Fixed | e744 |
Fri May 18 13:28:24 2018 |
Ruxandra | EXL-PPT | Software | 60Co | new calibration |
The new calibration coefficients for the PPT detector (run 339): E=0.526*ch+32.57 |
Fixed | |
Fri May 18 15:17:19 2018 |
Megumi and Louis | OTHER | Hardware | | TAC range setting and time calibration |
1 mm-cats 1us
2 mm-hf 200ns
3 mm-exl 1us
4 mm-ppt 1us
5 cats-hf 200ns
6 exl-cats 1us
7 ppt-cats 1us
8 mm-mcp 1us
9 mm-hfv 200ns
10 mcp-hf 200ns
time calibrator setting
period 0.02us
range 1.28us
run340 for test
run341 for calibration of mm-cats start
run343 for calibration of mm-hf
run344 for calibration of mm-exl, mm-ppt, mm-mcp, mm-hf, mcp-hfv
added on 21/05/2018
we found validation point was missing for some channels
run349 for calibraion cats_hf
run350 for calibration ppt_cats
run351 for calibraion cats_hf again
run352 for calibration exl_cats
calibration parameter (in ns)
1 mm-cats 0.0495255
2 mm-hf 0.0102974
3 mm-exl 0.0500575
4 mm-ppt 0.0510095
5 cats-hf 0.010322
6 exl-cats 0.0547923
7 ppt-cats 0.051581
8 mm-mcp 0.0507524
9 mm-hfv 0.0107085
10 mcp-hf 0.01041 |
Fixed | e744 |
Fri May 18 17:38:21 2018 |
Francois and Pedja | OTHER | Hardware | | translator nim to ecl |
There was a problem with the scalers, not all of them were working. After several tests, we changed the electronic module, from Lecroy ECl-NIM-ECL to module LPC translator NIM-ECL. We checked all channels, they work. For a while channel N°4 was not working. We changed the nim cable, it works now. |
Attachment 1: scalers.png
|
|
Fixed | e744 |
Sun May 20 18:53:44 2018 |
iulian Jaromir Cristina | EXL-PPT | General | N/A | Run #347 Background run EXL & PPT triggers |
Run #347
Background run EXL & PPT triggers
start @18h53 20/05
stop @11h14 21/05 |
Attachment 1: PPT_run347.png
|
|
Attachment 2: EXL_T_run347.png
|
|
Attachment 3: EXL_E_run347.png
|
|
Fixed | e744 |
Mon May 21 11:42:16 2018 |
iulian cristina | MUST2 | Software | N/A | run 348 test comunication with Must. Pulser T1 channel 11 |
|
Attachment 1: MUST2_Pulser_T1_E11_run348.png
|
|
Fixed | e744 |
Tue May 22 10:44:30 2018 |
iulian Fairouz Valerie | MUST2 | Software | N/A | Run #353 Calimero T1-T4 strips X |
Started run #353
Calimero T1-T4 strips X
Time Calibrator 320 ns ; period 20 ns |
Fixed | e744 |
Tue May 22 11:37:50 2018 |
Valérie | MUST2 | Software | N/A | calimero config GECO |
Calimero T1-T4 strips X
Time Calibrator 640 ns ; period 20 ns |
Attachment 1: PlotTel1strxT101calimOk.png
|
|
Attachment 2: PlotXTel1tXT88Probmate.png
|
|
Attachment 3: PlotXTel1to4XTProbmate.png
|
|
Attachment 4: PlotXTel1to4XTcalimero.png
|
|
Fixed | e744 |
Tue May 22 11:59:06 2018 |
Valérie | MUST2 | Software | N/A | Calimero config GECO |
|
Attachment 1: configGecotelstrYTcalimero.png
|
|
Attachment 2: StartCalimero6%.png
|
|
Attachment 3: ConfigGecostartCalimero.png
|
|
Attachment 4: startRccbrefCalim.png
|
|
Attachment 5: RccbeforeStartcalimero.png
|
|
Attachment 6: ConfigGecobeforeStart.png
|
|
Attachment 7: NarvalConfigGecoCal.png
|
|
Attachment 8: MUST2_calimGecoTX.png
|
|
Fixed | e744 |
Tue May 22 12:09:27 2018 |
iulian Fairouz Valerie | MUST2 | Software | N/A | Run #354 Calimero T1-T4 strips X |
Started run #353
Calimero T1-T4 strips X
Time Calibrator 640 ns ; period 20 ns |
Attachment 1: PlotXTel1tXT88Probmate.png
|
|
Attachment 2: PlotTel1strxT101calimOk.png
|
|
Attachment 3: PlotXTel1to4XTProbmate.png
|
|
Attachment 4: PlotXTel1to4XTcalimero.png
|
|
Attachment 5: PlotXTel1to4XTcalimero.png
|
|
Fixed | e744 |
Tue May 22 12:10:27 2018 |
iulian Fairouz Valerie | MUST2 | Software | N/A | Run #355 Calimero T1-T4 strips Y |
Started run #353
Calimero T1-T4 strips Y
