MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUST2, Page 45 of 48  Not logged in MUST2
Fixed, e738, Thu Jul 12 07:05:53 2018, Shumpei, Freddy, Megumi, Valérie, RUN, General, Cryo-LH2, RUN 357 7x
RUN 357

start: 6h39
stop: 8h13
e748 Exxx 3,2G 12 juil. 08:13 /data/e748X/e755/acquisition/run/run_0357.dat.12-07-18_06h38m56s
12C primary : 2.0 uA
CATS1: 220 kHz
CATS2: 186 kHz
PL : 192 kHz
Fixed, e738, Thu Jul 12 07:43:25 2018, Shumpei, OFFLINE, General, Cryo-LH2, Analysis of 9C beam setting c9b8pd2.pngbe7li6pt.pngc9b8pt.png
Used data : run0312-0355

For (p,t) reaction, the value of the mass for N=1 isotone is taken from proposal, so it can be shifted.
3He coincidence gate is applied to remove the beam triton background.
Fixed, e738, Thu Jul 12 08:16:21 2018, Daisuke, Megumi, Valérie, RUN, General, Cryo-LH2, RUN 358 20x
RUN 358
start: 8h14 stop: 9h48

e748 Exxx 3,3G 12 juil. 09:49 /data/e748X/e755/acquisition/run/run_0358.dat.12-07-18_08h12m26s


12C primary : 1.95 uA
CATS1: 224 kHz ; CATS2: 194 kHz
PL : 199 kHz

CATS1 670 V I ~ 0.16 ľA ; CATS2 650V I ~ 0.12V
Efficiencies estimated from the integrals in the plots IC_dE1_TCATS1HF
cf run358_ICdeTcatsHFtot.png run358_ICdeTcatsHFeff.png
CATS 1: (2.8 10**4)/(3.1 10**4) =0.90 -->90% ; CATS2 (2.7 10**4)/(3.06 10**4)==> 88%
Fixed, e738, Thu Jul 12 09:52:07 2018, Daisuke, Megumi, Valérie, RUN, General, Cryo-LH2, RUN 359 6x
RUN 359

start: 9h48 stop: 11h25
748 Exxx 3,2G 12 juil. 11:25 /data/e748X/e755/acquisition/run/run_0359.dat.12-07-18_09h48m08s
12C primary : 1.98 uA
CATS1: 230 kHz ; CATS2: 198 kHz
PL : 205 kHz

CATS1 670 V I ~ 0.16 ľA ; CATS2 650V I ~ 0.12V
Fixed, e738, Thu Jul 12 11:28:46 2018, Daisuke, Megumi, Valérie, RUN, General, Cryo-LH2, RUN 360 10x
RUN 360
e748 Exxx 3,4G 12 juil. 13:08 /data/e748X/e755/acquisition/run/run_0360.dat.12-07-18_11h24m44s
start: 11h25 stop: 13h08
12C primary : 1.95 uA
CATS1: 230 kHz ; CATS2: 198 kHz
PL : 205 kHz

CATS1 670 V I ~ 0.16 ľA ; CATS2 650V I ~ 0.12V
Stable conditions for beam and CATS during the run.
Series of plots for the raw spectra of CATS1 strips.
Fixed, e738, Thu Jul 12 13:09:14 2018, Daisuke, Megumi, Valérie, RUN, General, Cryo-LH2, RUN 361 6x
RUN 361
e748 Exxx 3,5G 12 juil. 15:01 /data/e748X/e755/acquisition/run/run_0361.dat.12-07-18_13h07m40s
start: 13h08 stop: 14h59
12C primary : 1.8 uA
CATS1: 200 kHz ; CATS2: 175 kHz
Plast : 178 kHz

CATS1 670 V, I ~ 0.14 ľA ; CATS2 650V, I ~ 0.10V sometimes up Ovc to 0.38 ľA but rather stable.
Ongoing, e738, Thu Jul 12 13:49:38 2018, takeshi, TARGET, Hardware, Cryo-LH2, temperature log of the target heatshield1207.pngcell1207.png
Fixed, e738, Thu Jul 12 15:07:43 2018, Daisuke, Louis, Marlčne, Megumi, Valérie, RUN, General, Cryo-LH2, RUN 362-TRIGGER TESTS 11x
RUN 362 WITH CHANGE OF TRIGGERS FOR TESTS see below
e748 Exxx 1,7G 12 juil. 16:09 /data/e748X/e755/acquisition/run/run_0362.dat.12-07-18_15h08m50s
start: 15h08 stop: 16h08
12C primary : between 1.7 and 1.9 uA
CATS1: 200 kHz ; CATS2: 175 kHz
Plast : 178 kHz FAG ~ 650/s

CATS1 670 V I ~ 0.14 ľA ; CATS2 650V I ~ 0.10V sometimes up Ovc to 0.38 ľA but rather stable.

