Fixede768s Mon Jul 22 16:54:33 2019 Marlène, Franck, François & NicolasBEAMGeneral Cheking beam transmission from G1 profiler to VAMOS
Goal:
Estimate beam transmission from profiler to VAMOS (FP)

Method:
1. Beam intensity is reduced with the pepper pot 1/9 and by closing slits after the 15O target production and the accelerators so that the beam quality (emittance, etc...) is not affected.
2. VAMOS plate is removed to allow the direct beam detection
3. LiF (1.25 mg/cm2, number 1) is in position

Results:
1. beam intensity at G1 profiler ~ 4e4 pps
2. VAMOS FP ~ 2e4 pps

Discussion:
1. There is a factor of two loss between the profiler and VAMOS. Taken this factor into account in the g-ray yield estimate (see post ..) the numbers of 1297- and 1340-keV g-rays is in line with what is expected from cross-section calculation.
2. Loosing a factor of two means a substantial amount of 15O is implanted in the pipe/target frame/MUST2 masks. A dedicated AGATA activation measurement could be performed to cross-check this factor of two.
Fixede768s Mon Jul 22 17:07:39 2019 Marlène, Franck, François & NicolasBEAMGeneral Profiler conditions for measuring beam intensity
We checked that the measured beam intensity (present case 1.5e7 pps) is independent on the voltage on the profiler. Values from 93 V up to 177 V gave similar results.
Fixed  Mon Jul 22 19:13:45 2019 KseniiaAGATAGeneral Current stats on 19Ne
Runs 112..125, still no joy...
Attachment 1: 112-125.png
112-125.png
Fixede768s Tue Jul 23 04:31:34 2019 Jenn & ChristianBEAMGeneral Diamond Scaler screenshot
Attachment 1: diamondscaler_run126_3.png
diamondscaler_run126_3.png
Fixede768s Tue Jul 23 04:34:03 2019 Jenn & ChristianBEAMGeneral Beam intensity and profile check
VAMOS OR: ~1200
BEAM INTENSITY ~ 2E7
Attachment 1: photo_2019-07-23_03-26-41.jpg
photo_2019-07-23_03-26-41.jpg
Fixede768s Tue Jul 23 06:49:54 2019 Jenn & Christian & KseniiaBEAMGeneral Beam intensity and profile check
VAMOS OR ~ 870
BEAM INTENSITY ~ 1.4E7
Attachment 1: photo_2019-07-23_05-50-19.jpg
photo_2019-07-23_05-50-19.jpg
Fixede768s Tue Jul 23 09:03:07 2019 Kseniia, Gilbert, YorickBEAMGeneral Beam intensity and screenshots
The beam has been pretty stable since yesterday 
Attachment 1: E1248B05-DA9D-43EF-A7B4-4FAD97F7D0B1.jpeg
E1248B05-DA9D-43EF-A7B4-4FAD97F7D0B1.jpeg
Attachment 2: Screenshot_from_2019-07-23_09-14-34.png
Screenshot_from_2019-07-23_09-14-34.png
Attachment 3: Screenshot_from_2019-07-23_09-14-15.png
Screenshot_from_2019-07-23_09-14-15.png
Attachment 4: Screenshot_from_2019-07-23_09-13-55.png
Screenshot_from_2019-07-23_09-13-55.png
Attachment 5: Screenshot_from_2019-07-23_09-13-35.png
Screenshot_from_2019-07-23_09-13-35.png
