Status |
ExpNbr |
Date |
Author |
Type |
Category |
Target-Source |
Subject |
Fixed | e786s |
Fri Jun 28 04:21:48 2019 |
Daniele Mengoni | RUN | General | Cryo-3He | run 74 |
screenshots run 74 attached
frequent error message on the FP DC Wi1 --> defqlt. The elog said so, card tripped, but spectra are still present .. puzzling |
Attachment 1: Q_vs_Xf_K1.png
|
|
Attachment 2: Q_vs_Xf_K2.png
|
|
Attachment 3: Q_vs_Xf_K3.png
|
|
Attachment 4: Q_vs_Xf_K4.png
|
|
Attachment 5: Q_vs_Xf_K5.png
|
|
Attachment 6: Q_vs_Xf_K6.png
|
|
Attachment 7: Q_vs_Xf_K7.png
|
|
Attachment 8: Q_vs_Xf_K8.png
|
|
Attachment 9: Q_vs_Xf_K9.png
|
|
Attachment 10: Q_vs_Xf_K10.png
|
|
Attachment 11: Q_vs_Xf_K11.png
|
|
Attachment 12: Q_vs_Xf_K12.png
|
|
Attachment 13: Q_vs_Xf_K13.png
|
|
Attachment 14: Q_vs_Xf_K14.png
|
|
Attachment 15: Q_vs_Xf_K15.png
|
|
Attachment 16: Q_vs_Xf_K16.png
|
|
Attachment 17: Q_vs_Xf_K17.png
|
|
Attachment 18: Q_vs_Xf_K18.png
|
|
Fixed | e786s |
Fri Jun 28 08:54:50 2019 |
Freddy | RUN | General | Cryo-3He | run 74 |
Daniele Mengoni wrote: | screenshots run 74 attached
frequent error message on the FP DC Wi1 --> defqlt. The elog said so, card tripped, but spectra are still present .. puzzling |
Put the voltage back at 8h00 when i arrived. |
Fixed | e793s |
Wed Mar 17 10:23:44 2021 |
Cyril | RUN | General | CD2 | run 64 screenshots at 10h05 - End of e793s |
The beam/acq has been stopped around 10h05. |
Attachment 1: run64_10h05_Scaler4.png
|
|
Attachment 2: run64_10h05_VAMOS.png
|
|
Attachment 3: run64_10h05_TACS.png
|
|
Attachment 4: run64_10h05_MM_DeE.png
|
|
Attachment 5: run64_10h05_MG_TOF.png
|
|
Attachment 6: run64_10h05_MM_TOF.png
|
|
Attachment 7: run64_10h05_Scaler4.png
|
|
Attachment 8: run64_10h05_MMImpact.png
|
|
Fixed | e793s |
Wed Mar 17 07:52:37 2021 |
Laetitia, Charlie | RUN | General | CD2 | run 64 screenshots at 07h45 |
|
Attachment 1: run64_07h45_Scaler2.png
|
|
Attachment 2: run64_07h45_Scaler1.png
|
|
Attachment 3: run64_07h45_ElabThetalab.png
|
|
Attachment 4: run64_07h45_MM_DeE.png
|
|
Attachment 5: run64_07h45_MG_TOF.png
|
|
Attachment 6: run64_07h45_MM_TOF.png
|
|
Attachment 7: run64_07h45_TACS.png
|
|
Attachment 8: run64_07h45_VAMOS.png
|
|
Attachment 9: run64_07h45_MMImpact.png
|
|
Attachment 10: run64_07h45_MGImpact.png
|
|
Fixed | e793s |
Wed Mar 17 05:49:02 2021 |
Laetitia, Charlie | RUN | General | CD2 | run 63 screenshots at 05h45 |
|
Attachment 1: run63_05h45_Scaler2.png
|
|
Attachment 2: run63_05h45_Scaler1.png
|
|
Attachment 3: run63_05h45_MM_DeE.png
|
|
Attachment 4: run63_05h45_MG_TOF.png
|
|
Attachment 5: run63_05h45_MM_TOF.png
|
|
Attachment 6: run63_05h45_TACS.png
|
|
Attachment 7: run63_05h45_VAMOS.png
|
|
Attachment 8: run63_05h45_MMImpact.png
|
|
Attachment 9: run63_05h45_MGImpact.png
|
|
Fixed | e793s |
Sat Mar 13 12:07:32 2021 |
Lynda, Laetitia, Chloe | ONLINE | General | CD2 | run 62 screenshots at 11h55 |
|
Attachment 1: run62_Theta_E_11h55.png
|
|
Attachment 2: run62_MG_IMPACT_11h55.png
|
|
Attachment 3: run62_MM_IMPACT_11h55.png
