Pending | e793s |
Thu Mar 11 09:02:20 2021 |
mugast | BEAM | Hardware | CD2 | beam interruption | power supply trip on the accelerator beam is off for about 30min |
Pending | e793s |
Thu Mar 11 11:14:04 2021 |
Adrien | OFFLINE | General | CD2 | TODO LIST for Offline analysis | Miscelanous
- Cleanup branches in npanalysis (remove old vamos branch and add MW TDC ones)
- Check the MW TDC, make an offset calibration to align them
- Implement the calibration in the analysis and out a MW_T branch with all the time
- Allow for multiplicity in EDC
A clean macro for offline analysis producing a standardise Tcanvas to be put in the elog:
- Ex And Ex normalised with coinc gamma overlay (Tac conditions for both) Indicate the Sn on this plot.
- Kinematic plot with kinematic all line overlay (add fonctions to add kine line for 47Kdp 47Tidp 47Kpp 47Kdd 47Kdt 47KCC)
- Ex vs EDC
- Ex vs ThetaLab
- Gamma-Gamma
Angular distribution
- Add a separate macro with gate on gamma to produce ThetaLab distribution for each states
- Do a simulation and produce a Solid angle for the g.s. And apply it to the distribution
- Overlay the ADWA calculation with appropriate scaling (on a reference angle bin to be chosen)
|
Pending | e793s |
Sat Mar 13 05:11:08 2021 |
Faïrouz, Juan Lois | AGATA | Software | CD2 | Problem with AGATA acquisition started at 4h40 |
There is a problem with AGATA Acquisition. All the bars in the global status & control are yellow!!!
This problem started at 4h40.
We called the acquisition astreinte but he doesn't know how to fix the problem.
We called Emmanuel but we got on the answering machine. We will call him again at 7 am.
The count rate of AGATA in the scaler looks ok: ~ 1200 |
Attachment 1: run61-scalers_5h19.png
|
|
Pending | e793s |
Mon Mar 15 00:21:18 2021 |
Sarah, Louis, Charlie | BEAM | General | CD2 | Source is running out | The beam intensity spiked to 9x10^9, so Louis called PCP to ask them to reduce it. He was told that the source was almost depleted, and was presented with two options:
A) Keep pushing the source and run at 5x10^5, but run out completely soon (probably during the night shift)
B) Let the rates slowly drop and open the slits as much as possible, to make the source last longer (perhaps another day). This option is strongly preferred by the PCP team, and is the current plan we are following.
Changing the source would apparently take ~12 hrs. There will be a large meeting of all of the technical staff - including the beam and source experts - tomorrow morning at 09:00, so there may be more information about the best course of action after that.
Wilton was consulted as one of the spokespeople of the experiment, and he approves of this action. |
Attachment 1: run62_Scaler_Mar14_00h40.png
|
|
Pending | e793s |
Wed Mar 17 08:03:15 2021 |
Laetitia, Charlie | AGATA | Hardware | CD2 | Agata issue | There is a problem on Agata (the dewars are not being filled), and the Ge's start to warm up. The beam has been stopped at ~08h05. |
Ongoing | TEST |
Tue Jan 22 18:52:23 2019 |
Valérian-Marlène | GRIT | General | 3-alpha | 3 alpha source run on MG10C & MG3T | 3 alpha source (not centered)
MG10C and MG3T ON
MG10C is connected to MG5T HT and BT and seen on MUVI1 channel2 -> online name=MM2
MG3T is connected to MG3T HT and BT and seen on MUVI1 channel3 -> online name=MM3
Attached : current and bias of the detectors |
Ongoing | e744s |
Sat Apr 13 21:57:48 2019 |
Adrien Matta | RUN | General | N/A | Inconsistent info in the run message |
Adrien Matta wrote: | Be careful to fill in correctly all info in the start run message.
There is inconsistency in run 121 comment (target 2) and list (target 3).
