MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 7 of 132  Not logged in MUGAST
Status ExpNbrdown Date Author Type Category Target-Source Subject
Fixede793s Fri Mar 12 22:31:46 2021 julien, nishuVAMOSHardwareCD2VAMOS tripped @ ~ 22:24 ans again ~ 22:45
back again to normal after restarting the HV
(22:39)
Attachment 1: vamos_nc.png
vamos_nc.png
Attachment 2: Screenshot_from_2021-03-12_22-33-47.png
Screenshot_from_2021-03-12_22-33-47.png
Attachment 3: Screenshot_from_2021-03-12_22-47-38.png
Screenshot_from_2021-03-12_22-47-38.png
Fixede793s Fri Mar 12 23:16:16 2021 julien, nishuRUNHardwareCD2vamos count rate goes down to 0 once again
We reduced the high voltage limit to 520v from 525v.
please see the screen shots of the focal planes attached below:.
Attachment 1: IMG20210312231358.jpg
IMG20210312231358.jpg
Attachment 2: IMG20210312230429.jpg
IMG20210312230429.jpg
Fixede793s Fri Mar 12 23:27:58 2021 julien, nishuRUNGeneralCD2summary of run 61
below attached the screenshots of scalers and spectra
Attachment 1: Screenshot_from_2021-03-12_23-31-23.png
Screenshot_from_2021-03-12_23-31-23.png
Attachment 2: Screenshot_from_2021-03-12_23-31-07.png
Screenshot_from_2021-03-12_23-31-07.png
Attachment 3: Screenshot_from_2021-03-12_23-30-43.png
Screenshot_from_2021-03-12_23-30-43.png
Attachment 4: Screenshot_from_2021-03-12_23-30-22.png
Screenshot_from_2021-03-12_23-30-22.png
Attachment 5: Screenshot_from_2021-03-12_23-29-52.png
Screenshot_from_2021-03-12_23-29-52.png
Fixede793s Fri Mar 12 23:54:52 2021 julien, nishu, FaïrouzVAMOSHardwareCD2Vamos tripped @ 23h35 & fixed
Fixede793s Sat Mar 13 01:41:21 2021 Juan Lois & faïrouzRUNGeneralCD2run61 screenshots at 1h40
Attachment 1: run61_TAC_01h35.png
run61_TAC_01h35.png
Attachment 2: run61_MG_ToF_01h35.png
run61_MG_ToF_01h35.png
Attachment 3: run61_MM_DeE_01h35.png
run61_MM_DeE_01h35.png
Attachment 4: run61_MM_ToF_01h35.png
run61_MM_ToF_01h35.png
Attachment 5: run61_MG_IMPACT_MATRIX_01h35.png
run61_MG_IMPACT_MATRIX_01h35.png
Attachment 6: run61_Vamos-MW_01h35.png
run61_Vamos-MW_01h35.png
Attachment 7: run61_MM_IMPACT_MATRIX_01h35.png
run61_MM_IMPACT_MATRIX_01h35.png
Attachment 8: run61_Scalers_01h34.png
run61_Scalers_01h34.png
Fixede793s Sat Mar 13 02:31:33 2021 Faïrouz, Juan LoisVAMOSHardwareCD2Vamos tripped @ 2h32 & fixed
Fixede793s Sat Mar 13 03:50:54 2021 Faïrouz, Juan LoisRUNGeneralCD2run61 screenshots at 3h44
Attachment 1: Run61_scalers_03h44.png
Run61_scalers_03h44.png
Attachment 2: Run61_MG_IMPACT_03h44.png
Run61_MG_IMPACT_03h44.png
Attachment 3: Run61_MM_ToF_03h44.png
Run61_MM_ToF_03h44.png
Attachment 4: Run61_MM_DeE_03h44.png
Run61_MM_DeE_03h44.png
Attachment 5: Run61_NG_TOF_03h44.png
Run61_NG_TOF_03h44.png
Attachment 6: Run61_TAC_03h44.png
Run61_TAC_03h44.png
Attachment 7: Run61_VAMOS_MW_03h44.png
Run61_VAMOS_MW_03h44.png
Attachment 8: Run61_MM_IMPACT_03h44.png
Run61_MM_IMPACT_03h44.png
Pendinge793s Sat Mar 13 05:11:08 2021 Faïrouz, Juan LoisAGATASoftwareCD2Problem with AGATA acquisition started at 4h40

There is a problem with AGATA Acquisition. All the bars in the global status & control are yellow!!!
This problem started at 4h40.
We called the acquisition astreinte but he doesn't know how to fix the problem.
We called Emmanuel but we got on the answering machine. We will call him again at 7 am.

