MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUST2, Page 45 of 48  Not logged in MUST2
Ongoing, e744, Sat May 26 20:23:52 2018, Cristina Iulian, OFFLINE, General, , PPT gamma spectrum runs_target4_proton_prompt_PPT.pdf
Attached is the gamma spectrum from PPT detector with a condition on prompt (T_MM_PPT), proton detected in MUST and multiplicity event equal to 1
Fixed, e744, Fri May 25 00:09:03 2018, Cristina , RUN, General, CH2, run 1003 (target no 6 -CH2 180 um) 6x
run 1003

14N

start@ 23:18 24/05/2018

stop@ 000:33
Ongoing, e744, Fri May 25 15:19:38 2018, Cristina, OFFLINE, General, , isomers present in the 14N runs 14Nruns_isomers_EXL.pdf14Nruns_isomers_PPT.pdf

Attached are some pictures of the energy_vs_time spectra for EXL and PPT detectors recorded during 14N runs

Still to do: time alignment of EXL detectors
Ongoing, e744, Tue Apr 17 17:36:37 2018, Cherry Mx Team, MUST2, Software, N/A, GRU SUCKS 
Pretty much today's summary in the title.

https://i.pinimg.com/originals/a8/d4/12/a8d41206e8bd9c189397fb4907656981.jpg
Fixed, e755, Wed Jun 20 16:45:00 2018, Cherry Mx Team, TARGET, Hardware, 3-alpha, Cryo target is filled targefilled.png
We dont see any alpha in detector 1-6m but we see them in detector 7-8.


Data are in run 3038 in e744 data folder.
Fixed, e755, Mon Jun 25 17:23:49 2018, Cherry Mx Team, CATS, General, Cryo-LH2, Confirmation of CATS reconsctruction 
Gating on T1 to T4 and looking at the the CATS reconstruction on Target we confirm that;
- CATS is correlated to MUST
- Telescope Position matches CATS Position, so Telescope numerotation is correct and there is no inversion in CATS.

In attached picture you can see CATS position on target gated by:
- Red: T1
- Green: T2
- Violet: T3
- Noir: T4
Fixed, e755, Mon Jun 25 17:27:59 2018, Cherry Mx Team, CATS, General, Cryo-LH2, Confirmation of CATS reconsctruction CATS_TELESCOPE.pngCATS_TELESCOPE_M2.png
Gating on T1 to T4 and looking at the the CATS reconstruction on Target we confirm that;
- CATS is correlated to MUST
- Telescope Position matches CATS Position, so Telescope numerotation is correct and there is no inversion in CATS.

In attached picture you can see CATS position on target gated by:
- Red: T1
- Green: T2
- Violet: T3
- Noir: T4

Second plot with MUST reconstruct coordinate. All is consistent.
Fixed, e744, Tue Apr 17 09:18:00 2018, Cherry MX team, MUST2, Hardware, N/A, T5 is back and running strong T5.pdf
T5 telescope appears to be fully functional again. It shows both Energy and Time with excellent resolution.
Fixed, e744, Wed May 23 15:40:58 2018, Charles Houarner Valérie, MUST2, Software, N/A, TESTS MUVI internal pulser T1-T4 strips X testGeneInterneT4plotCh5Mate.pngtestGeneInterneT4plotCh5MatePied.pngTestMust2tel4matePulserInt.pngTestMust2tel4PulserIntMate5.pngTestMust2telT1toT3andT5mat5onT4.png
To check the signals and trigger on T4 we used the internal pulser on T4,
from the SYNCHROTEST pulser at 150 Hz.
The signal was set on the channel 5 of each Mate of the MufeeX card of the T4 telescope:
MM4_STRX_E5, E21, E37, E53, E69, E85, e101, E117.
Figure testGeneInterneT4plotCh5Mate.png
The signal of the pulser can be seen around 9400 to 9850 on the several channels (see the zooms). the pedestal is large, around 8085 to 8100.
on the other channels see for instance the zoom on Strx_E118.
Pulser on channel 5 of 4 Mate of MufeeX TestMust2tel4matePulserInt.png

With the pulser only on channel 5 of Mate-1 of T4,
see the thinner pedestals on the other channels, in the plots:
TestMust2tel4PulserIntMate5.png
Checking the pedestals on the other telescopes: TestMust2telT1toT3andT5mat5onT4.png
Fixed, e755, Tue Jun 26 14:39:52 2018, Charles H, Fred, Valérie, RUN, General, Cryo-LH2, RUN 147 6x
Run 147 -- 26th June start@ 14h34

We make some checkings of DT during the beam stop due to the source disjonction.
Triggers: all MM + CATS DIV. FAG is around 260/s due to the source counting in Must2.
Fixed, e748, Thu Jul 6 10:39:09 2017, BéaF & ValérieL - shift U, OFFLINE, General, , Test of merge on Run 339 run339_ICEdiff_LTSDIff_TrigM2.pngrun339_TAC_LTSDIff_TrigM2.png
spectra of the merge checked by Béa.
The merge is correct on the correlations cf plots below.

To Check Merger use:

ssh -X e748chio@ganp976
cd Merge_CHIO
./ConvertRun.sh runnumber

Roughly 30% of events are merge.

Note that we had to include a new condition in the loop of Merge.C in order to avoid the loop to get stuck with the multihits at the end of the run.

