Fixed | TEST |
Thu Feb 7 09:52:08 2019 |
Freddy, Marlène | RUN | General | 3-alpha | RUN10 : 3 alpha source on MG1-3-4-5-7 |
Test run with alpha source in front of trapezoid detectors.
First run with new LABELS updated from "MM" to "MG0" in GECO.
name : run_0010.dat.06-02-19_19h08m11s (1.4 GB)
date: 06/02/2019
start: 19:08
stop: ~22:00 |
Fixed | |
Thu Feb 14 08:50:58 2019 |
Marlene | GRIT | General | 3-alpha | RUN 27 |
RUN 27 on 14th of feb.
3 alpha source (one in front of trap., one in front of square detectors)
MG10, MG4, MG7 and MG5 are OK.
MG1 shows alpha but at lower energy and with bad resolution (particularly on X). We suspect it is not well depleted. We tried to increase the bias up to 200V but did not see any change.
It is due to a non working SA/SL external cable that produces this error (we tested it with several detectors, we also tried to change the MUVI channel and it still does the same)
Conclusion :
We have tested all trapezoid detectors (MG1, MG3, MG4, MG5, MG7) the annular (MG11) and the thin square (MG10) they are all OK
Detectors (order in the GMT) :
MG10
MG1
MG4
-
MG7
MG5
-
-
attached : bias of detectors ( be careful : MG10 bias = MG11A)
spectra for all detectors |
Fixed | |
Thu Feb 14 15:30:14 2019 |
Marlene Daniele Jenny | GRIT | General | 3-alpha | RUN 28 |
RUN 28 14th Feb
3 alpha source on MG11A, MG3T, MG1T
GMT channels
MG11
-
MG3
-
-
MG1
- |
Fixed | TEST |
Thu Feb 21 09:08:49 2019 |
Freddy | RUN | General | 3-alpha | RUN 30 - alpha on MG 1,3,5,7 |
Run 30
Alpha source run for trapezoid detectors during the night with pedestals aligned and zero suppression. Cooling at 15°C.
Thresholds at 500 keV.
Run size: 1.3 G
On MUVI1:
1--> MG5 (150V)
2--> --
3--> MG3 (160V)
4--> --
On MUVI2:
5--> MG7 (120V)
6--> MG1 (140V)
7--> --
8--> --
TRIGGERS are:
GMT ch1 --> OR MUVI1
GMT ch2 --> OR MUVI2
GMT ch3 --> OR MUVI3 (nothing on it)
Plots are now represented using the correspondance map which means that it correspond to the geometrical ordering on the detector. With this representation groups of strips missing or disfunctionning on a given region of the detector is more obvious. |
Fixed | TEST |
Thu Feb 21 14:37:26 2019 |
iulian, valerian &Freddy | GRIT | Hardware | 3-alpha | Results Calibration 3-alpha RUN30 MG1, MG3,MG5 & MG7 |
Results of the calibration run 30
In order:
Telescope1 X,Y (MG5 X,Y)
Telescope4 X,Y (MG3 X,Y)
Telescope7 X,Y (MG7 X,Y)
Telescope9 X,Y (MG1 X,Y) |
Fixed | TEST |
Thu Feb 21 17:32:42 2019 |
iulian, freddy | GRIT | Hardware | 3-alpha | run 33 3 alfa with MG1,MG3,MG4,MG5,MG7,MG9 short run |
First test of the repaired 2xcables. The repaired cables were placed between MUVI1_2 <=> MG4 & MUVI1_4 <=>MG11 (longer cable)
Correspondance for the rest is the same as for run 30. The mapping is the same for all detectors (including anulaire detector).
Alpha source run for trapezoid detectors + anulaire detector with pedestals aligned and zero suppression. Cooling at 15°C.
Thresholds at 500 keV.
