Fixed, e768s, Fri Jul 26 13:48:41 2019, Marlène, Faïrouz, Bertrand & Nicolas, BEAM, General, , G1 profiler & VAMOS efficiency 
Factor of two between the profiler and VAMOS FP determination of incident beam is now understood.

Several effects were at work simultaneously and checked at 2 beam intensities (9e3 pps and 5e4 pps):
Fixed, , Sun Jul 21 16:51:30 2019, Marlène, Christian, Ksenia, BEAM, General, , trial of tracking of the implantation of 15O 
Looking at AGATA count rate, we have seen that the highest counting rate for 511 keV line is seen in the inner ring (0*A clusters).
The outer ring detectors (10B for example) counts 10 times less.
Fixed, e768s, Sun Jul 21 16:39:38 2019, Marlène, Antoine, BEAM, General, , IMPORTANT :How to send scalers to PCP 
Go to PCP

Open a terminal
Fixed, e744s, Tue Apr 9 18:08:22 2019, Marlène and François, DAQ, Software, , new inputs in the ADC Mugast 
new inputs in the ADC mugast
T_Mugast_HF = input N°2
E_AGATA =  agata 42 = input N°3
Fixed, , Fri Apr 5 19:05:56 2019, Marlène Nicolas, GRIT, General, , Brho beam & 17O 
Brho of  16O @ 6 MeV/A after 1mg/cm2 = 0.68906
Brho of 17O for g.s. after 1 mg/cm2=0.71065
Brho of 17O @ 871 keV after 1mg/cm2=0.70831
Fixed, , Mon Feb 24 22:34:43 2020, Marlène Nicolas, RUN, General, , Screenshots run 126 (target CD2+Au #2 rotated 180 deg) run126_theta_Elab.pngrun126_Tof.png
 
Fixed, e775s, Thu Feb 27 12:35:39 2020, Marlène Nicolas, GRIT, General, , RUN128-137_138 analysis Elab_Thetalab_RUN128_137_138.pngExEg_RUN128_137_139.pngEx_RUN128_137_138.png
Kinematic lines for the full statistics up to now
States above E*>7 MeV are cut in the angles >140 because of energy threshold.
Fixed, e786s, Mon Jun 24 03:52:11 2019, Marlène Matthieu, RUN, General, Cryo-3He, 3He target filled with 700 mb 
The 3He cryo target is filled with a pressure of 700 mb
The temperature on the target is 5.98K
Fixed, TEST, Thu Mar 7 14:02:03 2019, Marlène Iulian, MUST2, General, N/A, test of MUST2 electronics  
Test with pulser at 20 kHz
The OR of MUVI was scaled down by a factor of 100, to provoke some resets
Fixed, e744s, Thu Apr 11 05:08:20 2019, Marlène Iulian, RUN, General, CD2, Screenshots from RUN 74 8x
RUN 74 with 14N beam and CD2 104um target
First screenshots from MUGAST
Fixed, e775s, Sun Mar 1 17:04:43 2020, Marlène François, RUN, General, , Preliminary analysis of runs from 128 up to now (169) 6x
Preliminary analysis with all stat up to now.
Fixed, , Thu Jul 11 11:40:18 2019, Marlène & Franco, OTHER, Hardware, , Some changes in the electronic scheme  
- We checked the TACs with a time calibrator, all of them are working. Maybe to check the ADCs if some timings are not there during the exp.

- We set the possibility to have MUST2 divided by 100 (on a separate chain) in case of high rate on GMT channel number 7.
Fixed, TEST, Tue Dec 3 18:14:45 2019, Marlène & Franco, GRIT, Hardware, , Pulser test of new board 
The new board that will be installed for MG2 has been tested and validated with a pulser outside of the chamber.
Fixed, , Wed Jan 22 18:39:54 2020, Marlène & Franco, GRIT, Hardware, , Tests of the new box box_pin_map.pdf
The new box has been tested together with the new CAEN SY4527 power supply. We connected a laptop with Linux to the power supply by means of an ethernet
cable and communicated with it by opening a browser (http://192.168.0.1, then username: admin and pwd: admin). To control the channels we open a terminal,
go to the folder where we saved the file ChannelsController.jar and run the command: java -jar ChannelsController.jar 192.168.0.1 admin admin
    Fixed, TEST, Fri Feb 14 10:49:22 2020, Marlène & Franco, RUN, General, 3-alpha, MG11 issues 
We analyzed the behaviour of the annular. We realized that the 4-peaks behaviour on X was already present last year but only on one sector, now it's on
two sectors. We noticed that there is a correlation between the malfunctioning sectors and the missing strips on Y. Last year there was only a missing
strip on Y, affecting only one sector on X; now 3 strips are missing on Y, affecting 2 sectors on X. We believe that the fact that the boundings are missing
Fixed, e775s, Tue Feb 25 11:50:16 2020, Marlène & Franco, OFFLINE, General, , PID with gate on T_FPMW_HF MG1_gatetime_run128.pngannular_gatetime_run128.png
The black points are what remains after putting the condition abs(T_FPMW_HF-13000)<1000, for MG1 and the annular.
Fixed, TEST, Wed Mar 11 18:12:40 2020, Marlène & Franco, GRIT, Hardware, N/A, Update on Mugast 
(Probably) final update of the status of the detector before the experiment:

- MG5 does not trigger on Y (no pulser, no alpha source). This can explain why in the last experiment we could not run the time calibration of its Y.
Fixed, TEST, Thu Mar 12 10:14:02 2020, Marlène & Franco, GRIT, Hardware, N/A, Debugging the cooling system 
We realized that the increase of temperature of MG7 and MG9 during run 600 (alpha calibration) was due to the fact that a loop of the cooling system was
not working since the pipes were not properly connected. We re-connected them and switched on the low voltage without making the vacuum just to monitor
the evolution of the temperature.
Fixed, TEST, Thu Mar 12 14:47:59 2020, Marlène & Franco, GRIT, Hardware, N/A, Moving MG9 square to MUVI2 
- Since we are not using MG6 anymore (card burnt), we replaced it with MG9 on slot 3 of MUVI2. MUVI4 is now unused. We changed the actions parameters accordingly.
- We cannot run the time calibration for MG9 because the pulser does not go through neither on X nor on Y.
Fixed, e793s, Thu Mar 12 14:54:13 2020, Marlène & Franco, GRIT, General, 3-alpha, Run 0601 - alpha source calibration HV_run601.png
We start the alpha source calibration run with pedestal suppression (except for MG7 and MG9) and cooling at 0 deg.
The HV for MUST2 are in attachment. It is not clear whether they will remain unchanged during the experiment since some currents are high.