>>while( TMath::Abs(t1-t2-ftime_center) <= ftime_window && i1<Entries_numexo && i2<Entries_vme)



Then to check the correlations:
Go to:

cd Merger/merge
Open with ROOT the last file that was created with the runnumber and Draw
merger->Draw("TPLCHIO:LTSDiff>>a(400,-150.,150.,400,10.,8000.)","","colz");

This is to verify that the Driftime is correlated

Also Draw
merger->Draw("IC_E-IC_E_ANA:LTSDiff>>a(400,-150.,120.,400,-18000.,1500.)","","colz");
This is to verify that the difference gives a horizontal line centred at zero.

Save the plots and put them in the elog.


Validated by Antoine Lemasson
Fixed, e738, Mon Jul 9 00:03:24 2018, Benoit, Megumi, Takeshi, Valérie, RUN, General, Cryo-LH2, RUN 298 run298_ScalersAt0h10.pngscaler_2018-07-09_02h01m34s.pngvisu_2018-07-09_02h01m41s.pngvisu_2018-07-09_02h01m44s.pngvisu_2018-07-09_02h01m52s.png
Run 298 start 0h (9th July) stop at 02h02
e748 Exxx 2,9G 9 juil. 02:02 run_0298.dat.08-07-18_23h59m40s
Beam stop occurred from 01h06 to 01h09 (rotation speed failing of the LISE target)

MISSING STRIP, Y12 of CATS1 since run 296.
CATS 1 640V I ~ 0.14 ľA
CATS2 670V I ~ 0.14 ľA
Beam 10C around 8. 10**4 and 10**5/s
Dead time between 7 and 13%
Fixed, e738, Mon Jul 9 02:16:19 2018, Benoit, Megumi, RUN, General, Cryo-LH2, RUN 299 scaler_2018-07-09_02h20m00s.pngvisu_2018-07-09_04h04m37s.pngvisu_2018-07-09_04h04m53s.png
Run 299 start 02h01 (9th July) stop at 04h03
e748 Exxx 2,9G 9 juil. 04:03 run_0299.dat.09-07-18_02h01m29s

MISSING STRIP, Y12 of CATS1 since run 296.
CATS 1 640V I ~ 0.12 ľA
CATS2 670V I ~ 0.12 ľA
Beam 10C around 8. 10**4 and 10**5/s
Dead time between 7 and 13%
Fixed, e738, Mon Jul 9 04:22:21 2018, Benoit, Megumi, RUN, General, Cryo-LH2, RUN 300 scaler_2018-07-09_04h17m03s.pngvisu_2018-07-09_05h55m54s.pngvisu_2018-07-09_05h56m57s.pngvisu_2018-07-09_06h00m30s.png
Run 300 start 04h02 (9th July) stop at 06h04
e748 Exxx 2,8G 9 juil. 06:05 run_0300.dat.09-07-18_04h02m08s

MISSING STRIP, Y12 of CATS1 since run 296.
CATS 1 640V I ~ 0.12 ľA
CATS2 670V I ~ 0.12 ľA
Beam 10C around 8. 10**4 and 10**5/s
Dead time between 7 and 13%
Fixed, e738, Mon Jul 9 06:21:54 2018, Benoit, Megumi, RUN, General, Cryo-LH2, RUN 301 scaler_2018-07-09_06h13m13s.pngscaler_2018-07-09_07h51m37s.png
Run 301 start 06h04 (9th July) stop at 7h51


MISSING STRIP, Y12 of CATS1 since run 296.
CATS 1 640V I ~ 0.12 ľA
CATS2 670V I ~ 0.12 ľA
Beam 10C around 8. 10**4 and 10**5/s
Dead time between 7 and 13%
Fixed, e755, Sun Jul 1 09:43:58 2018, Benoit, Franck, RUN, General, Cryo-LH2, Run 209 9x
Sun 1 July 2018

Start 9:26
Stop 11:00

STOP BECAUSE OF SOURCE PROBLEM (No beam)

CATS1: 17 kcps
CATS2: 17 kcps
Plastic: 27 kcps

Trigger: MUST2, CATS1div, CATS2div

37Ca beam

Continue
Fixed, e755, Sun Jul 1 11:52:27 2018, Benoit, Franck, RUN, General, Cryo-LH2, Run 210 7x
Sun 1 July 2018

Start 11:46
Stop 11:57

(-> PCP retunes the beam)

37Ca continue

CATS1: 20 kcps
CATS2: 20 kcps
Plastic: 29 kcps
Fixed, e755, Mon Jun 25 16:09:24 2018, Benoit Louis, RUN, General, Cryo-LH2, Beam and CHIO tunning Run126_deTOF.pngRun127_BeamProfil.pngRun126_Scaler.png
During run 126:
Beam was tuned to be centered and homogeneous in the target
efficiency and resolution of the CHIO was improved
Fixed, e748, Thu Jul 6 14:07:20 2017, Bea, OFFLINE, General, CD2, Convert & Merge Numexo Runs in ganp166 
To Convert run DAQ_general

offline
npp e748
cd offline
Offline /data/e748X/e748acquisition/run/name of run to conver /data/e748X/e748/acquisition/root/name of run after


Then to convert Numexo runs: We have set up a new directory to convert Numexo runs without having to log in into e748chio account.

For regular conversions please use the following commands:

Go to:

/home/e748/analysis/nptool/Projects/e748/GRU_NUMEXO

GRU+

GRU "GruConvTree.C(run_number)"

To Merge:

Go to :

/home/e748/analysis/nptool/Projects/e748/merger


root -l "Merge.C(run_number_numexo,run_number_daq_normal)"
Fixed, e738, Fri Jul 13 16:41:49 2018, Antoine, Maurycy, OFFLINE, Social, Cryo-LH2, Multiparticles from (p,t) up to 5 Light Charged Particles in Must ! 15x