MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 51 of 132  Not logged in MUGAST
Status ExpNbrdown Date Author Type Category Target-Source Subject
Fixede744s Thu Apr 11 18:08:12 2019 MarlèneOTHERHardware changes in the electronic

Valérian, Chloé and François wrote:
Changes made in the electronic during the break between 14N and 14O

1) T_VAMOS-CATS1
it was not connected. We added 500 ns delay in CATS1


2) T_CATS1_HF
range changed from 2µs to 200 ns


T_VAMOS-CATS1 is now connected in T_CATS1_CATS2 as posted this night. We have tried to add 500ns more delay (800ns in total) to CATS and it did not work, so we suspected the TAC or ADC. This is why we inverted T_VAMOS-CATS1 with T_CATS1_CATS2. For this TAC, CATS is the START (and is not delayed by 500ns).
Fixede744s Thu Apr 11 18:20:27 2019 Valerian and DiegoOFFLINESoftwareCH2Timing issue between vamos and mugast
A cut is seen on the TAC between MUGAST and the focal plane multiwire of VAMOS when one select XF or DC0_Y>-1500


Edit: This must be a reconstruction problem as it is not seen in the Raw files (using DC0_X for example)

Edit: This is not only a reconstruction problem as the cut appear again if one select DC0_Y

Edit: We changed VAMOS FAG validation window from 500ns to 800ns
Attachment 1: T_MUGAST_FPMW.png
T_MUGAST_FPMW.png
Attachment 2: T_MUGAST_FPMW1.png
T_MUGAST_FPMW1.png
Fixede744s Thu Apr 11 18:36:34 2019 BeyhanBEAMGeneral before optimization : beam reconstructed profil with VAMOS and assoicated settings
target : hole
cats2 out
about 4000 pps triggering VAMOS
Attachment 1: IMG_1583.jpg
IMG_1583.jpg
Attachment 2: IMG_1582.jpg
IMG_1582.jpg
Fixede744s Thu Apr 11 19:09:35 2019 OttoRunRUNGeneralEmptyRun81
*************************
Run : 81
*************************

Start at 2019-04-11 19:09:20.734153

Run Conditions:
  • Operator: Adrien
  • Trigger: VAMOS %1
  • Target/Source: no target
  • Beam: 14O
  • Intensity: ~ 5e+3
  • Brho = 0.69186 Tm
  • !CATS/VAMOS (CATS Efficiency)! : XXXX

Comment:

Aligning beam, daq on for online monitoring

Stop at 2019-04-11 19:49:20.042737
Stop comment:
Fixede744s Thu Apr 11 19:19:21 2019 Iulian BeyhanRUNGeneralN/AChange target CH2 to no target
Target hole is now removed -> No target
Fixede744s Thu Apr 11 19:47:15 2019 MarlèneOFFLINEGeneral CATS charges in RUN76 and RUN77
Offline analysis to reconstruct the charges in X and Y for CATS.
The charges are lower in RUN77 because bias was reduced to 580V.
Wires seem to be all OK.
Attachment 1: CATS_stripY_run76.pdf
CATS_stripY_run76.pdf
Attachment 2: CATS_stripX_run76.pdf
CATS_stripX_run76.pdf
Attachment 3: CATS_stripY_run77.pdf
CATS_stripY_run77.pdf
Attachment 4: CATS_stripX_run77.pdf
CATS_stripX_run77.pdf
Attachment 5: CATS_RUN77.pdf
CATS_RUN77.pdf
Fixede744s Thu Apr 11 19:49:29 2019 OttoRunRUNGeneralEmptyRun82
*************************
Run : 82
*************************

Start at 2019-04-11 19:49:24.212733

Run Conditions:
  • Operator: Adrien
  • Trigger: VAMOS %1
  • Target/Source: no target
  • Beam: 14O
  • Intensity: ~ 5e+3
  • Brho = 0.69186 Tm
  • !CATS/VAMOS (CATS Efficiency)! : XXXX

Comment:

Aligning beam, daq on for online monitoring

Stop at 2019-04-11 19:54:51.148174
Stop comment: 5 PI emittance for 14O beam
Fixede744s Thu Apr 11 19:55:18 2019 OttoRunRUNGeneralEmptyRun83
*************************
Run : 83
*************************

Start at 2019-04-11 19:54:53.297399

Run Conditions:
  • Operator: Adrien
  • Trigger: VAMOS %1
  • Target/Source: no target
  • Beam: 14O
  • Intensity: ~ 5e+3
  • Brho = 0.69186 Tm
  • !CATS/VAMOS (CATS Efficiency)! : XXXX

Comment:

Aligning beam, daq on for online monitoring
2Pi beam emittance 14O

Stop at 2019-04-11 20:56:43.227260
Stop comment:
Fixede744s Thu Apr 11 20:55:17 2019 mugastONLINEGeneral screenshots RUN77
Attachment 1: run77_tab1.png
run77_tab1.png
Attachment 2: run77_tab11.png
run77_tab11.png
Attachment 3: run77_tab10.png
run77_tab10.png
Attachment 4: run77_tab9.png
run77_tab9.png
Attachment 5: run77_tab8.png
run77_tab8.png
Attachment 6: run77_time_calibrator.png
run77_time_calibrator.png
Attachment 7: run77_tab6.png
run77_tab6.png
Attachment 8: run77_tab5.png
run77_tab5.png
Attachment 9: run77_tab4.png
run77_tab4.png
Attachment 10: run77_tab3.png
run77_tab3.png
Attachment 11: run77_tab2.png
run77_tab2.png
Attachment 12: run77_tab0.png
run77_tab0.png
Fixede744s Fri Apr 12 00:29:11 2019 FrançoisBEAMGeneral no beam
They aligned the beam from CIME to G1 with 15N.
They are now working with 14O.
No beam on G1.
Fixede744s Fri Apr 12 01:27:41 2019 OttoRunRUNGeneralEmptyRun84
*************************
Run : 84
*************************

