Status |
ExpNbr |
Date |
Author |
Type |
Category |
Target-Source |
Subject |
Text |
 |
Fixed | e768s |
Fri Jul 26 13:48:41 2019 |
Marlène, Faïrouz, Bertrand & Nicolas | BEAM | General | | G1 profiler & VAMOS efficiency | Factor of two between the profiler and VAMOS
FP determination of incident beam is now
understood.
|
|
Fixed | |
Sun Jul 21 16:51:30 2019 |
Marlène, Christian, Ksenia | BEAM | General | | trial of tracking of the implantation of 15O | Looking at AGATA count rate, we have seen
that the highest counting rate for 511 keV
line is seen in the inner ring (0*A clusters).
|
|
Fixed | e768s |
Sun Jul 21 16:39:38 2019 |
Marlène, Antoine | BEAM | General | | IMPORTANT :How to send scalers to PCP | Go to PCP
Open a terminal
|
|
Fixed | e744s |
Tue Apr 9 18:08:22 2019 |
Marlène and François | DAQ | Software | | new inputs in the ADC Mugast | new inputs in the ADC mugast
T_Mugast_HF = input N°2
E_AGATA = agata 42 = input N°3
|
|
Fixed | |
Fri Apr 5 19:05:56 2019 |
Marlène Nicolas | GRIT | General | | Brho beam & 17O | Brho of 16O @ 6 MeV/A after 1mg/cm2 = 0.68906
Brho of 17O for g.s. after 1 mg/cm2=0.71065
Brho of 17O @ 871 keV after 1mg/cm2=0.70831 |
|
Fixed | |
Mon Feb 24 22:34:43 2020 |
Marlène Nicolas | RUN | General | | Screenshots run 126 (target CD2+Au #2 rotated 180 deg) | |
 |
Fixed | e775s |
Thu Feb 27 12:35:39 2020 |
Marlène Nicolas | GRIT | General | | RUN128-137_138 analysis | Kinematic lines for the full statistics up
to now
States above E*>7 MeV are cut in the angles |
  |
Fixed | e786s |
Mon Jun 24 03:52:11 2019 |
Marlène Matthieu | RUN | General | Cryo-3He | 3He target filled with 700 mb | The 3He cryo target is filled with a pressure
of 700 mb
The temperature on the target is 5.98K
|
|
Fixed | TEST |
Thu Mar 7 14:02:03 2019 |
Marlène Iulian | MUST2 | General | N/A | test of MUST2 electronics | Test with pulser at 20 kHz
The OR of MUVI was scaled down by a factor
of 100, to provoke some resets |
|
Fixed | e744s |
Thu Apr 11 05:08:20 2019 |
Marlène Iulian | RUN | General | CD2 | Screenshots from RUN 74 | RUN 74 with 14N beam and CD2 104um target
First screenshots from MUGAST
|
8x |
Fixed | e775s |
Sun Mar 1 17:04:43 2020 |
Marlène François | RUN | General | | Preliminary analysis of runs from 128 up to now (169) | Preliminary analysis with all stat up to now. |
6x |
Fixed | |
Thu Jul 11 11:40:18 2019 |
Marlène & Franco | OTHER | Hardware | | Some changes in the electronic scheme | - We checked the TACs with a time calibrator,
all of them are working. Maybe to check the
ADCs if some timings are not there during |
|
Fixed | TEST |
Tue Dec 3 18:14:45 2019 |
Marlène & Franco | GRIT | Hardware | | Pulser test of new board | The new board that will be installed for MG2
has been tested and validated with a pulser
outside of the chamber. |
|
Fixed | |
Wed Jan 22 18:39:54 2020 |
Marlène & Franco | GRIT | Hardware | | Tests of the new box | The new box has been tested together with
the new CAEN SY4527 power supply. We connected
a laptop with Linux to the power supply by |
|
Fixed | TEST |
Fri Feb 14 10:49:22 2020 |
Marlène & Franco | RUN | General | 3-alpha | MG11 issues | We analyzed the behaviour of the annular.
We realized that the 4-peaks behaviour on
X was already present last year but only |
|
Fixed | e775s |
Tue Feb 25 11:50:16 2020 |
Marlène & Franco | OFFLINE | General | | PID with gate on T_FPMW_HF | The black points are what remains after putting
the condition abs(T_FPMW_HF-13000)<1000,
for MG1 and the annular. |
 |
Fixed | TEST |
Wed Mar 11 18:12:40 2020 |
Marlène & Franco | GRIT | Hardware | N/A | Update on Mugast | (Probably) final update of the status of the
detector before the experiment:
|
|
Fixed | TEST |
Thu Mar 12 10:14:02 2020 |
Marlène & Franco | GRIT | Hardware | N/A | Debugging the cooling system | We realized that the increase of temperature
of MG7 and MG9 during run 600 (alpha calibration)
was due to the fact that a loop of the cooling |
|
Fixed | TEST |
Thu Mar 12 14:47:59 2020 |
Marlène & Franco | GRIT | Hardware | N/A | Moving MG9 square to MUVI2 | - Since we are not using MG6 anymore (card
burnt), we replaced it with MG9 on slot 3
of MUVI2. MUVI4 is now unused. We changed |
|
Fixed | e793s |
Thu Mar 12 14:54:13 2020 |
Marlène & Franco | GRIT | General | 3-alpha | Run 0601 - alpha source calibration | We start the alpha source calibration run
with pedestal suppression (except for MG7
and MG9) and cooling at 0 deg.
|
|