Fixed | e744s |
Thu Apr 11 18:20:27 2019 |
Valerian and Diego | OFFLINE | Software | CH2 | Timing issue between vamos and mugast |
A cut is seen on the TAC between MUGAST and the focal plane multiwire of VAMOS when one select XF or DC0_Y>-1500
Edit: This must be a reconstruction problem as it is not seen in the Raw files (using DC0_X for example)
Edit: This is not only a reconstruction problem as the cut appear again if one select DC0_Y
Edit: We changed VAMOS FAG validation window from 500ns to 800ns |
Attachment 1: T_MUGAST_FPMW.png
|
|
Attachment 2: T_MUGAST_FPMW1.png
|
|
Fixed | e744s |
Thu Apr 11 18:08:12 2019 |
Marlène | OTHER | Hardware | | changes in the electronic |
Valérian, Chloé and François wrote: | Changes made in the electronic during the break between 14N and 14O
1) T_VAMOS-CATS1
it was not connected. We added 500 ns delay in CATS1
2) T_CATS1_HF
range changed from 2µs to 200 ns |
T_VAMOS-CATS1 is now connected in T_CATS1_CATS2 as posted this night. We have tried to add 500ns more delay (800ns in total) to CATS and it did not work, so we suspected the TAC or ADC. This is why we inverted T_VAMOS-CATS1 with T_CATS1_CATS2. For this TAC, CATS is the START (and is not delayed by 500ns). |
Fixed | e744s |
Thu Apr 11 17:43:44 2019 |
OttoRun | RUN | General | Empty | Run80 |
*************************
Run : 80
*************************
Start at 2019-04-11 17:42:44.178301
Run Conditions:
- Operator: Adrien
- Trigger: VAMOS %1
- Target/Source: hole
- Beam: 14O
- Intensity: ~ 5e+3
- Brho = 0.69186 Tm
- !CATS/VAMOS (CATS Efficiency)! : XXXX
Comment:
After removing division on Cats and Vamos.
Aligning beam, daq on for online monitoring
Stop at 2019-04-11 18:50:52.122268
Stop comment: Stop to remove target hole |
Fixed | e744s |
Thu Apr 11 17:38:28 2019 |
iulian | RUN | Hardware | Empty | Setting trigger Vamos and Cats |
Removed division %40000 on Cats and %1000 on Vamos.
Removed Cats2 from the beam |
Fixed | e744s |
Thu Apr 11 17:38:24 2019 |
mugast (beyhan) | BEAM | General | | 14O8+ beam characteristics |
Brho 14O8+ = 0.69186 Tm
update Adrien:
La dream team de Division (non) Opération n'arrive pas a mesurer l’énergie en sortie de CIME (WTF??!!)
Du coup ils ont mis une énergie fausse mais avec une super précision de 1.8 Mev/A dans la fiche.
#onlyatganil |
Attachment 1: image1.jpeg
|
|
Attachment 2: IMG_1581.jpg
|
|
Fixed | e744s |
Thu Apr 11 17:37:29 2019 |
Valérian, Chloé and François | OTHER | Hardware | | changes in the electronic |
Changes made in the electronic during the break between 14N and 14O
1) T_VAMOS-CATS1
it was not connected. We added 500 ns delay in CATS1 (CATS2 in reality) (48+150ns before)
2) T_CATS1_HF
range changed from 2µs to 200 ns |
Fixed | e744s |
Thu Apr 11 17:35:21 2019 |
François | BEAM | General | | 14O beam is in G1 since 17H00 |
alignment of the beam in VAMOS
intensity ~ 5x10^3 pps |
Fixed | TEST |
Thu Apr 11 17:30:58 2019 |
OttoH.V. | OTHER | Hardware | N/A | HVM_ERROR |
*** hvm watcher error ***
FP_DC1_w1 Slot: 2 (A1821HMP) Channel:0 Vmon: 674.25 V Imon: 0 uA Status: OVER CURRENT
|
Fixed | e744s |
Thu Apr 11 17:08:39 2019 |
OttoRun | RUN | General | Empty | Run79 |
*************************
Run : 79
*************************
Start at 2019-04-11 17:08:21.837538
Run Conditions:
- Operator: Adrien
- Trigger: VAMOS
- Target/Source: hole
- Beam: 14O
- Intensity: ~ 3e+3
- Brho = ?? Tm
- !CATS/VAMOS (CATS Efficiency)! : XXXX
Comment:
Aligning beam, daq on for online monitoring
Stop at 2019-04-11 17:31:48.540765
Stop comment: Entering to Remove CATS2 and remove VAMOS %1000 |
Fixed | e744s |
Thu Apr 11 16:56:34 2019 |
