MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 44 of 132  Not logged in MUGAST
Fixed, e744s, Wed Apr 10 09:51:36 2019, OttoRun, RUN, General, N/A, Run24 
*************************
Run : 24
*************************

Start at 2019-04-10 09:50:23.123384

Run Conditions:
  • Operator: A. Lemasson
  • Trigger: Vamos Pb with DC 63
  • Target/Source: none
  • Beam: VAMOS Closed
  • Intensity:

Comment:
Test ACQ


Stop at 2019-04-10 10:10:11.417806
Stop comment:
Fixed, e786s, Sun Jun 23 17:32:14 2019, OttoRun, RUN, General, Cryo-empty, Run24 
*************************
Run : 24
*************************

Start at 2019-06-23 17:32:11.627642

VAMOS tunning Continued with 2.5 10*4
After Fixing the IC preamp issue
Brho0=1.05 Tm

DC pressure 6 mbar
IC pressure 120 mbar

CATS2 inserted


TRIGGER
VAMOS %1000
MUGAST
MUS2
CATS 2%100

Stop at 2019-06-23 17:34:31.534959
Stop comment:
Fixed, e768s, Tue Jul 16 09:32:06 2019, OttoRun, RUN, General, N/A, Run24 
*************************
Run : 24
*************************

Start at 2019-07-15 22:58:34.115263

fsc / calibT run

internal pulser frequency now set to 300 Hz (was about 1 kHz before)

GMT: MUGAST (square detector not biased) + MUST2

Time calibrator:
range -> 640 ns
freq. -> 20 ns

fsc modified to wait 60s instead of 10s

Stop at 2019-07-16 09:32:07.900270
Stop comment:

Stop at 2019-07-16 09:32:08.186319
Stop comment:

Stop at 2019-07-16 09:32:08.404715
Stop comment:

Stop at 2019-07-16 09:32:08.595475
Stop comment:

Stop at 2019-07-16 09:32:08.836126
Stop comment:

Stop at 2019-07-16 09:32:09.032791
Stop comment:
Fixed, com2019, Fri Apr 5 11:49:41 2019, OttoRun, RUN, General, CD2, Run238 
*************************
Run : 238
*************************

Start at 2019-04-05 11:48:16.391241

Run 238

TRigger Vamos
1e3
Fixed, com2019, Fri Apr 5 08:53:05 2019, OttoRun, RUN, General, N/A, Run236 
*************************
Run : 236
*************************

Start at 2019-04-05 08:51:52.531911

Run Conditions:
  • Operator: Diego
  • Trigger: CATS1
  • Target/Source: CD2 1 mg
  • Beam: 16O
  • Intensity: 1e3
  • CAST2
Numexo DAQ issue ----
Comment:
FINGER VAMOS MOVEMENT
Fixed, com2019, Fri Apr 5 08:40:05 2019, OttoRun, RUN, General, N/A, Run235 
*************************
Run : 235
*************************

Start at 2019-04-05 08:38:48.511954

Run Conditions:
  • Operator: Diego
  • Trigger: CATS1
  • Target/Source: CD2 1 mg
  • Beam: 16O
  • Intensity: 1e3
  • CATS1 & CATS2 in place
Numexo DAQ issue ----
Comment:
VAMOS TACS Test
Fixed, TEST, Thu Jun 13 09:47:39 2019, OttoRun, RUN, General, 3-alpha, Run235 
*************************
Run : 235
*************************

Start at 2019-06-13 09:47:19.893278

Discard run
alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!

Stop at 2019-06-13 14:23:23.430392
Stop comment: type your message here
Fixed, com2019, Fri Apr 5 08:28:36 2019, OttoRun, RUN, General, N/A, Run234 
*************************
Run : 234
*************************

Start at 2019-04-05 08:28:09.050268

Run Conditions:
  • Operator: A. Lemasson
  • Trigger: VAMOS
  • Target/Source: CD2 1 mg
  • Beam: 16O
  • Intensity: 1e3
  • CATS1 & CATS2 in place
Numexo DAQ issue ----
Comment:
First run with MUGAST trigger and target
Fixed, TEST, Thu Jun 13 09:46:40 2019, OttoRun, RUN, General, 3-alpha, Run234 
*************************
Run : 234
*************************

Start at 2019-06-13 09:45:49.383074

Discard run
alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!

Stop at 2019-06-13 09:47:06.006149
Stop comment: Discard this run
Fixed, com2019, Fri Apr 5 07:15:04 2019, OttoRun, RUN, General, N/A, Run233 
*************************
Run : 233
*************************

Start at 2019-04-05 07:13:06.173030

Run Conditions:
  • Operator: A. Lemasson
  • Trigger: VAMOS
  • Target/Source: CD2 1 mg
  • Beam: 16O
  • Intensity: 1e3
  • Only CATS2 in place
Numexo DAQ issue ----
Comment:
First run with MUGAST trigger and target
Fixed, TEST, Wed Jun 12 18:40:57 2019, OttoRun, RUN, General, 3-alpha, Run233 
*************************
Run : 233
*************************

Start at 2019-06-12 18:40:34.500264

MUGAST 3alpha run pedestal supression
temperature -1 deg (trap) & -4.5 deg (must2) consigne 5 deg

without pedestal suppression

alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!

