Fixed, e738, Sat Jul 14 00:32:55 2018, Shumpei, OFFLINE, General, Cryo-empty, empty target analysis for 9C beam setting   
|
H2 filled: run0312-0355
empty: run0370-0372
Scale factor for empty run is 10.0
Gate for target position and recoil pid (deuteron or triton)
Gate for theta_cm < 60 deg.
For (p,t) reaction, proton elastic scattering of beam triton contamination cannot be subtracted.
See entry 864
https://elog.lpc-caen.in2p3.fr/MUST2/864 |
Fixed, e744, Wed May 16 17:03:18 2018, Ruxandra, EXL-PPT, Software, , calibration
|
Calibration coefficients for the PPT detector using the 60Co and 40K lines: E=0.5514*ch - 83.322. Internal radioactivity at 1444 keV, not separated from 40K
Efficiency: During the run (21h46min40s) the number of gammas emitted from the 60Co source is 66796800; for the Co pics the integrals are: 2.71e05 for the 1173.2 and 1.87e05 for the 1333, respectively (0.4% and 0.3%) without geometrical factor |
Fixed, e744, Fri May 18 10:47:37 2018, Ruxandra, EXL-PPT, Hardware, 60Co, Co60 caliibration
|
Thu May 17 . run 339
60Co calibration
start @19h26 17/05/18
stop @7h39 18/05/18
EXL + PPT trigger |
Fixed, e744, Fri May 18 13:28:24 2018, Ruxandra, EXL-PPT, Software, 60Co, new calibration
|
The new calibration coefficients for the PPT detector (run 339): E=0.526*ch+32.57 |
Fixed, e755, Tue Jun 26 07:24:59 2018, Pierre, Shumpei, Yorick, Valérie, RUN, General, Cryo-LH2, RUN 139  
|
start: 07h23
stop: 8h52
748 Exxx 3003199093 26 juin 08:52 /data/e748X/e755/acquisition/run/run_0139.dat.26-06-18_07h22m11s
Trigger: all MM + CATS DIV
Beam= 1.6x10^5 pps fluctuating down to 1.0 10**5/s at 8h06
Scalers at 8h42 beam 1.3 10**5/s |
Fixed, e755, Tue Jun 26 00:09:40 2018, Pierre, Shumpei, RUN, General, Cryo-LH2, RUN 134  
|
Start: 26/06 at 00h05
Stop: 00h40 for PCP to improve beam tuning
Beam is back after some tunning.
Trigger: MM+CATS DIV
Beam intensity about 2x10^5 pps |
Fixed, e755, Tue Jun 26 01:02:42 2018, Pierre, Shumpei, RUN, General, Cryo-LH2, RUN 135 
|
start: 01h02
stop: 01h39
Trigger: MM+CATS DIV
Beam = 2x10^5; pps |
Fixed, e755, Tue Jun 26 01:46:56 2018, Pierre, Shumpei, RUN, General, Cryo-LH2, RUN 136  
|
start: 01h41
stop: 03h37
!!!!!!!!!
New signal from CAVIAR has been shifted by 50 ns.
=> TAC_CATS_CAVIAR is now qlso shifted => new cut condition will have to be applied for run 136 |
Fixed, e755, Tue Jun 26 03:40:15 2018, Pierre, Shumpei, RUN, General, Cryo-LH2, RUN 137  
|
start: 03h38
stop: 05h30
Same condition as before
Trigger: MM + CATS DIV
Beam = 2x10^5 pps |
Fixed, e755, Tue Jun 26 05:32:35 2018, Pierre, Shumpei, RUN, General, Cryo-LH2, RUN 138 
|
start: 05h31
stop: 07h23
Trigger: MM + CATS DIV
Beam = 2x10^5 pps |
Fixed, e755, Mon Jul 2 16:50:10 2018, Pierre, Louis, RUN, General, Cryo-LH2, Run 231 
|
start: 16h47
stop: 18h48
Beam= 26000 pps |
Fixed, e738, Mon Jul 9 15:43:50 2018, Pierre Valérie, RUN, General, Cryo-empty, Run 309 
|
start: 15h36
stop: 16h10
Trigger: OR MM, CATS1 DIV and PLASTIC%
large beam interuption between 15h40 and 16h06 |
Fixed, e755, Sat Jun 23 00:47:19 2018, Pierre Valerian, CATS, General, N/A, NEW GOOD mask reconstruction 
|
New shift for CATS after fixing Z position of CATS detectors.
