MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 4 of 132  Not logged in MUGAST
Status ExpNbr Date Author Type Category Target-Source Subjectdown
Fixede793s Sun Mar 14 15:52:36 2021 Faïrouz, Juan Lois, WiltonRUNGeneralCD2run62 screenshots at 15h52 (14/03/2021)
Fixede793s Sun Mar 14 12:02:10 2021 Faïrouz, Juan LoisRUNGeneralCD2run62 screenshots at 11h57 (14/03/2021)
Fixede793s Sun Mar 14 02:24:22 2021 Léo, Serge, NishuONLINEGeneralCD2run62 screenshots at 02:25
Fixede793s Sun Mar 14 00:12:41 2021 Léo, Serge, NishuONLINEGeneralCD2run62 screenshots at 00h13
Fixede793s Sat Mar 13 05:33:41 2021 Faïrouz, Juan LoisRUNGeneralCD2run61 screenshots at 5h30
Fixede793s Sat Mar 13 03:50:54 2021 Faïrouz, Juan LoisRUNGeneralCD2run61 screenshots at 3h44
Fixede793s Sat Mar 13 01:41:21 2021 Juan Lois & faïrouzRUNGeneralCD2run61 screenshots at 1h40
Fixede793s Thu Mar 11 23:30:27 2021 Faïrouz, FrancoRUNGeneralCD2run56 screenshots at 23h20
Fixede793s Thu Mar 11 20:23:58 2021 Faïrouz, FrancoRUNGeneralCD2run56 screenshots at 20h18
Fixede793s Fri Mar 12 06:23:44 2021 Taras, Nishu, Chloe, CharlieRUNGeneralCD2run56 screenshots at 06h15
Screenshots attached
Fixede793s Fri Mar 12 04:19:51 2021 Taras, Nishu, Chloe, CharlieRUNGeneralCD2run56 screenshots at 04h18
Images attached.
Fixede793s Fri Mar 12 02:01:52 2021 Taras, Nishu, Chloe, CharlieRUNGeneralCD2run56 screenshots at 02h10
screenshots of run 56 attached below
Fixede793s Thu Mar 11 11:07:05 2021 mugastOFFLINEGeneralCD2run53 - Egamma vs Ex
FixedTEST Tue Mar 5 18:18:46 2019 iulian MarleneMUST2Hardware3-alpharun 97 3a run Muvi3 night run
Run 87

stop@ 5:42 am

3a run Muvi3 (MM)
-4 deg
pedestals aligned and suppressed
FixedTEST Thu Mar 7 09:29:47 2019 MarleneGRITGeneral run 96 double alpha source run shifted 60mm
RUN 96

double alpha source
MG1
MG3
MG4
MG5

MG7
-
-
MG11

MM1
MM2
MM3
MM4

Missing strips in Y the same as in run 40

Problem with the annular MG11 -> no current when bias is put... --> to be checked

Problem with the insertion of the target holder --> solved

I also checked again MG9 (still noisy)
and MG10 with 140V I=0,52uA and counting rate is the same order of magnitude as trapezoid. Spectra are OK (not saved).
Fixede744s Fri Apr 12 08:01:12 2019 valerianRUNGeneralCH2run 88 screenshots
FixedTEST Tue Mar 5 17:48:38 2019 iulian MarleneMUST2SoftwareN/Arun 82 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM4 X . good run
Run 82

Calimero on Muvi3 MM4(X)

Gamme 600 ns
FixedTEST Tue Mar 5 17:11:29 2019 iulian MarleneMUST2SoftwareN/Arun 81 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM3 X . good run
Run 81

Run 80 bad
Calimero on Muvi3 MM3(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
FixedTEST Tue Mar 5 16:42:56 2019 iulian MarleneMUST2SoftwareN/Arun 79 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM2 X . good run
Run 79

Run 78 bad
Calimero on Muvi3 MM2(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
FixedTEST Tue Mar 5 16:20:11 2019 iulian MarleneMUST2SoftwareN/Arun 77 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X . good run
Run 76


Calimero on Muvi3 MM1(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)

ELOG Home Page