OttoRun | RUN | General | Empty | Run78 |
*************************
Run : 78
*************************
Start at 2019-04-11 16:55:28.720017
Run Conditions:
- Operator: Adrien
- Trigger: Mugast
- Target/Source: hole
- Beam: 14O
- Intensity: ~ 3e+3
- Brho = ?? Tm
- !CATS/VAMOS (CATS Efficiency)! : XXXX
Comment:
Alignign beam, daq on for online monitoring
Stop at 2019-04-11 17:07:49.331973
Stop comment: changing trigger to vamos |
Fixed | e744s |
Thu Apr 11 16:07:59 2019 |
Valerian and Louis | OFFLINE | Software | CH2 | HF resolution |
Using the value of the HF=83,4ns. We can obtain the calibrated plot of T_CATS2_HF which is joined and gives us roughly a 2-3 ns resolution for the HF |
Attachment 1: T_CATS2_HFns.png
|
|
Fixed | e744s |
Thu Apr 11 15:16:10 2019 |
Adrien, Valerian, Louis | CATS | General | CH2 | run77 CATS efficiency |
CATS efficiency has been computed for run 77 using T_FagHF an T_Si_Cats1 with different triggers:
Trigger GATCONF_MASTER<8 (MUST+Mugast) : 94%
Trigger GATCONF_MASTER=4 (MUST) : 99.4%
Trigger GATCONF_MASTER=2 (square) : 85%
Trigger GATCONF_MASTER=1 (trapezoidal+annular) : 22%
The 22% resolution is due to the fact that the trapezoidal detectors are noisy. Doing the same on an analysed run we obtain ~ 85% efficiency |
Fixed | e744s |
Thu Apr 11 14:03:45 2019 |
Valerian | CATS | Hardware | N/A | In case of a CATS2 tripped |
Currently, CATS2 is inserted, if the window :
CATS2 has tripped appear, one should do the following :
- Open a terminal
- Type: HV
- Username is : admin
- Password is : admin
- Press enter twice
The list of channels is going to appear.
- Using the arrow of the keyboard, go down until you reach cats2
- then go right until you reach Off
- press space, the dectector should start ramping up
- If it trips again, you can try to put voltage to cats1 (right above cats2)
Two possibility :
- If cats1 trip it is the gas bottle that is empty call : 07 83 34 30 81 (Sebastien)
- If cats1 doesn't trip try cats2 again, and if it trips lower the voltage of cats2
To lower the voltage of cats2 : go to the left until you reach the first cell after cats2 name and lower the voltage 10 volts and try to start cats2 again.
THEN CLOSE THE TERMINAL |
Fixed | TEST |
Thu Apr 11 13:55:54 2019 |
OttoH.V. | OTHER | Hardware | N/A | HVM_ERROR |
*** hvm watcher error ***
cats2 Slot: 13 (A1835 ) Channel:1 Vmon: 193.6 V Imon: 0 uA Status: INTERNAL TRIP
|
Fixed | TEST |
Thu Apr 11 13:51:54 2019 |
OttoH.V. | OTHER | Hardware | N/A | HVM_ERROR |
*** hvm watcher error ***
cats2 Slot: 13 (A1835 ) Channel:1 Vmon: 170.3 V Imon: 0 uA Status: INTERNAL TRIP
|
Fixed | TEST |
Thu Apr 11 13:50:22 2019 |
OttoH.V. | OTHER | Hardware | N/A | HVM_ERROR |
*** hvm watcher error ***
cats2 Slot: 13 (A1835 ) Channel:1 Vmon: 0 V Imon: 0 uA Status: INTERNAL TRIP
|
Fixed | TEST |
Thu Apr 11 13:47:10 2019 |
OttoH.V. | OTHER | Hardware | N/A | HVM_ERROR |
*** hvm watcher error ***
cats2 Slot: 13 (A1835 ) Channel:1 Vmon: 200.7 V Imon: 0 uA Status: INTERNAL TRIP
|
Fixed | e744s |
Thu Apr 11 13:43:26 2019 |
Valerian | OFFLINE | Software | CH2 | Target position for 14N runs |
|
Attachment 1: XY_run14N.png
|
|
Attachment 2: Xshift_run14N.png
|
|
Attachment 3: Yshift_run14N.png
|
|
Fixed | TEST |
Thu Apr 11 13:43:20 2019 |
OttoH.V. | OTHER | Hardware | N/A | HVM_ERROR |
*** hvm watcher error ***
cats2 Slot: 13 (A1835 ) Channel:1 Vmon: 189.8 V Imon: 0 uA Status: INTERNAL TRIP
|
Fixed | TEST |
Thu Apr 11 13:43:10 2019 |
OttoH.V. | OTHER | Hardware | N/A | HVM_ERROR |
*** hvm watcher error ***
cats2 Slot: 13 (A1835 ) Channel:1 Vmon: 541.2 V Imon: 0.9 uA Status: OVER CURRENT
|