MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 62 of 132  Not logged in MUGAST
Fixed, , Tue Mar 19 19:35:53 2019, marlene, GRIT, General, , RUN 98 3alpha source all detectors 
RUN 98
3 alpha source
all detectors (trap,square, MUST2,annular)
Temp=15deg
Acquisition mugast
Fixed, TEST, Tue Feb 18 10:00:40 2020, mugast, RUN, General, 3-alpha, RUN 501 - missing info 
RUN 501

alpha run with cooling during the night from 13-14/02/2020
Fixed, TEST, Thu Feb 21 09:08:49 2019, Freddy, RUN, General, 3-alpha, RUN 30 - alpha on MG 1,3,5,7  3alpha_source_MG_1357_20Feb.pngLowV_bis_run30.pngLowV_run30.pngHV_run30.png
Run 30

Alpha source run for trapezoid detectors during the night with pedestals aligned and zero suppression. Cooling at 15°C.
Thresholds at 500 keV.
Run size: 1.3 G

On MUVI1:

1--> MG5 (150V)
2--> --
3--> MG3 (160V)
4--> --

On MUVI2:
5--> MG7 (120V)
6--> MG1 (140V)
7--> --
8--> --

TRIGGERS are:
GMT ch1 --> OR MUVI1
GMT ch2 --> OR MUVI2
GMT ch3 --> OR MUVI3 (nothing on it)

Plots are now represented using the correspondance map which means that it correspond to the geometrical ordering on the detector. With this representation groups of strips missing or disfunctionning on a given region of the detector is more obvious.
Fixed, , Thu Feb 14 15:30:14 2019, Marlene Daniele Jenny, GRIT, General, 3-alpha, RUN 28 3alpha_sourceMG03XY._OKpng3alpha_sourceMG11XY.png3alpha_sourceMG01XY_OK.png
RUN 28 14th Feb
3 alpha source on MG11A, MG3T, MG1T

GMT channels
MG11
-
MG3
-
-
MG1
-
Fixed, , Thu Feb 14 08:50:58 2019, Marlene, GRIT, General, 3-alpha, RUN 27 8x
RUN 27 on 14th of feb.
3 alpha source (one in front of trap., one in front of square detectors)

MG10, MG4, MG7 and MG5 are OK.
MG1 shows alpha but at lower energy and with bad resolution (particularly on X). We suspect it is not well depleted. We tried to increase the bias up to 200V but did not see any change.
It is due to a non working SA/SL external cable that produces this error (we tested it with several detectors, we also tried to change the MUVI channel and it still does the same)

Conclusion :
We have tested all trapezoid detectors (MG1, MG3, MG4, MG5, MG7) the annular (MG11) and the thin square (MG10) they are all OK




Detectors (order in the GMT) :
MG10
MG1
MG4
-
MG7
MG5
-
-

attached : bias of detectors ( be careful : MG10 bias = MG11A)
spectra for all detectors
Fixed, e744s, Thu Apr 18 09:28:12 2019, Marlène, MUST2, General, 3-alpha, RUN 206 (mugast) 3Alpha calib after e744s experiment run206_calibMMY.pngrun206_calibMGY.pngrun206_calibMMX.pngrun206_calibMGX.png
RUN 206
3 alpha source all detectors with no change in the configuration (low thresholds in MUST2)
From 17/04 18h13 -> 18/04 9h31
Fixed, TEST, Mon Mar 18 18:31:18 2019, valerian marlène, GRIT, General, 3-alpha, RUN 20 (com2019) 3 alpha overnight RUN20_HV_MG_3lapha.pngRUN20_3alpha_HV_MUST2.png
RUN20

Run with double 3alpha source MUST2+MG1_7+MG11
temperature -4 deg on both cooling

Acquisition with VAMOS+MUGAST coupled (com2019)

Time range =600ns
pedestal aligned and suppressed

run start at 18h23 monday 18

run stop at 8h34 tuesday 19

about 12Go of data
    Fixed, e775s, Fri Feb 28 14:00:08 2020, Marlène, RUN, General, , RUN 151 beam stop 