Thèse information will be used for the analysis. |
--> Well.. in entry 4175 it is written target 3...
... ... |
Ongoing | |
Sat Jul 20 09:32:02 2019 |
Kseniia, Christian, Nikola, | OTHER | General | | summary of the morning shift | timeline of run 116
09:15 resync AGATA crustal 13C, AGATA was in back-preassure for ~7 min
10:30 resync AGATA crustal 13C, AGATA was in back-preassure for ~3 min
10:40 run 116 stopped (PCP called, they have a problem with primary beam high intensity reader, they called in the on-call people) |
Ongoing | e768s |
Sun Jul 28 19:58:57 2019 |
Nicolas | RUN | General | | Update on possible restart | Beam is now at the exit of CSS2 though We may have beam around midnight.... |
|
Ongoing | e768s |
Mon Jul 29 22:50:43 2019 |
Sylvain & Freddy | ONLINE | General | LiF | Screenshots run 158 @ 22h05 |
|
Attachment 1: Vamos_tab9.png
|
|
Attachment 2: Vamos_tab8.png
|
|
Attachment 3: Vamos_tab7.png
|
|
Attachment 4: Vamos_tab6.png
|
|
Attachment 5: Vamos_tab5.png
|
|
Attachment 6: Vamos_tab4.png
|
|
Attachment 7: Vamos_tab3.png
|
|
Attachment 8: Vamos_tab2.png
|
|
Attachment 9: Vamos_tab1.png
|
|
Attachment 10: Mugast_tab10.png
|
|
Attachment 11: Mugast_tab9.png
|
|
Attachment 12: Mugast_tab8.png
|
|
Attachment 13: Mugast_tab7.png
|
|
Attachment 14: Mugast_tab6.png
|
|
Attachment 15: Mugast_tab5.png
|
|
Attachment 16: Mugast_tab4.png
|
|
Attachment 17: Mugast_tab3.png
|
|
Attachment 18: Mugast_tab2.png
|
|
Attachment 19: Mugast_tab1.png
|
|
Ongoing | e775s |
Sat Feb 29 03:26:52 2020 |
mugast | RUN | General | N/A | Continues RUN 153 |
Run 153 (target #2 CD2+Au)
Beam Intensity 4*10⁵
**** MUST2 crate A:
+10v/mg Slot: 5 (A1516B ) Channel:5 Vmon: 10 V Imon: 0.406 A Status: On
+10v/t14mg Slot: 2 (A1516 ) Channel:5 Vmon: 0 V Imon: 0.001 A Status: Off
+2.5V/MG10 Slot: 1 (A1518B ) Channel:2 Vmon: 0 V Imon: 0.015 A Status: Off
+2.5v/MG9 Slot: 1 (A1518B ) Channel:1 Vmon: 0 V Imon: 0.013 A Status: Off
+2.5v/mg1 Slot: 4 (A1518B ) Channel:2 Vmon: 2.5 V Imon: 1.264 A Status: On
+2.5v/mg11 Slot: 3 (A1518B ) Channel:2 Vmon: 2.49 V Imon: 1.197 A Status: On
+2.5v/mg3 Slot: 3 (A1518B ) Channel:0 Vmon: 2.49 V Imon: 1.241 A Status: On
+2.5v/mg4 Slot: 4 (A1518B ) Channel:1 Vmon: 2.49 V Imon: 1.24 A Status: On
+2.5v/mg5 Slot: 3 (A1518B ) Channel:1 Vmon: 2.49 V Imon: 1.253 A Status: On
+2.5v/mg7 Slot: 4 (A1518B ) Channel:0 Vmon: 2.49 V Imon: 1.269 A Status: On
+2.5v/t1 Slot: 0 (A1518B ) Channel:0 Vmon: 0 V Imon: 0.007 A Status: Off
+2.5v/t2 Slot: 0 (A1518B ) Channel:1 Vmon: 0 V Imon: 0.006 A Status: Off
+2.5v/t3 Slot: 0 (A1518B ) Channel:2 Vmon: 0 V Imon: 0 A Status: Off
+2.5v/t4 Slot: 1 (A1518B ) Channel:0 Vmon: 0 V Imon: 0.011 A Status: Off
