Fixede793s Mon Mar 15 10:18:37 2021 Taras, cyril, LéoONLINEHardwareCD2Cats2 trippedCats2 tripped with vamos, had to be reset
Beam is back
 scalers10h18.png 
Fixede793s Mon Mar 15 12:08:49 2021 Taras, cyril, LéoONLINEHardwareCD2 vamos dipole tripped at 12h05Vamos dipole tripped at around 12h05, had
to be reset
  
Pendinge793s Wed Mar 17 08:03:15 2021 Laetitia, CharlieAGATAHardwareCD2Agata issueThere is a problem on Agata (the dewars are
not being filled), and the Ge's start to
warm up. The beam has been stopped at ~08h05.
  
Fixede793s Wed Mar 17 09:53:38 2021 CyrilAGATAHardwareCD2Agata issueAgata problem has been solved, a valve was
closed by mistake.
Beam returned at 9h40.
  
FixedTEST Wed Jan 16 16:33:28 2019 MarlèneGRITGeneral3-alphaStatus of detectors on 16/01/19MG1 has some issue with the HT cable (for
the BT bias). There is a shortcut only under
vacuum. We checked the connector but we have
 Detectors_status_160119.xlsxCapture_d’eÌcran_2019-01-16_aÌ€_17.05.57.png 
FixedTEST Wed Jan 16 17:36:30 2019 Valérian-MarlèneGRITGeneralN/ACalimero tests - How to configure Calimero in GECOTests of Calimero time calibration with MUVI1
channel 2 (MM2)
 Calimero_MUVI1_MM2_voie3_allmates_bidim.pngCalimero_MUVI1_MM2_voie1_allmates_bidim.pngCalimero_MUVI1_MM3_voie1_allmates_bidim_withchannelsenabled.pngCalimero_MUVI1_MM3_voie3_allmates_bidim_withchannelsenabled.pngCalimero_MUVI1_MM2_allchannels_allmates_bidim_stat100000_withchannelsenabled.png 
FixedTEST Fri Jan 18 15:32:50 2019 Adrien ValérianGRITGeneralN/ACalimero tests Muvi2 MM7 MG10CAfter disabling nptool Multithreading, online
is working fine again.
 Calimero_MUVI2_MM7_allchannels_allmates__stat100000_withchannelsenabled.pngCalimero_MUVI2_MM7_allchannels_allmates_withchannelsenabled.png 
Fixed  Tue Jan 22 11:15:15 2019 MarlèneGRITGeneralN/ASquare detectorsMG9C (thin square detector coupled with thick
square detector unplugged) noisy
On X channels --> no noise problem ( about
 MG9C_disabled_channels__no_crazy_ORD.pngMG9C_disabled_channels__no_crazy_ORD_moredetailed.pngMG9C_piedestal_of_disabled_channels_Y.png 
OngoingTEST Tue Jan 22 18:52:23 2019 Valérian-MarlèneGRITGeneral3-alpha3 alpha source run on MG10C & MG3T3 alpha source (not centered)
MG10C and MG3T ON
MG10C is connected to MG5T HT and BT and
  
FixedTEST Wed Jan 23 09:11:26 2019 Valérian-MarlèneGRITGeneral3-alphaRUN9 : 3 alpha source run on MG10C & MG3T[B]RUN 9 (over night 22/01 ->23/01)[/B]
3 alpha source (not centered)
MG10C and MG3T ON
 3alpha_sourceMG10C-MM2_MG3T-MM3.pngHV_squares_plugged_on_MG4T-MG5T_andMG3T.pngHV_squares_plugged_on_MG4T-MG5T_andMG3T_afternightrun.pngMG10C_7missingstrips.png 
Fixed  Wed Jan 23 12:24:10 2019 MarlèneGRITGeneral Change cable MG1T - Exchange of detectors named MG9C & MG10C23/01/19

Change of the HT/BT cable of MG1T (the one
  
Fixed  Fri Jan 25 17:02:25 2019 MarlèneGRITGeneral Detectors numbering   MUGAST_detectors.pdf 
FixedTEST Thu Feb 7 09:52:08 2019 Freddy, MarlèneRUNGeneral3-alphaRUN10 : 3 alpha source on MG1-3-4-5-7Test run with alpha source in front of trapezoid
detectors.
 alpha_source_5trap_rawhistos_2.pngalpha_source_5trap_rawhistos.png 
Fixed  Thu Feb 14 08:50:58 2019 MarleneGRITGeneral3-alphaRUN 27RUN 27 on 14th of feb.
3 alpha source (one in front of trap., one
in front of square detectors)
 8x 
Fixed  Thu Feb 14 15:30:14 2019 Marlene Daniele JennyGRITGeneral3-alphaRUN 28RUN 28 14th Feb
3 alpha source on MG11A, MG3T, MG1T
 3alpha_sourceMG03XY._OKpng3alpha_sourceMG11XY.png3alpha_sourceMG01XY_OK.png 
Fixed  Fri Feb 15 15:31:28 2019 MarlèneGRITGeneralN/ADetectors bias    
Fixedcom2019 Mon Feb 18 10:41:36 2019 MarleneGRITGeneralN/ABias & current of detectors   bias.png 
FixedTEST Thu Feb 21 09:08:49 2019 FreddyRUNGeneral3-alphaRUN 30 - alpha on MG 1,3,5,7 [U][B]Run 30[/B][/U]

Alpha source run for trapezoid detectors
 3alpha_source_MG_1357_20Feb.pngLowV_bis_run30.pngLowV_run30.pngHV_run30.png 
FixedTEST Fri Mar 1 12:07:44 2019 FreddyGRITGeneral3-alphaEfficiency reconstructionblue: reconstructed efficiency using alpha
run 40 (arbitratily normalized)
red: simulated efficiency
 comparison.pdfcomparison.Calpha_eff.root 
FixedTEST Mon Mar 4 12:28:26 2019 Valerian and IulianMUST2General3-alphaNo DECS and acq error run 47-48-49. Das Saved after problem fixedFixed. Solution:
1) CENTRUM not configured on GENTRUM Master
2) GMT matrix on GMT3 (2nd slave) was all
 run49_error_acq.png 