|
|
Attachment 4: run62_Vamos-MW_11h55.png
|
|
Attachment 5: run62_FSC_11h55.png
|
|
Attachment 6: run62_TAC_11h55.png
|
|
Attachment 7: run62_MG_ToF_11h55.png
|
|
Attachment 8: run62_MM_DeE_11h55.png
|
|
Attachment 9: run62_MM_ToF_11h55.png
|
|
Attachment 10: run62_scaler_graph_13h44.png
|
|
Fixed | e793s |
Mon Mar 15 06:59:02 2021 |
Sarah, Louis, Charlie | RUN | General | CD2 | run 62 screenshots at 06h55 |
Screenshots attached. Note that the fluctuation issues have returned after the "electrostatic trip"(?) but the beam pps remain pretty high. |
Attachment 1: run62_scaler_06h55.png
|
|
Attachment 2: run62_MM_DeE_06h55.png
|
|
Attachment 3: run62_MM_ToF_06h55.png
|
|
Attachment 4: run62_MG_TOF_06h55.png
|
|
Attachment 5: run62_TAC_06h55.png
|
|
Attachment 6: run62_VAMOS_06h55.png
|
|
Attachment 7: run62_MM_IMPACT_06h55.png
|
|
Attachment 8: run62_MG_IMPACT_06h55.png
|
|
Fixed | e793s |
Mon Mar 15 04:57:32 2021 |
Sarah, Louis, Charlie | RUN | General | CD2 | run 62 screenshots at 04h50 |
The beam intensity has been surprisingly high so far, considering the news we were given about the source a few hours ago. It has just taken a dip to ~4.5e5 in the last 10 mins or so. |
Attachment 1: run62_scaler_04h50.png
|
|
Attachment 2: run62_MM_IMPACT_04h50.png
|
|
Attachment 3: run62_MM_ToF_04h50.png
|
|
Attachment 4: run62_MM_DeE_04h50.png
|
|
Attachment 5: run62_MG_TOF_04h50.png
|
|
Attachment 6: run62_TAC_04h50.png
|
|
Attachment 7: run62_VAMOS_04h50.png
|
|
Attachment 8: run62_MG_IMPACT_04h50.png
|
|
Fixed | e786s |
Wed Jun 26 20:46:20 2019 |
Sylvain, Louis | RUN | General | Cryo-3He | run 62 screenshots |
|
Attachment 1: Vamos10.png
|
|
Attachment 2: Vamos9.png
|
|
Attachment 3: Vamos8.png
|
|
Attachment 4: Vamos7.png
|
|
Attachment 5: Vamos6.png
|
|
Attachment 6: Vamos5.png
|
|
Attachment 7: Vamos4.png
|
|
Attachment 8: Vamos3.png
|
|
Attachment 9: Vamos2.png
|
|
Attachment 10: Vamos1.png
|
|
Attachment 11: Vamos.png
|
|
Attachment 12: MGToD3.png
|
|
Attachment 13: MGToD2.png
|
|
Attachment 14: MGToD1.png
|
|
Attachment 15: MMToD.png
|
|
Attachment 16: MGImpact.png
|
|
Attachment 17: MMImpact.png
|
|
Attachment 18: scaler.png
|
|
Fixed | e793s |
Fri Feb 12 08:52:30 2021 |
mugast | RUN | General | 3-alpha | run 613: long alpha run |
all Muvi in trigger, mugast stand alone. all detector Bias on. |
Fixed | TEST |
Tue Mar 5 12:13:41 2019 |
iulian Marlene | MUST2 | Software | N/A | run 60 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X |
RUNS 58,59,60 are bad runs
Run 61
Calimero on Muvi3 MM1(X)
Gamme 600 ns
Procedure:
0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco) |
Attachment 1: run60_calimero_mm1X.png
|
|
Attachment 2: run60_calimero_mm1X_spectra.png
|
|
Fixed | TEST |
Tue Mar 5 10:16:03 2019 |
iulian | MUST2 | Software | N/A | run 57 Gamme 300 ns Time Calibration run (Calimero) on MUVI 3 (MUST) X sectors more stats (10000) |
RUNS 53,54,55,56 are bad runs
Run 57
Calimero on Muvi3 MM1(X)
Gamme 300 ns
Procedure:
0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco) |