+3.3v/mg Slot: 5 (A1516B ) Channel:2 Vmon: 3.29 V Imon: 0.973 A Status: On
+3.3v/t14mg Slot: 2 (A1516 ) Channel:2 Vmon: 0.03 V Imon: 0 A Status: Off
+5v/mg3511 Slot: 5 (A1516B ) Channel:3 Vmon: 5 V Imon: 0.546 A Status: On
+5v/mg741 Slot: 5 (A1516B ) Channel:4 Vmon: 5 V Imon: 0.555 A Status: On
+5v/t123 Slot: 2 (A1516 ) Channel:3 Vmon: 0.02 V Imon: 0.001 A Status: Off
+5v/t4mg Slot: 2 (A1516 ) Channel:4 Vmon: 0 V Imon: 0.001 A Status: Off
-2.5V/MG10 Slot: 1 (A1518B ) Channel:5 Vmon: 0 V Imon: 0.009 A Status: Off
-2.5V/MG9 Slot: 1 (A1518B ) Channel:4 Vmon: 0 V Imon: 0 A Status: Off
-2.5v/mg1 Slot: 4 (A1518B ) Channel:5 Vmon: 2.49 V Imon: 1.797 A Status: On
-2.5v/mg11 Slot: 3 (A1518B ) Channel:5 Vmon: 2.48 V Imon: 1.787 A Status: On
-2.5v/mg3 Slot: 3 (A1518B ) Channel:3 Vmon: 2.49 V Imon: 1.782 A Status: On
-2.5v/mg4 Slot: 4 (A1518B ) Channel:4 Vmon: 2.49 V Imon: 1.781 A Status: On
-2.5v/mg5 Slot: 3 (A1518B ) Channel:4 Vmon: 2.49 V Imon: 1.807 A Status: On
-2.5v/mg7 Slot: 4 (A1518B ) Channel:3 Vmon: 2.5 V Imon: 1.806 A Status: On
-2.5v/t1 Slot: 0 (A1518B ) Channel:3 Vmon: 0 V Imon: 0 A Status: Off
-2.5v/t2 Slot: 0 (A1518B ) Channel:4 Vmon: 0 V Imon: 0 A Status: Off
-2.5v/t3 Slot: 0 (A1518B ) Channel:5 Vmon: 0 V Imon: 0.006 A Status: Off
-2.5v/t4 Slot: 1 (A1518B ) Channel:3 Vmon: 0 V Imon: 0.007 A Status: Off
-5v/mg3511 Slot: 5 (A1516B ) Channel:0 Vmon: 4.99 V Imon: 0.357 A Status: On
-5v/mg741 Slot: 5 (A1516B ) Channel:1 Vmon: 5 V Imon: 0.345 A Status: On
-5v/t123 Slot: 2 (A1516 ) Channel:0 Vmon: 0 V Imon: 0.002 A Status: Off
-5v/t4mg Slot: 2 (A1516 ) Channel:1 Vmon: 0.01 V Imon: 0 A Status: Off
CsI/t1 Slot: 12 (A1837 ) Channel:0 Vmon: 0 V Imon: 0 uA Status: Off
CsI/t2 Slot: 12 (A1837 ) Channel:2 Vmon: 0 V Imon: 0 uA Status: Off
CsI/t3 Slot: 12 (A1837 ) Channel:4 Vmon: 0 V Imon: 0 uA Status: Off
CsI/t4 Slot: 12 (A1837 ) Channel:6 Vmon: 0 V Imon: 0 uA Status: Off
CsI/t5 Slot: 13 (A1837 ) Channel:0 Vmon: 0 V Imon: 0 uA Status: Off
CsI/t6 Slot: 13 (A1837 ) Channel:2 Vmon: 0 V Imon: 0 uA Status: Off
CsI/t7 Slot: 13 (A1837 ) Channel:4 Vmon: 0.02 V Imon: 0 uA Status: Off
CsI/t8 Slot: 13 (A1837 ) Channel:6 Vmon: 0 V Imon: 0 uA Status: Off
DSSD/t1 Slot: 12 (A1837 ) Channel:1 Vmon: 0 V Imon: 0 uA Status: Off
DSSD/t2 Slot: 12 (A1837 ) Channel:3 Vmon: 0 V Imon: 0 uA Status: Off
DSSD/t3 Slot: 12 (A1837 ) Channel:5 Vmon: 0 V Imon: 0 uA Status: Off
DSSD/t4 Slot: 12 (A1837 ) Channel:7 Vmon: 0 V Imon: 0 uA Status: Off
MG11A Slot: 13 (A1837 ) Channel:5 Vmon: 79.9 V Imon: 0.2 uA Status: On
MG1T Slot: 13 (A1837 ) Channel:11 Vmon: 139.82 V Imon: 0.7 uA Status: On
MG2T Slot: 12 (A1837 ) Channel:10 Vmon: 80.06 V Imon: 2.75 uA Status: On