On MUVI1:
1--> MG5 (150V)
2--> MG4 (95V)
3--> MG3 (160V)
4--> MG11 (60V)
On MUVI2:
5--> MG7 (120V)
6--> MG1 (140V)
7--> --
8--> --
TRIGGERS are:
GMT ch1 --> OR MUVI1
GMT ch2 --> OR MUVI2
GMT ch3 --> OR MUVI3 (nothing on it) |
Fixed | TEST |
Thu Feb 21 18:14:48 2019 |
iulian | GRIT | Hardware | 3-alpha | ACQ problem run 34 3 alfa with MG1,MG3,MG4,MG5,MG7,MG9 |
Alpha source run for trapezoid detectors + anulaire detector with pedestals aligned (with HV ) and zero suppression. Cooling at 15°C.
Thresholds at 500 keV.
On MUVI1:
1--> MG5 (150V)
2--> MG4 (95V)
3--> MG3 (160V)
4--> MG11 (60V)
On MUVI2:
5--> MG7 (120V)
6--> MG1 (140V)
7--> --
8--> --
TRIGGERS are:
GMT ch1 --> OR MUVI1
GMT ch2 --> OR MUVI2
GMT ch3 --> OR MUVI3 (nothing on it)
PROBLEM ACQ , MUVI2 has no validations (DECS). MUVI1 OK!. Looks like a problem on the catchslave actor of Narval as no events seems to pass trough
The problem was caused by the cable used to triger the GMTSlave from the FanIn FanOut. |
Fixed | TEST |
Fri Feb 22 11:56:53 2019 |
iulian | GRIT | Hardware | 3-alpha | run 38 3-alpha with reordering of the MUGAST cables for the logic and analogic signals |
Run 38
Alpha source run for trapezoid detectors during the night with pedestals aligned (with HV on) and zero suppression. Cooling at 15°C.
Thresholds at 500 keV.
Important:
The MUGAST signal cables (logic & analogic) have been reordered:
On MUVI1
Channel | Spectra Label | Detector | Voltage
|
1 | MG1 | MG1 | 140V
|
2 | MG3 | MG3 | 160V
|
3 | MG4 | MG4 | 95V
|
4 | MG5 | MG5 | 150V
|
On MUVI2:
Channel | Spectra Label | Detector | Voltage
|
1 | MG7 | MG7 | 120V
|
2 | MG9 | MG11 | 60V
|
3 | MG10 | -- | --
|
4 | MG11 | -- | --
|
TRIGGERS are:
GMT ch1 --> OR MUVI1
GMT ch2 --> OR MUVI2
GMT ch3 --> OR MUVI3 (nothing on it) |
Fixed | TEST |
Fri Feb 22 14:36:35 2019 |
iulian | MUST2 | Hardware | 3-alpha | test new cable for Must2 detectors. Cables OK! |
Cables put on Muvi3.
Testing procedure. Use the numbered cables to conect the annular detector (MG11) and check if ok: GECO writting on MUVI3 , ORD Present, DECS present, STOP present.
Couldn't check the spectra as they do not exist.
MUVI3
| Entry 1 | Entry 2 | Entry 3 | Entry 4 |
| Cable 1 | Cable 2 | Cable 3 | Cable 4 |
GECO writting | OK | OK | OK | OK |
ORD Present | OK | OK | OK | OK |
DECS Present | NO | NO | NO | NO |
STOP Present | OK | OK | OK | OK |
|
|
Fixed | TEST |
Fri Feb 22 15:54:30 2019 |
iulian | GRIT | Hardware | 3-alpha | run 40 3-alpha night run |
Run 40
Alpha source run for trapezoid detectors during the night with pedestals aligned (with HV off) and zero suppression. Cooling at 15°C.
Thresholds at 500 keV.