The triggers can be seen on Run362_TriggerMMallCats.png
There are: IN10- CATS1div and IN11 -CATS2div, IN16 Plastic %
IN12 and IN14.

NB: the entry IN12 noted "CATS" in the snapshot was MM2all div, with div =10
In Gatconfmaster the value is 2**(12-1) = 2048.
The "IN14" trigger input is:
MM2all*CATS (hardware coincidence between MUST2 and CATS) in IN14
GMT trigger in GATCONFmaster : 2**(14-1) = 2**13 = 8192.
Fixed, e738, Thu Jul 12 15:27:56 2018, daiuske, MUST2, General, N/A, MUST2*CATS coincidence trigger 2018-07-09_message_adrien.pdfScreenshot_2018-07-09_()_-_document.png
Message from Adrien.
Fixed, e738, Thu Jul 12 16:22:43 2018, Frank, Louis, RUN, General, Cryo-LH2, RUN 363 10x
RUN 363

start: 16h09 stop: 18h10
12C primary I : between 1.7 and 1.8 uA
CATS1: 194 kHz ; CATS2: 170 kHz
Plast : 170 kHz
FAG ~ 650/s

CATS1 670 V I ~ 0.12 ľA ; CATS2 650V I ~ 0.10V sometimes up Ovc to 0.38 ľA but rather stable.
Fixed, e738, Thu Jul 12 18:12:33 2018, Frank, Louis, RUN, General, Cryo-LH2, RUN 364 12x
RUN 364

start: 18h11 stop: 20h45

12C primary I : 1.8 uA
CATS1: 215 kHz ; CATS2: 171 kHz
Plast : 190 kHz
FAG ~ 730/s

CATS1 670 V I ~ 0.14 ľA ; CATS2 650V I ~ 0.10V
Pinned, e738, Thu Jul 12 18:47:32 2018, Nori, BEAM, General, N/A, Summary of Brho2 values 
Beam Brho2 RMN (Tm) Brho2 L++ (Tm)
9C 1.6352 1.6343
10C 1.8345 1.8335
40Ca (high) 2.2448 2.2432
37Ca 1.9242 1.9228
38Ca 1.9860 1.9846
40Ca (low) 2.1672 (?) 2.1656 (?)

 

Fixed, e738, Thu Jul 12 20:47:32 2018, Frank, Louis, RUN, General, Cryo-LH2, RUN 365 10x
RUN 365

start: 20h46 stop: 22h43

12C primary I : 1.8 uA
CATS1: 200 kHz ; CATS2: 185 kHz
Plast : 177 kHz
FAG ~ 755/s

CATS1 670 V I ~ 0.14 ľA ; CATS2 650V I ~ 0.10V
Fixed, e738, Thu Jul 12 22:45:50 2018, Frank, Louis, RUN, General, Cryo-LH2, RUN 366 15x
RUN 366

start: 22h44 stop: 00h42

12C primary I : 1.8 uA
CATS1: 200 kHz ; CATS2: 185 kHz
Plast : 177 kHz
FAG ~ 755/s

CATS1 670 V I ~ 0.14 ľA ; CATS2 650V I ~ 0.10V
Fixed, e738, Fri Jul 13 00:45:45 2018, Pierre, RUN, General, Cryo-LH2, Run 367 
start: 00h44
Fixed, e738, Fri Jul 13 00:47:23 2018, Pierre, RUN, General, Cryo-LH2, Run 367 scaler_2018-07-13_00h46m09s.pngvisu_2018-07-13_00h51m01s.pngvisu_2018-07-13_00h50m40s.png
RUN 367

start: 00h44
stop: 02h44

CATS1: 670 V & 0.10 uA
CATS2: 650 V & 0.08 uA
Fixed, e738, Fri Jul 13 02:45:44 2018, Pierre, RUN, General, Cryo-LH2, Run 368 7x
RUN 368

start: 02h44
stop: 04h46

CATS1: 670V & 0.14 uA
CATS2: 650V & 0.10 uA
Fixed, e738, Fri Jul 13 04:09:10 2018, takeshi, TARGET, Hardware, Cryo-LH2, temperature of the target cell1307.png
We found the target temperature suddenly increased at around 3:40 13th.
The online beam image is still indicating the target is filled up to now.
Fixed, e738, Fri Jul 13 04:48:19 2018, Pierre, RUN, General, Cryo-LH2, Run 369 7x
RUN 369

start: 04h47
stop:06h41

CATS1: 670V & 0.10 uA
CATS2: 650V & 0.08 uA

The run is stopped due to some issues from the cyclotron (some tripping).
Could take up to 2 hours to have the beam back.
Fixed, e738, Fri Jul 13 06:02:31 2018, takeshi, pierre, TARGET, General, Cryo-LH2, target is half filled target1307.pdfcell1307_2.png
We found the liquid hydrogen target is filled only half during the run 368, around 4:10.
We will continue the measurement for now.

ELOG Home Page