MG3T Slot: 13 (A1837 ) Channel:1 Vmon: 159.96 V Imon: 0.5 uA Status: On
MG4T Slot: 13 (A1837 ) Channel:9 Vmon: 95.12 V Imon: 0.8 uA Status: On
MG5T Slot: 13 (A1837 ) Channel:3 Vmon: 149.94 V Imon: 1 uA Status: On
MG6T Slot: 12 (A1837 ) Channel:11 Vmon: 80.02 V Imon: 1.1 uA Status: On
MG7T Slot: 13 (A1837 ) Channel:7 Vmon: 119.98 V Imon: 1.2 uA Status: On
MG9C_MUST2 Slot: 12 (A1837 ) Channel:9 Vmon: 0 V Imon: 0 uA Status: Off
spare1 Slot: 12 (A1837 ) Channel:8 Vmon: 0 V Imon: 0 uA Status: Off
spare3 Slot: 13 (A1837 ) Channel:8 Vmon: 0 V Imon: 0 uA Status: Off
spare4 Slot: 13 (A1837 ) Channel:10 Vmon: 0 V Imon: 0 uA Status: Off
**** MUST2 crate B:
+10V/t1234 Slot: 3 (A1516B ) Channel:5 Vmon: 9.99 V Imon: 0.159 A Status: On
+2.5V/t1 Slot: 1 (A1518B ) Channel:0 Vmon: 2.49 V Imon: 1.294 A Status: On
+2.5V/t2 Slot: 1 (A1518B ) Channel:1 Vmon: 2.49 V Imon: 0.058 A Status: On
+2.5V/t3 Slot: 1 (A1518B ) Channel:2 Vmon: 2.49 V Imon: 0.064 A Status: On
+2.5V/t4 Slot: 2 (A1518B ) Channel:0 Vmon: 2.49 V Imon: 0.006 A Status: On
+2.5V/t5 Slot: 2 (A1518B ) Channel:1 Vmon: 2.49 V Imon: 0.004 A Status: On
+2.5V/t6 Slot: 2 (A1518B ) Channel:2 Vmon: 2.49 V Imon: 1.21 A Status: On
+3.3V/t123 Slot: 3 (A1516B ) Channel:2 Vmon: 3.29 V Imon: 0.327 A Status: On
+5V/t123 Slot: 3 (A1516B ) Channel:3 Vmon: 4.99 V Imon: 0.182 A Status: On
+5V/t456 Slot: 3 (A1516B ) Channel:4 Vmon: 4.91 V Imon: 0.576 A Status: On
-2.5V/t1 Slot: 1 (A1518B ) Channel:3 Vmon: 2.49 V Imon: 0.06 A Status: On
-2.5V/t2 Slot: 1 (A1518B ) Channel:4 Vmon: 2.49 V Imon: 0.06 A Status: On
-2.5V/t3 Slot: 1 (A1518B ) Channel:5 Vmon: 2.49 V Imon: 1.825 A Status: On
-2.5V/t4 Slot: 2 (A1518B ) Channel:3 Vmon: 2.49 V Imon: 0.003 A Status: On
-2.5V/t5 Slot: 2 (A1518B ) Channel:4 Vmon: 2.49 V Imon: 0.005 A Status: On
-2.5V/t6 Slot: 2 (A1518B ) Channel:5 Vmon: 2.49 V Imon: 1.798 A Status: On
-5V/t123 Slot: 3 (A1516B ) Channel:0 Vmon: 5 V Imon: 0.116 A Status: On
-5V/t456 Slot: 3 (A1516B ) Channel:1 Vmon: 5 V Imon: 0.252 A Status: On
**** VAMOS crate:
Slot: 8 (A1821H ) Channel:1 Vmon: 0 V Imon: 0 uA Status: Off
CHANNEL00 Slot: 9 (A1821H ) Channel:0 Vmon: 0 V Imon: 0 uA Status: Off
CHANNEL01 Slot: 9 (A1821H ) Channel:1 Vmon: 0 V Imon: 0 uA Status: Off
CHANNEL02 Slot: 9 (A1821H ) Channel:2 Vmon: 0 V Imon: 0 uA Status: Off
CHANNEL03 Slot: 9 (A1821H ) Channel:3 Vmon: 0 V Imon: 0 uA Status: Off
CHANNEL04 Slot: 9 (A1821H ) Channel:4 Vmon: 0 V Imon: 0 uA Status: Off
CHANNEL05 Slot: 9 (A1821H ) Channel:5 Vmon: 0 V Imon: 0 uA Status: Off
CHANNEL06 Slot: 9 (A1821H ) Channel:6 Vmon: 0 V Imon: 0 uA Status: Off
CHANNEL07 Slot: 9 (A1821H ) Channel:7 Vmon: 0 V Imon: 0 uA Status: Off
CHANNEL08 Slot: 9 (A1821H ) Channel:8 Vmon: 0 V Imon: 0 uA Status: Off
