  MUGAST, Page 108 of 132  Not logged in MUGAST
Fixed, , Thu Feb 27 15:54:09 2020, Marlène , RUN, General, , RUN 141 junk (problem at start) 
 
Fixed, TEST, Wed Jan 16 16:33:28 2019, Marlène, GRIT, General, 3-alpha, Status of detectors on 16/01/19 Detectors_status_160119.xlsxCapture_d’eÌcran_2019-01-16_aÌ€_17.05.57.png
MG1 has some issue with the HT cable (for the BT bias). There is a shortcut only under vacuum. We checked the connector but we have to do vacuum again to
check it.
Fixed, , Tue Jan 22 11:15:15 2019, Marlène, GRIT, General, N/A, Square detectors MG9C_disabled_channels__no_crazy_ORD.pngMG9C_disabled_channels__no_crazy_ORD_moredetailed.pngMG9C_piedestal_of_disabled_channels_Y.png
MG9C (thin square detector coupled with thick square detector unplugged) noisy
On X channels --> no noise problem ( about 100 ORD in the dark)
On Y channels --> MATE 10 & 11 noisy ! (see picture below with noisy Y piedestals)
Fixed, , Wed Jan 23 12:24:10 2019, Marlène, GRIT, General, , Change cable MG1T - Exchange of detectors named MG9C & MG10C 
23/01/19

Change of the HT/BT cable of MG1T (the one that had a short cut before when the chamber was closed)
Fixed, , Fri Jan 25 17:02:25 2019, Marlène, GRIT, General, , Detectors numbering MUGAST_detectors.pdf
 
Fixed, , Tue Mar 12 10:21:17 2019, Marlène, MUST2, General, , Debug Calimero Y test_GECO_masquage_mate2&10activees_seulement.png
Debug of Calimero

test of channels (mate) masking : OK with DAS, OK with GECCO (see attachment) when done in the "disable channels" of GECCO
Fixed, , Thu Apr 4 17:55:47 2019, Marlène, GRIT, General, , Scalers 
The scalers are

Scaler 1:
Fixed, com2019, Fri Apr 5 00:47:16 2019, Marlène, OTHER, General, N/A, Game plan gameplan_comMUGAST.pdf
 
Fixed, e744s, Thu Apr 11 02:53:20 2019, Marlène, RUN, General, , TAc VAMOS-CATS 
Just before RUN 68
TAC VAMOS-CATS is now in TAC CATS1-CATS2 in order to check if we see something in the online spectra.
TCATS total delay = 300 + 500 ns=800 ns
Fixed, , Thu Apr 11 05:03:13 2019, Marlène, RUN, Hardware, CH2, TO DO list 
- 2 TACS are not working: 
[LIST]
[*] T_Si_CATS1 (stop should be CATS n FAG --> to be checked)
    Fixed, e744s, Thu Apr 11 18:08:12 2019, Marlène, OTHER, Hardware, , changes in the electronic 
[quote="Valérian, Chloé and François"]Changes made in the electronic during the break between 14N and 14O

1) T_VAMOS-CATS1
Fixed, e744s, Thu Apr 11 19:47:15 2019, Marlène, OFFLINE, General, , CATS charges in RUN76 and RUN77 CATS_stripY_run76.pdfCATS_stripX_run76.pdfCATS_stripY_run77.pdfCATS_stripX_run77.pdfCATS_RUN77.pdf
Offline analysis to reconstruct the charges in X and Y for CATS.
The charges are lower in RUN77 because bias was reduced to 580V.
Wires seem to be all OK.
Fixed, e744s, Fri Apr 12 08:19:31 2019, Marlène, OFFLINE, General, , CATS profiles RUN87 CATS_pos_RUN87.pngCATS_chargeX_run87.pngCATS_chargeX_2_run87.pngCATS_chargeY_run87.pngCATS_ChargeY_2_run87.png
 
Fixed, e744s, Fri Apr 12 09:25:10 2019, Marlène, ONLINE, General, , TAC Capture_du_2019-04-12_09-24-34.png
 
Fixed, e744s, Mon Apr 15 09:59:32 2019, Marlène, RUN, General, CH2, change thresholds in DSSD MUST2 and suppression CsI 
From RUN126
change of MUST2 DSSD thresholds : thresholds are 583 keV now
change of CsI suppression : before it was suppressed up to 8210 now it is 8200 
Fixed, e744s, Tue Apr 16 11:01:04 2019, Marlène, RUN, General, , statistics on each target 
Cumulated statistics (evaluated from CATS trigger) on each target is : 
-target #3 : 7.547e5 (x 30000) 
-target #5 : 2.02e5 (x30000)
Fixed, e744s, Thu Apr 18 09:28:12 2019, Marlène, MUST2, General, 3-alpha, RUN 206 (mugast) 3Alpha calib after e744s experiment run206_calibMMY.pngrun206_calibMGY.pngrun206_calibMMX.pngrun206_calibMGX.png
RUN 206
3 alpha source all detectors with no change in the configuration (low thresholds in MUST2)
From 17/04 18h13 -> 18/04 9h31
Fixed, e786s, Fri May 31 09:53:14 2019, Marlène, GRIT, General, , Alpha source --> status of detectors 
MG3T has been fixed by changing the X kapton.

I have removed aluminium foil on square detectors, the noise is slightly higher.
Fixed, e786s, Mon Jun 3 18:00:10 2019, Marlène, GRIT, General, , MUGAST detectors status --> MG4T  
The problem of MG4T ha sbeen solved by changing the feedthrough.
The pulser on Y was not working and we were seeing no events (no ORD) because there was a problem with the bias of board. The issue is due to the feedthrough.
So now MG4T (HT &BT) are plugged in a free feedthrough (below MG1T) and I see pulser signals on Y. 
Fixed, e786s, Mon Jun 24 03:08:49 2019, Marlène, RUN, General, Cryo-3He, Brho issue Brho_vamos.png
We observe a change in the Brho of VAMOS : the charged states are divided in 2 !!!

Édit: it is now understood as ice water cracking on the upstream of the target.