Time Calibrator 640 ns ; period 20 ns |
Fixed | e744 |
Tue May 22 13:40:03 2018 |
Fairouz Valerie | MUST2 | Software | N/A | Run #356 Calimero T1-T4 strips Y |
Run #356
Time Calibrator Calimero stripYT T1to T4 640ns period 20ns |
Fixed | e744 |
Tue May 22 19:36:50 2018 |
Fairouz, Valérie | MUST2 | Software | N/A | Run #359 Calimero T1-T4 strips X |
Run #359
amplitude of the gene is now set to 10 MeV
Time Calibrator Calimero stripsX T1to T4 640ns period 20ns |
Attachment 1: CalimeroRunXTampli10MeV.png
|
|
Attachment 2: PlotStripsXTcalimero.png
|
|
Attachment 3: PlotStripsT1XT67calim1.png
|
|
Attachment 4: PlotStripsT2XT27calim1.png
|
|
Attachment 5: PlotStripsT4XT28calim.png
|
|
Attachment 6: PlotStripsT4XT44calim.png
|
|
Attachment 7: PlotEXPulserProbT4.png
|
|
Attachment 8: PlotStripsT1toT4XT_run359.png
|
|
Fixed | e744 |
Tue May 22 19:40:48 2018 |
Valérie | MUST2 | General | N/A | snapshots for steps in Calimero calibration |
|
Attachment 1: Calimero-step1.png
|
|
Attachment 2: Calimero-step2.png
|
|
Attachment 3: Calimero-step3.png
|
|
Attachment 4: Calimero-step4.png
|
|
Attachment 5: Calimero-step5.png
|
|
Attachment 6: Calimero-step6.png
|
|
Attachment 7: Calimero-step7prog.png
|
|
Attachment 8: Calimero-step8proginit.png
|
|
Attachment 9: Calimero-step9startRCC.png
|
|
Attachment 10: Calimero-step10startcalim.png
|
|
Attachment 11: CalimeroRunXTampli10MeV.png
|
|
Fixed | e744 |
Tue May 22 20:06:15 2018 |
Fairouz, Valerie, edited by Iulian Megumi | MUST2 | Software | N/A | Run #360 Calimero T1-T4 strips Y |
Started run #360
calimero stripsYT T1 to T4 640 ns; period 20 ns
amplitude pulser 10 MeV
There was a problem on T2 as the channel 2 on GMT MASTER was set on TTL. |
Attachment 1: Calimero_GMTproblem.png
|
|
Fixed | e744 |
Tue May 22 21:43:31 2018 |
iulian & megumi | MUST2 | Software | N/A | run 361 |
run 361
start 20h38
stop 21h41 |
Attachment 1: T1T4_Xstrips_run361.png
|
|
Fixed | e744 |
Tue May 22 21:57:51 2018 |
iulian & megumi | MUST2 | Software | N/A | Run #362 Calimero T1-T4 strips X again with solution |
Changed the Thresholds on T4. No longer double peaks...
I guess there is some noise on the T4 and he triggers on the noise from time to time => double peaks
Also T1-T3 have have strip 16 near pidestal who is counting a lot (red) |
Attachment 1: Thresolds_T4_run362.png
|
|
Attachment 2: T1T4_Xstrips_run362.png
|
|
Fixed | e744 |
Tue May 22 22:33:09 2018 |
e744 | MUST2 | Software | N/A | Run #363 Calimero T1-T4 strips X again with threshold at 50% (same as 40%) |
Run #363
NB: Thresholds are the ones on the pulser amplitude to inject the signal (Calimero Pulser interface on Geco). see snapshots on the Calimero summary.
Calimero with 50% thresholds on T4. Same spectra as with the thresholds @40%. I think T4 is triggering on cross-talk, so when the thresholds it's high it's ok. |
Attachment 1: T1T4_Xstrips_run363.png
|
|
Fixed | e744 |
Tue May 22 22:50:47 2018 |
e744 | MUST2 | Software | N/A | Run #365 Calimero strips Y T1-T4 100% Threshold |
Run #365
Calimero strips Y T1-T4
100% threshold on T4
Something is wrong with the masking?! |
Attachment 1: T1T4_Ystrips_run364.png
|
|
Fixed | e744 |
Wed May 23 00:30:34 2018 |
e744 | EXL-PPT | General | N/A | RUN #402, EXL PPT trigger |
RUN #402
start exl ppt run for checking coincidence
exl ch 17 put in to adc2 ch 32 |
Fixed | e744 |
Wed May 23 10:52:28 2018 |
Valérie | EXL-PPT | General | N/A | RUN #405, EXL PPT trigger |
RUN #405
start exl ppt run for checking coincidence
|
Attachment 1: GMTtriggerRun405.png
|
|
Attachment 2: ScalersRun405.png
|
|
Attachment 3: EXL_E1to16duringRun405.png
|
|
Attachment 4: EXL_T1to16duringRun405.png
|
|