CHANNEL09 Slot: 9 (A1821H ) Channel:9 Vmon: 0 V Imon: 0 uA Status: Off
CHANNEL10 Slot: 9 (A1821H ) Channel:10 Vmon: 0 V Imon: 0 uA Status: Off
CHANNEL11 Slot: 9 (A1821H ) Channel:11 Vmon: 0 V Imon: 0 uA Status: Off
FP_CATS0 Slot: 2 (A1821HMP) Channel:6 Vmon: 729.75 V Imon: 0.733 uA Status: On
FP_CATS1 Slot: 2 (A1821HMP) Channel:7 Vmon: 639.75 V Imon: 0 uA Status: On
FP_CATS2 Slot: 2 (A1821HMP) Channel:8 Vmon: 649.75 V Imon: 0.005 uA Status: On
FP_CATS3 Slot: 2 (A1821HMP) Channel:9 Vmon: 649.75 V Imon: 0.019 uA Status: On
FP_CATS4 Slot: 2 (A1821HMP) Channel:10 Vmon: 664.5 V Imon: 0.16 uA Status: On
FP_CATS5 Slot: 2 (A1821HMP) Channel:11 Vmon: 649.5 V Imon: 0.04 uA Status: On
FP_DC1_cat Slot: 3 (A1821HM ) Channel:0 Vmon: 0 V Imon: 0 uA Status: Off
FP_DC2_cat Slot: 3 (A1821HM ) Channel:1 Vmon: 0 V Imon: 0 uA Status: Off
FP_DC_Wi1 Slot: 6 (A1821H ) Channel:0 Vmon: 0 V Imon: 0 uA Status: Off
FP_DC_Wi2 Slot: 6 (A1821H ) Channel:1 Vmon: 0 V Imon: 0 uA Status: Off
FP_DC_Wi3 Slot: 6 (A1821H ) Channel:2 Vmon: 0 V Imon: 0 uA Status: Off
FP_DC_Wi4 Slot: 6 (A1821H ) Channel:3 Vmon: 0 V Imon: 0 uA Status: Off
FP_IC_Cath Slot: 10 (AG5262 ) Channel:0 Vmon: 7 V Imon: 0.02 uA Status: Off
FP_IC_PAD Slot: 7 (A1821H ) Channel:1 Vmon: 0 V Imon: 0 uA Status: Off
FP_IC_Wi Slot: 7 (A1821H ) Channel:0 Vmon: 0 V Imon: 0 uA Status: Off
FP_IC_casq Slot: 10 (AG5262 ) Channel:1 Vmon: 5 V Imon: 0.02 uA Status: Off
FP_diamond Slot: 8 (A1821H ) Channel:0 Vmon: 0 V Imon: 0.001 uA Status: Off
FP_mob_pla Slot: 5 (A18332 ) Channel:0 Vmon: 0 V Imon: 0 uA Status: Off
Si1 Slot: 8 (A1821H ) Channel:5 Vmon: 0 V Imon: 0 uA Status: Off
Si2 Slot: 8 (A1821H ) Channel:6 Vmon: 0 V Imon: 0 uA Status: Off
Si3 Slot: 8 (A1821H ) Channel:7 Vmon: 0 V Imon: 0 uA Status: Off
Si4 Slot: 8 (A1821H ) Channel:8 Vmon: 0 V Imon: 0 uA Status: Off
Si5 Slot: 8 (A1821H ) Channel:9 Vmon: 0 V Imon: 0 uA Status: Off
Si6 Slot: 8 (A1821H ) Channel:10 Vmon: 0 V Imon: 0 uA Status: Off
Si7 Slot: 8 (A1821H ) Channel:11 Vmon: 0 V Imon: 0 uA Status: Off
TP_DC1 Slot: 3 (A1821HM ) Channel:6 Vmon: 0.5 V Imon: 0.02 uA Status: Off
TP_DC2 Slot: 3 (A1821HM ) Channel:7 Vmon: 0.25 V Imon: 0 uA Status: Off |
Attachment 1: RUN153_online0305.png
|
|
Attachment 2: RUN153_online0306.png
|
|
Attachment 3: RUN153_online0307.png
|
|
Attachment 4: RUN153_online0308.png
|
|
Attachment 5: RUN153_online0309.png
|
|
Ongoing | e775s |
Sat Feb 29 07:35:21 2020 |
mugast : Valerian | AGATA | General | CD2 | AGATA is not responding |
Since ~4:30am AGATA is not responding, everyhting was frozen. Unclear if it comes from disk problem or other... |
Ongoing | e775s |
Sat Feb 29 10:34:31 2020 |
mugast ; valerian irene | OFFLINE | Software | CD2-Au | Issues in the EX reconstruction |