ConfigCATS
ZPROJ -824
ZPROJ -1332
SHIFT_CATS CATS1X -2.3
SHIFT_CATS CATS1Y -3.9
SHIFT_CATS CATS2X 4.5
SHIFT_CATS CATS2Y -2.2 |
Fixed, e755, Wed Jun 27 22:11:20 2018, Pierre Louis, OFFLINE, General, Cryo-LH2, Offline analysis of 38Ca(p,d)37Ca 
|
Here are some plot summarizing the analysis status on the 38Ca(p,d) channel.
Run concerned: 128-156 (from 25/06 @ 17h11 to 27/06 @ 7h03)
Running time: 33 hours
Gates applied:
- Target Impact (from CATS)
- Deuton in dE-E
- Incoming 38Ca, TCATS2_HF in [200;1000]
- Outgoing (Z=20)Ca, dE-E (IC-PL)
We analysed statistics in the Ground States of 38Ca(p,d) and compared to statistics of the 38Ca(p,t) channel telescope by telescope
38Ca(p,t)36Ca:
T1: mean = -0.84
sigma = 0.608
Count number = 302
T2: mean = -1.06
sigma = 0.76
Count number = 238
T3: mean = -1.70
sigma = 0.77
Count number = 342
T4: mean = -1.57
sigma = 1.21
Count number = 271
TOTAL_t_GS = 1153
38Ca(p,d)37Ca:
T1: mean = 0.71
sigma = 0.73
Count number = 2868
T2: mean = 0.18
sigma = 0.66
Count number = 1828
T3: mean = 0.23
sigma = 0.81
Count number = 3536
T4: mean = 0.127
sigma = 0.88
Count number = 3240
TOTAL_d_GS = 11472
Remarks:
- see 10 times more statistics in the ground states of 37Ca than in 36Ca
- resolution of the t channel is 80% of the resolution of the d channel
- excited states in (p,d) channel are more populated than the ground state
|
Fixed, e755, Mon Jul 2 23:09:55 2018, Pierre Louis, OFFLINE, General, Cryo-LH2, 37Ca(p,d)36Ca
|
|
Fixed, e755, Fri Jun 22 21:29:05 2018, Pierre Fairouz Adrien Valerian, CATS, General, N/A, Mask reconstruction 
|
Mask reconstruction after adding the correction shift for CATS.
CATS1X = +1.2
CATS1Y = -0.1
CATS2X = 2.0
CATS2Y = -0.2 |
Fixed, e738, Mon Jul 9 08:55:13 2018, Pierre Daisuke, RUN, General, Empty, Run 303
|
start: 08h27
stop: 08h50
Run for background with no beam |
Fixed, e738, Mon Jul 9 08:56:50 2018, Pierre Daisuke, RUN, General, Cryo-empty, Run 304    
|
start: 08h54
stop: 10h00
Run with same trigger but empty target
TRIGGER: MMi, CATS1DIV, CATS2DIV, PLASTIC% |
Fixed, e744, Fri May 25 18:15:41 2018, Pierre, MUST2, General, CD2, new dEE PID using Shumpei calibration
|
|
Fixed, e755, Tue Jun 26 05:40:52 2018, Pierre, OFFLINE, General, Cryo-LH2, Kinematic lines and Ex for 38Ca(p,d) and 38Ca(p,t) 
|
-> 38Ca(p,d) uses run 128
-> 38Ca(p,t) uses run from 128 up to 136
The macro to plot them are located in /home/e748/analysis/nptool/Projects/e755/Analysis/macro
- ShowResult_pd.C
- ShowResult_pt.C |
|