Stop at 2019-06-13 09:06:36.277090
Stop comment:
    Fixed, TEST, Sun Jun 23 10:31:08 2019, OttoRun, RUN, General, 3-alpha, Run233 energy_cal_run0233.png

OttoRun wrote:
*************************
Run : 233
*************************

Start at 2019-06-12 18:40:34.500264

MUGAST 3alpha run pedestal supression
temperature -1 deg (trap) & -4.5 deg (must2) consigne 5 deg

without pedestal suppression

alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!

Stop at 2019-06-13 09:06:36.277090
Stop comment:
Fixed, com2019, Fri Apr 5 07:15:05 2019, OttoRun, RUN, General, N/A, Run232 
*************************
Run : 232
*************************

Start at 2019-04-05 07:02:55.391269

Run Conditions:
  • Operator: A. Lemasson
  • Trigger: VAMOS
  • Target/Source: CD2 1 mg
  • Beam: 16O
  • Intensity: 1e3
  • Only CATS2 in place
Numexo DAQ issue ----
Comment:
First run with MUGAST trigger and target
Fixed, TEST, Tue Jun 11 21:44:38 2019, OttoRun, RUN, General, 3-alpha, Run232 
*************************
Run : 232
*************************

Start at 2019-06-11 21:44:02.397316

MUGAST 3alpha run pedestal supression
temperature -1 deg (trap) & -4.5 deg (must2) consigne 5 deg

with pedestal suppression

alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!

Stop at 2019-06-12 10:05:20.279209
Stop comment: type your message here
Fixed, com2019, Fri Apr 5 06:46:44 2019, OttoRun, RUN, General, N/A, Run231 
*************************
Run : 231
*************************

Start at 2019-04-05 06:46:25.550428

Run Conditions:
  • Operator: A. Lemasson
  • Trigger: VAMOS
  • Target/Source: CD2 1 mg
  • Beam: 16O
  • Intensity: 1e3
  • Only CATS2 in place
Numexo DAQ issue ----
Comment:
First run with MUGAST trigger and target
Fixed, TEST, Tue Jun 11 18:59:36 2019, OttoRun, RUN, General, 3-alpha, Run231 
*************************
Run : 231
*************************

Start at 2019-06-11 18:58:31.909158

MUGAST 3alpha run pedestal supression
temperature -1 deg (trap) & -4 deg (must2) consigne 5 deg

no pedestal suppression (but alignement)

alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!

Stop at 2019-06-11 21:40:22.107913
Stop comment:
Fixed, TEST, Thu May 30 15:58:20 2019, OttoRun, RUN, General, 3-alpha, Run230 
*************************
Run : 230
*************************

Start at 2019-05-30 15:57:54.390230

MUGAST 3alpha run pedestal supression
temperature -3 deg (trap) & -4 deg (must2) consigne 5 deg

no pedestal suppression (but alignement)
Unsual bias for MG4T =130V, I=4 uA because no trigger on Y.

same as before except that MG7 had a mate hidden on Y (now it is on)

Stop at 2019-05-30 15:59:04.950841
Stop comment: type your message here

Stop at 2019-05-30 16:01:48.801380
Stop comment:
Fixed, TEST, Thu May 30 16:04:09 2019, OttoRun, RUN, General, 3-alpha, Run230 
*************************
Run : 230
*************************

Start at 2019-05-30 16:04:07.241232

MUGAST 3alpha run pedestal supression
temperature -3 deg (trap) & -4 deg (must2) consigne 5 deg

no pedestal suppression (but alignement)
Unsual bias for MG4T =130V, I=4 uA because no trigger on Y.

same as before except that MG7 had a mate hidden on Y (now it is on)

Stop at 2019-05-30 16:06:21.092742
Stop comment: type your message here
Fixed, TEST, Thu May 30 16:08:39 2019, OttoRun, RUN, General, 3-alpha, Run230 
*************************
Run : 230
*************************

Start at 2019-05-30 16:08:37.332238

MUGAST 3alpha run pedestal supression
temperature -3 deg (trap) & -4 deg (must2) consigne 5 deg

no pedestal suppression (but alignement)
Unsual bias for MG4T =130V, I=4 uA because no trigger on Y.

same as before except that MG7 had a mate hidden on Y (now it is on)

Stop at 2019-05-30 16:09:38.806154
Stop comment: type your message here
Fixed, TEST, Thu May 30 16:10:18 2019, OttoRun, RUN, General, 3-alpha, Run230 run229_3lapha_mugast_calibre.pngrun229_3alpha_must2.pngrun229_3alpha.png
*************************
Run : 230
*************************

Start at 2019-05-30 16:09:44.953775

MUGAST 3alpha run pedestal supression
temperature -3 deg (trap) & -4 deg (must2) consigne 5 deg

no pedestal suppression (but alignement)
MG4T =95V

same as before except that MG7 had a mate hidden on Y (now it is on)

Stop at 2019-05-31 06:39:11.358615
Stop comment: type your message here

ELOG Home Page