We are currently working on the EX reconstruction which is off for detector 1, 3 and 7... One of our guess might be that there is an issue with the position of the detectors which gives error in theta... As the energy calibration is ok. Joined are :
- the Excitation energy for good detectors compared to all detectors,
- the theoretical kinematic line for the 3.57 state for each detectors,
- the excitation energy as a function of the laboratory angle (for this last spectra a clear dependency is visible for telescopes 1,3 and 7)
- the 3D view, which doesn't show clear problem in position... |
Attachment 1: Comparison_all_and_best_detector.png
|
|
Attachment 2: Kinematics_for_each_detectors.pdf
|
|
Attachment 3: ExvsTheta_each_detectors.pdf
|
|
Attachment 4: MUGAST_3D.pdf
|
|
Ongoing | e775s |
Sat Feb 29 14:27:48 2020 |
mugast ; Valerian | OFFLINE | Software | CD2 | Issues in the EX reconstruction |
mugast ; valerian irene wrote: | We are currently working on the EX reconstruction which is off for detector 1, 3 and 7... One of our guess might be that there is an issue with the position of the detectors which gives error in theta... As the energy calibration is ok. Joined are :
- the Excitation energy for good detectors compared to all detectors,
- the theoretical kinematic line for the 3.57 state for each detectors,
- the excitation energy as a function of the laboratory angle (for this last spectra a clear dependency is visible for telescopes 1,3 and 7)
- the 3D view, which doesn't show clear problem in position... |
Testing changing the beam impact :
-Kine for beam impact = (0,0,0) for reference
-Kine for beam impact = (0,0,-5mm)
-Kine for beam impact = (0,0,+5mm)
-Ex vs ThetaLab for beam impact = (0,0,0) for reference
-Ex vs ThetaLab for beam impact = (0,0,-5mm)
-Ex vs ThetaLab for beam impact = (0,0,+5mm)
Red lines corresponds to Ex = 3.570 MeV and Ex = 4.070 MeV
To try to correct the problem i also offseted the beam in X and Y, but this doesn't solve the problem quite the opposite as it make the result worse for other detectors also... The fact that it is better at lower angle and worse at higher angle might mean that the detectors angles are tilted compared to the
surveyors measurments... I also tried to use the CAO positions for MG3 (see Detectors/mugast.detector) but it didn't solve the problem. |
Attachment 1: Kinematics_for_each_detectorsBeamZ0.pdf
|
|
Attachment 2: Kinematics_for_each_detectorsBeamZminus5.pdf
|
|
Attachment 3: Kinematics_for_each_detectorsBeamZplus5.pdf
|
|
Attachment 4: ExvsTheta_each_detectorsZ0.pdf
|
|
Attachment 5: ExvsTheta_each_detectorsZminus5.pdf
|
|