Marlène wrote:
RUN 151

Change of target now : CD2+AU #2
Beam intensity : 4e+5pps



beam stopped due to unstable equipment in CSS1.
expected duration : 15 min
Fixed, e775s, Fri Feb 28 11:21:31 2020, Marlène, RUN, General, , RUN 151 
RUN 151

Change of target now : CD2+AU #2
Beam intensity : 4e+5pps
Fixed, , Fri Feb 28 11:20:22 2020, Marlène, GRIT, General, , RUN 150 
RUN 150
target CD2 300 ug/cm2
Beam intensity 4e+5pps

Note that there is no 144,145,146,147, 148 ans 149 runs
Fixed, , Thu Mar 28 18:29:04 2019, marlene valerian, GRIT, General, 3-alpha, RUN 147 3alpha source overnight all detectors -4deg 
RUN 147 start@ 18:30

3 alpha source over night

cooling at -4 deg on all detectors

warning !!! FIFO for BLK signal has been put in 4x4 instead of 2x8, BLK is only coming from the 3 VXI of MUGAST
    Fixed, e775s, Fri Feb 28 11:49:46 2020, mugast, RUN, General, CD2, RUN 142/143 preliminary analysis with gamma ExEg_RUN142full.png
    Fixed, , Thu Feb 27 16:25:45 2020, Marlène, RUN, General, , RUN 142-143 screenshots RUN143_ElabThetaLab.pngRUN143_impact.pngRUN143_tac.pngRUN143_tof.png

Marlène wrote:
RUN142 on AGATA and RUN143 on MUGAST/VAMOS

target CD2 (350ug/cm2)
Trigger MUGAST
Beam intensity : 5e+5pps



!!!! there is a shift in the run number on AGATA with respect to VAMOS/MUGAST
Fixed, , Thu Feb 27 15:56:09 2020, Marlène, RUN, General, , RUN 142 AGATA- 143 VAMOS-MUGAST 
RUN142 on AGATA and RUN143 on MUGAST/VAMOS

target CD2 (350ug/cm2)
Trigger MUGAST
Beam intensity : 5e+5pps



!!!! there is a shift in the run number on AGATA with respect to VAMOS/MUGAST
    Fixed, , Thu Feb 27 16:30:53 2020, Marlène, RUN, General, , RUN 142 -143 preliminary analysis  RUN143_ana_Elab_Thetalab.pngRUN143_ana_Ex_Eg.pngRUN143_ana_Ex.png

Marlène wrote:

Marlène wrote:
RUN142 on AGATA and RUN143 on MUGAST/VAMOS

target CD2 (350ug/cm2)
Trigger MUGAST
Beam intensity : 5e+5pps



!!!! there is a shift in the run number on AGATA with respect to VAMOS/MUGAST
Fixed, , Thu Feb 27 15:54:09 2020, Marlène , RUN, General, , RUN 141 junk (problem at start) 
Fixed, e775s, Thu Feb 27 15:51:03 2020, mugast, RUN, General, , RUN 140 
Target CD2 thin target
Beam intensity ~ 5e+5 pps


to check the gammas from 20O
    Fixed, e775s, Wed Feb 26 16:19:35 2020, Marlène, RUN, General, , RUN 137 (suite) Ex_Egamma_run137.png
Here is the E gamma versus Excitation energy spectra for run 137



Marlène wrote:
RUN 137
CD2+Au #2
beam intensity 6e+5pps

beam was retuned during the night, seems to be degraded although intensity is higher.
Fixed, , Wed Feb 26 14:31:47 2020, Fairouz, Nicolas, Marlène, GRIT, General, , RUN 137  
Run 137
CD2+Au #2
Beam intensity 6e+5pps on average

Beam was retuned during the night and the quality of the beam is probably degraded
Fixed, e775s, Wed Feb 26 15:02:38 2020, Marlène, RUN, General, , RUN 137 Ex_137.pdfEx_run128_137.pdf
RUN 137
CD2+Au #2
beam intensity 6e+5pps

beam was retuned during the night, seems to be degraded although intensity is higher.

ELOG Home Page