Fixed, e744, Fri May 25 21:37:45 2018, iulian nicolas valérie, RUN, General, N/A, run 1012 all triggers 14O beam in D4 
|
|
Fixed, e744, Fri May 25 22:03:12 2018, e744 Must2+LISE teams, RUN, General, CH2, RUNS 2000, 2001 6x
|
Beam 14O
Run 2000 start with all triggers, including LISE trigger, during 1 minute for tests.
start 22:00:58; stop
|
Fixed, e744, Fri May 25 22:46:26 2018, e744 Must2+LISE teams, edit Valérie, RUN, General, CH2, RUN 2004 6x
|
Run 2004 Date : 25/05/2018 22:45:58 25/05/2018 23:17:39
End of Run, error Narval see snapshot ProblemNarvalRCCEndRun2004.png
|
Fixed, e744, Fri May 25 23:26:23 2018, e744 Must2+LISE teams, edit Valérie, RUN, General, CH2, START NARVAL, new version
|
the NARVAL version is now reinstalled by Fred Saillant.
see plot ResolProblemNarvalEndofRun25may23h.png
|
Fixed, e744, Fri May 25 23:33:27 2018, iulian, OTHER, Hardware, CH2, changes after run 2004
|
Target was changed to target number 5 (CH2 100 um GOODFELLOW)
The T_EXL1718 ok
|
Fixed, e744, Sat May 26 00:24:50 2018, e744 valérie, DAQ, General, CH2, change of trigger and division
|
TRIGGER entry for CATS, presently IN8, seems to be not running. We put it on IN7(Gatconf =64).
divisions EXL 10**4
PPT 3000
|
Fixed, e744, Sat May 26 00:38:38 2018, e744 all night shift teams, (edit) Valérie, RUN, General, CH2, Run 2010 20x
|
RUN 2010
Date : 26/05/2018 00:37:44
Target n°5 CH2 100 µm
|
Fixed, e744, Sat May 26 01:41:19 2018, e744 night shift team, edit valérie, DAQ, General, CH2, summary GMT, DAS config since run2005   
|
Triggers: all except LISE
IN1 to IN5 : MM1 to MM5;
|
Fixed, e744, Sat May 26 02:39:18 2018, Mihai, Pedja, RUN, General, CH2, Run 2011, Target n°5 9x
|
RUN 2011
Date : 26/05/2018
Beam used: 14O
|
Fixed, e744, Sat May 26 04:24:16 2018, Mihai, Pedja, RUN, General, CH2, Run 2012, Target n°5
|
RUN 2012
Date : 26/05/2018
Beam used: 14O
|
Fixed, e744, Sat May 26 05:16:12 2018, Iulian Valerian, pedja, mihai, TARGET, General, CH2, Target change
|
Target No 4 is placed in-beam 5h15. |
Fixed, e744, Sat May 26 05:23:12 2018, Mihai, Pedja, RUN, General, CH2, Run 2013, Target n°4 
|
RUN 2013
Date : 26/05/2018
Beam used: 14O
|
Fixed, e744, Sat May 26 05:51:00 2018, Mihai, Pedja, RUN, General, CH2, Run 2014, Target n°4 9x
|
[U][B]RUN 2014[/B][/U]
Date : 26/05/2018
Beam used: 14O
|
Fixed, e744, Sat May 26 06:05:13 2018, Iulian Valerian Pedjia, TARGET, General, CH2, Target issue No5
|
In the attached figure it could be seen that with the target No5 (run 2010) lower energy resonances were visible (lower part of the figure), which is not
the case with the target No4 (upper part)from run 2013.
|
Fixed, e744, Sat May 26 09:33:48 2018, Giuseppe, Dieter, RUN, General, CH2, run 2015  
|
Missing strips:
x
MM1 1
|
Fixed, e744, Sat May 26 10:03:37 2018, Giuseppe, Dieter, RUN, General, , run 2016& 2017
|
Problems with the ACQ (merger), they were solved by making Breakup and Init
|
Fixed, e744, Sat May 26 11:19:37 2018, Giuseppe, Dieter, RUN, General, CH2, interruption due to to waterleak ibn CSS1
|
10:50 h Water leak at CSS1
cryo pumps have to be regenerated
continuation until the expert intervention starts is possible
|
Fixed, e744, Sat May 26 11:42:03 2018, Giuseppe, Dieter, RUN, General, CH2, run 2019 
|
Run 2019
|
Fixed, e744, Sat May 26 14:03:27 2018, Giuseppe, Dieter, J. Eric, RUN, General, CH2, Run 2020 9x
|
run 2020
|
Fixed, e744, Sat May 26 16:31:53 2018, Jaromir, Giuseppe, EXL-PPT, Software, N/A, EXL calibration, light leak    
|
Selector.C can be used to adjust energy bad reconstruction in the EXL detectors due to energy leak (see the PDF attached for more information about it)
once we already have the coefficients for the energy on EXL crystals. To operate with it, for first calibration coefficient must be deduced. Once it is
done the slopes and offset of the calibration have to be inserted in Selector.C file at lines 41 (a[19]) and at line 49 (b[19]). After thqt thresholds |
|