Attachment 6: ExvsTheta_each_detectorsZplus5.pdf
|
|
Ongoing | e775s |
Sat Feb 29 15:53:12 2020 |
mugast ; Valerian | OFFLINE | Software | CD2-Au | Issues in the EX reconstruction |
mugast ; Valerian wrote: |
mugast ; valerian irene wrote: | We are currently working on the EX reconstruction which is off for detector 1, 3 and 7... One of our guess might be that there is an issue with the position of the detectors which gives error in theta... As the energy calibration is ok. Joined are :
- the Excitation energy for good detectors compared to all detectors,
- the theoretical kinematic line for the 3.57 state for each detectors,
- the excitation energy as a function of the laboratory angle (for this last spectra a clear dependency is visible for telescopes 1,3 and 7)
- the 3D view, which doesn't show clear problem in position... |
Testing changing the beam impact :
-Kine for beam impact = (0,0,0) for reference
-Kine for beam impact = (0,0,-5mm)
-Kine for beam impact = (0,0,+5mm)
-Ex vs ThetaLab for beam impact = (0,0,0) for reference
-Ex vs ThetaLab for beam impact = (0,0,-5mm)
-Ex vs ThetaLab for beam impact = (0,0,+5mm)
Red lines corresponds to Ex = 3.570 MeV and Ex = 4.070 MeV
To try to correct the problem i also offseted the beam in X and Y, but this doesn't solve the problem quite the opposite as it make the result worse for other detectors also... The fact that it is better at lower angle and worse at higher angle might mean that the detectors angles are tilted compared to the
surveyors measurments... I also tried to use the CAO positions for MG3 (see Detectors/mugast.detector) but it didn't solve the problem. |
By looking at the EnergyCalibrator.C, I saw that the extrapolation used to find the dead layer of aluminum is commented... This might be an answer to the problem... |
Ongoing | e775s |
Sat Feb 29 17:18:02 2020 |
mugast ; Valerian, Marlène , Pierre | OFFLINE | Software | CD2-Au | Issues in the EX reconstruction |
mugast ; Valerian wrote: |
mugast ; Valerian wrote: |
mugast ; valerian irene wrote: | We are currently working on the EX reconstruction which is off for detector 1, 3 and 7... One of our guess might be that there is an issue with the position of the detectors which gives error in theta... As the energy calibration is ok. Joined are :
- the Excitation energy for good detectors compared to all detectors,
- the theoretical kinematic line for the 3.57 state for each detectors,
- the excitation energy as a function of the laboratory angle (for this last spectra a clear dependency is visible for telescopes 1,3 and 7)
- the 3D view, which doesn't show clear problem in position... |
Testing changing the beam impact :
-Kine for beam impact = (0,0,0) for reference