The count rate of AGATA in the scaler looks ok: ~ 1200
Attachment 1: run61-scalers_5h19.png
run61-scalers_5h19.png
Fixede793s Sat Mar 13 05:33:41 2021 Faïrouz, Juan LoisRUNGeneralCD2run61 screenshots at 5h30
Attachment 1: run61_scalers_05h30.png
run61_scalers_05h30.png
Attachment 2: run61_MM_IMPACT_05h30.png
run61_MM_IMPACT_05h30.png
Attachment 3: run61_VAMOS_MW_05h30.png
run61_VAMOS_MW_05h30.png
Attachment 4: run61_TAC_05h30.png
run61_TAC_05h30.png
Attachment 5: run61_MG_ToF_05h30.png
run61_MG_ToF_05h30.png
Attachment 6: run61_MM_DeE_05h30.png
run61_MM_DeE_05h30.png
Attachment 7: run61_MM_ToF_05h30.png
run61_MM_ToF_05h30.png
Attachment 8: run61_MG_IMPACT_5h30.png
run61_MG_IMPACT_5h30.png
Fixede793s Sat Mar 13 07:18:10 2021 OttoRunRUNGeneralCD2Run62
*************************
Run : 62
*************************

Start at 2021-03-13 07:17:49.732042

run62. ACQ DT


starting again after centering the beam on the target

Beam: 3.5e5 pps
target: CD2 0.5mg
Trigger are MUGAST/MM/CATS%

Stop at 2021-03-15 10:05:15.581255
Stop comment: stop run

Stop at 2021-03-17 07:17:25.231024
Stop comment: Vamos stopped writing data at 06:50
Fixede793s Sat Mar 13 07:26:40 2021 Faïrouz, Juan LoisRUNGeneralCD2End of run61 screenshots at 7h20
RUN 61 stopped to let Emmanuel fix the AGATA acquisition problem.
Attachment 1: run61_MG_IMPACT_7h20.png
run61_MG_IMPACT_7h20.png
Attachment 2: run61_MM_ToF_7h20.png
run61_MM_ToF_7h20.png
Attachment 3: run61_MM_DeE_7h20.png
run61_MM_DeE_7h20.png
Attachment 4: run61_MG_ToF_7h20.png
run61_MG_ToF_7h20.png
Attachment 5: run61_TAC_7h20.png
run61_TAC_7h20.png
Attachment 6: run61_Vamos-MW_7h20.png
run61_Vamos-MW_7h20.png
Attachment 7: run62_MM_IMPACT_7h20.png
run62_MM_IMPACT_7h20.png
Fixede793s Sat Mar 13 07:29:56 2021 Faïrouz, Juan LoisAGATASoftwareCD2AGATA acquisition problem fixed by Emmanuel
Fixede793s Sat Mar 13 07:31:13 2021 Faïrouz, Juan LoisRUNGeneralCD2Run 62 scalers
Attachment 1: run62_scaler_7h20.png
run62_scaler_7h20.png
Fixede793s Sat Mar 13 09:19:14 2021 Lynda, Laetitia, ChloeONLINEGeneralCD2run62 screenshots at 9h10
Online spectra reset at 9h20
Attachment 1: run62_MG_IMPACT_9h10.png
run62_MG_IMPACT_9h10.png
Attachment 2: run62_MM_ToF_9h10.png
run62_MM_ToF_9h10.png
Attachment 3: run62_MM_DeE_9h10.png
run62_MM_DeE_9h10.png
Attachment 4: run62_MG_ToF_9h10.png
run62_MG_ToF_9h10.png
Attachment 5: run62_TAC_9h10.png
run62_TAC_9h10.png
Attachment 6: run62_FSC_9h10.png
run62_FSC_9h10.png
Attachment 7: run62_Vamos-MW_9h10.png
run62_Vamos-MW_9h10.png
Attachment 8: run62_MM_IMPACT_9h10.png
run62_MM_IMPACT_9h10.png
Attachment 9: run62_Theta_E_9h10.png
run62_Theta_E_9h10.png
Attachment 10: run62_scaler_9h15.png
run62_scaler_9h15.png
Fixede793s Sat Mar 13 10:11:44 2021 AdrienOFFLINEGeneralCD2Offline analysis of run 51-61
Run 51 to 61 are converted and analysed.
I fixed the gamma/no gamma normalisation after binning change.