Attachment 6: Screenshot_from_2019-07-23_09-13-06.png
Screenshot_from_2019-07-23_09-13-06.png
Attachment 7: Screenshot_from_2019-07-23_09-12-49.png
Screenshot_from_2019-07-23_09-12-49.png
Attachment 8: Screenshot_from_2019-07-23_09-12-33.png
Screenshot_from_2019-07-23_09-12-33.png
Attachment 9: Screenshot_from_2019-07-23_09-12-15.png
Screenshot_from_2019-07-23_09-12-15.png
Attachment 10: Screenshot_from_2019-07-23_09-12-01.png
Screenshot_from_2019-07-23_09-12-01.png
Attachment 11: Screenshot_from_2019-07-23_09-10-57.png
Screenshot_from_2019-07-23_09-10-57.png
Attachment 12: Screenshot_from_2019-07-23_09-10-29.png
Screenshot_from_2019-07-23_09-10-29.png
Attachment 13: Screenshot_from_2019-07-23_09-10-11.png
Screenshot_from_2019-07-23_09-10-11.png
Attachment 14: Screenshot_from_2019-07-23_09-09-41.png
Screenshot_from_2019-07-23_09-09-41.png
Attachment 15: Screenshot_from_2019-07-23_09-09-15.png
Screenshot_from_2019-07-23_09-09-15.png
Attachment 16: Screenshot_from_2019-07-23_09-08-43.png
Screenshot_from_2019-07-23_09-08-43.png
Attachment 17: Screenshot_from_2019-07-23_09-06-56.png
Screenshot_from_2019-07-23_09-06-56.png
Attachment 18: Screenshot_from_2019-07-23_09-05-56.png
Screenshot_from_2019-07-23_09-05-56.png
Attachment 19: Screenshot_from_2019-07-23_09-05-37.png
Screenshot_from_2019-07-23_09-05-37.png
Attachment 20: Screenshot_from_2019-07-23_09-05-19.png
Screenshot_from_2019-07-23_09-05-19.png
Attachment 21: Screenshot_from_2019-07-23_09-05-03.png
Screenshot_from_2019-07-23_09-05-03.png
Attachment 22: Screenshot_from_2019-07-23_09-04-48.png
Screenshot_from_2019-07-23_09-04-48.png
Attachment 23: Screenshot_from_2019-07-23_09-04-34.png
Screenshot_from_2019-07-23_09-04-34.png
Attachment 24: Screenshot_from_2019-07-23_09-02-54.png
Screenshot_from_2019-07-23_09-02-54.png
Attachment 25: Screenshot_from_2019-07-23_09-02-27.png
Screenshot_from_2019-07-23_09-02-27.png
Fixede768s Tue Jul 23 11:02:54 2019 Kseniia, GilbertBEAMGeneral Beam profile
Attachment 1: FF0E1C03-CD2A-4BAB-8E82-26422F7EB866.jpeg
FF0E1C03-CD2A-4BAB-8E82-26422F7EB866.jpeg
Attachment 2: Screenshot_from_2019-07-23_11-04-17.png
Screenshot_from_2019-07-23_11-04-17.png
Attachment 3: Screenshot_from_2019-07-23_11-04-05.png
Screenshot_from_2019-07-23_11-04-05.png
Fixede768s Tue Jul 23 14:33:03 2019 Marlène, Yorick, Faïrouz & NicolasBEAMGeneral Cheking beam transmission from G1 profiler to VAMOS
Goal:
Estimate beam transmission from profiler to VAMOS (FP)
We question whether the factor of two found yesterday (VAMOS/Profiler ~ 0.5) could be due:
1. to an incorrect measurement of the beam intensity with the profiler
2. a charge state equilibration in the target (LISE++ calculation indicates 8+/7+ = 85%/15%)

