Fixed, e793s, Mon Mar 15 10:18:37 2021, Taras, cyril, Léo, ONLINE, Hardware, CD2, Cats2 tripped
|
Cats2 tripped with vamos, had to be reset
Beam is back |
Fixed, e793s, Mon Mar 15 12:08:49 2021, Taras, cyril, Léo, ONLINE, Hardware, CD2, vamos dipole tripped at 12h05
|
Vamos dipole tripped at around 12h05, had to be reset |
Pending, e793s, Wed Mar 17 08:03:15 2021, Laetitia, Charlie, AGATA, Hardware, CD2, Agata issue
|
There is a problem on Agata (the dewars are not being filled), and the Ge's start to warm up. The beam has been stopped at ~08h05. |
Fixed, e793s, Wed Mar 17 09:53:38 2021, Cyril, AGATA, Hardware, CD2, Agata issue
|
Agata problem has been solved, a valve was closed by mistake.
Beam returned at 9h40. |
Fixed, TEST, Wed Jan 16 16:33:28 2019, Marlène, GRIT, General, 3-alpha, Status of detectors on 16/01/19 
|
MG1 has some issue with the HT cable (for the BT bias). There is a shortcut only under vacuum. We checked the connector but we have to do vacuum again to
check it. |
Fixed, TEST, Wed Jan 16 17:36:30 2019, Valérian-Marlène, GRIT, General, N/A, Calimero tests - How to configure Calimero in GECO    
|
Tests of Calimero time calibration with MUVI1 channel 2 (MM2)
1- channel 3 of all mates selected (see screeshots)
|
Fixed, TEST, Fri Jan 18 15:32:50 2019, Adrien Valérian, GRIT, General, N/A, Calimero tests Muvi2 MM7 MG10C 
|
After disabling nptool Multithreading, online is working fine again.
Concerning Calimero, it is currently working both on MUVI1 and MUVI2 for the X at least.
|
Fixed, , Tue Jan 22 11:15:15 2019, Marlène, GRIT, General, N/A, Square detectors  
|
MG9C (thin square detector coupled with thick square detector unplugged) noisy
On X channels --> no noise problem ( about 100 ORD in the dark)
On Y channels --> MATE 10 & 11 noisy ! (see picture below with noisy Y piedestals)
|
Ongoing, TEST, Tue Jan 22 18:52:23 2019, Valérian-Marlène, GRIT, General, 3-alpha, 3 alpha source run on MG10C & MG3T
|
3 alpha source (not centered)
MG10C and MG3T ON
MG10C is connected to MG5T HT and BT and seen on MUVI1 channel2 -> online name=MM2
|
Fixed, TEST, Wed Jan 23 09:11:26 2019, Valérian-Marlène, GRIT, General, 3-alpha, RUN9 : 3 alpha source run on MG10C & MG3T   
|
[B]RUN 9 (over night 22/01 ->23/01)[/B]
3 alpha source (not centered)
MG10C and MG3T ON
|
Fixed, , Wed Jan 23 12:24:10 2019, Marlène, GRIT, General, , Change cable MG1T - Exchange of detectors named MG9C & MG10C
|
23/01/19
Change of the HT/BT cable of MG1T (the one that had a short cut before when the chamber was closed)
|
Fixed, , Fri Jan 25 17:02:25 2019, Marlène, GRIT, General, , Detectors numbering
|
|
Fixed, TEST, Thu Feb 7 09:52:08 2019, Freddy, Marlène, RUN, General, 3-alpha, RUN10 : 3 alpha source on MG1-3-4-5-7 
|
Test run with alpha source in front of trapezoid detectors.
First run with new LABELS updated from "MM" to "MG0" in GECO.
|
Fixed, , Thu Feb 14 08:50:58 2019, Marlene, GRIT, General, 3-alpha, RUN 27 8x
|
RUN 27 on 14th of feb.
3 alpha source (one in front of trap., one in front of square detectors)
|
Fixed, , Thu Feb 14 15:30:14 2019, Marlene Daniele Jenny, GRIT, General, 3-alpha, RUN 28  
|
RUN 28 14th Feb
3 alpha source on MG11A, MG3T, MG1T
|
Fixed, , Fri Feb 15 15:31:28 2019, Marlène, GRIT, General, N/A, Detectors bias
|
|
Fixed, com2019, Mon Feb 18 10:41:36 2019, Marlene, GRIT, General, N/A, Bias & current of detectors
|
|
Fixed, TEST, Thu Feb 21 09:08:49 2019, Freddy, RUN, General, 3-alpha, RUN 30 - alpha on MG 1,3,5,7   
|
[U][B]Run 30[/B][/U]
Alpha source run for trapezoid detectors during the night with pedestals aligned and zero suppression. Cooling at 15°C.
|
Fixed, TEST, Fri Mar 1 12:07:44 2019, Freddy, GRIT, General, 3-alpha, Efficiency reconstruction  
|
blue: reconstructed efficiency using alpha run 40 (arbitratily normalized)
red: simulated efficiency
|
Fixed, TEST, Mon Mar 4 12:28:26 2019, Valerian and Iulian, MUST2, General, 3-alpha, No DECS and acq error run 47-48-49. Das Saved after problem fixed
|
Fixed. Solution:
1) CENTRUM not configured on GENTRUM Master
2) GMT matrix on GMT3 (2nd slave) was all 0
|
|