Start at 2019-04-12 01:25:57.979942

Run Conditions:
  • Operator: Adrien
  • Trigger: VAMOS %1
  • Target/Source: hole target 1
  • Beam: 14O
  • Intensity: ~ 4e+4
  • Brho = 0.69186 Tm
  • !CATS/VAMOS (CATS Efficiency)! : XXXX

Comment:

Aligning beam in VAMOS
Fixede744s Fri Apr 12 01:37:24 2019 OttoRunRUNGeneralEmptyRun85
*************************
Run : 85
*************************

Start at 2019-04-12 01:37:03.553133

Run Conditions:
  • Operator: Adrien
  • Trigger: VAMOS %1
  • Target/Source: hole target 1
  • Beam: 14O
  • Intensity: ~ 4e+4
  • Brho = 0.69186 Tm
  • !CATS/VAMOS (CATS Efficiency)! : XXXX

Comment:

Aligning beam in VAMOS, restart after crash of the aquisition
Fixede744s Fri Apr 12 01:50:50 2019 OttoRunRUNGeneralEmptyRun86
*************************
Run : 86
*************************

Start at 2019-04-12 01:49:22.775057

Run Conditions:
  • Operator: Jenn
  • Trigger: VAMOS %1
  • Target/Source: hole target 1
  • Beam: 14O
  • Intensity: ~ 4e+4
  • Brho = 0.6912 Tm
  • !CATS/VAMOS (CATS Efficiency)! : XXXX

Comment:

Start CATS and target 1 hole

Stop at 2019-04-12 01:59:01.799964
Stop comment: type your message here
Fixede744s Fri Apr 12 01:56:35 2019 Jenn, Louis, Cyril,Daniele and ChloeBEAMGeneralEmptyBrho for the new 14O setting RUN86
CATS2 on and target position 1: hole
Beam intensity in CATS: 4E4
VAMOS intensity 1.3E4

CATS2 off and target position 1 = hole
Beam intensity in VAMOS = 4e4 pps (~100% transmission)
Attachment 1: Brho_14O_run86.png
Brho_14O_run86.png
Attachment 2: 20190412_193918.jpg
20190412_193918.jpg
Fixede744s Fri Apr 12 02:19:02 2019 Francois, Jenn, Cyril, LouisCATSHardwareN/ASet Division for CATS and Agata
Enter in G1 to set division on CATS and Agata
CATS divion: 1000
Agata division: 100
Fixede744s Fri Apr 12 02:23:57 2019 OttoRunRUNGeneralCH2Run87: First physics run of 14O
*************************
Run : 87
*************************

Start at 2019-04-12 02:20:23.557408

Run Conditions:
  • Operator: Jenn
  • Trigger: MG Backward, MG90, MUST2, CATS %1000 AGATA %100
  • Target/Source: CH2 target 2
  • Beam: 14O
  • Intensity: ~ 4e+4
  • Brho = 0.6912 Tm
  • !CATS/VAMOS (CATS Efficiency)! : XXXX

Comment:

Start Target 2: 104 um
VAMOS is closed.

Stop at 2019-04-12 06:23:53.544073
Stop comment: type your message here
Fixede744s Fri Apr 12 04:13:49 2019 François, ChloéOTHERHardware TAC to do list
TAC of
- MUGAST-HF : HFvalidated is suspected to come every 200ns after MUGAST, it suggests to go to a range of 500ns
- T_SI_CATS1 : CAST1 stop comes before T_SI start (roughly 100ns to be delayed)
Fixede744s Fri Apr 12 06:24:31 2019 OttoRunRUNGeneralCH2Run88
========================================
*************************
Run : 88
*************************

Start at 2019-04-12 06:24:07.821497

Run Conditions:
  • Operator: Louis
  • Trigger: MG Backward, MG90, MUST2, CATS %1000 AGATA %100
  • Target/Source: CH2 target 2
  • Beam: 14O
  • Intensity: ~ 3e+4
  • Brho = 0.6912 Tm
  • !CATS/VAMOS (CATS Efficiency)! : XXXX

Comment:

Start Target 2: 104 um
VAMOS is closed.

Stop at 2019-04-12 08:37:19.321598
Stop comment: Stop to setup VAMOS
Fixede744s Fri Apr 12 06:51:59 2019 LouisBEAMGeneralN/ABrho and energy of the 14O beam
Attachment 1: DSC_3261.JPG
DSC_3261.JPG
Fixede744s Fri Apr 12 07:26:29 2019 iulianRUNGeneralCH2Run88
Please add all the spectra from the visualization into the elog and all the scales.... see runs before



OttoRun wrote:
*************************
Run : 88
*************************

Start at 2019-04-12 06:24:07.821497

Run Conditions:
  • Operator: Louis
  • Trigger: MG Backward, MG90, MUST2, CATS %1000 AGATA %100
  • Target/Source: CH2 target 2
  • Beam: 14O
  • Intensity: ~ 3e+4
  • Brho = 0.6912 Tm
  • !CATS/VAMOS (CATS Efficiency)! : XXXX

Comment:

Start Target 2: 104 um
VAMOS is closed.

ELOG Home Page