On MUVI1
Channel | Spectra Label | Detector | Voltage
|
1 | MG1 | MG1 | 140V
|
2 | MG3 | MG3 | 160V
|
3 | MG4 | MG4 | 95V
|
4 | MG5 | MG5 | 150V
|
On MUVI2:
Channel | Spectra Label | Detector | Voltage
|
1 | MG7 | MG7 | 120V
|
2 | MG9 | MG11 | 60V
|
3 | MG10 | -- | --
|
4 | MG11 | -- | --
|
TRIGGERS are:
GMT ch1 --> OR MUVI1
GMT ch2 --> OR MUVI2
GMT ch3 --> OR MUVI3 (nothing on it)
[S]Problem MG5?![/S]---> No real problem, just spectra definition on plot was incorrect |
Fixed | TEST |
Fri Mar 1 12:07:44 2019 |
Freddy | GRIT | General | 3-alpha | Efficiency reconstruction |
blue: reconstructed efficiency using alpha run 40 (arbitratily normalized)
red: simulated efficiency
There is a difference in angle for trapezoids -> to be investigated
Remark: Simulation curve seems wrong (maybe a target in detector file not removed) |
Fixed | TEST |
Mon Mar 4 12:28:26 2019 |
Valerian and Iulian | MUST2 | General | 3-alpha | No DECS and acq error run 47-48-49. Das Saved after problem fixed |
Fixed. Solution:
1) CENTRUM not configured on GENTRUM Master
2) GMT matrix on GMT3 (2nd slave) was all 0
DAS saved after solution found |
Fixed | TEST |
Mon Mar 4 13:54:24 2019 |
iulian & Valerian | GRIT | Hardware | 3-alpha | run 50 3-alpha short run |
RUN 50
First 3a test with all Must2 detectors. HV on all set to 50 V for this test only to avoid any risks |
Fixed | TEST |
Mon Mar 4 15:30:31 2019 |
iulian Marlene & Valerian | MUST2 | Hardware | 3-alpha | run 51 3-alpha run with Pedestals Aligned @ 13 deg Measured on cooling system |
RUN 51
3a run with Pedestals Aligned but not suppressed |
Fixed | TEST |
Tue Mar 5 18:18:46 2019 |
iulian Marlene | MUST2 | Hardware | 3-alpha | run 97 3a run Muvi3 night run |
Run 87
stop@ 5:42 am
3a run Muvi3 (MM)
-4 deg
pedestals aligned and suppressed |
Fixed | TEST |
Wed Mar 6 15:02:51 2019 |
Valerian | MUST2 | Hardware | 3-alpha | Cooling temperature is relevant for the resolution |
Using run 52 -> 13deg and run 87 -> -5deg.
We see that the resolution improves at -5 degree. See attached picture as an example for MM2 X. |
Fixed | TEST |
Mon Mar 18 18:31:18 2019 |
valerian marlène | GRIT | General | 3-alpha | RUN 20 (com2019) 3 alpha overnight |
RUN20
Run with double 3alpha source MUST2+MG1_7+MG11
temperature -4 deg on both cooling
Acquisition with VAMOS+MUGAST coupled (com2019)
Time range =600ns
pedestal aligned and suppressed
run start at 18h23 monday 18
run stop at 8h34 tuesday 19
about 12Go of data |
Fixed | TEST |
Wed Mar 27 11:06:33 2019 |
Valerian and Cyril | GRIT | Hardware | 3-alpha | Evolution of resolution with temperature |
For the mugast telescope, changing the temperature from 13deg to minus 4deg doesn't seem to have an impact as strong as for the MUST2 telescopes.
Still, the resolution improve with a lower temperature, but the improvement is of only 0.1 channel... (See joined screenshots) |
Fixed | |
Thu Mar 28 18:29:04 2019 |
marlene valerian | GRIT | General | 3-alpha | RUN 147 3alpha source overnight all detectors -4deg |
RUN 147 start@ 18:30
3 alpha source over night
cooling at -4 deg on all detectors
warning !!! FIFO for BLK signal has been put in 4x4 instead of 2x8, BLK is only coming from the 3 VXI of MUGAST |
Fixed | com2019 |
Fri Mar 29 17:11:04 2019 |
OttoRun | RUN | General | 3-alpha | Run155 |
*************************
Run : 155
*************************
Start at 2019-03-29 17:10:50.177220
Run Conditions:
- Operator: test
- Trigger:
- Target/Source:
- Beam:
- Intensity:
Comment:
Stop at 2019-03-29 17:11:37.286667
Stop comment: stop |