Attachment 1: run57_calimero_config.png
|
|
Attachment 2: run57_calimero_mm1.png
|
|
Fixed | e793s |
Thu Mar 11 17:32:07 2021 |
Faïrouz, Franco, Sarah | RUN | General | CD2 | run 56 screenshots 17h30 |
|
Attachment 1: run56_MG_IMPACT_MATRIX_17h30.png
|
|
Attachment 2: run56_MM_ToF_17h30.png
|
|
Attachment 3: run56_MM_DeE_17h27.png
|
|
Attachment 4: run56_MG-ToF_17h27.png
|
|
Attachment 5: run56_TAC_17h27.png
|
|
Attachment 6: run56_Vamos_MW_17h26.png
|
|
Attachment 7: run56_MM_Impact_Matrix_17h26.png
|
|
Attachment 8: run56_scaler_17h25.png
|
|
Fixed | e793s |
Thu Mar 11 15:26:37 2021 |
Nicolas, Lynda, Charlie, Sarah | RUN | General | CD2 | run 56 screenshots 15h30 |
online histograms reset around 15h45 |
Attachment 1: run56_mg_impact_matrix_02.png
|
|
Attachment 2: run56_mm_de_e_02.png
|
|
Attachment 3: run56_mg_e_tof_02.png
|
|
Attachment 4: run56_tac_02.png
|
|
Attachment 5: run56_FSC_02.png
|
|
Attachment 6: run56_vamos_02.png
|
|
Attachment 7: run56_mm_impact_matrix_02.png
|
|
Attachment 8: run56_scaler_15h30.png
|
|
Fixed | e786s |
Mon Jun 24 20:39:54 2019 |
Freddy | RUN | General | Cryo-3He | run 56 screenshots |
|
Attachment 1: scalers_run55_24062019_20h37.png
|
|
Ongoing | e793s |
Thu Mar 11 11:27:28 2021 |
Nicolas, Lynda, Charlie | RUN | General | CD2 | run 56 screenshots |
Scaler only |
Attachment 1: run56_scalar_start.png
|
|
Fixed | e793s |
Thu Mar 11 12:39:01 2021 |
Nicolas, Lynda, Charlie, Sarah | RUN | General | CD2 | run 56 screenshots |
|
Attachment 1: run56_mm_impact_matrix_01.png
|
|
Attachment 2: run56_vamos_01.png
|
|
Attachment 3: run56_tac_01.png
|
|
Attachment 4: run56_mg_e_tof_01.png
|
|
Attachment 5: run56_mm_de_e_01.png
|
|
Attachment 6: run56_mm_e_tof_01.png
|
|
Attachment 7: run56_mg_impact_matrix_01.png
|
|
Fixed | e793s |
Thu Mar 11 10:25:30 2021 |
Nicolas, Lynda, Charlie | RUN | General | CD2 | run 55 screenshots |
Start of run, so scaler only - low stats in vigru plots |
Attachment 1: run54_start_Scalers.png
|
|
Fixed | e793s |
Thu Mar 11 01:24:47 2021 |
Ablaihan, Louis | RUN | General | CD2 | run 53 screenshots |
Screenshots of the run 53
Beam intensity is ~ 7.4 10^5 pps in CATS and 5.3 10^5 in VAMOS
~75% transmition
Dt is ~ 3% |
Attachment 1: run53_Scalers.png
|
|
Attachment 2: run53_MM_Impact.png
|
|
Attachment 3: run53_Times.png
|
|
Attachment 4: run53_VAMOS_MWpng
|
Attachment 5: run53_MM_TY.png
|
|
Attachment 6: run53_MM_TX.png
|
|
Attachment 7: run53_MG_TY.png
|
|
Attachment 8: run53_MG_TX.png
|
|
Attachment 9: run53_MM_ErawY.png
|
|
Attachment 10: run53_MM_ErawX.png
|
|
Attachment 11: run53_MG_ErawY.png
|
|
Attachment 12: run53_MG_ErawX.png
|
|
Attachment 13: run53_MG_E_TOF.png
|
|
Attachment 14: run53_MM_dEE.png
|
|
Attachment 15: run53_MM_E_TOF.png
|
|
Attachment 16: run53_MG_impact.png
|
|
Fixed | e793s |
Thu Mar 11 05:12:36 2021 |
Ablaihan, Louis | RUN | General | CD2 | run 53 new screenshots |
|
Attachment 1: run53_1_MMMG_T.png
|
|
Attachment 2: run53_1_MM_Impact.png
|
|
Attachment 3: run53_1_MG_impact.png
|
|