fixed | e786s |
Wed Jun 26 07:01:00 2019 |
Freddy | AGATA | General | N/A | AGATA validation problem - again - during run 59 | 30 min after Emmanuel left (around 6h00),
the AGATA problem of validation reappeared...
|
|
Fixed | e786s |
Wed Jun 26 04:54:07 2019 |
Freddy | AGATA | General | N/A | AGATA validation problem (solved) | Since about 4h20, we realized that AGATA validated
events and timestamps have suddenly dropped
on the AGATA display (yellow bars instead |
|
Ongoing | e775s |
Sat Feb 29 07:35:21 2020 |
mugast : Valerian | AGATA | General | CD2 | AGATA is not responding | Since ~4:30am AGATA is not responding, everyhting
was frozen. Unclear if it comes from disk
problem or other... |
|