Same conditions as yesterday:
1. Beam intensity is reduced with the pepper pot 1/9 and by closing slits after the 15O target production and the accelerators so that the beam quality (emittance, etc...) is not affected.
2. VAMOS plate is removed to allow the direct beam detection

Results:
We measure in VAMOS with the LiF target and with the target ladder totally removed. In the latter case, direct beam 15O(8+) is sent in VAMOS and no charge state effect enter into the game.
1. Beam intensity at G1 profiler ~ 6.5e4 pps
VAMOS (FP) + no target ~ 3.8e4 pps
2. Beam intensity at G1 profiler ~ 3.8e4 pps
VAMOS (FP) + thick LiF (nb 1) ~ 2.0e4 pps

Discussion:
We find the same results as yesterday, e.g. a factor of two attenuation between the profiler and VAMOS. If part of this loss would happen in the pipe/target/MUST2 mask, this should be observed in AGATA which is not the case. Could it be that the PPAC in VAMOS is not as efficient as announced at the beginning of the campaign (98%)? This must be checked when we get beam back later on.

Note:
While doing the above measurement, we observed a factor of 3 loss between profiler AR.Gaz13 and G1.Gaz31. Numbers follows:
L3.GAZ42 AR.GAZ13 G1.GAZ31
2.1e4 1.5e4 4.1e3
1.4e5 1.0e5 2.5e4
??? 1.8e5 6.5e4
Attachment 1: 714D4921-2A75-44B7-AD8D-87048BDA17A8.jpeg
714D4921-2A75-44B7-AD8D-87048BDA17A8.jpeg
Attachment 2: 6A77930D-8992-4118-B59C-DBAA37289533.jpeg
6A77930D-8992-4118-B59C-DBAA37289533.jpeg
Attachment 3: 3CD66D1D-D6E6-44B4-A15A-C947F048D2A0.jpeg
3CD66D1D-D6E6-44B4-A15A-C947F048D2A0.jpeg
Fixede768s Tue Jul 23 16:19:14 2019 MarlèneOTHERHardware VAMOS division
VAMOS division was set 10 instead of 100.
CATS 1 & CATS2 scalers are sent to the two last channels of the scaler
Fixed  Tue Jul 23 17:09:43 2019 UK refugeesOTHERSocial Kipper issues.
Attachment 1: 3948.jpg
3948.jpg
Fixed  Tue Jul 23 23:22:56 2019 UK refugeesOTHERSocial Kipper issues.
I would mail you some but the postage is nuts those days!
Fixed  Wed Jul 24 12:44:55 2019 UK refugeesOTHERGeneral Fire out thanks to de Piffle
Attachment 1: BorisJ_fireman.jpg
BorisJ_fireman.jpg
Fixede768s Wed Jul 24 16:24:18 2019 NicolasOTHERGeneral 19Ne / 19F analog connections
Attachment 1: ne19_f19_AnalogStates.pdf
ne19_f19_AnalogStates.pdf ne19_f19_AnalogStates.pdf
Fixede768s Wed Jul 24 17:21:27 2019 DiegoOTHERGeneral VAMOS Acceptance with different positions of the beam in the target
Attachment 1: VAMOS_acceptance_for_different_positions_at_the_target.png
VAMOS_acceptance_for_different_positions_at_the_target.png
Fixede768s Thu Jul 25 11:05:18 2019 Francois de OliveiraOTHERGeneral Previous Papers - alpha-branching ratios for states in the 4-MeV region
In the reaction 15N(7Li,t)19F, the ratio between the measured cross sections is R(3/2+/9/2-)>29. We should get the same ratio in 19Ne. In 19Ne, the 3/2+ state decays with Eg=4033 keV (80%), the second with 2632 keV (100%), gamma efficiency = 4% and 5.5%, this gives a ratio between the number of counts measured in the 2632 keV and in the 4033 keV > 29 x 100/80 x 5.5 / 4 ~ 50.
Fixede768s Fri Jul 26 02:27:48 2019 Marlène, Faïrouz, Franck, Bertrand, Sylvain, Diego, NicolasBEAMGeneral G1 profiler efficiency
G1 profiler efficiency was checked against CATS1 for two beam intensities. Agreement between the two measurements is within 20%

Profiler -> CATS1
2.3e4 pps -> 1.8e4 pps
1.4e5 pps -> 1.16e5 pps

Conclusion:
G1 profiler is measuring correctly the beam intensity for these intensities
Pending  Fri Jul 26 11:25:46 2019 UK refugeesOTHERGeneral Be positive and the experiment will work if we put lead in our pencil
Attachment 1: bojoonazipwire.jpg
bojoonazipwire.jpg
Fixede768s Fri Jul 26 13:48:41 2019 Marlène, Faïrouz, Bertrand & NicolasBEAMGeneral G1 profiler & VAMOS efficiency
Factor of two between the profiler and VAMOS FP determination of incident beam is now understood.

Several effects were at work simultaneously and checked at 2 beam intensities (9e3 pps and 5e4 pps):
1. Profiler is using energy loss of the crossing beam particles to deduce the beam intensity. But the profiler was set to 15N instead of 15O.
We checked that for the same beam intensity of 15O the profiler gives 1.2e4 pps / 9e3 pps when tuned on 15N / 15O respectively. This amouts for 75% difference

2. Profiler tends to overestimate the real beam intensity by about 20% when compared to CATS1
Profiler / CATS1: 9e3 pps / 7.3e3 pps (81%)
Profiler / CATS1: 5e4 pps / 3.9e4 pps (78%)

3. VAMOS counting rates are lower than CATS1 by 20% which is consistent with the charge state distribution q(8+)/q(7+) = 80% calculated from LISE++
CATS1 / VAMOS: 7.5e3 pps / 5.8e3 pps (78%)
CATS1 / VAMOS: 3.9e4 pps / 3.0e4 pps (77%)

4. VAMOS PPAC efficiency was checked to be 100%: the PPAC logic signal is always present when triggerd by the drift chamber logic signal

Results:
By combining all previous factors and for 15O beam at the beginning of the experiment we obtain between the profiler and VAMOS FP a reduction factor of: 0.75 * 0.81 * 0.78 = 47%, CQFD!

Discussion:
This factor of two explains the difference we always observed when comparing estimated and measured g-ray yields.
Fixede768s Fri Jul 26 15:48:12 2019 KseniiaAGATAGeneral Full stats with AGATA on 19F with the 15N beam
Charge state 8 (plus zooms) and charge state 7;
Statistics of all the 15N runs combined;
Identification in VAMOS;

Alas, no clear peaks at 3908 and 2354.
Attachment 1: 19F_Q8_tot.png
19F_Q8_tot.png
Attachment 2: 19F_Q8_zoom1.png
19F_Q8_zoom1.png
Attachment 3: F19_Q8_tot_zoom2.png
F19_Q8_tot_zoom2.png
Attachment 4: 19F_Q7_total.png
19F_Q7_total.png