Fixed, e744s, Thu Apr 11 08:01:12 2019, OttoRun, RUN, General, CH2, Run76
|
*************************
[SIZE=3][B]Run : 76[/B][/SIZE]
*************************
|
Fixed, e744s, Thu Apr 11 08:55:23 2019, OttoRun, RUN, General, CH2, Run77 23x
|
*************************
[SIZE=3][B]Run : 77[/B][/SIZE]
*************************
|
Fixed, e744s, Thu Apr 11 09:05:37 2019, mugast, CATS, General, N/A, CATS HV changed to 580
|
CATS HV changed from 600V to 580V. |
Fixed, e744s, Thu Apr 11 10:38:13 2019, Adrien and Valerian, BEAM, General, N/A, They are preparing 14O beam
|
News from (edit after talking to Pierre D.)
[LIST]
[*] They are preparing the 14O beam.
|
Fixed, e744s, Thu Apr 11 11:26:52 2019, Adrien, MUST2, General, CH2, TOF in MUST2 MM3 is now fixed  
|
The E-TOF plot in MM3 was showing a 5MeV threshold.
After investigation the 12 strip was not disable on MM2 to MM4, and particularly noisy on MM3, killing the counts at small angle where most of the proton
are.
|
Fixed, e744s, Thu Apr 11 15:16:10 2019, Adrien, Valerian, Louis, CATS, General, CH2, run77 CATS efficiency
|
CATS efficiency has been computed for run 77 using T_FagHF an T_Si_Cats1 with different triggers:
Trigger GATCONF_MASTER<8 (MUST+Mugast) : 94%
|
Fixed, e744s, Thu Apr 11 16:56:34 2019, OttoRun, RUN, General, Empty, Run78
|
*************************
[SIZE=3][B]Run : 78[/B][/SIZE]
*************************
|
Fixed, e744s, Thu Apr 11 17:08:39 2019, OttoRun, RUN, General, Empty, Run79
|
*************************
[SIZE=3][B]Run : 79[/B][/SIZE]
*************************
|
Fixed, e744s, Thu Apr 11 17:35:21 2019, François, BEAM, General, , 14O beam is in G1 since 17H00
|
alignment of the beam in VAMOS
intensity ~ 5x10^3 pps |
Fixed, e744s, Thu Apr 11 17:38:24 2019, mugast (beyhan), BEAM, General, , 14O8+ beam characteristics 
|
Brho 14O8+ = 0.69186 Tm
|
Fixed, e744s, Thu Apr 11 17:43:44 2019, OttoRun, RUN, General, Empty, Run80
|
*************************
[SIZE=3][B]Run : 80[/B][/SIZE]
*************************
|
Fixed, e744s, Thu Apr 11 18:36:34 2019, Beyhan, BEAM, General, , before optimization : beam reconstructed profil with VAMOS and assoicated settings 
|
target : hole
cats2 out
about 4000 pps triggering VAMOS |
Fixed, e744s, Thu Apr 11 19:09:35 2019, OttoRun, RUN, General, Empty, Run81
|
*************************
[SIZE=3][B]Run : 81[/B][/SIZE]
*************************
|
Fixed, e744s, Thu Apr 11 19:19:21 2019, Iulian Beyhan, RUN, General, N/A, Change target CH2 to no target
|
Target hole is now removed -> No target |
Fixed, e744s, Thu Apr 11 19:47:15 2019, Marlène, OFFLINE, General, , CATS charges in RUN76 and RUN77    
|
Offline analysis to reconstruct the charges in X and Y for CATS.
The charges are lower in RUN77 because bias was reduced to 580V.
Wires seem to be all OK. |
Fixed, e744s, Thu Apr 11 19:49:29 2019, OttoRun, RUN, General, Empty, Run82
|
*************************
[SIZE=3][B]Run : 82[/B][/SIZE]
*************************
|
Fixed, e744s, Thu Apr 11 19:55:18 2019, OttoRun, RUN, General, Empty, Run83
|
*************************
[SIZE=3][B]Run : 83[/B][/SIZE]
*************************
|
Fixed, e744s, Thu Apr 11 20:55:17 2019, mugast, ONLINE, General, , screenshots RUN77 12x
|
|
Fixed, e744s, Fri Apr 12 00:29:11 2019, François, BEAM, General, , no beam
|
They aligned the beam from CIME to G1 with 15N.
They are now working with 14O.
No beam on G1. |
Fixed, e744s, Fri Apr 12 01:27:41 2019, OttoRun, RUN, General, Empty, Run84
|
*************************
[SIZE=3][B]Run : 84[/B][/SIZE]
*************************
|
|