We clearly see :
a state at 2.5 MeV that decay directly to the g.s. and a state at around 5 MeV with a 3 gamma cascade.

To do list:
- Add an angle dist in AngDist.C macro for the states around 2MeV (you can remove the Ex plot)
- Put Logy in all the canvas for the angle dist;)
- create a new macro showing the EDC 1D histo gated on several sliding windows in Ex to see the the gamma peak changing height
- add arrow or line indication the gamma rays position on every plots
- build the level scheme based on this (there is a preliminary scheme posted earlier by Manu based on gamma-gamma).
Attachment 1: Plot51-61.png
Plot51-61.png
Attachment 2: Angle51-61.png
Angle51-61.png
Fixede793s Sat Mar 13 10:14:16 2021 Lynda, Laetitia, ChloeVAMOSGeneral Vamos tripped @ 10:10 & fixed
Vamos tripped around 10:10. we switched back 3 HV on
Online spectra reset to check
Attachment 1: run62_scaler_9h16.png
run62_scaler_9h16.png
Fixede793s Sat Mar 13 11:16:38 2021 Lynda, Laetitia, ChloeBEAMGeneral beam lost and back
we loose the beam around 11:12, it should be back in few minutes according to pcp.

11:18: the beam is back (3.5E5 pps)
Fixede793s Sat Mar 13 11:52:26 2021 Lynda, Laetitia, ChloeBEAMGeneralCD2Beam lost
A security problem occurs at a door near CSS1 ( AFR alarm ), need to call the security to fix the problem
The person needs to come so it will take ~1h
Fixede793s Sat Mar 13 12:07:32 2021 Lynda, Laetitia, ChloeONLINEGeneralCD2run 62 screenshots at 11h55
Attachment 1: run62_Theta_E_11h55.png
run62_Theta_E_11h55.png
Attachment 2: run62_MG_IMPACT_11h55.png
run62_MG_IMPACT_11h55.png
Attachment 3: run62_MM_IMPACT_11h55.png
run62_MM_IMPACT_11h55.png
Attachment 4: run62_Vamos-MW_11h55.png
run62_Vamos-MW_11h55.png
Attachment 5: run62_FSC_11h55.png
run62_FSC_11h55.png
Attachment 6: run62_TAC_11h55.png
run62_TAC_11h55.png
Attachment 7: run62_MG_ToF_11h55.png
run62_MG_ToF_11h55.png
Attachment 8: run62_MM_DeE_11h55.png
run62_MM_DeE_11h55.png
Attachment 9: run62_MM_ToF_11h55.png
run62_MM_ToF_11h55.png
Attachment 10: run62_scaler_graph_13h44.png
run62_scaler_graph_13h44.png
Fixede793s Sat Mar 13 13:16:04 2021 FrancoOFFLINEGeneralCD2Angular distributions
- Updated the macro for the angular distributions.
- For the angular distribution of the peak at 4 MeV excitation energy I think we need the calculation for the transfer 0f5/2.
- Added the gamma spectra gated on the 3 main peaks in Ex in the macro DrawPlots.C (arrows and labels of the gamma should be added).
Attachment 1: ang_distr_run51-61.png
ang_distr_run51-61.png
Attachment 2: gamma_gateEx.png
gamma_gateEx.png

ELOG Home Page