-Kine for beam impact = (0,0,-5mm)
-Kine for beam impact = (0,0,+5mm)
-Ex vs ThetaLab for beam impact = (0,0,0) for reference
-Ex vs ThetaLab for beam impact = (0,0,-5mm)
-Ex vs ThetaLab for beam impact = (0,0,+5mm)
Red lines corresponds to Ex = 3.570 MeV and Ex = 4.070 MeV
To try to correct the problem i also offseted the beam in X and Y, but this doesn't solve the problem quite the opposite as it make the result worse for other detectors also... The fact that it is better at lower angle and worse at higher angle might mean that the detectors angles are tilted compared to the
surveyors measurments... I also tried to use the CAO positions for MG3 (see Detectors/mugast.detector) but it didn't solve the problem. |
By looking at the EnergyCalibrator.C, I saw that the extrapolation used to find the dead layer of aluminum is commented... This might be an answer to the problem... |
WE HAVE A SOLUTION!!! It seems that the order of the detectors in mugast.detector matters!! Now we start with a dummy MG8, then MG1, MG2, ... And everything is fine!! We think that is because it was taking the wrong calibration files (to be checked).
Joined are (for run 128,137,138,139,151,152,153):
-Ex vs ThetaLab
-Ex for all detectors
-Egamma vs Ex |
Attachment 1: ExvsTheta_each_detectorsOK.pdf
|
|
Attachment 2: Ex_GOOD.pdf
|
|
Attachment 3: EgvsEx_GOOD.pdf
|
|
Ongoing | e793s |
Wed Mar 3 17:12:57 2021 |
Franco | S1 | Hardware | N/A | Schema of the new S1 |
The signals of the new S1 will be splitted among 6 Mugast boards, according to the scheme attached.
The HV for the S1 have been updated accordingly, with labels like MG11/CsIMGXX. |
Attachment 1: S1.PNG
|
|
Ongoing | e793s |
Wed Mar 10 21:09:25 2021 |
mugast (Adrien, Nigel et co) | BEAM | General | | Primary beam issues |
The secondary beam intensity has been fluctuating around 4 - 5 x 10**5 Hz most of the evening. 15 mins ago it was realised that it had dropped off to only 1 x 10**5 Hz. As suspected, PCP confirmed that primary beam intensity has dropped off. They are investigating ...
It would be good to have the TI (primary beam intensity) available to monitor here in the daq room. |
Ongoing | e793s |
Thu Mar 11 11:27:28 2021 |
Nicolas, Lynda, Charlie | RUN | General | CD2 | run 56 screenshots |
Scaler only |
Attachment 1: run56_scalar_start.png
|
|
Ongoing | e793s |
Fri Mar 12 12:58:10 2021 |
Emmanuel C | OFFLINE | General | | Gamma Gamma Analysis |
|
Attachment 1: 48K.pdf
|
|
Attachment 2: e793s_1.png
|
|
Attachment 3: e793s_142Gated.png
|
|
Attachment 4: e793s_280Gated.png
|
|
Attachment 5: e793s_451Gated.png
|
|
Attachment 6: e793s_1842Gated.png
|
|
Attachment 7: e793s_3523Gated.png
|
|